1. Introduction
The JINGYAN MESR-100 V2 is an auto-ranging capacitor ESR and Low Ohm Meter designed for measuring equivalent series resistance (ESR) and low resistance values. It supports in-circuit testing and uses a true 100 KHz sine wave for accurate ESR measurement, aligning with capacitor manufacturers' testing methods.
Unlike some techniques that use short pulse methods, which can yield varying values depending on capacitance, the MESR-100 V2 provides consistent readings.
What is ESR of a Capacitor?
ESR (Equivalent Series Resistance) refers to the internal series resistance within a capacitor. By using a 100kHz test frequency, the impedance 1/(2*π*F*C) becomes small, allowing for the accurate measurement of the true series resistance. A high ESR value in an electrolytic capacitor often indicates a faulty or aged component, leading to increased ripple and reduced filtering effectiveness. Generally, a large capacitor with an ESR greater than 3 ohms is considered problematic.
2. Safety Information
WARNING: DISCHARGE CAPACITORS BEFORE TESTING!
Always discharge capacitors before testing with the MESR-100/M6013. You can use a screwdriver to short circuit the capacitor terminals or connect a 10-ohm resistor in series and short for 5 to 10 seconds. While the meter has internal fast diodes to prevent high voltage damage, a large surge current and voltage from an undischarged capacitor can still damage the meter.
3. Features and V2 Improvements
The MESR-100 V2 model incorporates several improvements over its predecessor (V1):
- Test Signal: Changed from square wave to a true 100 KHz sine wave to reduce high-frequency components that could affect readings.
- Resolution: Higher resolution up to 0.001 ohm.
- Display: 128x64 dot matrix LCD for larger value display and more information.
- ESR Reference Table: Embedded 25V capacitor table on the LCD for quick reference to determine if a capacitor is good or bad based on common 25V electrolytic capacitor values.
- Design: New plastic case with a curved design for comfortable hand-carrying. Includes a new stand for a 60-degree viewing angle on a desk.
- Power: Uses 2x AA 1.5V batteries, offering more convenience and longer battery life compared to 9V batteries.
- External Power: Supports external 5V power via a standard micro-USB port.
4. Setup
Battery Installation
The device requires 2x AA 1.5V batteries (not included). Open the battery cover (located at the bottom rear of the device) and insert the batteries, observing correct polarity.
External Power (Optional)
The meter can be powered via a standard micro-USB cable (5V) without batteries. Connect the micro-USB cable to the port on the side of the device.
Using the Stand
The device features a built-in plastic stand for convenient desktop use. To deploy the stand, pull it out from the back. To lock it back in place, press and push it into the lock position.
5. Operating Instructions
5.1 Power On/Off
- To power on, press and hold the orange ON/OFF button for 2 to 3 seconds.
- To shut down the meter, press and release the ON/OFF button.
5.2 Backlight
The LCD backlight will be on continuously when the device is powered on.
5.3 Zeroing the Test Leads (1 Key Zero)
- Before taking measurements, short circuit the test leads' terminals (red and black clips).
- Press and release the 'ZERO' button. The LCD will display 'ZERO' and then clear. This cancels the wire resistance of the test leads.
- If using the meter array socket, use a short pin to short circuit the socket for zeroing.
5.4 Auto/Manual Ranging
- Auto Range: Press and release the 'MODE RANGE' button. The first line of the LCD will show 'AUTO:'. In auto mode, the meter automatically selects the best range for detection.
- Manual Range: Press and release the 'MODE RANGE' button repeatedly to scroll through manual ranges: 0-1R, 1-10R, and 10-100R. The LCD will show 'MANUAL' on the first line and the selected range on the second line.
5.5 Taking Measurements
After zeroing and selecting the desired range, connect the test clips to the component you wish to measure. Ensure the component is discharged if it's a capacitor.
6. Specifications
| Parameter | Value |
|---|---|
| Measuring Range | 0.001 to 100.0R |
| Accuracy (0.000 to 1.000R) | 1% + 2 Digits |
| Accuracy (1.000 to 10.00R) | 1% + 1 Digit |
| Accuracy (10.00 to 100.0R) | 2% + 1 Digit |
| Refresh Time (Manual Mode) | ~0.4s |
| Refresh Time (Auto Mode) | 0 to 2s (depends on value) |
| Wide Measurement Range | >1uF (error larger for 0.1uF at 100KHz) |
| High Resolution | 4 digit, or 0.001 Ohm @ 1 Ohm range |
| Measuring Voltage (TESTING VOLTAGE) | <~40mV RMS |
| Clamping Voltage (Open Voltage) | ~0.15V |
| Battery | 2x AA 1.5V battery (not included) |
| External Power | 5V micro USB |
| Operating Current | 0.02A |
| Battery Life Time | >80 Hours |
Note on Accuracy: Accuracy may be affected by the test lead's length and contact point resistance. Standard test leads are 15cm. For longer test leads, special shielding may be required to mitigate EMI interference.
7. ESR Reference Table
The MESR-100 V2 features an embedded ESR reference table on its casing for quick comparison. This table provides typical worst-case ESR values for electrolytic capacitors at various voltages and capacitances. It serves as a general guide; actual good ESR values may vary slightly between manufacturers.
To use the table, find the capacitance and voltage rating of the capacitor you are testing. Compare your measured ESR value to the corresponding value in the table. If your measured ESR is significantly higher (e.g., several times larger) than the table's reference, the capacitor may be faulty.
| Capacitance | 10V | 16V | 25V | 35V | 63V | 100V | 250V |
|---|---|---|---|---|---|---|---|
| 1uF | 14 | 16 | 18 | 20 | |||
| 2.2uF | 6.0 | 8.0 | 10 | 10 | 18 | ||
| 4.7uF | 15 | 7.5 | 4.2 | 2.3 | 5.0 | ||
| 10uF | 8.0 | 5.3 | 3.2 | 2.4 | 3.0 | 2.5 | |
| 22uF | 5.4 | 3.6 | 2.1 | 1.5 | 1.5 | 1.5 | 1.8 |
| 47uF | 2.2 | 1.6 | 1.2 | 0.68 | 0.56 | 0.7 | 0.8 |
| 100uF | 1.2 | 0.7 | 0.32 | 0.32 | 0.3 | 0.15 | 0.8 |
| 220uF | 0.6 | 0.33 | 0.23 | 0.17 | 0.16 | 0.09 | 0.5 |
| 470uF | 0.24 | 0.18 | 0.12 | 0.09 | 0.09 | 0.05 | 0.3 |
| 1000uF | 0.12 | 0.09 | 0.08 | 0.07 | 0.05 | 0.06 | |
| 4700uF | 0.23 | 0.20 | 0.12 | 0.08 | 0.04 | ||
| 10000uF | 0.12 | 0.08 | 0.06 | 0.04 |
8. Troubleshooting
Overflow (OL) Display
If the display shows 'OL' (Overflow), it indicates that the measured value is out of the meter's range. First, ensure that the test leads are properly zeroed. If the issue persists, the component's value might be too high for the selected range, or the component is open/damaged.
Auto Sleep
The device will automatically shut down after approximately 10 hours of inactivity (no testing) to conserve battery power.
9. Maintenance
Battery Replacement
When the battery indicator on the LCD shows low power, replace the 2x AA 1.5V batteries. Open the battery cover on the back of the device to access the battery compartment.
Cleaning
Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
Storage
Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
10. User Tips
- Short Test Leads: For optimal accuracy, especially when measuring very low ESR values, keep the test leads as short as possible. Longer leads may introduce resistance and EMI interference.
- In-Circuit Testing: The MESR-100 V2 is designed for in-circuit testing, making it highly efficient for diagnosing faulty capacitors without desoldering them. This is particularly useful for repairing circuit boards in TVs, LCDs, and audio equipment.
- Compare with Known Good Capacitors: Since ESR values can vary by manufacturer, it's a good practice to test a known good capacitor of the same type and rating to establish a baseline for comparison.
- Small Capacitors: The included clips might be too large for very small capacitors. Consider using finer probes or alternative connection methods for these components.
11. Warranty and Support
For detailed warranty information and technical support, please refer to the official documentation provided by meterk. A user manual in PDF format is available for download from the manufacturer's website.
PDF Manual Reference: The official user manual (MESR100 V2 User Guide V1.0) is available in PDF format. For comprehensive details, please consult this document.