GVDA Smart Multimeter User Manual

Models: GD136A / GD136B

1. Introduction

This manual provides instructions for the safe and effective operation of your GVDA Smart Digital Multimeter (Models GD136A and GD136B). This intelligent true RMS digital multimeter features a gear display and an analog bar for multiple display readings.

Key Features:

  • 4000 Counts Display
  • Smart Measurement (Auto Ranging)
  • AC/DC Voltage Measurement
  • Resistance Measurement
  • Continuity Test
  • Frequency Measurement
  • Non-Contact AC Voltage Detection (NCV)
  • Live Wire Detecting
  • True RMS
Diagram showing key features of the GVDA Smart Multimeter, including 4000 counts, smart measurement, AC/DC voltage, resistance, continuity, frequency, NCV, and live wire detection.
Figure 1: Key Features and Included Accessories

2. Safety Information

Please read this manual carefully and pay special attention to all safety warnings before using the multimeter. Failure to follow these instructions may result in electric shock, personal injury, or damage to the meter or equipment.

2.1. General Safety Guidelines

  • Operate the meter according to this manual; otherwise, the protection function provided by the instrument may be compromised.
  • Do not use the meter if it appears abnormal or damaged.
  • Before use, check for cracks or damaged plastic parts on the instrument shell and probes. Replace damaged probes with the same model and electrical specification.
  • Wear personal protective equipment (PPE) such as approved rubber gloves and flame-retardant clothing when working with dangerous live conductors.
  • When measuring, connect the neutral or ground wire first, then the live wire. When disconnecting, disconnect the live wire first, then the neutral or ground wire.
  • Remove the probes from the meter before opening the case or battery cover. Do not use the meter when disassembled or with the battery cover open.
  • The meter conforms to IEC61010-1 CAT III 600V overvoltage safety standard and pollution level 2.

2.2. Warnings

  • Take special care when measuring values that exceed 60V DC, 30V AC RMS, or 42V peak, as these voltages pose a risk of electric shock.
  • Do not measure voltage higher than the rated value between terminals or between terminals and ground.
  • Do not use the meter in environments with explosive gas, steam, or high humidity.
  • When using probes, keep your fingers behind the probe finger guards.
  • Replace the battery immediately when the "BAT" symbol is displayed to prevent inaccurate measurements.

3. Product Overview

3.1. Components

Front view of the GVDA Smart Multimeter GD138B with red and black test leads connected.
Figure 2: GVDA Smart Multimeter (GD138B model shown)
Contents of the product package, including the multimeter, user manual, and test leads.
Figure 3: Package Contents

The package typically includes:

  • GVDA Smart Digital Multimeter
  • Test Leads (Original pen)
  • User Manual
  • Gift Box

3.2. Meter Layout

The multimeter features a clear digital display, function buttons, and input jacks:

  • Display: Digital Display with 4000 counts and an analog bar.
  • Buttons: Power On/Off (Power Symbol), Backlight (Light Bulb Symbol), Data Hold (H), Function (FUNC).
  • Input Jacks: COM (Common), INPUT.
  • NCV Sensor Area: Located at the top of the meter for non-contact voltage detection.

4. Setup

4.1. Battery Installation

The multimeter requires 2 x AAA batteries (not included).

  1. Ensure the meter is powered off and test leads are disconnected.
  2. Locate the battery compartment on the back of the meter.
  3. Unscrew the battery cover and remove it.
  4. Insert two 1.5V AAA batteries, observing the correct polarity (+/-).
  5. Replace the battery cover and secure it with the screw.
Rear view of the multimeter with the battery compartment open, showing two AAA batteries and a flashlight.
Figure 4: Battery Compartment and Flashlight

5. Operating Instructions

5.1. Power On/Off

  • Press and hold the Power Symbol key for approximately 2 seconds to turn the meter on or off.

5.2. Smart Measurement (AUTO)

This is the default mode when the meter is powered on. In this mode, the meter automatically identifies and measures DC voltage, AC voltage, resistance, and continuity.

  1. Power on the meter. It will display "Auto" and enter intelligent measurement mode.
  2. Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
  3. Contact the probes with both ends of the circuit or component to be measured.
  4. Read the measurement from the display.

Note: The minimum measurable voltage in this mode is 0.8V.

Multimeter performing a smart measurement of voltage from a power strip.
Figure 5: Smart Measurement in action

5.3. Manual Mode (Function Selection)

From Smart Measurement mode, press the FUNC key to cycle through manual measurement modes. Press and hold the FUNC key to return to Smart Measurement mode.

5.3.1. DC Voltage Measurement

  1. Press FUNC key to select the DC Voltage (V) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the DC power supply.
  4. Read the results from the display.

5.3.2. AC Voltage Measurement

  1. Press FUNC key to select the AC Voltage (V~) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the AC power supply.
  4. Read the results from the display.

5.3.3. Resistance Measurement

  1. Press FUNC key to select the Resistance (Ω) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the measured resistance.
  4. Read the results from the display.

5.3.4. Continuity Test

  1. Press FUNC key to select the Continuity (Ω))) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the circuit or component.
  4. If the resistance value is less than 50Ω, the buzzer will sound and the alarm indicator will be on.

5.3.5. Capacitance Measurement

  1. Press FUNC key to select the Capacitance (nF/µF/mF) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the measured capacitance.
  4. Read the results from the display.

5.3.6. Frequency Measurement

  1. Press FUNC key to select the Frequency (Hz) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the probes with both ends of the measured power supply.
  4. Read the results from the display.

5.3.7. Diode Test

  1. Press FUNC key to select the Diode (Diode Symbol) gear.
  2. Insert the red probe into "INPUT" jack and the black probe into "COM" jack.
  3. Contact the red probe to the anode of the diode and the black probe to the cathode.
  4. Read the results from the display. If the probe polarity is opposite to the diode polarity, the display will show "OL".

5.4. Non-Contact AC Voltage Detection (NCV)

  1. Power on the meter.
  2. Press FUNC key to select the NCV gear.
  3. Gradually bring the NCV sensor area (top of the meter) close to the conductor.
  4. When a weak electric field signal is detected, the buzzer will sound slowly and the green light will turn on.
  5. When a strong electric field signal is detected, the buzzer will sound quickly, the red light will turn on, and "H" will be displayed.
Multimeter performing NCV measurement near a power strip, indicating non-contact voltage detection.
Figure 6: NCV Measurement

5.5. Live Wire Detecting

  1. Power on the meter.
  2. Press FUNC key to select the Live gear.
  3. Insert the red probe into the "INPUT" jack and remove the black probe.
  4. Use the red probe to contact the conductor.
  5. When a weak electric field signal is detected, the buzzer will sound slowly and the green light will turn on.
  6. When a strong electric field signal is detected, the buzzer will sound quickly, the red light will turn on, and "H" will be displayed.
Multimeter performing live wire detection with a single red probe inserted into a wall socket.
Figure 7: Live Wire Detecting

5.6. Data Hold

  • Press the H key to turn on or off data holding. (Not available in NCV/Live mode).

5.7. Backlight

  • Press the Light Bulb Symbol key to turn on or off the backlight.

5.8. Auto Power Off

  • Auto power off is enabled by default. If there is no key operation for about 15 minutes, the meter will automatically shut down to save battery energy.
  • To disable auto power off, press and hold the FUNC key while powering on the meter. The auto power off symbol will not be displayed.

6. Specifications

6.1. General Technical Specifications

  • Environment Condition: CAT III 600V, Pollution level 2, Altitude < 2000m
  • Operating Temperature: 0 - 40 ℃
  • Storage Temperature: -10 - 60 ℃
  • Display: Digital Display, 4000 counts
  • Sampling Rate: Approximately 3 times/second
  • Over Range Indication: "OL"
  • Low Battery Indication: Displayed on screen
  • Power Requirement: 2 x 1.5V AAA batteries
  • Dimensions: 126 * 19 * 65mm
  • Certifications: CE, FCC

6.2. Accuracy Specifications

Accuracy is guaranteed for one year after calibration, at 18°C to 28°C and relative humidity no more than 80%.

Measurement Range Accuracy
DC Voltage (GD136A) 4V, 40V, 400V, 600V ±(0.5% + 3)
AC Voltage 4V, 40V, 400V, 600V ±(0.8% + 3)
Resistance (GD136A) 4000 Ω, 40kΩ, 400kΩ, 4MΩ ±(1.0% + 5)
Resistance (GD136A) 40MΩ ±(1.5% + 10)
Capacitance (GD136B) 4nF, 40nF, 400nF, 4µF, 40µF, 400µF, 4mF ±(4.0% + 5)
Frequency 40Hz, 400Hz, 4kHz, 40kHz, 400kHz, 4MHz Not specified in provided data
Impedance (Voltage) Approx 10MΩ
Overload Protection (Resistance/Diode/Continuity) 250V

7. Maintenance

7.1. Cleaning

  1. Turn off the meter and remove the test leads.
  2. Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
  3. Clean the contacts in each input socket with a clean swab soaked in alcohol.

7.2. Battery Replacement

Refer to Section 4.1 for detailed battery replacement instructions. Always replace with new 1.5V AAA batteries, observing correct polarity.

7.3. Storage

  • If the meter will not be used for a long period, remove the batteries to prevent leakage and damage to the product.
  • Keep the inside of the meter clean and dry to prevent electric shock or damage.

8. Troubleshooting

  • Meter not powering on: Check battery installation and ensure batteries are not depleted. Replace if necessary.
  • Inaccurate readings or intermittent operation: Check battery level (replace if low), ensure test leads are securely connected and not damaged. Verify the correct measurement mode is selected.
  • Continuity test beeps incorrectly: Ensure test leads are not damaged or shorted internally. Check for proper contact with the circuit. If the issue persists, the unit may be faulty.
  • Physical damage during battery replacement: Exercise caution when opening the battery compartment. Do not overtighten screws.

If you encounter persistent issues not covered here, please contact customer support.

9. User Tips

  • Utilize Smart Measurement: The auto-ranging and smart measurement function can simplify operation and help prevent common errors, especially for beginners.
  • Read the Display Carefully: Pay attention to units (e.g., Ω vs. kΩ) and decimal places, especially when measuring resistance, to correctly interpret readings.
  • Handle with Care: While compact, the multimeter may benefit from careful handling to prevent damage, as some users have noted the absence of a protective rubber cover.
  • Display Readability: The large LCD screen and analog bar are designed for easy reading. Ensure adequate lighting for optimal visibility.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the documentation provided at the time of purchase or contact your retailer/manufacturer directly. Keep your purchase receipt as proof of purchase.

A digital version of this user manual is available for download at: GVDA Smart Multimeter User Manual (PDF)

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.