1. Introduction

This manual provides detailed instructions for the proper use and maintenance of the SNDWAY SW-523 and SW-524 Sound Level Meters. Please read this manual thoroughly before operating the device to ensure optimal performance and accurate measurements.

This device can be used in noise projects, quality control, healthy prevention and treatment, and various circumstances such as industrial, office, traffic, home, and acoustic environments.

Note: The manufacturer reserves the right to update and amend the design specifications and user manual without prior notice. Materials and specifications are subject to change.

2. Safety Precautions and Attentions

  • Do not use the device in high temperature or humid environments.
  • Ensure that the ambient noise level is below the noise level being tested for accurate readings.
  • Remove batteries if the device will not be used for an extended period to prevent electrolyte leakage and potential damage to the product.
  • When using the meter outdoors, attach the sponge ball (windshield) to the microphone to prevent interference from other noises.
  • Optimal operating environment conditions: Altitude below 2000 meters, relative humidity ≤ 80% RH, ambient temperature 0°C to 40°C (indoor use).
  • Dispose of used batteries according to local laws and regulations.

3. Key Features

  • Complies with IEC651 TYPE2 of The National Committee and The American National standard ANSI S1.4 TYPE2.
  • Measurement range: 30dB to 130dB.
  • Accuracy: ±1.5dB.
  • Two characteristic choices: A-weighting (simulates human ear response) and C-weighting (for mechanical noise) response.
  • Two reaction rate choices: FAST and SLOW.
  • AC and DC analog signal output can be connected to a frequency analyzer or X-Y axis recorder for statistical data analysis.
  • Automatic shutdown after approximately ten minutes of inactivity.
  • 1.62 inch display.
  • Double injection molding case for durability.
  • **SW-524 Specific Features:**
    • USB cable connection for computer communication.
    • Supports downloading and storing records.
    • Real-time sampling, printing icons and data.
    • Data storage: 60 groups per minute, continuous storage for up to 17 hours, coexistence of data up to 61200 groups.
    • Time display.
    • Alarm upper limit can be set.

4. Product Overview and Parts Description

Refer to Figure 1 for a visual representation of the sound level meter's components.

Labeled diagram of the SNDWAY SW-524 Sound Level Meter showing its components and dimensions.
Figure 1: SNDWAY SW-524 Sound Level Meter Components and Dimensions

This diagram illustrates the main parts of the device, including the sponge ball, condenser microphone sensor, 1.62-inch display, various control keys (Max/Min, A/C, Fast/Slow, Backlight, Power), and output ports. Dimensions are also indicated.

  1. Capacitor microphone
  2. Display screen
  3. Power on/off button
  4. Shift button: For selecting measurement ranges (30-80dB, 50-100dB, 80-130dB, 30-130dB)
  5. Max/Min button
  6. Frequency weighting (A/C) button: 'A' for sound volume felt by humans, 'C' for mechanical noise.
  7. Time weighting (FAST/SLOW) button: 'FAST' (125ms) for most situations, 'SLOW' (1s) for more stable readings.
  8. Backlight control key
  9. AC analog signal output port
  10. DC analog signal output port
  11. Sponge ball (windshield)
  12. Hole for adjustment (calibration potentiometer)
  13. Battery cover

LCD Display Illustration

Refer to Figure 2 for an explanation of the LCD display elements.

LCD display of the sound level meter showing various indicators.
Figure 2: LCD Display Elements (Example)

This image shows the digital display with indicators for measurement range, power, sound value, unit (dB), A/C weighting, analog scale, OVER/UNDER warning signals, FAST/SLOW response, and MAX/MIN values.

  1. Measuring Range
  2. Power indicator
  3. Sound value
  4. Unit (dB)
  5. A/C weight indicator
  6. Analog scale
  7. OVER warning signal (value beyond max of scale)
  8. SLOW (react slowly)
  9. FAST (react fastly)
  10. UNDER warning signal (value below min of scale)
  11. MAX value
  12. MIN value

5. Setup and Preparation

5.1 Battery Installation

  1. Before first use, please open the battery cover on the back of the product.
  2. Insert three AAA batteries correctly, observing polarity.
  3. Replace the battery cover securely.
  4. When the power signal on the LCD screen indicates low battery, replace the batteries promptly.
Back view of the sound level meter with battery compartment open, showing battery installation.
Battery Compartment (Example)

This image shows the general location of the battery compartment on the back of the device for battery installation.

5.2 Calibration (Adjustment)

The product is factory-calibrated before shipping. A recalibration period of one year is recommended. To adjust:

  1. Set the device to A-weighting, FAST response, and 30-130dB shift.
  2. Carefully insert the microphone head into the hole of a 94dB@1KHz standard sound source.
  3. Turn on the sound source.
  4. Using a slot-type screwdriver, adjust the potentiometer located inside the small hole at the bottom (as shown in Figure 1, item 13) until the LCD displays "94.0".

6. Operation

  1. Press the power key to turn on the device.
  2. Use the ▲ or ▼ buttons to select a suitable measurement range for the current noise level. This prevents "UNDER" or "OVER" indications on the display.
  3. Choose 'dBA' for noise perceived by humans or 'dBC' for mechanical noise, using the A/C button.
  4. Select 'FAST' for current readings or 'SLOW' for average readings, using the FAST/SLOW button.
  5. Press the 'MAX/MIN' button to display the maximum or minimum measured value.

6.1 Data Logging and Computer Communication (SW-524 Only)

The SW-524 model supports data logging and communication with a computer via a USB cable.

USB cable connection to computer for data logging with the SW-524 Sound Level Meter.
USB Cable Connection and Data Logging Features

This image illustrates the USB connection for the SW-524 model, enabling data download, real-time sampling, and record printing. The device can store 60 groups of data per minute, continuously for up to 17 hours, with a total capacity of up to 61200 data groups.

Connect the SW-524 to your computer using a USB cable. The device supports downloading and storing measurement records, real-time sampling, and printing data. It can store data every minute, offering continuous storage for up to 17 hours, with a total capacity for up to 61200 data groups.

Note: Software for data logging may be required and is typically provided by the manufacturer or available for download. Refer to the official product page or contact support for software details.

7. Multi-Domain Applications

The SNDWAY Sound Level Meter is versatile and can be used in various settings for noise detection and analysis:

Grid of images showing various applications of the sound level meter.
Figure 3: Multi-Domain Applications

This image displays the diverse uses of the sound level meter, including sound insulation material inspection, office noise detection, library noise detection, home appliance noise detection, car noise detection, and machining noise detection.

  • Sound insulation material inspection
  • Office noise detection
  • Library noise detection
  • Home appliance noise detection
  • Car noise detection
  • Machining noise detection
  • And other acoustic measurement needs.

8. Technical Specifications

Below is a detailed comparison of the specifications for the SW-523 and SW-524 models.

Comparison table of specifications for SNDWAY SW-523 and SW-524 Sound Level Meters.
Figure 4: SW-523 vs. SW-524 Specifications

This image provides a side-by-side comparison of the technical specifications for both models, highlighting their similarities and differences, particularly regarding data logging and alarm features.

Feature SW-523 SW-524
Accuracy ±1.5dB ±1.5dB
Frequency Response 31.5Hz ~ 8.5KHz 31.5Hz ~ 8.5KHz
AC, DC Signal Output Yes Yes
Measurement Range 30 ~ 130dB 30 ~ 130dB
A/C Weighting Yes Yes
Resolution 0.1dB 0.1dB
Sampling Frequency 2 times/second 2 times/second
Microphone 6mm condenser microphone 6mm condenser microphone
Battery 3x 1.5V AAA 3x 1.5V AAA
USB Cable Connection No Yes
Data Storage No Yes
Time Display No Yes
Alarm Upper Limit Setting No Yes
Dynamic State Range 50dB 50dB
Numerical Digit Display 4 digits, resolution 0.1dB 4 digits, resolution 0.1dB
Dynamic Scale Display 1 scale = 2dB or 1dB 1 scale = 2dB or 1dB
Dynamic Scale Shift 30-130dB, 30-80dB, 50-100dB, 80-130dB 30-130dB, 30-80dB, 50-100dB, 80-130dB
Overload Display "OVER" and "UNDER" "OVER" and "UNDER"
AC Signal Output 0.65Vrms for each shift and full scale, impedance ~600ohm 0.65Vrms for each shift and full scale, impedance ~600ohm
DC Signal Output 10mV/dB, impedance ~100ohm 10mV/dB, impedance ~100ohm
Working Temperature & Humidity 0°C-40°C, 10%-80%RH 0°C-40°C, 10%-80%RH
Storage Temperature & Humidity -10°C-60°C, 10%-70%RH -10°C-60°C, 10%-70%RH
Is Smart Device Yes Yes
Battery Included No No
High-concerned chemical None None
Certification CE, FCC, RoHS, UL, KC, pse, weee CE, FCC, RoHS, UL, KC, pse, weee

9. Maintenance

  • Keep the device clean and free from dust and debris. Use a soft, dry cloth for cleaning.
  • Avoid exposing the device to extreme temperatures, humidity, or direct sunlight.
  • Store the device in a cool, dry place when not in use.
  • As mentioned in the "Safety Precautions" section, remove batteries if storing for extended periods to prevent leakage.
  • Regularly check the battery compartment for any signs of corrosion.
  • Recalibration is recommended annually to maintain accuracy.

10. Troubleshooting

Problem Possible Cause Solution
Device does not turn on Low or dead batteries, incorrect battery installation. Replace batteries with new AAA batteries, ensure correct polarity.
"UNDER" or "OVER" displayed Measurement range is too narrow or too wide for the current noise level. Adjust the measurement range using the ▲ or ▼ buttons until the reading is within the displayable range.
Inaccurate readings Interference from wind (outdoor use), device needs calibration, high humidity/temperature. Use the sponge ball (windshield) outdoors. Perform calibration as described in Section 5.2. Ensure operating conditions are within specified limits.
SW-524: Data logging not working Incorrect USB connection, missing or incompatible software. Ensure USB cable is securely connected. Install the correct data logging software for your computer (check manufacturer's website or contact support).

11. User Tips

  • For consistent and reliable measurements, especially in professional settings, consider annual calibration.
  • If using the SW-524 model for data logging, ensure you have the appropriate software installed on your PC. This software is essential for downloading, analyzing, and managing the recorded data.
  • When comparing noise levels, always use the same weighting (A or C) and response time (FAST or SLOW) settings for consistency.

12. Frequently Asked Questions (FAQ)

Q: Does the device come with a calibration certification?
A: The product is factory-calibrated before shipment. While a separate certification might not be included with every unit, the device is adjusted to standard specifications. Annual recalibration is recommended to maintain accuracy.
Q: Is software included for the PC for data logging (SW-524 model)?
A: The SW-524 model supports USB connection for data logging. Specific software for PC communication and data analysis is typically required. Please check the manufacturer's official website or contact customer support for information on obtaining the compatible software.

13. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the SNDWAY Official Store or your authorized dealer. Keep your purchase receipt as proof of purchase.

The manufacturer reserves the right to make changes to the product design and specifications without prior notice.

Certifications: CE, FCC, RoHS, UL, KC, pse, weee.

14. Product Video Overview

Video: SNDWAY Sound Level Meter Product Overview

This video provides a visual overview of the SNDWAY Sound Level Meter, demonstrating its design, features, and general appearance.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.