DSO3D12 3-in-1 Digital Oscilloscope

Model: DSO3D12 | Brand: meterk

Dual Channel 120Mhz Bandwidth 250MSa/s Sampling Rate Handheld Oscilloscope, Multimeter, and Signal Generator

1. Introduction

The meterk DSO3D12 is a versatile 3-in-1 handheld device integrating a digital oscilloscope, a high-precision multimeter, and a signal generator. Designed for a wide range of applications including basic R&D, education, car repair, and audio/radio amateur use, it offers dual-channel operation, 120MHz bandwidth, and a 250MSa/s real-time sampling rate. This manual provides detailed instructions for its setup, operation, and maintenance.

meterk DSO3D12 3-in-1 Digital Oscilloscope
The meterk DSO3D12 device, showcasing its oscilloscope display, control buttons, and input ports.

For a comprehensive guide, you may also refer to the official user manual PDF: DSO3D12 User Manual (PDF).

2. Safety Precautions

Please read and understand the following safety precautions before operating the device to prevent injury or damage to the instrument.

  • High Voltage: The x10 range of the probe measures up to 220V. An x100 probe is required for higher voltages.
  • Probe Attenuation: Before measuring voltages greater than 40V, switch the probe to the X10 position.
  • Charging: The charging voltage is 5V. Use a computer or mobile phone charger.
  • Charging and Operation: When measuring high-voltage or non-isolated circuits, the device cannot be used while charging (powered by a built-in battery).
  • Multimeter Use: Do not use the multimeter while charging.
  • Current Measurement: Never insert the red probe into the current hole to measure voltage; this will blow the fuse or damage the machine. After measuring current, remember to change the red probe to the voltage hole.
  • Current Measurement Duration: When the multimeter measurement current is greater than 6A, the duration should not exceed 10 seconds to avoid line damage caused by heating.
  • Simultaneous Use: When measuring high voltages, do not use the oscilloscope and multimeter at the same time.
  • Probe Switch: Always switch the probe to the 10X position before connecting high voltage signals to prevent damage to the oscilloscope's internal circuit.

3. Setup

3.1 Powering On and Charging

Press and hold the power button to turn the device on or off.

When the battery is low, connect the device to a 5V adapter using the USB cable for charging. The charging current is approximately 700mA. It is recommended to use a computer or mobile phone charger. The LED indicator will light red during charging and turn green when fully charged.

3.2 Probe Calibration

The probes included with this device are pre-compensated. If you use a new probe, it must be calibrated before use to ensure accurate measurements (applies only to 10X/100X probes).

  1. Switch the probe to X10.
  2. Connect the probe to a 1KHz square wave with an amplitude of approximately 2V-3V. You can use the built-in signal generator for this purpose.
  3. Adjust the horizontal and vertical scales of the oscilloscope until the waveform is stably displayed at a suitable position on the screen.
  4. Use a screwdriver to rotate the adjustable capacitor on the probe until you achieve a waveform similar to the correctly calibrated example shown below.
Probe calibration diagram
Diagram illustrating correctly calibrated (middle) and incorrectly calibrated (left, right) probe waveforms.

Probe Bandwidth

The 1X probe typically has a bandwidth within 5MHz due to its large input capacitance. The 10X probe offers the standard bandwidth of the oscilloscope. For signals with bandwidths greater than 5MHz, always use the 10X position.

Probe Withstand Voltage

The maximum withstand voltage is usually marked on the probe, typically 600Vpk-pk.

4. Operating Instructions

4.1 Button Functions

The device features various buttons for controlling its functions. Below is a table summarizing their primary and secondary (Shift + button) functions.

Button Default Function Shift + Button Function
OKConfirms menu selections.N/A
AutoAutomatic waveform adjustment.Auto2: Enter real-time auto (CH1 only).
StopRun / Stop waveform acquisition.Zoom: Turn on/off zoom mode.
SingleSingle trigger mode.Ref: Reference waveform.
AC/DCSwitch DC/AC coupling (CH1).Switch DC/AC coupling (CH2).
CH1/CH2Change CH1/CH2 priority.CH2: Open/close CH2.
MenuOpen/close the menu.TgMode: Change trigger mode (Auto / Normal).
Shift↑Activates secondary functions of other buttons.N/A
x1x10Probe attenuation setting (CH1).Probe attenuation setting (CH2).
GenTurn on/off signal generator interface.TgEdge: Change trigger edge (rising/falling).
SaveSave a screenshot of the waveform.View: View waveform screenshots.
50%Quick centering (CH1).Quick centering (CH2).
DMMTurn multimeter on/off.Full: Multimeter displayed in full screen.
SelectMultimeter function selection.N/A
mV/VAdjust vertical sensitivity.N/A
s/nsAdjust time base.N/A

4.2 Quick Start - Auto Mode

After connecting the probe to a signal, press the [Auto] button. The oscilloscope will automatically detect the signal and adjust the range. If no waveform is displayed, ensure the probe is correctly connected and a signal is present.

Oscilloscope in Auto mode
The oscilloscope display in fully automatic mode, showing a sine wave.

4.3 Sensitivity and Time Base Adjustment

Use the "mV"/"V" buttons to scale the waveform vertically (adjust vertical sensitivity). The "s"/"ns" buttons scale the waveform horizontally (adjust time base). Vertical sensitivity indicates the voltage per grid, and time base represents the time per grid.

4.4 Signal Generator

To use the built-in signal generator:

  1. Connect the probe to the generator output pin.
  2. Press the [Gen] key to open the generator interface.
  3. Press the [Stop] key to change the generator to the [ON] state.
  4. Use the [ns] key to switch between sine, square, and triangle waveforms.
  5. Press the [Single] key to reset the frequency to 1.00KHz.
  6. Press [OK] or [Gen] to confirm and exit.

The signal amplitude is 2.5V. Maximum frequency for sine wave is 5MHz, and for other waveforms is 1MHz. Square waves allow duty cycle adjustment. Note: For frequencies above 1MHz, square wave signals may exhibit small horizontal jitter.

Signal generator display
Digital display showing signal generator frequency (001.00 KHz) and duty cycle (50.0%).

4.5 Multimeter Functions

To use the multimeter:

  1. Ensure the red probe is in the voltage hole.
  2. Press the [Select] key to open the multimeter function table.
  3. Select the desired function and press the [OK] key.

For current measurement, insert the red probe into the current hole, ensuring it is in series with the power line. The multimeter does not support operation while charging.

Multimeter display
The oscilloscope displaying the multimeter function with a voltage reading of 12.01V.
Multimeter function selection screen
Multimeter function selection screen, showing options for voltage, current, resistance, and capacitance.

4.6 Dual Channel Operation

Press [Shift] + [CH1/CH2] to turn Channel 2 on or off. Channel 1 is always on. When not using two channels, turn off CH2 for higher sampling rates and reduced power consumption. In dual-channel mode, press [CH1/CH2] to switch priority. The channel with high priority can be moved using the up/down buttons.

Dual channel display
Oscilloscope screen showing two distinct waveforms for dual channel operation.

4.7 Trigger System

Triggering stabilizes the waveform on the screen and allows capturing specific segments. The DSO3D12 supports edge triggering (rising or falling edge).

Trigger Level

The trigger level is automatically adjusted by default. To adjust manually:

  1. In Menu -> Trigger, change "Trigger Level" to "Manual".
  2. Press the [Shift] key, then use the up and down keys to move the trigger level.
Trigger level indicator
Oscilloscope screen showing the trigger level indicator (red arrow) on a square wave.

Trigger Sources

The small arrow on the right indicates the trigger source (yellow for CH1, other colors for CH2). Only one channel can be used as a trigger source at a time.

Trigger Mode

Options include Automatic, Normal, and Single trigger. Automatic is commonly used. Normal mode is suitable for discontinuous waveforms, requiring manual trigger level adjustment. Single trigger captures one frame of waveform per press, useful for sporadic pulses.

Trigger settings screen
Oscilloscope screen showing trigger settings for mode, level, edge, and source.

4.8 Cursors

Press [Shift] + [s] to toggle horizontal cursors. Press [Shift] + [ns] to toggle vertical cursors. Use the left/right arrow keys to move horizontal cursors. After pressing [Shift], press horizontal arrow keys again to move the second horizontal cursor. The same applies to vertical cursors.

For vertical cursors, Y1 is the voltage at the upper dashed line, Y2 is the voltage at the lower dashed line, and dY = Y1 - Y2 is the voltage difference. Vertical cursors measure only one channel, the one with the highest priority.

Oscilloscope display with cursors
Oscilloscope display showing horizontal and vertical cursors for precise waveform measurements.

4.9 Screenshots and Image Viewing

To capture the current screen, click the [Save] button.

To view saved images, press [Shift] + [Save] to enter the Image View window. Here you can browse or delete historical screenshots. Click [x1x10] to delete, [OK] to zoom in/out, and [Save] to return to the oscilloscope window.

6. Specifications

6.1 Oscilloscope Parameters

Parameter Value Parameter Value
ModelDSO3D12CouplingAC/DC
Channels2AUTOSupport
Screen size3.2 inchMeasurement14 types
Sampling rate250MSa/sDC Offset±2%
Bandwidth120MHzXY modeSupport
Equivalent sample500MScreenshotSupport
Rise time<3nsFrequency±0.01%
Storage depth128KbSingleSupport
Impedance1M ΩFFTSupport
Time base5ns - 10sWave out amp2.5V ±0.05
Ver sensitivity10mV/div-10V/divWave out freq5MHz(sin) 1MHz(other)
Max voltage±40V (x1), ±400V (x10)Shell size145*86*33mm
Trigger modeAuto/Normal/SingleLanguageCN/EN
Trigger typerise/fallChargerTYPE-C / 5V
Display modeYT / RollCharge current700mA
PersistenceNone/1s/∞Weight266g

6.2 Multimeter Parameters

Function Range Precision
DC Voltage600mV/6.00V/60.0V/600V/750V±(0.5%+3)
AC Voltage600mV/6.00V/60.0V/600V±(1%+3)
DC Current600mA/10A±(2%+5)
AC Current600mA/10A±(3%+5)
Pay attention:600mA hole with 1A fuse, 10A hole without fuse
Resistance600.0Ω±(1.5%+3)
6.000kΩ/60.00kΩ/600.0kΩ±(1%+3)
6.000MΩ±(1.5%+5)
60.00MΩ±(3%+3)
Capacitance60.00nF/600.0nF/6.000μF±(10%+5)
60.00μF/600.0μF±(15%+5)
Diode0.0V – 3.3V, Display "OL" above 3.3V
ContinuitySound at 50Ω and below

Multimeter Calibration

If you have a higher accuracy multimeter, you can calibrate the DSO3D12 for improved accuracy:

  1. Enter DMM full-screen mode and select DC1000V. Measure the same voltage with both the DSO3D12 and your reference multimeter.
  2. Quickly click "50%->50%->Save" to enter calibration mode. Adjust the value using the up/down keys until it matches your reference meter. Click "OK" to confirm.
  3. Repeat for resistor, diode, capacitor, current, and mV functions. AC parameters do not require separate calibration.

7. Troubleshooting (Q&A)

7.1 How do I reset the device to its initial state?

Answer: Open Menu -> Set, then long press the "Auto" button to reset.

7.2 What is the difference between probe 1X and 10X?

Answer: The 1X probe does not attenuate the signal, while the 10X probe attenuates it by 10 times. Use 10X for voltages above 40V. The 1X bandwidth is limited to about 5MHz. Generally, use 10X, and 1X for signals with amplitudes less than a few hundred millivolts. Ensure the probe setting in the oscilloscope menu matches the physical probe setting.

7.3 Why does the waveform refresh become slower when the time base is >50ms?

Answer: A time base of 50ms means each grid represents 50ms. With 12 horizontal grids, one full screen sampling takes 50ms * 12 = 600ms, leading to a slower refresh rate.

7.4 How to measure DC power voltage?

Answer: Set coupling to "DC", probe to 10X (match menu setting). Connect probe to power output, then press [Auto]. Observe the "Mean" value.

7.5 How to measure 220V mains?

Answer: First, remove the USB cable (for isolation). Adjust probe to 10X, vertical sensitivity to 100V, and time base to 10ms. Carefully connect the probe to the neutral and live wires.

7.6 How to measure audio signals?

Answer: Set coupling to "AC", probe to 10X. Connect probe to audio output. Adjust time base to 1-10ms and voltage sensitivity to about 500mV (adjust based on volume).

7.7 How to measure main board communication or bus signals?

Answer: Connect the probe to the signal. Set coupling to DC, probe to 10X (match menu setting). Press [Auto]. If the waveform is incorrect, manually adjust the time base using the [s]/[ns] buttons.

7.8 How to measure power supply ripple?

Answer: Set coupling to "AC" in the menu, probe to 1X (match menu setting). Adjust time base to about 200us. Adjust vertical sensitivity to 10mV. If the waveform exceeds the screen, click the [V] key to reduce it until all ripples are visible.

7.9 Can you use the signal generator and the oscilloscope at the same time?

Answer: Yes, the device is designed to function as a 3-in-1 tool, allowing simultaneous use of the signal generator and oscilloscope features.

7.10 Is it possible to download screenshots to a computer via the USB-C interface?

Answer: The manual indicates the ability to save screenshots internally and view them. For direct download to a computer via USB-C, consult the latest firmware documentation or manufacturer support for specific instructions, as this functionality may depend on software updates.

8. User Tips

  • The device offers good accuracy for non-professional use and is highly regarded for its convenience, compactness, and capabilities, especially considering its price point.
  • Many users find the device to be an excellent value for money, performing well as advertised.
  • The oscilloscope functions very well up to 25MHz. While rated for 120MHz, performance may be less effective at higher frequencies.
  • The multimeter provides accurate measurements comparable to higher-end devices.
  • The device is highly recommended for hobbyist use due to its comprehensive features and portability.
  • Consider updating the firmware to access potential new features and improvements.
  • The included accessories are generally of good quality.

9. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the documentation provided at the time of purchase or contact the retailer/manufacturer directly. Keep your proof of purchase for any warranty claims.