GVDA GD128PLUS

Contents

GVDA Smart Rechargeable Digital Multimeter

Model: GD128PLUS

1. Introduction

The GVDA GD128PLUS is an intelligent True RMS digital multimeter designed for both intelligent and professional measurement functions. It features a full-function display with gear indication, an analog bar, and multiple display modes. This device is capable of measuring DC/AC Voltage, DC/AC Current, Resistance, Capacitance, Frequency/Duty Cycle, and Temperature, along with Non-Contact Voltage (NCV) detection and Live Wire detection.

The multimeter is rechargeable via USB at 5V, eliminating the need for frequent battery replacement.

GVDA GD128PLUS Smart Digital Multimeter front view with color LCD display

Figure 1: GVDA GD128PLUS Smart Digital Multimeter

2. Safety Information

Please read this manual carefully before using the instrument and save it for future reference. Observe all safety warnings and instructions to prevent electric shock, personal injury, or damage to the meter or equipment.

2.1. Safety Statements

  • Caution: Indicates an operation that may cause damage to the meter or equipment.
  • Warning: Indicates an operation that may cause danger to users.

2.2. Safety Instructions

The meter conforms to the IEC61010-1 international electrical safety standard. Its design and manufacture comply with IEC61010-1 CAT.III 1000V, CAT.IV 600V overvoltage safety standard and pollution level 2.

  • Always read this manual carefully and pay special attention to safety warnings before use.
  • Operate the meter strictly according to the manual. Failure to do so may damage or weaken the instrument's protection function.
  • Exercise special care when measuring values exceeding 60VDC, 30VAC RMS, or 42V, as these voltages pose a risk of electric shock.
  • Do not measure voltage higher than the rated value between terminals or between terminals and ground.
  • Before each use, check the meter's functionality by measuring a known voltage. If it appears abnormal or damaged, do not use it.
  • Inspect the instrument shell for cracks or damaged plastic parts before use. If found, do not use the meter.
  • Check the probes for cracks or damage. If damaged, replace them with probes of the same model and electrical specification.
  • Use the meter according to the measurement category, voltage, or current rating specified in the meter or manual.
  • Observe all local and national safety regulations. Wear personal protective equipment (e.g., approved rubber gloves, masks, flame-retardant clothing) when dangerous live conductors are exposed to prevent injury from electric shock or arc flash.
  • When the low battery symbol (Low battery icon) is displayed, replace the battery promptly to prevent measurement errors.
  • Do not use the meter in environments with explosive gas, steam, or high humidity.
  • When using probes, always hold your fingers behind the probe finger guard.
  • When measuring, connect the null or ground wire first, then the live wire. When disconnecting, disconnect the live wire first, then the null or ground wire.
  • Remove the probes from the meter before opening the case or battery cover. Do not operate the meter when it is disassembled or its battery cover is open.
  • The meter must only be used with the provided probes or probes that meet the same safety standard requirements.

3. Product Overview

The GVDA GD128PLUS is an intelligent true RMS digital multimeter with a color LCD display, 9999 counts, and various measurement capabilities. It features a robust design and a built-in LED flashlight for convenience.

3.1. Meter Layout

Diagram of GVDA GD128PLUS Multimeter with numbered parts

Figure 2: Multimeter Components

  1. Power On/Off Button
  2. Charge Jack (USB-C)
  3. NCV (Non-Contact Voltage) Sensor Area
  4. Warning Indicator
  5. Display (Color LCD)
  6. Flashlight Key
  7. Input Jack Indicator
  8. 10A Jack (for high current measurements)
  9. mA Jack (for milliampere current measurements)
  10. COM Jack (Common terminal for all measurements)
  11. Jack other than current and NCV (Input for voltage, resistance, capacitance, etc.)
  12. Flashlight

3.2. Package Contents

Diagram showing the contents of the GVDA GD128PLUS package: gift box, manual, bag, multimeter, original test leads, charging cable, and thermocouple probe.

Figure 3: Package Contents

  1. Gift box
  2. User Manual
  3. Carrying Bag
  4. GVDA GD128PLUS Multimeter
  5. Original Test Leads
  6. USB Charging Cable
  7. K-type Thermocouple Probe

4. Setup

4.1. Initial Inspection

Before first use, carefully inspect the multimeter and its accessories for any signs of damage. Ensure all components listed in the package contents are present and in good condition.

4.2. Charging the Battery

The GD128PLUS is equipped with a rechargeable lithium battery. To charge:

  1. Connect the USB charging cable to the Charge Jack (2) on the multimeter.
  2. Connect the other end of the USB cable to a 5V USB power source (e.g., a power adapter, computer USB port, or power bank).
  3. The display will show a charging indicator.
GVDA GD128PLUS Multimeter being charged via a USB power bank

Figure 4: Charging the Multimeter

4.3. Connecting Test Leads

For most measurements, insert the red test lead into the appropriate input jack (e.g., INPUT, mA, or 10A) and the black test lead into the COM jack (10).

5. Operating Instructions

5.1. Power On/Off

Press and hold the Power On/Off button (1) for approximately 2 seconds to turn the meter on or off.

5.2. Range Selection

Press the RANGE key to enter manual range mode. Press it repeatedly to cycle through available ranges. Press and hold the RANGE key for about 2 seconds to return to automatic range mode.

  • Note 1: This function is invalid in intelligent measurement mode.
  • Note 2: Only voltage, resistance, and mA gears are valid for manual range selection.

5.3. Gear Selection

Press the FUNC key to enter manual shift mode. Press it repeatedly to select the desired measurement function (gear). Press and hold the FUNC key for about 2 seconds to return to the intelligent (auto) measurement mode. The meter defaults to intelligent measurement mode upon power-on.

For multiple functions within one gear, press the SEL key to switch between them.

5.4. Max/Min Measurement

Press the MAX MIN key to enter maximum/minimum measurement mode. Press it again to display the maximum and minimum values in a cycle. Press and hold the MAX MIN key for about 2 seconds to return to normal measurement.

  • Note 1: This function is invalid in capacitance, frequency/duty cycle, temperature, NCV/Live gear.
  • Note 2: In Max/Min measurement, the instrument automatically enters manual range mode.

5.5. Relative Value Measurement

Press and hold the HOLD REL key for about 2 seconds to turn on or off relative value measurement.

  • Note 1: This function is invalid in continuity, frequency/duty cycle, temperature, NCV/Live modes.
  • Note 2: In relative value measurement, the meter automatically enters manual range mode.

5.6. Data Hold

Press the HOLD REL key to turn on or off data holding. This function is invalid in NCV/Live modes.

5.7. Flashlight

Press the SEL key to turn on or off the built-in flashlight (12).

Back of the GVDA GD128PLUS Multimeter showing the LED flashlight

Figure 5: LED Flashlight

5.8. Automatic Power Off

After power on, the auto power off function is enabled by default, indicated by the Auto power off icon symbol on the display. If there is no key operation for approximately 15 minutes, the meter will automatically shut down to save battery energy.

To turn off the automatic shutdown function, press the SEL key. The automatic shutdown function is off when the Auto power off icon symbol is no longer displayed.

5.9. Input Jack Indicator

When the measurement gear is changed, the corresponding input light will flash 5 times to prompt the user to insert the probe into the correct jack.

5.10. Measurement Operations

Warning: Do not measure voltage higher than DC1000V or AC750V, otherwise the meter may be damaged. Pay special attention to safety when measuring high voltage to avoid electric shock or personal injury. Before use, test the known voltage or current with the meter to confirm that the meter is in good condition.

5.10.1. Smart (AUTO) Measurement

This is the default mode when the meter is powered on. In this mode, the meter automatically identifies and measures DC voltage, AC voltage, resistance, and continuity.

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  3. Contact the probes with both ends of the measured power supply or resistance (in parallel). The meter will automatically recognize the signal.
  4. Read the results from the display.

5.10.2. AC/DC Voltage Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "V" gear. The "DC" symbol will be displayed for DC voltage measurement. Press the SEL key to display the "AC" symbol for AC voltage measurement.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured power supply (in parallel).
  5. Read the results from the display.

5.10.3. Resistance Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "Ω" gear.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured resistance (in parallel).
  5. Read the results from the display.

5.10.4. Continuity Test

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "Continuity icon))" gear.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured resistance or circuit.
  5. If the resistance value is less than 50 Ω, the buzzer will sound, and the alarm indicator will be on.
  6. Read the results from the display.

5.10.5. Diode Test

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "Diode icon" gear.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. The red probe contacts the anode of the diode, and the black probe contacts the cathode of the diode.
  5. If the probe polarity is opposite to the diode polarity, the display will show "OL".
  6. Read the results from the display.

5.10.6. Capacitance Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "Capacitance icon" gear.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured capacitance (in parallel).
  5. Read the results from the display.

5.10.7. AC/DC mV Voltage Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "mV" gear. The "DC" symbol will be displayed for DC mV voltage measurement. Press the SEL key to display the "AC" symbol for AC mV voltage measurement.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured power supply.
  5. Read the results from the display.

5.10.8. Frequency/Duty Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "Hz%" gear.
  3. Insert the red probe into the INPUT jack (11) and the black probe into the COM jack (10).
  4. Contact the probes with both ends of the measured power supply.
  5. Read the results from the display.

5.10.9. Temperature Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "°C/°F" gear.
  3. Insert the positive pole of the K-type thermocouple into the INPUT jack (11) and the negative pole into the COM jack (10).
  4. The thermocouple probe contacts the measured object.
  5. Read the results from the display.
GVDA GD128PLUS Multimeter performing a temperature test with a K-type probe in a glass of water

Figure 6: Temperature Test

5.10.10. Non-Contact AC Voltage (NCV) Detection

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "NCV/Live" gear. The "NCV" symbol will be displayed.
  3. Gradually move the NCV sensor area (3) close to the conductor.
  4. When a weak electric field signal is detected, the display will show "---L", and the buzzer will sound slowly with the green light on.
  5. When a strong electric field signal is detected, the display will show "---H", and the buzzer will sound quickly with the red light on.

5.10.11. Live Wire Detecting

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "NCV/Live" gear. Press the SEL key to display the "LIVE" symbol.
  3. Insert the red probe into the INPUT jack (11) and remove the black probe.
  4. Use the red probe to contact the conductor.
  5. When a weak electric field signal is detected, the display will show "---L", and the buzzer will sound slowly with the green light on.
  6. When a strong electric field signal is detected, the display will show "---H", and the buzzer will sound quickly with the red light on.
GVDA GD128PLUS Multimeter performing live wire detection

Figure 7: Live Wire Detecting

5.10.12. mA Current Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "mA" gear. Alternatively, insert the red probe into the mA jack (9) to automatically select the "mA" gear.
  3. The "DC" symbol indicates DC current measurement. Press the SEL key to display the "AC" symbol for AC current measurement.
  4. Insert the red probe into the mA jack (9) and the black probe into the COM jack (10).
  5. Connect the meter in series with the measured power supply.
  6. Read the results from the display.

Caution: Do not measure current > 630 mA in this gear, otherwise the mA fuse will be burnt out.

5.10.13. Ampere (A) Current Measurement

  1. Press the Power On/Off button (1) to start the meter. The display will show "Auto" and enter intelligent measurement mode.
  2. Press the FUNC key to select the "A" gear. Alternatively, insert the red probe into the 10A jack (8) to automatically select the "A" gear.
  3. The "DC" symbol indicates DC current measurement. Press the SEL key to display the "AC" symbol for AC current measurement.
  4. Insert the red probe into the 10A jack (8) and the black probe into the COM jack (10).
  5. Connect the meter in series with the measured power supply.
  6. Read the results from the display.

Caution: Do not measure current > 10A in this gear, otherwise the 10A fuse will be burnt out.

6. Maintenance

6.1. Cleaning

To clean the meter, follow these steps:

  1. Turn off the meter power and remove the probes.
  2. Wipe the case with a damp cloth or mild detergent. Do not use abrasives or solvents.
  3. Wipe the contacts in each input socket with a clean swab soaked in alcohol.

Warning: Always keep the inside of the meter clean and dry to prevent electric shock or damage to the meter.

6.2. Battery Replacement

The GD128PLUS uses a rechargeable lithium battery, so regular replacement is not needed. However, if the battery needs to be accessed or replaced by a qualified technician:

  1. Turn off the meter power and remove the probes.
  2. Remove the screw fixing the battery cover and remove the battery cover.
  3. Remove the old battery and replace it with a new one of the same specification. Pay attention to the battery polarity.
  4. Install the battery cover back to its original position, and fix and lock the battery cover with screws.

Warning: To avoid electric shock or personal injury caused by wrong readings, replace the battery immediately when the battery is low. Do not discharge the battery by shorting it or reversing its polarity. To operate and maintain the meter safely, take out the battery when it is not used for a long time to prevent battery leakage from damaging the product.

6.3. Fuse Replacement

If a fuse is burnt out, the symbol "FUSE" will be displayed along with the fuse icon. Replace the fuse immediately.

  1. Turn off the meter power and remove the probes.
  2. Remove the screw fixing the back cover and remove the back cover.
  3. Remove the burnt-out fuse, replace it with a new one of the same specification (mA: F630mA/250V fuse; 10A: F10A/250V fuse). Ensure the fuse is installed in the safety clip and clamped tightly.
  4. Install the back cover and fix it with screws.

Warning: To prevent electric shock, personal injury, or damage to the meter, please use a fuse of the same specification or specified specification.

7. Troubleshooting

  • "LEAd" displayed: This indicates that the meter is switched manually to a current gear (A-jack or mA-jack) but no test lead is inserted. Insert the test lead or switch to a different measurement mode.
  • Fuse burnt out: If the fuse symbol and "FUSE" are displayed, the fuse for the current measurement range is burnt. Refer to section 6.3 for fuse replacement.
  • Low battery indication: If the low battery symbol (Low battery icon) is displayed, recharge the multimeter.
  • Inaccurate or intermittent readings: Ensure test leads are securely connected to the meter and the circuit. Loose connections can cause unreliable measurements.
  • "OL" displayed: This indicates an over-range condition (measurement value exceeds the meter's maximum range) or an open circuit.

8. Specifications

8.1. General Technical Specifications

  • Environment Condition: CAT. IV 600V; CAT. III 1000V; Pollution level 2, Altitude < 2000m
  • Working Temperature and Humidity: 0~40°C (<80% RH, <10°C non-condensing)
  • Storage Temperature and Humidity: -10~60°C (<70% RH, remove the battery)
  • Temperature Coefficient: 0.1 × accuracy /°C (<18°C or >28°C)
  • MAX. Voltage between terminals and earth ground: DC1000V/AC750V
  • Fuse Protection:
    • mA: F630mA/250V fuse
    • 10A: F10A/250V fuse
  • Sampling Rate: Approx. 3 times/second
  • Display: 9999 counts
  • Over Range Indication: "OL"
  • Low Battery Indication: Low battery icon symbol
  • Input Polarity Indication: Display "-"
  • Power Requirement: 3.7V Li batteries (rechargeable)
  • Dimensions: 166*83*29mm
  • DIY Supplies: Electrical
  • Certification: CE, FCC
  • Is Smart Device: No
  • Hign-concerned Chemical: None

The accuracy applies within one year after calibration. Reference condition: environment temperature 18°C to 28°C, relative humidity no more than 80%, accuracy: ± (% reading + word).

8.2. Measurement Accuracy

Range Resolution Accuracy
DC Voltage
99.99mV0.01mV±(0.5%+3)
999.9mV0.1mVImpedance: Approx. 10MΩ
9.999V0.001V
99.99V0.01V
999.9V0.1V
AC Voltage
99.99mV0.01mV±(0.8%+3)
999.9mV0.1mVImpedance: Approx. 10MΩ
Frequency Response: 40Hz~1kHz; TRMS
9.999V0.001V
99.99V0.01V
750V0.1V
AC/DC Current
9.999mA0.001mA±(0.8%+3)
99.99mA0.01mA
600.0mA0.1mA
9.999A0.001A±(1.2%+3)
Overload protection: mA: F630mA/250V fuse; 10A: F10A/250V fuse.
NOTE: The time to measure ampere current must be less than 15 seconds.
Diode/Continuity
Diode iconDisplay diode voltage drop
Continuity icon))<Approx. 50Ω, Buzzer will sound and the indicator light will be on.
Resistance
999.9Ω0.1Ω±(1.0%+5)
9.999KΩ0.001 KΩ
99.99 KΩ0.01V KΩ
999.9 KΩ0.1V KΩ
9.999MΩ0.001 MΩ
99.99 MΩ0.01 MΩ±(2.0%+10)
Overload protection: 250V
Capacitance
9.999nF0.001nF±(4.0%+3)
99.99nF0.01nF±(5.0%+5)
999.9nF0.1nF
9.999µF0.001µF
99.99µF0.01µF
999.9µF0.1µF
9.999mF0.001mF
99.99mF0.01mF±(5.0%+5)
Overload protection: 250V
Frequency/Duty
9.999Hz0.001Hz±(1.0%+3)
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999KHz0.001KHz
99.99kHz0.01kHz±(1.0%+3)
999.9kHz0.1kHz
9.999MHz0.001MHz
1.0~99.0%0.1%±(1.0%+3)
Temperature (use K-type thermocouple probe)
-40°C~0°C1°C±5.0% or ±3°C
0°C~400°C1°C±1.0% or ±2°C
400°C~1000°C1°C±2.0%
-40°F~32°F1°F±5.0% or ±6°F
32°F~52°F1°F±1.0% or ±4°F
752°F~1832°F1°F±2.0%

9. User Tips

  • Rechargeable Battery: The multimeter features a built-in rechargeable lithium battery, eliminating the need for frequent battery replacements. Simply charge it via a 5V USB power source when the low battery indicator appears.
  • Clear Display: Users appreciate the clear screen and solid case design, which enhances readability and durability.
  • Check Test Lead Connections: If you experience intermittent or inaccurate measurements, ensure your test leads are firmly seated in the input jacks. Loose connections can affect readings.

10. Warranty and Support

For detailed warranty information and technical support, please refer to the official GVDA website or contact your retailer.

A comprehensive user manual in PDF format is available for download:

View User Manual (PDF)