SUNSHINE DT-17N

Contents

SUNSHINE DT-17N Fully Automatic Digital Multimeter User Manual

Model: DT-17N

1. Introduction

The SUNSHINE DT-17N is a 3 5/6 digit automatic digital multimeter, offering stable performance, high precision, and high reliability. Powered by batteries, this instrument features a 21mm high LCD display for clear readings, approximately 15 seconds of backlight delay, high voltage alarm, and overload protection, making it convenient to use.

This series of instruments can be used to measure DC voltage, AC voltage, DC current, AC current, resistance, capacitance, diode, temperature, on-off test, True RMS measurement, LED high-voltage output measurement, and frequency. It is an ideal tool for laboratories, factories, radio enthusiasts, and home use.

SUNSHINE DT-17N Digital Multimeter front view with key features highlighted
Figure 1.1: SUNSHINE DT-17N Digital Multimeter with highlighted features like intelligent detection, precise measurement, fuse protection, and auto range frequency measurement.

2. Safety Precautions

WARNING: Please read all safety instructions carefully before use to prevent electric shock or damage to the instrument.

  • When measuring voltage, do not input voltage exceeding DC 1000V or AC 700V RMS.
  • Voltages below 36V are considered safe. When measuring above 36V DC and 25V AC, ensure test leads have reliable contact, are correctly connected, and insulation is proper to avoid electric shock.
  • When changing functions and ranges, ensure test leads are disconnected from the test points.
  • Always select the correct function and range. Although the instrument has full-range protection, careful operation is essential for safety.
  • When measuring current, do not input current exceeding 10A.
  • Safety Symbol Description:
    • Δ: Indicates dangerous voltage.
    • Δ: Operator must refer to the instruction manual.
    • ⊝: Ground.
    • ■■: Double insulation.
    • ☀: Low voltage symbol.
Close-up of a fuse, illustrating fuse protection
Figure 2.1: The multimeter features built-in fuse protection to prevent damage from misoperation.

3. Product Features

  • Display: LCD with 5999 (3 5/6) counts, automatic polarity display.
  • Measurement Method: Double integral A/D conversion.
  • Sampling Rate: Approximately 3 times per second.
  • Overrange Display: "OL" symbol.
  • Low Voltage Display: Battery symbol appears.
  • Operating Environment: 0~40°C, relative humidity <80%.
  • Power Supply: 9V battery (not included).
  • Dimensions: 184 x 90 x 46 mm (LWH).
  • Weight: Approximately 320g (including 1.5V battery).
  • Accessories: User manual, certificate, test leads, K-type thermocouple TP01 (for temperature measurement).
Close-up of a high-speed chip, representing intelligent detection
Figure 3.1: High-speed chip for intelligent and precise measurements.

4. Setup

4.1 Battery Installation

The DT-17N requires a 9V battery (not included). When the low battery symbol (☀) appears on the screen, replace the battery immediately.

  1. Unscrew the screws that secure the battery cover and remove the cover.
  2. Remove the old battery and insert a new 9V battery, ensuring correct polarity. Alkaline batteries are recommended for extended use.
  3. Replace the battery cover and tighten the screws.

4.2 Test Lead Connection

Connect the black test lead to the "COM" jack and the red test lead to the appropriate function jack (e.g., "V/Ω/Hz" for voltage, resistance, frequency, or "10A" for high current).

Multimeter showing integrated test lead storage slots
Figure 4.1: The multimeter features convenient slots for organizing test leads.

4.3 Kickstand Usage

The multimeter is equipped with a kickstand on the back for convenient angled viewing during use.

Multimeter with its kickstand extended for angled viewing
Figure 4.2: Extend the kickstand for a 45° reading display, improving readability.

5. Operating Instructions

5.1 Overview of Operation Panel

The DT-17N features a clear LCD display and a rotary switch for selecting functions. Buttons include HOLD (data hold/backlight), RANGE, RELΔ (relative measurement), Hz% (frequency/duty cycle), and MAX/MIN.

Multimeter displaying '100.0 mV APO' with feature callouts
Figure 5.1: Front panel layout of the DT-17N Digital Multimeter.

5.2 DC Voltage Measurement (DCV)

  1. Insert the black test lead into the "COM" jack and the red test lead into the "V/Ω/Hz" jack.
  2. Rotate the function switch to the "V" position (DCV mode).
  3. Connect the test leads across the circuit or component to be measured. The display will show the DC voltage value and the polarity of the red test lead.

Note:

  • Do not input voltage exceeding DC 1000V.
  • Exercise caution when measuring high voltage circuits.
  • Disconnect test leads from the circuit after measurement.

5.3 AC Voltage Measurement (ACV)

  1. Insert the black test lead into the "COM" jack and the red test lead into the "V/Ω/Hz" jack.
  2. Rotate the function switch to the "V" position (ACV mode).
  3. Connect the test leads across the circuit or component to be measured. The display will show the AC voltage value.

Note:

  • Do not input voltage exceeding AC 750V RMS.
  • Exercise caution when measuring high voltage circuits.
  • Disconnect test leads from the circuit after measurement.

Video 5.1: Product demonstration showing various measurement functions, including switching between buzzer and diode modes.

5.4 DC Current Measurement (DCA)

  1. Insert the black test lead into the "COM" jack.
  2. For currents up to 600mA, insert the red test lead into the "mA/µA" jack. For currents up to 10A, insert the red test lead into the "10A" jack.
  3. Rotate the function switch to the "A" position (DCA mode). Use the "SELECT" button to switch between AC and DC current.
  4. Turn off power to the circuit, then connect the multimeter in series with the circuit.
  5. Turn on power and read the DC current value. The display will also show the polarity of the red test lead.

Note:

  • Always turn off circuit power before connecting the multimeter in series.
  • If the current range is unknown, start with the highest range (10A) and decrease as needed. "OL" indicates overrange.
  • Maximum input current is 600mA or 10A. Do not measure currents exceeding 10A. For 10A measurements, limit duration to less than 10 seconds to prevent overheating.
  • Do not connect test leads in parallel to a circuit when measuring current.
  • Do not input more than 36V DC or 25V AC between the current jack and "COM" jack.

5.5 AC Current Measurement (ACA)

  1. Follow steps 1-2 from DC Current Measurement.
  2. Rotate the function switch to the "A" position (ACA mode). Use the "SELECT" button to switch between AC and DC current.
  3. Turn off power to the circuit, then connect the multimeter in series with the circuit.
  4. Turn on power and read the AC current value.

Note:

  • All notes from DC Current Measurement apply.

5.6 Resistance Measurement (Ω)

  1. Insert the black test lead into the "COM" jack and the red test lead into the "V/Ω/Hz" jack.
  2. Rotate the function switch to the "Ω" position. The multimeter will automatically select the resistance range.
  3. Connect the test leads across the component to be measured.

Note:

  • "OL" indicates an open circuit or over-range.
  • For resistances above 1MΩ, readings may take a few seconds to stabilize.
  • For low resistance measurements, short the test leads first to measure lead resistance, then subtract this value from the measurement.
  • Always ensure the circuit is powered off and all capacitors are fully discharged before measuring resistance.
  • Do not input voltage into the resistance range.

5.7 Capacitance Measurement (C)

  1. Insert the black test lead into the "COM" jack and the red test lead into the "V/Ω/Hz" jack.
  2. Rotate the function switch to the "C" position. The multimeter will automatically select the capacitance range.
  3. Connect the test leads across the capacitor to be measured.

Note:

  • For 10nF range, there might be residual readings due to test lead capacitance; subtract this value for accurate readings.
  • Large capacitance measurements may show unstable values or take time to stabilize.
  • Always fully discharge capacitors before measurement to prevent damage.
  • Capacitance units: 1F = 1000mF, 1mF = 1000µF, 1µF = 1000nF, 1nF = 1000pF.

5.8 Frequency Measurement (Hz)

  1. Insert the test leads or shielded cable into the "COM" and "V/Ω/Hz" jacks.
  2. Rotate the function switch to the "Hz" position.
  3. Connect the test leads or cable across the signal source or load.

Note:

  • Input sensitivity: 1V RMS.
  • Overload protection: 550V DC or AC peak (not more than 10 seconds).
  • Do not input voltage symbols into this range.

5.9 Diode and Continuity Test

  1. Insert the black test lead into the "COM" jack and the red test lead into the "V/Ω/Hz" jack (note: red test lead is positive "+").
  2. Rotate the function switch to the "Diode" position. Use the "SELECT" button to switch between Diode and Continuity test modes.
  3. For Diode Test: Connect the test leads across the diode. The display shows the forward voltage drop. "OL" indicates an open circuit or reverse polarity.
  4. For Continuity Test: Connect the test leads across the circuit. A buzzer will sound if the resistance is below approximately 50Ω.

Note:

  • Do not input more than 250V DC or AC peak voltage.

5.10 Data Hold and Backlight

  • Press the "HOLD" button briefly to hold the current reading on the display. Press again to release.
  • Press and hold the "HOLD" button for 3 seconds to turn the backlight ON. Hold for another 3 seconds to turn it OFF. The backlight will automatically turn off after 15 seconds if not manually turned off.

5.11 Auto Power Off (APO)

  • The multimeter will automatically power off after approximately 15 minutes of inactivity to save battery life.
  • To restart, turn the function switch to "OFF" and then to any other measurement function.
  • To disable APO, press and hold the "SELECT" button while turning the power switch ON. The "APO" symbol will disappear from the display.

6. Maintenance

The DT-17N is a precision instrument and should not be modified internally.

  • Keep the instrument waterproof, dustproof, and protected from drops.
  • Avoid storing or using the instrument in high temperature, high humidity, flammable, explosive, or strong magnetic field environments.
  • Clean the exterior with a damp cloth and mild detergent. Do not use abrasive cleaners or strong solvents like alcohol.
  • If not used for a long time, remove the battery to prevent leakage.

6.1 Fuse Replacement

When replacing the fuse, use only the specified type and rating.

7. Troubleshooting

If the instrument does not function correctly, refer to the table below for common issues. If the problem persists, contact customer support.

Failure Phenomenon Check and Methods
Not shown (display blank) Battery not connected / Replace the battery
Battery symbol appears Replace the battery
Current is not input Replace fuse
Display error Replace the battery

8. Technical Specifications

Accuracy is specified at (reading data of a% + least significant digits) under guaranteed accuracy environment temperature: (23.5)°C, Relative humidity <75%, calibration guarantee period from the date of manufacture for one year.

8.1 DC Voltage (DCV)

Range Accuracy Resolution
600mV±(0.5%+3)0.1mV
6V0.001V
60V0.01V
600V0.1V
1000V ±(0.8%+10)1V

Input impedance: 10MΩ. Overload protection: 200mV range is 550V DC or AC peak; the rest is 1000V DC or 750V AC peak.

8.2 AC Voltage (ACV) True RMS

Range Accuracy Resolution
600mV±(0.8%+5)0.1mV
6V1mV
60V10mV
600V100mV
750V ±(1.2%+10)1V

Input impedance: 10MΩ. Frequency response: 40Hz-1kHz for standard sine and triangular waves; 40Hz-200Hz for other waveforms.

8.3 DC Current (DCA)

Range Accuracy Resolution
600µA±(0.8%+10)0.1µA
6000µA0.001mA
60mA±(2.0%+30)0.01A
600mA0.1A
10A ±(2.0%+30)1A

Maximum measured pressure drop: 600mV. Overload protection: 600mA range with 600mA/250V fast-blow glass fuse; 10A range with 10A/250V ceramic fast-blow fuse.

8.4 AC Current (ACA) True RMS

Range Accuracy Resolution
600µA±(0.8%+10)0.1µA
6000µA0.001mA
60mA±(2.0%+30)0.01A
600mA0.1A
10A ±(2.0%+30)1A

Maximum measured pressure drop: 600mV. Overload protection: 600mA range with 600mA/250V fast-blow glass fuse; 10A range with 10A/250V ceramic fast-blow fuse. Frequency response: 40Hz-1kHz for sine and triangular waves; 40Hz-200Hz for other waveforms.

8.5 Resistance (Ω)

Range Accuracy Resolution
600Ω±(0.8%+5)0.1Ω
6kΩ1Ω
60kΩ10Ω
600kΩ±(0.8%+3)100Ω
6MΩ1kΩ
60MΩ ±(1.0%+25)10kΩ

Open circuit voltage: less than 3V. Overload protection: 550V DC or AC peak.

8.6 Capacitance (C)

Range Accuracy Resolution
60nF±(3.5%+20)10pF
600nF100pF
6µF1nF
60µF10nF
600µF±(5.0%+10)100nF
6000µF1µF

Overload protection: 550V DC or AC peak.

8.7 Frequency (Hz)

Range Accuracy Resolution
10Hz±(0.1%+3)0.01Hz
100Hz0.1Hz
1kHz1Hz
10kHz10Hz
100kHz100Hz
1MHz/20MHz1kHz/10kHz

Input sensitivity: 1V RMS. Overload protection: 550V DC or AC peak (not more than 10 seconds).

8.8 Temperature (℃/℉)

Range Accuracy Resolution
-20 to 1000℃±(1.0%+5) <400℃
±(1.5%+15) ≥400℃
1℃
0 to 1832℉±(0.75%+5) <750℉
±(1.5%+15) ≥750℉
1℉

9. User Tips

  • The DT-17N is an automatic digital multimeter, meaning it features auto-ranging for most measurements. This simplifies operation as you don't always need to manually select the range.
  • Remember that the 9V battery is not included with the multimeter. Ensure you have one ready for immediate use.
  • For electronics repair, this multimeter is highly recommended due to its precision and ease of use.
  • The integrated test lead slots on the side of the multimeter are useful for keeping your workspace organized.

10. Frequently Asked Questions (FAQ)

Q: Is the multimeter new and what does the package include?
A: Yes, the product is new. The package typically includes the multimeter, test leads, a user manual, and a certificate. A 9V battery is not included.
Q: Does it come with the cables/tips?
A: Yes, the multimeter usually comes with a pair of standard test leads (cables/tips).
Q: What is the measurement range for capacitance?
A: The capacitance measurement ranges from 60nF to 6000µF (6mF). Please refer to the detailed specifications in Section 8.6 for specific ranges and resolutions.

11. Warranty and Support

This product is designed and manufactured to high standards. For any issues not covered in the troubleshooting section or for warranty inquiries, please contact your retailer or the manufacturer's customer service. Refer to the product packaging or included documentation for specific contact information.

The manufacturer, Guangzhou Sunshine Electronics Technology Co.,Ltd., holds all rights reserved for this product. Production standards meet industry and SUNSHINE quality inspection standards.

Service Hotline: 400-038-0255

E-Mail: sunshine01@exinxun.com


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.