1. Introduction
The ASAIR AMC2100 gas mass flow controller is a device designed for precise measurement and control of gas mass flow. It incorporates a self-developed MEMS mass flow chip, offering high precision, stability, resistance to temperature variations, good linearity, and fast response time. Each AMC2100 unit undergoes rigorous testing and calibration for its measure range, accuracy, repeatability, and response time before delivery.
This mass flow controller is suitable for monitoring and controlling dry, clean, non-corrosive gases such as air, nitrogen (N2), oxygen (O2), carbon dioxide (CO2), and argon (Ar). It is widely applied in various fields including universities, scientific research, fire protection, environmental monitoring, tobacco, smart agriculture, food, and medicine.

2. Setup and Installation
2.1 Product Structure
The AMC2100 unit comprises a sensor air duct, which includes a control valve, flow measurement components, and gas inlet/outlet interfaces. The housing features an arrow indicating the required gas flow direction. The device also includes a digital display and a DB9 communication interface.

2.2 Interface Definition and Wiring
The communication interface is a DB9 male connector. The wiring uses a DB9 female single-ended cable. The transparent wire in the cable is the shielded wire.
DB9 Male Connector Pin Definition
| Pin | Function | Description |
|---|---|---|
| 1 | Analog voltage input | 0~5V, effective when controller works in analog signal input mode |
| 2 | Analog voltage output | 0~5V |
| 3 | RS485B | RS485 B |
| 4 | RS485A | RS485 A |
| 5 | +24V | Power positive |
| 6 | GND | Power common terminal 0V |
| 7, 8, 9 | NC | Test pin, user does not need to connect |
DB9 Female Cable Pin Colors
- Orange: 0~5V flow setting
- Brown: 0~5V flow test
- Green: RS485 B
- Yellow: RS485 A
- Red: Power +24V
- Grey, Purple, Blue, Black: NC (Not Connected)
Ensure proper connection of the DB9 female cable to the AMC2100's DB9 male connector according to the pin definitions and color codes provided. Incorrect wiring can damage the device or lead to improper operation.
3. Operating Instructions
The AMC2100 features a digital display and control buttons (S, Up, Down) for user interaction.
3.1 Normal Working Interface Display
In normal operation, the controller displays the instantaneous flow rate, which updates in real-time. Pressing the Up or Down key will adjust the set flow rate value.
- To set flow rate: Short press the 'S' (setting) key to enter the flow setting interface. The display will flash at 1Hz. Use the Up or Down keys to increase or decrease the flow rate value. The setting updates in real-time. Short press 'S' again to exit and return to the normal working interface.
3.2 Setting Modbus Communication Address
- From the normal working interface, press and hold the 'S' key for 2 seconds to enter the menu page (CH01).
- Press 'S' again to enter the sub-page.
- Use the Up or Down keys to set the communication address.
- Press 'S' to return to the menu page (CH01).
3.3 Setting Communication Baud Rate
- From the menu page (CH01), press the Down key once to navigate to the menu page (CH02).
- Press 'S' once to enter the sub-page.
- Use the Up or Down keys to set the desired baud rate.
- Press 'S' to return to the menu page (CH02).
3.4 Checking or Clearing Cumulative Flow
- From the menu page (CH02), press the Down key once to navigate to the menu page (CH03).
- Press 'S' once to enter the sub-page, which displays the cumulative flow.
- Press the Up or Down keys to clear the cumulative flow.
- Press 'S' to return to the menu page (CH03).
3.5 Setting the Measurement Gas
The factory default measurement gas is Air. To select other gases:
- From the menu page (CH03), press the Down key once to navigate to the menu page (CH04).
- Press 'S' again to enter the sub-page.
- Use the Up or Down keys to select the measurement gas from the available options: Air, Nitrogen (N2), Oxygen (O2), Carbon Dioxide (CO2), Argon (Ar).
- Press 'S' to return to the menu page (CH04).
3.6 Setting Display Brightness Level
- From the menu page (CH04), press the Down key once to navigate to the menu page (CH05).
- Press 'S' once to enter the sub-page.
- Use the Up or Down keys to set the display brightness level (1-8 levels).
- Press 'S' to return to the menu page (CH05).
3.7 Exiting Parameter Setting
- From the menu page (CH05), press the Down key once to navigate to the menu page (End).
- Press 'S' once to exit to the normal working interface.
4. Maintenance
To ensure optimal performance and longevity of your AMC2100 mass flow controller, adhere to the following maintenance guidelines:
- Gas Quality: The AMC2100 is designed for dry, clean, non-corrosive gases. Ensure that the gas supply meets these requirements to prevent damage to internal components.
- Cleaning: Keep the exterior of the device clean and free from dust and debris. Use a soft, dry cloth for cleaning. Do not use abrasive cleaners or solvents.
- Environmental Conditions: Operate the device within its specified environmental conditions (temperature, humidity) to prevent malfunction.
- Regular Checks: Periodically check all connections, especially the DB9 communication cable and gas inlet/outlet, for secure fit and absence of leaks.
- Calibration: While the device is factory calibrated, professional re-calibration may be required over time depending on application and usage frequency to maintain accuracy.
5. Troubleshooting
This section provides basic troubleshooting steps for common issues. For complex problems or issues related to Modbus communication protocol, please refer to the full PDF manual or contact technical support.
5.1 Communication Issues
The AMC2100 uses standard Modbus RTU communication via RS485. If communication fails:
- Check Wiring: Verify that the DB9 cable is correctly wired according to the pin definitions (Table 1) and that all connections are secure.
- Baud Rate and Address: Ensure the Modbus communication address and baud rate settings on the AMC2100 (refer to sections 3.2 and 3.3) match those of your host system.
- Abnormal Response Codes: If the host receives an abnormal response, check the function code being sent. The AMC2100 only supports function codes 03 (Read keeping register) and 06 (Preset single register). An incorrect function code (e.g., 07) will result in an error.
5.2 Inaccurate Flow Readings
- Gas Selection: Verify that the correct measurement gas is selected on the device (refer to section 3.5). Incorrect gas selection will lead to inaccurate readings.
- Gas Quality: Ensure the gas being measured is dry, clean, and non-corrosive as specified. Contaminants can affect sensor accuracy.
- Work Pressure: Confirm that the gas work pressure is within the specified range of 0.15~0.35MPa.
5.3 No Display or Incorrect Display
- Power Supply: Check that the device is receiving a stable 24V DC power supply.
- Brightness: Adjust the display brightness level (refer to section 3.6) to ensure it's not set too low to be visible.
6. Specifications
6.1 Technical Parameters
| Parameter | Description |
|---|---|
| Model Variants | AMC2000R00, AMC2000R01, AMC2000R02, AMC2000R03, AMC2000R04 |
| Range | 100mL/min, 300mL/min, 500mL/min, 1L/min, 5L/min (depending on model) |
| Accuracy | ±3%F.S. |
| Repeatability | 0.5%F.S. |
| Response time | <2s |
| Power supply | 24V DC |
| Output | RS485/Analog voltage |
| Work pressure | 0.15~0.35MPa |
| Calibrated gas | Air (standard pressure) |
| Communication interface | DB9 male |
| Power consumption | <3W |
| Connector | 2-RC1/4 |
| Weight | 269g |
Note: Standard pressure means 0°C, 1 atm.
6.2 Dimensions
The overall dimensions of the AMC2100 are approximately 73±1 mm (height) x 50±0.5 mm (width) x 80±0.5 mm (depth), with a tolerance of ±0.2 mm. The device features a DB9 connector and 2-G 1/4 connectors for gas inlet/outlet.

6.3 Packing List
| Product | Quantity |
|---|---|
| AMC2000 mass flow controller | 1 |
| Cable | 1 |
| Datasheet | 1 |
| Certificate | 1 |
7. User Tips
- Gas Compatibility: The AMC2100 is versatile and can be used for various gases including Air, N2, O2, CO2, and Ar. Remember to select the correct gas type in the settings for accurate measurements.
- Control and Measurement: This device not only measures gas flow but also provides precise control, making it suitable for applications requiring both monitoring and regulation.
- Modbus RTU: For advanced integration, familiarize yourself with the Modbus RTU protocol details provided in the comprehensive PDF manual.
8. Warranty and Support
8.1 Quality Assurance
Guangzhou Aosong Electronics Co., Ltd. provides the following quality assurance from the date of shipment:
- AMC2000 mass flow controller: 12 months warranty
- Cable and other accessories: 6 months warranty
The company is responsible for defective products used under conditions that meet the product's technical specifications. The warranty does not cover applications in special scenarios not recommended by the company or reliability promises for products applied to unsupported circuits.
8.2 Warning and Personal Injury
Do not use this product in safety protection devices or emergency stop devices, or in any other application where personal injury may occur due to product failure, unless there is a specific purpose and authorization for use. Failure to follow recommendations may result in death or serious personal injury. The company will not be liable for compensation for personal injury and death resulting therefrom, and waives all claims against the company's managers, employees, agents, and distributors.
This manual is subject to change without prior notice. The final right of interpretation of this product belongs to Guangzhou Aosong Electronic Co., Ltd.
8.3 Further Information
For detailed communication protocols, register definitions, and code examples, please refer to the complete user manual PDF: Download Full User Manual (PDF)