DSO3D12 Digital Oscilloscope User Manual

Model: DSO3D12 | Brand: meterk

1. Introduction

The meterk DSO3D12 is a versatile 3-in-1 handheld device integrating a Digital Oscilloscope, a Multimeter, and a Signal Generator. Designed for hobbyists, students, and professionals, it offers dual-channel oscilloscope functionality with a 120MHz bandwidth and 250MSa/s sampling rate, a high-precision multimeter for various electrical measurements, and a signal generator for testing circuits. Its compact design and comprehensive features make it an ideal tool for R&D, education, car repair, power, audio, and radio amateur applications.

DSO3D12 Smart Oscilloscope (3 in 1) with screen showing a sine wave, buttons, and input jacks.

Figure 1.1: DSO3D12 Smart Oscilloscope (3 in 1) - Dual channel + DMM + Generator, 120MHz BW/Fully Auto, XY/ZOOM/Ref/Cursor.

2. Safety Precautions

Please read all safety information carefully before using the device to prevent injury or damage.

  • High Voltage: The x10 range of the probe measures up to 220V. For voltages exceeding this, an x100 probe is required.
  • Probe Attenuation: Before measuring voltages greater than 40V, always switch the probe to the X10 position.
  • Charging: Use a 5V charger (computer USB or mobile phone charger). Do not use the device to measure high-voltage or non-isolated circuits while it is charging (powered by its built-in battery).
  • Multimeter Use: Do not use the multimeter while charging.
  • Current Measurement: Never insert the red probe into the current hole to measure voltage; this will blow the fuse or damage the device. After measuring current, always return the red probe to the voltage hole.
  • Current Duration: If multimeter current measurement exceeds 6A, the duration should not exceed 10 seconds to prevent line damage from heating.
  • Simultaneous Use: When measuring high voltages, do not use the oscilloscope and multimeter functions simultaneously.
  • Probe Connection: Always switch the probe to the 10X position before connecting to a high voltage signal to prevent damage to the oscilloscope's internal circuit.

3. Package Contents

Carefully unpack the box and ensure all items are present:

  • DSO3D12 3-in-1 Device
  • Oscilloscope Probes (x2)
  • Multimeter Test Leads (Red and Black)
  • USB Type-C Charging Cable
  • Small Components/Accessories (e.g., probe adjustment tool, color rings)
  • User Manual (this document)
DSO3D12 device with all included accessories: oscilloscope, two probes, multimeter leads, USB-C cable, and small components.

Figure 3.1: DSO3D12 and its accessories.

4. Device Overview

4.1 Front Panel and Display

The front panel features a 3.2-inch display and control buttons. The display provides real-time waveform data, measurement readings, and menu options.

DSO3D12 Smart Oscilloscope (3 in 1) with screen showing a sine wave, buttons, and input jacks.

Figure 4.1: Front view of the DSO3D12.

Display Elements:

  1. Status Indicator: Green (running), Flashing Cyan (waiting for trigger), Red (stopped).
  2. Channel Priority: Yellow 1 (CH1 high priority), CH1 waveform movable.
  3. Signal Generator Output: Yellow (ON), Gray (OFF).
  4. Time Base Window Position: Shows current time base window in memory.
  5. Trigger Mode: Auto (automatic), Normal (normal).
  6. Trigger Edge: Rising or falling edge trigger.
  7. Battery Level: Indication of remaining battery charge.
  8. Trigger Level Position: Visual indicator of the trigger level.
  9. Probe Attenuation (X1/X10): Matches probe setting to oscilloscope.
  10. DC/AC Coupling: Input coupling type.
  11. Vertical Sensitivity: Voltage per grid division.
  12. Timebase: Time per grid division.
  13. Zero Arrow: Baseline position (0V).

4.2 Top Panel

The top panel includes BNC connectors for oscilloscope probes, switches, and the USB-C port.

Top view of the DSO3D12 showing BNC connectors, switches, USB-C port, and power button.

Figure 4.2: Top view of the DSO3D12.

4.3 Rear Panel

The rear panel typically houses the battery compartment.

Rear view of the DSO3D12 showing the battery compartment.

Figure 4.3: Rear view of the DSO3D12.

5. Setup and Initial Operation

5.1 Charging the Device

Before first use, fully charge the device. Connect the USB Type-C cable to the device and a 5V USB power adapter (e.g., computer USB port or mobile phone charger). The charging current is approximately 700mA. The LED indicator will be red during charging and turn green when fully charged.

5.2 Power On/Off

Press and hold the power button (located on the top panel) to turn the device on or off.

5.3 Reset to Initial State

To reset the device to its factory default settings: Navigate to Menu > Set, then long-press the Auto button.

6. Operating Modes

6.1 Oscilloscope Operation

The oscilloscope function allows for waveform visualization and analysis.

Play video
Watch video online

Video 6.1: Demonstration of DSO3D12 oscilloscope, multimeter, and signal generator functions.

6.1.1 Automatic Setup (Auto)

After connecting the probe to the signal, press the Auto button. The oscilloscope will automatically detect the signal and adjust the range for optimal display. If no waveform appears, check probe connection and signal presence.

DSO3D12 screen showing a waveform in fully automatic mode.

Figure 6.1: Fully automatic mode for quick waveform adjustment.

6.1.2 Manual Adjustment (Sensitivity & Timebase)

  • Vertical Sensitivity: Use the mV/V buttons to adjust the vertical scale (voltage per grid). Press mV to decrease sensitivity (for smaller voltages) and V to increase sensitivity (for larger voltages).
  • Timebase: Use the s/ns buttons to adjust the horizontal scale (time per grid). Press ns for high-frequency signals and s for low-frequency signals.

6.1.3 Trigger System

The trigger system stabilizes waveforms and captures specific signal segments.

DSO3D12 screen showing trigger settings for Auto, Normal, Single trigger types, and Rise/Fall edges.

Figure 6.2: Perfect Trigger settings.

  • Trigger Level: Adjust the trigger level manually via Menu > Trigger > Trigger Level > Manual. Use the Shift key and up/down arrows to move the trigger level.
  • Trigger Source: Select CH1 or CH2 as the trigger source.
  • Trigger Mode: Choose from Auto (most common, forces trigger if none detected), Normal (waits for trigger, suitable for intermittent signals), or Single (captures one waveform frame then stops).
  • Single Trigger: Connect signal, set sensitivity/timebase, set trigger level to manual, press Single. The oscilloscope waits for the signal to cross the trigger level, captures it, and stops. Press Stop to cancel.

6.1.4 Dual Channel Operation

Press Shift + CH1/CH2 to turn CH2 on/off. CH1 is always on. To achieve a higher sampling rate and reduce power consumption, turn off CH2 if not needed. In dual-channel mode, press CH1/CH2 to switch priority; the channel with priority can be moved with up/down buttons.

6.1.5 Cursor Measurements

Press Shift + s to toggle horizontal cursors. Press Shift + ns to toggle vertical cursors. Use left/right arrows to move horizontal cursors and up/down arrows to move vertical cursors. The display shows voltage (Y1, Y2) and time (T1, T2) differences (dY, dT).

6.1.6 Display Options

  • Persistence: Keeps historical waveforms on screen (None, 1s, Infinite).
  • Roll Mode: Displays waveform while sampling (active when time base ≥ 200ms).
  • XY Mode: Displays CH1 on the X-axis and CH2 on the Y-axis, useful for Lissajous figures.
  • Zoom: Use Shift + Zoom to enter/exit zoom mode.
  • FFT: Fast Fourier Transform can be enabled via Menu > Settings to display the frequency spectrum of CH1. Options include logarithmic, linear, and music spectrum.

6.1.7 Screenshots and Image Viewing

Press Save to capture the current screen. Press Shift + Save to enter Image View mode, where you can browse or delete screenshots. Use x1x10 to delete, OK to zoom, and Save to return to oscilloscope view.

6.2 Multimeter Operation

The multimeter function allows for precise electrical measurements.

DSO3D12 screen showing a digital multimeter reading of 12.01V.

Figure 6.3: High-precision multimeter function.

  • Activate: Press the DMM button to turn the multimeter on/off. Press Shift + DMM for full-screen multimeter display.
  • Function Selection: Press the Select button to cycle through measurement functions (Voltage, Current, Resistance, Diode, Capacitance, Continuity).
  • Probe Connection: For current measurement, insert the red test lead into the current hole (10A or 600mA) and connect in series with the circuit. For voltage, resistance, diode, capacitance, and continuity, use the voltage hole.
  • Continuity Test: The device produces a sound (buzzer) when resistance is below 50Ω.

6.3 Signal Generator Operation

The built-in signal generator can output various waveforms for testing.

  • Activate: Press the Gen button to open the signal generator interface.
  • Output Control: Press Stop to turn the generator output on/off.
  • Waveform Selection: Use the s/ns buttons to switch between sine wave, square wave, and triangle wave.
  • Frequency Adjustment: Press Single to reset frequency to 1.00KHz. Adjust frequency as needed. The maximum frequency for sine wave is 5MHz, and for other waves is 1MHz.
  • Amplitude: The signal amplitude is 2.5V.
  • Duty Cycle: For square waves, the duty cycle can be adjusted.
  • Exit: Press OK or Gen to confirm settings and exit.

6.4 Probe Calibration

Probes should be calibrated for accurate measurements, especially new ones or if using different probes.

  1. Switch the probe to X10.
  2. Connect the probe to a 1KHz square wave (2V-3V amplitude), using the built-in generator.
  3. Adjust oscilloscope horizontal and vertical scales to display a stable waveform.
  4. Use a screwdriver to rotate the adjustable capacitor on the probe until the waveform is a perfect square wave (not rounded or spiked).

7. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
  • Battery Care: Recharge the battery when low. Avoid fully discharging the battery for extended periods.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
  • Self-Calibration: If zero offset occurs due to temperature changes or aging, remove probes and perform self-calibration via Menu > Settings > Calibration.
  • Auto Shutdown: The device can be set to automatically shut down after 15 minutes of inactivity (Menu > Settings > Auto Shutdown).

8. Troubleshooting

  • No Waveform Display: Check probe connection, ensure signal is present, and try the Auto function.
  • Slow Waveform Refresh (Timebase > 50ms): This is normal due to the longer sampling time required for larger time base settings.
  • Measuring DC Power Voltage: Set coupling to DC, probe to 10X (match menu setting), connect probe to power output, press Auto. Observe the 'Mean' value.
  • Measuring 220V Mains: Disconnect USB cable (for isolation). Set probe to 10X, vertical sensitivity to 100V, time base to 10ms. Connect probe to neutral and live wires (exercise extreme caution).
  • Measuring Audio Signal: Set coupling to AC, probe to 10X. Connect to audio output. Adjust time base (1-10ms) and voltage sensitivity (approx. 500mV).
  • Measuring Main Board/Bus Signals: Connect probe, set coupling to DC, probe to 10X (match menu setting), press Auto. Manually adjust time base (s/ns) if needed.
  • Measuring Power Supply Ripple: Set coupling to AC, probe to 1X (match menu setting). Adjust time base to 200us. Start with 10mV vertical sensitivity and adjust with V button if waveform exceeds screen.
  • Lissajous Figures: Yes, the device supports displaying Lissajous figures using XY mode.
  • Waveform Image Download to PC: The device saves screenshots internally. Direct download to PC is not explicitly supported via the manual, but images can be viewed on the device.
  • LCR Measurement: No, this device is a Digital Multimeter (DMM) and does not have LCR meter functionality.

9. Specifications

9.1 General Parameters

ParameterSpecification
ModelDSO3D12
Channels2
Screen Size3.2 inch
Shell Size145*86*33mm
Weight266g
ChargerTYPE-C / 5V
Charge Current700mA

9.2 Oscilloscope Parameters

ParameterSpecification
Bandwidth120MHz
Real Time Sampling Rate250MSa/s
Equivalent Sample Rate500M
Rise Time<3ns
Storage Depth128Kb
Impedance1M Ω
Time Base Range5ns - 10s
Vertical Sensitivity10mV/div - 10V/div
Max Voltage (x1)±40V
Max Voltage (x10)±400V
Trigger ModeAuto/Normal/Single
Trigger TypeRise/Fall
Display ModeYT / Roll
PersistenceNone/1s/∞

9.3 Multimeter Parameters

FunctionRangePrecision
DC Voltage600mV/6.00V/60.0V/600V/750V±(0.5%+3)
AC Voltage600mV/6.00V/60.0V/600V±(1%+3)
DC Current600mA/10A±(2%+5)
AC Current600mA/10A±(3%+5)
Resistance600.0Ω to 60.00MΩ±(1.5%+3) to ±(3%+3)
Capacitance60.00nF to 600.0μF±(10%+5) to ±(15%+5)
Diode0.0V - 3.3VDisplay "OL" above 3.3V
ContinuitySound at 50Ω and below-

9.4 Signal Generator Parameters

ParameterSpecification
Waveform Output Amplitude2.5V ± 0.05V
Waveform Output Frequency5MHz (Sine), 1MHz (Other)

10. User Tips

  • Hobbyist Use: This device is excellent for hobbyist and non-professional use, offering great value for its price and capabilities.
  • Portability: Its compact size and battery power make it highly convenient and portable for various tasks.
  • Probe Quality: Users appreciate the quality of the included probes with BNC connectors.
  • Firmware Update: Consider updating the firmware for potential new features and improvements.
  • Battery Life: The autonomy is generally good, and the device is reported to be stable.
  • Automotive Use: Suitable for testing sensor signals in automotive applications.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the documentation provided at the time of purchase or contact your retailer. Keep your proof of purchase for warranty claims.