Contents

ASAIR AS200 MEMS MFC Gas Mass Flow Controller

Instruction Manual

1. Introduction

The ASAIR AS200 series Mass Flow Controller (MFC) is a high-precision instrument designed for accurate and reliable measurement and control of gas mass flow. Developed using advanced MEMS technology by Aosong Electronic, the AS200 series offers good stability, high accuracy, low leakage rate, and high pressure resistance. It supports both digital and analog signals and is suitable for a wide range of applications.

Video: Overview of the ASAIR AS200 Mass Flow Controller, highlighting its features, specifications, and diverse applications.

1.1. Safety Statement

This manual provides detailed information for the correct and safe use of the AS200 series products. Please read it carefully before use. The manufacturer is not responsible for losses or injuries resulting from failure to follow the instructions in this manual. This manual is essential for installation, maintenance, and troubleshooting.

1.2. Safety Precautions

  • Do not replace or modify: Do not replace any parts or make unauthorized modifications to the product. Safety labels must remain intact.
  • Contact professional staff: Any technical support or part replacement must be performed by authorized technicians.
  • Hazardous gases: When using hazardous gases, take precautions and purge the product completely if conditions permit. Ensure moist hazardous gases do not react chemically with product materials.
  • Blowing and washing products: Before and after installation, purge the entire system and remaining gas residues with dry gas. Perform this under a ventilation hood with appropriate personal protective equipment (gloves).
  • Explosive environments: This product is not suitable for use in flammable or explosive gases unless it has valid certification.
  • Connectors: Use appropriate connectors and follow relevant tightening procedures as per the product manual.
  • Leakage check: Carefully check all component joints for leaks after installation.
  • Safe inlet pressure: Do not exceed the maximum working pressure specified in the product data sheet.
  • System cleanliness: Avoid using gas containing pollutants (dust, dirt, fiber, glass fragments, iron filings).
  • Preheat before use: Preheat the product before use, especially for dangerous gases. Use an absolutely closed valve to prevent wrong flow during preheating.

2. Product Features

The AS200 series Mass Flow Controller incorporates advanced MEMS technology to deliver superior performance and reliability. Key features include:

  • High stability and accuracy.
  • Low leakage rate (1x10-10 Pa·m³/sec He).
  • High pressure resistance.
  • Supports digital (RS485 or EtherCAT) and analog (1-5V or 4-20mA) signals.
  • PC software available for control and debugging.
  • Multiple measure ranges for various gases (Air, O₂, N₂, Ar, CO₂).
  • Automatic fault alarm.
  • Multi-gas and multi-range capabilities.
  • Robust construction with 316L stainless steel or aluminum alloy.

3. Specifications

The following table details the technical specifications for the ASAIR AS200 Mass Flow Controller:

AS200 Technical Indicators
ParameterSpecification
ModelAS200
Valve TypeNormally Closed
Flow specifications (Air, O₂, N₂)1~100, 2~200, 3~300 SCCM (0.03~3, 0.05~5, 0.1~10, 0.2~20, 0.3~30, 0.5~50)SLM
Flow specifications (Ar, CO₂)1~100, 2~200, 3~300 SCCM (0.03~3, 0.05~5, 0.1~10, 0.2~20, 0.3~30)SLM
Accuracy±1.0% S.P. (≥ 35% F.S.) / ±0.35% F.S. (< 35% F.S.)
Linearity±0.5% F.S.
Repeatability±0.2% F.S.
Response time≤2 sec
Working pressure difference range0.05~0.35MPa (0 < flow specification ≤10SLM); 0.1~0.35MPa (10 < flow specification ≤30SLM); 0.2~0.45MPa (30 < flow specification ≤50SLM)
Temperature CoefficientZero point: ≤±0.05% FS/°C; When adjusting: ≤±0.1% FS/°C Flow specification ≤30SLM; When adjusting: ≤±0.2% FS/°C Flow specification > 30SLM
Compressive strength3MPa
Leak rate1×10-10 Pa·m³/sec He
Sealing materialFluororubber
Working temperature5~45°C
Input signalDigital: RS485 or EtherCAT; Analog: (1~5) VDC or (4~20) mA
Output signalDigital: RS485 or EtherCAT; Analog: (1~5) VDC or (4~20) mA
Zero DriftWithout zero point correction <0.6%FS/year
Power supply+15~+28VDC
Standard connector typeDouble ferrule φ 6; Double ferrule 1/4"; Double ferrule 3/8"; VCR1/4" male connector; Double ferrule φ 3; Double ferrule 1/8"; VCR 1/4"; φ 6 (inner) x 1 hose connector; φ 5 (inner) x 1.5 hose connector; φ 4 (inner) x 1 hose connector; double ferrule φ 10mm; A-type sealing connector
Power connector typeD-type 9-pin male connector; EtherCAT connector
MaterialMetal, Stainless steel
TheoryFlow controller
UseFlow controller
Measuring mediumAir, N2, O2

4. Installation

4.1. Overview

Proper installation is crucial for the accurate and safe operation of the AS200 MFC. Ensure the air is dry and clean, and avoid external impact or vibration during installation. It is recommended to install a filter in the upstream pipeline to prevent contamination. The product should be installed horizontally for best accuracy, though vertical installation is possible with zero point calibration.

4.2. Unpacking

AS200 series products are packaged in clean vacuum bags. Remove the outer plastic bag before entering a clean room. To minimize contamination, open the inner clean bag in a clean environment.

4.3. Mechanical Installation

When installing the MFC, ensure the direction of the arrow on the product channel matches the gas flow direction. Do not remove the protective caps on the connectors until actual installation.

ASAIR AS200 Mass Flow Controller side view
Figure 1: Side view of the AS200 Mass Flow Controller, showing the flow direction indicator.
ASAIR AS200 Mass Flow Controller with EtherCAT protocol
Figure 2: AS200-EtherCAT protocol product appearance, showing the top view with RJ45 ports.
ASAIR AS200 Mass Flow Controller 100/200L model
Figure 3: AS200-100/200L product appearance, showing a different configuration.

4.3.1. Connector Dimensions

The following diagrams illustrate the dimensions for various connector types:

Double ferrule 1/4 inch product dimensions diagram
Figure 4: Double ferrule 1/4" product dimensions (unit: mm). This diagram shows the top and side views with measurements for the double ferrule connector.
VCR 1/4 inch male connector product dimensions diagram
Figure 5: VCR 1/4" male connector product dimensions (unit: mm). This diagram shows the top and side views with measurements for the VCR male connector.
AS200-100/200L dimensions diagram
Figure 6: AS200-100/200L Dimensions (Tolerance for unmarked items is ±1mm). This diagram provides detailed measurements for the 100/200L model, including height with and without cable plug.

4.3.2. VCR Connector Installation

For VCR connectors, both non-positioned and position-based VCR washers are used. For non-positioned washers, place them into the female nut. For position-based washers, fix them at the joint of the protrusions. First, hand-tighten the nut and joint, then use two wrenches: one to clamp the joint and the other to rotate the nut to the appropriate position. This two-wrench method is also crucial for disassembly to prevent loosening and affecting the seal.

Schematic diagram of VCR joint installation without positioning VCR gasket
Figure 7: Schematic diagram of VCR joint installation without positioning VCR gasket.
Schematic diagram of VCR joint installation with VCR gasket positioned
Figure 8: Schematic diagram of VCR joint installation with VCR gasket positioned.

4.3.3. Double Ferrule Fitting Installation

After installing the front ferrule, rear ferrule, and nut, hand-tighten them. Then, use two wrenches to tighten it further. One wrench should hold the joint, and the other should rotate the nut to the appropriate position to ensure no leakage. This method is also essential when disassembling the pipe.

Schematic diagram of double ferrule joint installation
Figure 9: Schematic diagram of double ferrule joint installation.

4.4. Electrical Installation

The AS200 series products use a switching power supply and support a wide voltage input of +15~28VDC. Electrical connectors include D-type male connectors and EtherCAT connectors. The 9-pin D-type connector is SEMI standard and supports 4-20mA input and output. The AS200 also supports RS485 serial communication and EtherCAT protocol.

4.4.1. Connection Diagrams

The following diagrams illustrate the connection interfaces:

9-pin D-type connector connection diagram
Figure 10: 9-pin D-type connector connection diagram. This diagram shows the pinout for 1-5V Set, Signal GND, 15 to 28V, Power GND, and 4-20mA Flow set.
RS485 connection diagram
Figure 11: RS485 connection diagram. This diagram shows the pinout for Signal GND, 485-, and 485+.
EtherCAT product RJ45 connector connection diagram
Figure 12: EtherCAT product RJ45 connector connection diagram. This diagram shows the pinout for NC, 15-28V, and GND.
EtherCAT product power interface pin diagram
Figure 13: EtherCAT product power interface pin diagram.

5. Setup

5.1. Calibration

The AS200 series can be calibrated according to user requirements. If no specific working conditions are provided, calibration will be performed under standard conditions (25°C, 101 kPa air pressure). The standard installation position for MFC is horizontal. For highest accuracy, users should specify the installation method when purchasing. After manufacturing, each MFC undergoes a 24-hour stabilization period and precise calibration, with accuracy, dynamic response, and pressure stability checked twice before shipment.

5.2. Zero Adjustment

Customers can correct the zero drift of the flow meter by zeroing. Before zeroing, ensure no gas is passing through the product. Zeroing can be performed via software commands or a dedicated button. If using the button, press it for more than 5 seconds until the green LED indicator flashes, then release. The green LED will stop flashing once zeroing is complete.

6. Operation

6.1. Valve Control

AS200 series products support direct closing or opening of the electromagnetic regulating valve via digital communication or analog voltage. For more details, refer to the official website: www.aosong.com.

6.2. Valve Type

The solenoid control valve of the AS200 series MFC is a normally closed valve type. When the MFC is not powered on, the solenoid valve remains closed.

6.3. Total Flow Measurement

The AS200 series products can record and output the cumulative value of gas flow. The unit for cumulative value is standard milliliters (SCC). For example, a reading of 3000 SCC means 3000 standard milliliters of gas have passed through the MFC.

7. Maintenance

7.1. Overview

The MFC does not require routine maintenance, only occasional cleaning and recalibration. If the controller is used with ultra-clean and non-corrosive gases, it should be cleaned and recalibrated every 3 to 4 years. For low-cleanliness or corrosive gases, cleaning and recalibration should occur every 1 to 2 years.

7.2. Media Usage Requirements

The gas used must be purified, free from dust, liquid, and oil. If necessary, install a filter in the gas line. If the MFC outlet is connected to a liquid source, a one-way valve must be installed at the outlet to prevent liquid backflow and product damage.

7.3. Valve Port Sealing

The solenoid valve of the MFC is a regulating valve, not a stop valve. If a stop valve is required, it should be purchased separately. For dangerous gases, a stop valve should be added to both the inlet and outlet of the MFC to ensure safety. The internal leakage rate of the MFC valve port is within 1%FS, which is considered normal.

8. Troubleshooting

Before troubleshooting, perform a preliminary inspection:

  • Check if the gas source and gas line to the MFC are open.
  • Ensure power and control signals are properly delivered to the electrical connectors.
  • Verify the communication line is connected correctly.
Troubleshooting Guide
Fault PhenomenonPossible Cause of FailureApproach
When there is no ventilation, the displayed value is not zeroGas is still flowingCheck whether the stop valve is closed
Zero driftUse the Zero Adjustment Function
Other faultsContact "Aosong Electronics"
Unable to control trafficIncorrect gas connectionCheck whether the MFC is correctly connected to the gas line
The pressure difference is not within the required rangeCheck the pressure
MFC works in other modesCheck and change control mode using software
Power IssuesCheck the power supply and whether the power connection is correct.
Incorrect signal settingCheck the set voltage or current signal
MFC is contaminatedContact "Aosong Electronics"
Unable to communicateSensor damageContact "Aosong Electronics"
Circuit board problemContact "Aosong Electronics"
Valve body damageContact "Aosong Electronics"
General communication issuesPower IssuesCheck the power supply and whether the power connection is correct.
Communication connection problemCheck whether the communication line is connected correctly
Address ConflictCheck whether the address of MFC conflicts with other MFCs
The baud rate setting is incorrectCheck the baud rate setting of the MFC

9. User Tips

  • For detailed technical specifications or specific application guidance, always refer to the official product data sheet.
  • If you encounter issues not covered in this manual or require immediate assistance, contact ASAIR customer support.

10. Warranty and Support

10.1. Quality Assurance

Guangzhou Aosong Electronics Co., Ltd. provides quality assurance for its products to direct purchasers. The warranty period is calculated from the date of shipment, adhering to the technical specifications outlined in the Aosong Electronics product manual. If a product is proven defective during the warranty period, the company will provide free repair or replacement services.

Warranty Period
Product CategoryWarranty Period
AS200 Series Gas Mass Flow Controller12 months

The company is responsible only for defective products used according to technical requirements. No guarantees are made for applications in unrecommended scenarios or for reliability when integrated with unsupported products or circuits.

10.2. Further Support

This manual is subject to change without prior notice. The final right of interpretation of this product belongs to Guangzhou Aosong Electronics Co., Ltd.

For a complete and detailed manual, please refer to the official user manual PDF: ASAIR AS200 User Manual (PDF)