1. Introduction
The HoldPeak HP-870N is a portable, 6000-digit auto-range and auto-power-off digital AC/DC clamp multimeter. It is designed for safe and accurate measurement of various electrical parameters. This instrument is ideally suited for professional use in fields, laboratories, and workshops, as well as for general home applications.
2. Safety Information
To ensure maximum personal safety during operation, please observe the following safety guidelines:
- When measuring voltage, ensure the instrument is not switched to current, resistance, diode, continuity, capacitance, or temperature ranges.
- Use extreme care when measuring voltages above 50V, especially from high-energy sources.
- Avoid making connections to 'live' circuits whenever possible.
- Before making resistance, diode, continuity, capacitance, or temperature tests, ensure the circuit under test is de-energized.
- Always ensure that the correct function and range are selected for the measurement.
- Exercise extreme care when using the instrument with a current transformer if an open circuit occurs.
- Ensure that the test leads and probes are in good condition with no damage to the insulation.
- Do not exceed the overload limits specified in the specifications.
- Before opening the battery cabinet cover, disconnect the test leads from any external circuit and set the selector switch to the 'OFF' position.
- Keep fingers behind the protection ring when measuring with the instrument leads.
- Change the battery when the 'BAT' symbol appears on the display to avoid incorrect data.
3. Package Contents
The package includes the following items:
- 1 x HoldPeak HP-870N Digital Clamp Meter
- 1 x Carrying Case (Protecting mask)
- 1 x Pair of Test Leads (Red and Black probes)
- 1 x Thermocouple (Yellow probe for temperature measurement)
- 1 x User Manual
- 1 x Packing Box
4. Product Overview
Familiarize yourself with the components of your HP-870N Digital Clamp Meter:
- Pliers Head (Clamp Jaw): Used for non-contact current measurement. Guaranteed to shut down in place to read the LCD measurement results.
- Function Rotary Switch: Selects the desired measurement function (e.g., Voltage, Resistance, Capacitance, Frequency, Current, Temperature).
- DH (Data Hold) / Backlight Key: Holds the current display reading or activates the backlight.
- MAX/MIN Keys: Displays the maximum or minimum measured value.
- SELECT Key: Toggles between different measurement modes within a single rotary switch position (e.g., AC/DC voltage, resistance/diode/continuity, frequency/duty cycle, °F/°C).
- REL (Relative Value) / Inrush Current Key: Used for relative measurements or to measure inrush current.
- Hz/Duty Key: Selects frequency or duty cycle measurement.
- LCD Monitor: Large liquid crystal display for clear readings.
- Input Jacks (T+, T-COM, VΩHz): Connection points for test leads and thermocouple.
5. Specifications
| Parameter | Range | Resolution | Accuracy | Frequency |
|---|---|---|---|---|
| General Specifications | ||||
| Display Type | Digital Only (6000 counts) | |||
| Range Select | AUTO | |||
| Jaw Caliber | 45mm | |||
| Dimensions | 225 x 97 x 40 mm | |||
| Weight | 450g (without batteries) | |||
| Operating Temperature | -20~1000°C | |||
| Data Hold | Yes | |||
| Back Light | Yes | |||
| True RMS | Yes | |||
| Relative Value | Yes | |||
| Diode Test | Yes | |||
| Inrush Current Test | Yes | |||
| Battery Included | No | |||
| High-concerned chemical | None | |||
| DC Voltage | ||||
| 600mV | 0.1mV | Rdg + 5digits 1.0% | ||
| 6V | 1mV | Rdg + 5digits 1.0% | ||
| 60V | 10mV | Rdg + 5digits 1.0% | ||
| 600V | 100mV | Rdg + 5digits 1.0% | ||
| 1000V | 1V | Rdg + 5digits 1.0% | ||
| AC Voltage (TRUE RMS) | ||||
| 600mV | 0.1mV | Rdg + 5basamak% 3.0 | 40~100Hz | |
| 6V | 1mV | Rdg + 5basamak% 1.5 | 40~400Hz | |
| 60V | 10mV | Rdg + 5basamak% 1.5 | 40~400Hz | |
| 600V | 100mV | Rdg + 5basamak% 1.5 | 40~400Hz | |
| 750V | 1V | Rdg + 5basamak% 2.5 | 40~400Hz | |
| AC Current | ||||
| 600A | 0.1A | 3.0% of rdg + 10 digits | 50-60Hz | |
| 1000A | 1A | 3.5% of rdg + 10 digits | 50-60Hz | |
| Resistance | ||||
| 600Ω | 0.1Ω | 1.0% of rdg + 8 digits | ||
| 6KΩ | 1Ω | 1.0% of rdg + 8 digits | ||
| 60KΩ | 10Ω | 1.0% of rdg + 8 digits | ||
| 600KΩ | 100Ω | 1.0% of rdg + 8 digits | ||
| 6MΩ | 1KΩ | 2.5% of rdg + 15 digits | ||
| 60MΩ | 10KΩ | 2.5% of rdg + 15 digits | ||
| Capacitance | ||||
| 9.999nF | 1pF | 2.5% of rdg + 15 digits | ||
| 99.99nF | 10pF | 2.5% of rdg + 15 digits | ||
| 999.9nF | 100pF | 2.5% of rdg + 15 digits | ||
| 9.999µF | 1nF | 2.5% of rdg + 10 digits | ||
| 99.99µF | 10nF | 2.5% of rdg + 10 digits | ||
| 999.9µF | 100nF | 3.0% of rdg + 20 digits | ||
| 9.999mF | 1µF | 5.0% of rdg + 25 digits | ||
| 99.99mF | 10µF | 5.0% of rdg + 25 digits | ||
| Frequency | 9.999Hz--9.999MHz | |||
6. Setup
6.1. Battery Installation
The HP-870N requires a 9V battery (not included) for operation. To install or replace the battery:
- Ensure the meter is turned OFF and all test leads are disconnected.
- Locate the battery compartment on the back of the meter.
- Unscrew the screw securing the battery compartment cover.
- Remove the cover.
- Insert a new 9V battery, observing the correct polarity.
- Replace the battery compartment cover and fasten the screw securely.
6.2. Test Lead Connection
- For most voltage, resistance, capacitance, frequency, diode, and continuity measurements, connect the black test lead to the 'T-COM' input jack and the red test lead to the 'VΩHz' input jack.
- For temperature measurements, connect the thermocouple to the 'T+' and 'T-' input jacks, ensuring correct polarity.
7. Operating Instructions
To perform a measurement, turn the rotary switch to the desired function. Use the 'SELECT' button to toggle between modes if a position has multiple functions.
7.1. Function Rotary Switch
The main rotary switch allows you to select the measurement type:
- OFF: Turns the meter off.
- V~: AC Voltage measurement.
- V=: DC Voltage measurement.
- Ω / Diode / Continuity: Resistance, Diode Test, and Continuity. Use the 'SELECT' button to switch between these.
- CAP: Capacitance measurement.
- Hz: Frequency measurement. Use the 'Hz/Duty' button to switch to Duty Cycle.
- 600A~ / 1000A~: AC Current measurement (via clamp).
- 600A= / 1000A=: DC Current measurement (via clamp).
- °F/°C: Temperature measurement. Use the 'SELECT' button to switch between Fahrenheit and Celsius.
7.2. Button Functions
- DH (Data Hold) / Backlight:
- Press briefly to freeze the current reading on the display. The 'DH' symbol will appear. Press again to release.
- Press and hold for more than 2 seconds to turn the display backlight ON or OFF. The backlight will automatically turn off after approximately 15 seconds.
- SELECT: In ranges with multiple functions (e.g., Ω/Diode/Continuity, V~/V=, °F/°C), press this button to cycle through the available modes.
- MAX/MIN: Press to enter MAX/MIN recording mode. The meter will display the maximum ('MAX') or minimum ('MIN') value measured since the mode was activated. Press again to toggle between MAX and MIN. Hold for 2 seconds to exit.
- Hz/Duty: When the rotary switch is in an AC Voltage or AC Current range, or the dedicated Hz range, press this button to switch between Frequency (Hz) and Duty Cycle (%) measurement.
- REL (Relative Value) / Inrush Current:
- Relative Value: Press briefly to store the current display value as a reference. The display will then show the difference between subsequent measurements and this reference. This is useful for comparing values.
- Inrush Current: In AC/DC Current measurement mode, press and hold for more than 2 seconds to activate the Inrush Current test. This measures the peak current during motor startup or similar events.
8. Maintenance
- Keep the meter clean and dry. Use a soft, damp cloth for cleaning; do not use abrasives or solvents.
- Store the meter in its carrying case when not in use to protect it from dust and moisture.
- If the meter is not used for a long period, remove the battery to prevent leakage.
- Regularly inspect test leads and probes for any damage to the insulation. Replace immediately if damaged.
9. Troubleshooting
- Display shows 'BAT': The battery is low. Replace the 9V battery as described in the Setup section.
- Display shows 'OL' or 'OVERLOAD': The measured value exceeds the selected range. Switch to a higher range or check if the circuit is within the meter's capabilities.
- Inaccurate readings for resistance, diode, continuity, capacitance, or temperature: Ensure the circuit under test is de-energized before measurement.
- No display or erratic readings: Check battery installation and charge. Ensure test leads are properly connected and undamaged.
- Measuring small DC currents (<1A) with the clamp: For better accuracy, wrap the wire multiple times through the clamp jaw and divide the displayed reading by the number of turns. For very small currents, a traditional multimeter (without clamp) might offer higher precision.
- Constant 6A reading or other unexpected current values: Ensure the clamp jaw is fully closed and there are no external magnetic fields interfering. If persistent, a 'zeroing' function might be needed, though not explicitly detailed for this model, it's a common feature on some clamp meters.
10. User Tips
- The 'REL' (Relative Value) function is very convenient for comparing components. Measure a known good component, press 'REL', then measure others to see the difference.
- When measuring inrush currents on a car, the meter can be very useful. It can also be used to check modern inverters or test plasma cutters.
- For large current measurements, avoid prolonged contact (more than 10-15 seconds) to prevent potential damage to the meter.
11. Warranty and Support
For any questions, technical assistance, or warranty claims, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.