1. Introduction
This manual provides detailed instructions for the safe and effective use of your GVDA Digital Clamp Meter (Models GD112A, GD112B, and GD112C). Please read this manual thoroughly before operation and keep it for future reference.
2. Safety Information
Warning: Improper use of this meter can cause damage, electric shock, injury, or death. Read and understand all safety information before operating the meter.
2.1 General Safety Guidelines
- Always ensure the meter is in good working condition. Do not use the meter if it appears damaged or if the battery cover is open.
- Do not apply more than the rated voltage (600V AC/DC) between terminals or between any terminal and ground.
- Use extreme caution when working with voltages above 60V DC or 30V AC RMS, as these pose a shock hazard.
- Keep fingers behind the probe barriers during measurements.
- Disconnect the test leads from the circuit before changing functions.
- Replace the battery immediately when the low battery indicator appears.
- Do not operate the meter in explosive gas, vapor, or dust environments.
- Adhere to local and national safety codes.
2.2 Safety Symbols
- ⚠ Warning: Important safety signs.
- ⚡ Danger: High voltage.
- ☑ Double Insulation: Protection against electric shock.
- CAT III 600V: Overvoltage protection category III, rated for 600 Volts.
3. Product Overview
The GVDA Digital Clamp Meter is a portable, professional measuring instrument with a digital display. It features auto-ranging, True RMS measurement, NCV detection, and a flashlight.

Figure 3.1: GVDA GD112B Digital Clamp Meter with test leads.
3.1 Instrument Panel Description
- Clamp Jaw: For AC/DC current measurement.
- NCV Sensor: Non-contact AC voltage detection.
- Flashlight: Illuminates the measurement area.
- LCD Display: Shows measurement readings and indicators.
- Function Rotary Switch: Selects measurement functions.
- MAX/MIN Button: Toggles between maximum and minimum value display.
- FUNC Button: Selects sub-functions (e.g., AC/DC, Hz/Duty).
- HOLD/Backlight Button: Freezes the display reading; long press activates/deactivates backlight.
- NCV/Flashlight Button: Activates NCV detection; long press activates/deactivates flashlight.
- COM Jack: Common input for test leads (black).
- INPUT Jack: Positive input for test leads (red).
3.2 Key Features (Model Dependent)

Figure 3.2: GD112A Basic Model features including 6000 counts, AC/DC Voltage, AC Current, Resistance, Max/Min, Continuity, Auto Power Off, Data Hold, Flashlight, Double Backlight.
- GD112A: 6000 Counts, AC/DC Voltage, AC Current, Resistance, Max/Min, Continuity, Auto Power Off, Data Hold, Flashlight, Double Backlight.
- GD112B & GD112C: Include all GD112A features plus DC Current, Capacitance, Frequency, Duty Cycle, Temperature Measurement, and NCV Detection. GD112C has higher current range.
4. Setup
4.1 Battery Installation
Your clamp meter requires two 1.5V AAA batteries (not included). Ensure correct polarity when installing.

Figure 4.1: Battery compartment for two 1.5V AAA batteries.
- Turn off the meter and disconnect all test leads.
- Use a screwdriver to open the battery compartment cover on the back of the meter.
- Insert two 1.5V AAA batteries, observing the correct polarity (+ and -).
- Replace the battery cover and secure it with the screw.
5. Operating Instructions
Before any measurement, ensure the rotary switch is set to the OFF position and test leads are correctly connected.
5.1 AC/DC Voltage Measurement
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "V~" (AC Voltage) or "V–" (DC Voltage) position. Use the FUNC button to switch between AC and DC if combined.
- Connect the test probes in parallel to the circuit or component under test.
- Read the voltage value on the LCD display.
Caution: Do not measure voltages exceeding the maximum input rating (600V). Be extremely careful when measuring high voltages.
5.2 AC/DC Current Measurement
- Set the rotary switch to the "600A" or "60A" (AC/DC Current) position. Use the FUNC button to switch between AC and DC.
- Open the clamp jaw and enclose only one conductor of the circuit. Ensure the conductor is centered within the jaw.
- Read the current value on the LCD display.
Warning: Do not clamp around multiple conductors when measuring current, as this will result in an incorrect reading.
5.3 Resistance Measurement
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "Ω" (Resistance) position.
- Connect the test probes across the component to be measured. Ensure the circuit is de-energized before measuring resistance.
- Read the resistance value on the LCD display.
Caution: Do not measure resistance on live circuits.
5.4 Continuity Test
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "Ω" (Resistance/Continuity) position and press FUNC until the continuity symbol (♫) appears.
- Connect the test probes across the circuit or component.
- If the resistance is below approximately 50Ω, the buzzer will sound, indicating continuity.
5.5 Diode Test
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "Ω" (Resistance/Diode) position and press FUNC until the diode symbol (△) appears.
- Connect the red probe to the anode and the black probe to the cathode of the diode.
- The display will show the forward voltage drop. Reverse the probes; an open circuit (OL) reading indicates a good diode.
5.6 Capacitance Measurement (GD112B/C only)
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "CAP" (Capacitance) position.
- Connect the test probes across the capacitor. Ensure the capacitor is fully discharged before measurement.
- Read the capacitance value on the LCD display.
Warning: Discharged capacitors can still hold a charge. Always discharge them before testing.
5.7 Non-Contact AC Voltage Detection (NCV) (GD112B/C only)

Figure 5.1: NCV detection in progress, indicating the presence of AC voltage without contact.
- Press the NCV button. The NCV indicator will light up.
- Move the top part of the meter (NCV sensor) close to the AC voltage source (e.g., wire, outlet).
- If AC voltage is detected, the meter will beep and the LED indicator will flash. The intensity of the beeping and flashing increases with stronger signals.
5.8 Temperature Measurement (GD112B/C only)

Figure 5.2: Temperature measurement using a K-type thermocouple probe.
- Set the rotary switch to the "Temp" (Temperature) position.
- Insert the K-type thermocouple probe into the INPUT (positive) and COM (negative) jacks, observing polarity.
- Place the tip of the thermocouple on or in the object whose temperature is to be measured.
- Read the temperature on the LCD display. Press FUNC to switch between Celsius (℃) and Fahrenheit (℉).
5.9 Frequency/Duty Cycle Measurement (GD112B/C only)
- Insert the red test lead into the INPUT jack and the black test lead into the COM jack.
- Set the rotary switch to the "Hz%" (Frequency/Duty Cycle) position.
- Connect the test probes across the signal source.
- Read the frequency (Hz) or duty cycle (%) on the LCD display. Press FUNC to toggle between them.
5.10 Flashlight and Backlight

Figure 5.3: Integrated LED flashlight for illuminating work areas.
- Flashlight: Long press the NCV/Flashlight button to turn the LED flashlight on or off.
- Backlight: Long press the HOLD/Backlight button to turn the LCD backlight on or off.
5.11 Data Hold
Press the HOLD button briefly to freeze the current reading on the display. Press it again to release the hold function.
5.12 Auto Power Off
The meter will automatically power off after approximately 15 minutes of inactivity to conserve battery life. Press any button or turn the rotary switch to wake it up.
6. Maintenance
6.1 General Maintenance
- Keep the meter dry. If it gets wet, wipe it dry immediately.
- Use a soft, dry cloth to clean the case. Do not use abrasives or solvents.
- Store the meter in a dust-free environment away from direct sunlight and extreme temperatures.
- Remove batteries if the meter is not used for an extended period to prevent leakage.
6.2 Battery Replacement
When the low battery indicator appears on the display, replace the batteries as described in Section 4.1. Failure to do so may result in inaccurate readings.
7. Troubleshooting
If the meter does not function correctly, check the following:
- No display or dim display: Check battery installation and replace batteries if necessary.
- Incorrect readings: Ensure test leads are properly connected, the correct function is selected, and the circuit is de-energized (for resistance/capacitance).
- "OL" displayed: Indicates an over-range condition or an open circuit.
- Meter does not respond: Turn the meter off and then on again. If the issue persists, remove and reinsert batteries.
If problems persist, contact customer support.
8. Specifications
| Parameter | Specification |
|---|---|
| Maximum Display | 6000 Counts |
| DC Voltage (GD112B/C) | 6V/60V/600V |
| AC Voltage (GD112B/C) | 600mV/6V/60V/600V |
| DC Current (GD112C) | 6A/60A/600A |
| AC Current | 6A/60A/600A |
| Resistance Range | 600Ω/6KΩ/60KΩ/600KΩ/6MΩ/60MΩ |
| Capacitance Range | 10nF/100nF/1uF/10uF/100uF/1mF/10mF/100mF |
| Operating Temperature | 0 - 40 ℃ |
| Storage Temperature | -10 - 50 ℃ |
| Power Supply | 2 x 1.5V AAA Batteries |
| Dimensions | 193mm x 73mm x 34mm |
| Jaw Size | 26mm |
| Safety Rating | EN61010-1, -2-030, EN61010-2-033, EN61326-1 CATIII 600V |
| Certifications | CE, FCC |
9. User Tips
- For temperature measurements, ensure you are using a K-type thermocouple probe, which is typically included with GD112B and GD112C models.
- The auto-off feature activates after 15 minutes of inactivity, which is great for battery saving. Just press any button to reactivate.
- When measuring high currents (up to 600A on GD112C), always ensure the single conductor is properly centered within the clamp jaw for accuracy.
- Remember to discharge capacitors before testing them to prevent damage to the meter or inaccurate readings.
10. Warranty and Support
This product is covered by a standard manufacturer's warranty. For warranty claims, technical support, or service inquiries, please refer to your purchase documentation or contact the seller directly. Keep your proof of purchase for any warranty-related requests.