1. Introduction
The ZOYI ZT-702S handheld oscilloscope multimeter combines the functions of an oscilloscope and a multimeter into one compact device. It features double-injection molding technology for durability and a comfortable grip. Designed for ease of use, it includes a 2.8-inch IPS full-view color screen and a clear, intuitive function button menu interface. With 9999 counts for multimeter readings, superior performance, and powerful functions, the ZT-702S is suitable for various measurement scenarios to meet diverse user needs.
For a detailed overview of the product's capabilities, you can refer to the official user manual PDF: ZOYI ZT-702S User Manual (PDF)

Image: ZOYI ZT-702S highlighting its dual functionality as an oscilloscope and multimeter, along with features like Type-C charger, 10MHz bandwidth, 9999 counts, True RMS, adjustable shutdown time, 2.8-inch full-view color screen, and one-click automatic adjustment.
2. Safety Information
WARNING: To avoid possible electric shock, fire, and personal injury, please read all safety precautions before use. Use the product only for its intended purpose; otherwise, the protection provided may be impaired.
- Always check the case for cracks or plastic defects before using the product. Double-check the insulator near the input port.
- Follow this User Manual, use the correct input port, and set the correct gear. Measure only within the specified range.
- Do not use this product near explosive gases, vapors, or in humid environments.
- Keep your fingers behind the guard of the test lead probe.
- When connected to a circuit, do not touch unused input ports.
- Disconnect test leads and the circuit before changing the measuring mode.
- Avoid electric shock: DC voltages higher than 36V or AC voltages higher than 25V can cause serious injury.
- Select the correct test gear and range to prevent instrument damage or personal injury.
- Do not operate with the front or back cover opened.
- Charge the battery promptly when low to maintain measurement accuracy.
- The probe's ground wire is connected to GND. Do not connect it to a high-voltage power supply when charging via USB, as this can damage the product or cause injury.
- When measuring voltages higher than AC25V or DC36V with an oscilloscope probe, ensure the USB protective cover is tightly closed to prevent contact with exposed metal parts.
3. Setup
3.1. Package Contents

Image: The ZOYI ZT-702S device shown with its full set of accessories, including the main unit, red and black test leads, an oscilloscope probe with various tips, a USB cable, and a protective carrying case.
The package typically includes:
- ZOYI ZT-702S Digital Oscilloscope Multimeter
- Test Leads (Red and Black)
- Oscilloscope Probe
- USB Type-C Charging Cable
- Carrying Case
- User Manual (this document and a physical copy)
3.2. Charging the Battery
The ZT-702S is powered by a rechargeable 2000mAh Lithium battery (18650 type). When the battery symbol appears in the upper right corner of the screen, it's time to charge.
- Connect the Type-C data cable to the device.
- Connect the other end of the cable to a DC 5V output adapter or a computer's USB port.
- During charging, the screen will display a charging symbol.
- When fully charged, the screen will display a full battery symbol.
- The built-in red light on the power button will be on during charging. It will flash if no battery is installed.

Image: An exploded view of the ZOYI ZT-702S, illustrating its internal structure, including the 2.8-inch full-view color screen, reinforced structure, double injection molding, and the 2000mAh Lithium rechargeable battery.
3.3. Input/Output Ports

Image: Detailed diagram showing the input/output ports of the ZOYI ZT-702S. This includes the 10A current input, mA current input, COM (common) port, VΩHz input, oscilloscope probe input (CH1), square wave signal output, and USB Type-C port for charging and data transfer.
The device features several input/output ports:
- 10A Input Port: For current measurements up to 10A.
- mA Input Port: For current measurements up to 200mA.
- COM Port: Common (return) port for all measurements.
- VΩHz Input Port: For voltage, resistance, capacitance, diode, continuity, frequency, and temperature measurements.
- CH1 (Oscilloscope Probe Input): For oscilloscope measurements. Max input voltage: 1X < 150V, 10X < 300V (DC+AC peak value).
- Square Wave Signal Output: Metal port for outputting a constant 3V/1KHz signal. Do not touch this terminal with bare hands during measurement.
- USB Port: Type-C port for communication with a computer and charging.
4. Operating Instructions
4.1. General Operation

Image: Front panel of the ZOYI ZT-702S, showing the 2.8-inch IPS full-view color screen displaying a waveform, and the button layout including F1-F4, REL, AUTO RANGE, HOLD SAVE, MODE, MENU, and directional keys. Dimensions are also indicated.
Power On/Off
Press and hold the REL (Power) button for 2 seconds to switch the device on or off.
Mode Switching
Press the MODE key to switch between Oscilloscope mode and Multimeter mode.
Menu Navigation
Press the MENU key to open the system function expansion menu. Use the directional keys (up, down, left, right) to adjust settings and navigate menu options. The F1-F4 keys correspond to the functions displayed on the screen.
Video: A short demonstration of the ZOYI ZT-702S, showcasing its physical features, the process of switching between oscilloscope and multimeter modes, and performing basic measurements like DC voltage, continuity, and AC voltage. It also highlights the rechargeable battery and input jack overload protection.
4.2. Oscilloscope Mode
4.2.1. LCD Display Overview
The 2.8-inch IPS full-view color screen displays waveform data and measurement parameters. Key elements include:
- RUN/WAIT/T'D/STOP: Waveform acquisition status.
- Time Base: Current time base position and scale.
- Voltage Scale: Current vertical voltage scale.
- Trigger Level: Set trigger voltage value.
- Battery Power Status: Current battery level and charging status.
- Horizontal/Vertical Cursor: Position indicators.
- Waveform: Display of the collected signal.
- FreQ: Frequency of the collected signal.
- VPP: Peak-to-peak voltage.
- Vmax: Maximum voltage.
4.2.2. Probe Check and Compensation
Before using the oscilloscope probe, ensure safety by keeping fingers behind the safety ring and not touching metal parts when connected to high voltage. The measured voltage must not exceed probe specifications.

Image: Illustrates the safety ring on an oscilloscope probe and examples of normal, overcompensated, and undercompensated square waveforms for probe adjustment.
Manual Probe Offset:
- After powering on, connect the probe to the signal input terminal.
- Input a 4V/1KHz square wave signal.
- Press the AUTO key to check the waveform display status.
- If necessary, adjust the capacitance on the probe using the provided adjustment tool to achieve normal compensation.
4.2.3. Probe Attenuation Setting
The probe attenuation coefficient affects the vertical scale reading. Ensure the attenuation switch on the probe matches the setting in the oscilloscope's system settings (X1 or X10).

Image: Close-up of an oscilloscope probe showing the X1/X10 adjustment switch.
Remarks: When the probe is set to X1, a 6MΩ/X1 probe will limit the oscilloscope's bandwidth to 6MHz. To use the full bandwidth, set the switch to X10 or use a higher specification probe.
4.2.4. Auto Settings
For uncertain waveforms or to avoid manual settings, press the AUTO button. The oscilloscope will automatically identify the waveform type and adjust controls for accurate display.
4.2.5. Vertical System
Sets waveform voltage amplitude, scale, and position.
- Vertical Voltage/Scale Setting: In oscilloscope mode, press F1 to access the voltage time base menu. Use up/down directional keys to adjust voltage settings.
- Vertical Position: Press F2 for waveform movement. Use up/down directional keys to move the waveform vertically.
4.2.6. Horizontal System
Sets horizontal scale (time base) and position.
- Horizontal Scale: Press F1 to the voltage time base menu. Use left/right directional keys to change the horizontal scale. Right key zooms out, left key zooms in.
- Horizontal Position: Select waveform movement menu. Use left/right arrow keys to move the waveform horizontally. Press and hold MENU to return the horizontal cursor to the center.
- Rolling Mode: For slow signals (<100ms/div), the oscilloscope enters rolling mode, displaying waveforms scrolling from left to right.
4.2.7. Triggering System
Used to stabilize and capture specific waveforms.
- Trigger Cursor Setting: Press F3 to access the trigger cursor menu. Use up/down directional keys to adjust the trigger level.
- Trigger Mode: Press MENU, then F2 to select:
- Auto: Continuously acquires and refreshes waveforms without retaining them.
- Normal: Locks and displays the waveform when the signal reaches the trigger level. Continues to collect and updates the waveform upon subsequent triggers.
- Single: Locks and displays the waveform once the signal reaches the trigger level, then stops acquisition (STOP state). Press HOLD to re-enable triggering.
- Trigger Edge: Press MENU, then F3 to select rising or falling edge triggering.
4.2.8. Measurement
- Automatic Measurement: Press AUTO to automatically identify and adjust waveform amplitude and time base for unknown signals.
- Manual Measurement: Manually set parameters like voltage, time base, cursor position, trigger, coupling mode, and probe attenuation.
- Measured Values: Press MENU, then navigate to the next menu with the right directional key, and press F4 to show/hide values. FreQ, VPP, Vmax are always displayed. Vmin, RMS, Prd can be toggled.
4.2.9. Saving and Viewing Waveforms
To save a measured waveform, press and hold HOLD/SAVE for 2 seconds. The oscilloscope saves the waveform data as a picture in its internal flash memory.
To view saved waveforms, connect the device to a computer via the Type-C USB cable. The device will appear as a USB disk. Open the 'pic' folder to view or download the waveform images.
4.2.10. Additional Functions (Oscilloscope Mode)
- Language Settings: Press MENU, navigate to the next menu, press F1 to choose Simplified Chinese or English.
- Automatic Shut-down: Press MENU, navigate to the next menu, press F2 to set auto-shutdown time (15, 30, 60, 120 minutes, or unlimited).
- Backlight Brightness: Press MENU, navigate to the next menu, press F3 to adjust brightness (30%, 50%, 80%, 100%). 30% is recommended for indoor use.
- Baseline Calibration: Press MENU, navigate to the last page menu, press F1. Do not connect probes or input signals during calibration. Do not perform other operations.
- Reset: Press MENU, navigate to the last page menu, press F2 to format and reset to defaults. The device will shut down. Press the power button to turn it back on.
4.3. Multimeter Mode
4.3.1. LCD Display Overview

Image: The ZOYI ZT-702S in multimeter mode, showing the LCD display with a large numerical reading, analog pointer, battery status, measurement mode indicators (DC, AC, resistance, capacitance, diode, buzzer), and min/max/avg values.
The multimeter display shows:
- HOLD: Indicates data hold is active.
- Battery Power Status: Current battery level.
- Analog Pointer: Changes with measurement data, indicating scale position.
- Measurement Symbol: AC, DC, resistance, capacitance, diode, buzzer symbols.
- Negative: Displays a negative sign for negative values.
- Main Display: Measured value (max 9999 counts).
- Unit Symbol: Unit of the measured data.
- Measurement Mode: Auto or Manual range.
- Max/Min: Maximum/Minimum readings.
- AVG/Hz/Temp: Average value (DC voltage, resistance, capacitance), AC frequency (Hz), or Fahrenheit (°F) for temperature.
4.3.2. Multimeter Input Terminals

Image: Diagram detailing the multimeter input terminals: 10A (for high current), mA (for low current), COM (common), and VΩHz (for voltage, resistance, frequency, temperature, diode, continuity).
Connect test leads to the appropriate terminals:
- 10A: For current measurement (≤9.999A).
- mA: For current measurement (≤99.99mA).
- COM: Common (return) port for all measurements.
- VΩHz: For AC/DC voltage, resistance, electrical appliances, frequency, temperature, on/off, and diode measurements.
4.3.3. Measuring AC and DC Voltage
- Insert the black test lead into the COM terminal and the red test lead into the VΩHz terminal.
- For voltages less than 99.99mV, press F4 to enter mV range, then press it twice for AC mV.
- For voltages more than 99.99mV, press F1 for DC voltage range, then press it twice for AC voltage range.
- Contact the probe to the test point on the circuit and read the value.
WARNING: Do not exceed the rated maximum voltage. Avoid touching high voltage circuits.
4.3.4. Measuring AC and DC Current
- Insert the black test lead into the COM port. Insert the red test lead into the 10A port or the mA port (choose based on expected current).
- Press F3 to enter the current mode menu. F4 corresponds to mA current.
- Press the corresponding menu key again to switch between DC and AC.
WARNING: Do not exceed the rated maximum current. If the current is unknown, start with the 10A port. It is strictly forbidden to input voltage in current measurement mode.
4.3.5. Measuring Resistance
- Insert the black test lead into the COM port and the red test lead into the VΩHz port.
- Press F2 to enter the resistance gear.
- Touch the probe to the test point of the circuit and read the value.
WARNING: Before measuring resistance, ensure all power supplies are off and capacitors are discharged. It is strictly forbidden to input voltage in this mode.
4.3.6. Measuring Continuity
- Insert the black test lead into the COM port and the red test lead into the VΩHz port.
- In resistance mode, press F2 to enter continuity mode.
- Connect the probes to the two points of the circuit. A buzzer sound indicates a short circuit.
WARNING: It is strictly forbidden to input voltage in continuity mode. Disconnect power and discharge high voltage capacitors before testing.
4.3.7. Measuring Diode
- Press F2 in the on-off gear to enter diode gear.
- Connect the red test lead to the positive of the diode and the black test lead to the negative. Read the forward voltage. If polarity is opposite or diode is damaged, the screen will display "OL".
WARNING: It is strictly forbidden to input voltage in diode mode. Disconnect power and discharge high voltage capacitors before testing.
4.3.8. Measuring Capacitance
- Insert the black test lead into the COM port and the red test lead into the VΩHz port.
- Press F2 in diode gear to enter electronic gear.
- Connect the red test lead to the positive pole of the capacitor and the black test lead to the negative pole.
- After the reading stabilizes, read the capacitance value.
WARNING: It is strictly forbidden to input voltage in capacitance mode. Disconnect power and discharge high voltage capacitors before testing.
4.3.9. Measuring Temperature
- Connect the black test lead to the COM terminal and the red lead to the VΩHz terminal.
- Push F4 three times to enter Temperature Mode. Both displays will show room temperature; the primary display shows Celsius, secondary shows Fahrenheit.
- Touch the probes to the desired test points and read the temperature.
WARNING: Do not input voltage in this mode.
5. Maintenance
5.1. Cleaning the Product
Wipe the product with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture in the terminals can affect readings. Always remove input signals before cleaning.
5.2. Battery Storage
If the instrument is not used for more than 6 months, charge it to 50%-70% capacity, then remove the battery and store it in a cool, dry environment. If the lithium battery shows signs of rust, leakage, or bulging, discard it immediately.
5.3. Replacing the Batteries
The lithium battery is rechargeable but consumable. If standby time significantly shortens, replace it with the same type of 18650 lithium battery. Ensure correct positive and negative polarities when installing.
5.4. Replacing the Fuses
If a fuse is blown or not working properly, replace it as follows:
- Remove the test leads and turn off the product.
- Loosen the four screws on the back cover and the screw on the battery door. Remove the battery door and back cover.
- Replace the fuse with a new fuse of the same type.
- Place the back cover and battery door back and fasten the screws.
6. Troubleshooting
- Device freezes or crashes: If the device freezes or crashes during use, please restart it.
- Low battery affecting accuracy: If the battery voltage is low, it may affect the accuracy of test results. Charge the device promptly.
- Inaccurate oscilloscope readings: Check probe compensation. Uncompensated or improperly compensated probes can cause measurement errors. Perform manual probe offset as described in section 4.2.2.
- Battery not holding charge: If the built-in battery does not hold its charge for an extended period (e.g., less than 24 hours when off), it may need to be replaced. Refer to section 5.3.
- No display or incorrect readings in multimeter mode: Ensure test leads are connected to the correct input ports for the desired measurement. Verify the correct measurement gear is selected.
- Overload protection warnings: Pay attention to overload protection warnings. Current input jacks may not have overload protection, while voltage input jacks typically have protection up to 950V.
7. Specifications
7.1. General Specifications
| Characteristic | Value |
|---|---|
| Display | 9999 Counts (LCD) |
| Ranging | Auto/Manual |
| Material | ABS+TPE |
| Update Rate | 3 Times/Second |
| True RMS | Yes |
| Data Hold | Yes |
| Backlight | Yes |
| Low Battery Indication | Yes |
| Auto Power Off | Yes |
7.2. Mechanical Specifications
| Characteristic | Value |
|---|---|
| Dimension | 177*89*40mm |
| Weight | 345g (NO BATTERY) |
| Battery Type | 18650 Battery * 1 |
7.3. Environmental Specifications
| Category | Characteristic | Value |
|---|---|---|
| Operating | Temperature | 0~40°C |
| Humidity | <75% | |
| Storage | Temperature | -20~60°C |
| Humidity | <80% |
7.4. Multimeter Specifications
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| DC Voltage (V) | 999.9mV | 0.1mV | ±(0.5%+3) |
| 9.999V | 0.001V | ||
| 99.99V | 0.01V | ||
| 999.9V | 0.1V | ||
| DC Voltage (mV) | 9.999mV | 0.001mV | |
| 99.99mV | 0.01mV | ||
| AC Voltage (V) | 999.9mV | 0.1mV | ±(1.0%+3) |
| 9.999V | 0.001V | ||
| 99.99V | 0.01V | ||
| 750.0V | 0.1V | ||
| AC Voltage (mV) | 9.999mV | 0.001mV | |
| 99.99mV | 0.01mV | ||
| *Frequency response of ACV: 40Hz-1kHz | |||
| DC Current (A/mA) | 9.999A | 0.001A | ±(1.0%+3) |
| 999.9mA | 0.1mA | ||
| DC Current (µA/mA) | 99.99mA | 0.01mA | ±(0.8%+3) |
| 9999µA | 1µA | ||
| AC Current (A) | 9.999A | 0.001A | ±(1.2%+3) |
| 999.9mA | 0.1mA | ||
| AC Current (µA/mA) | 99.99mA | 0.01mA | ±(1.0%+3) |
| 9999µA | 1µA | ||
| Frequency response of AC Current: 40Hz-1kHz | |||
| Resistance | 99.99Ω | 0.01Ω | ±(1.0%+3) |
| 999.9Ω | 0.1Ω | ||
| 9.999KΩ | 0.001KΩ | ±(0.5%+3) | |
| 99.99KΩ | 0.01KΩ | ||
| 999.9KΩ | 0.1KΩ | ||
| 9.999MΩ | 0.001MΩ | ±(1.5%+3) | |
| 99.99MΩ | 0.01MΩ | ±(3.0%+5) | |
| Capacitance | 9.999nF | 0.001nF | ±(5.0%+20) |
| 99.99nF | 0.01nF | ±(2.0%+5) | |
| 999.9nF | 0.1nF | ||
| 9.999µF | 0.001µF | ||
| 99.99µF | 0.01µF | ±(5.0%+5) | |
| 9.999mF | 0.001mF | ||
| 99.99mF | 0.01mF | ||
| Frequency (Measures only to 100kHz under ACV setting) | 99.99Hz | 0.01Hz | ±(0.1%+2) |
| 999.9Hz | 0.1Hz | ||
| 9.999kHz | 0.001kHz | ||
| 99.99kHz | 0.01kHz | ||
| 999.9kHz | 0.1kHz | ||
| Temperature | (-20~1000)°C | 1°C | ±(2.5%+5) |
| (-4~1832)°F | 1°F | ||
| Diode | Yes | ||
| Continuity | Yes | ||
7.5. Oscilloscope Specifications
| Characteristic | Instructions |
|---|---|
| Bandwidth | 10MHz |
| Sampling method | Real-time sampling |
| Real-time sample rate | 48MSa/s |
| Channels | 1 |
| Input coupling | AC/DC |
| Input resistance | 1MΩ, @16pF |
| Probe attenuation | X1, X10 |
| Max input voltage | X1<150V, X10<300V (DC+AC peak) |
| Sample rate range | 1.5Sa/s - 48MSa/s |
| Waveform interpolation | (sinx)x |
| Sweep speed range | 100ns/div - 20s/div |
| Time Base Accuracy | 20ppm |
| Record length | Max 64Kbyte |
| Sensitivity | 20mV/div - 10V/div |
| Displacement range | 4 cells (positive and negative) |
| Analog Bandwidth | 10MHz |
| Low frequency response | >10Hz |
| Rise Time | <10ns |
| DC Gain Accuracy | ±3% |
| Automatic measurement | Period, frequency, peak-to-peak value, maximum value, minimum value, effective value |
| Trigger method | Auto, Normal, Single |
| Trigger edge | Rising edge, Falling edge |
8. Warranty and Support
8.1. Limited Warranty
Customers enjoy a one-year warranty from the date of purchase. This warranty does not cover:
- Fuses or disposable batteries.
- Damage from misuse, accident, neglect, alteration, contamination, or abnormal operating/handling conditions.
- Failures caused by use outside of the product's specifications.
- Normal wear and tear of mechanical components.
8.2. Support
Beyond replacing batteries and fuses, do not attempt to repair or service the product unless you are qualified and have the relevant calibration, performance test, and service instructions.
9. User Tips
- Read Instructions Carefully: Several users emphasize the importance of reading the manual to avoid damaging the device and to understand its full capabilities.
- Consider Two-Channel: While this is a single-channel device, some users suggest considering a two-channel version if your work requires it.
- Component Testing: The multimeter functions are effective for checking components like diodes, capacitors, and resistors on electronic boards.
- Battery Management: Be aware that the built-in battery might not hold a charge for extended periods when off. It may need charging before each use, or consider replacing it if its capacity significantly degrades.
- Correct Connections: Always ensure test leads are connected to the correct input ports and the appropriate measurement mode/gear is selected to prevent damage and ensure accurate readings.
- Voltage Input Caution: Never input voltage into current, resistance, continuity, diode, capacitance, or temperature measurement modes, as this can damage the device.