1. Introduction
This manual provides essential instructions for the safe and effective use, maintenance, and sterilization of the Xceldent Dental Bur Holder Block Sterilizer Case. This product is designed for professional dental use only.
Disclaimer: This item may be intended for use by professional health care providers only. The seller may ask you to provide evidence that you are a qualified health care provider and are entitled to buy and use certain items in your country.
For a detailed user manual in PDF format, please refer to the official document: User manual (PDF)
2. Safety Warnings
- Only skilled dentists or qualified health care providers are allowed to use this product.
- Always sterilize the product before each use.
- Use this product strictly in accordance with its intended purpose as a dental instrument sterilizer case.
- Wear safety glasses to protect your eyes from potential hazards during handling or sterilization.
- Wear a dust protective mask if necessary to prevent inhaling dust.
- Dispose of the product if it is damaged or contaminated. Treat as medical waste when disposing.
3. Product Overview
The Xceldent Dental Bur Holder Block Sterilizer Case is designed for the disinfection and organization of various small dental tools, including burs, reamers, H-files, K-files, and gutta percha points.
3.1. Features
- Designed as disinfection trays for various small dental tools.
- Autoclavable under high temperature and pressure.
- Constructed from durable Aluminum material.
- Available in Silver, Blue, and Red colors.
3.2. Components
The product consists of a single dental sterilizer case.


4. Specifications
| Attribute | Value |
|---|---|
| Product Name | Dental Sterilizer Case |
| Model Name | Dental Bur Holder |
| Brand Name | Xceldent |
| Material | Aluminum |
| Colors Available | Silver, Blue, Red |
| Commodity Quality Certification | CE |
| Size | Moderate (Dimensions: Length 18cm, Width 23cm, Height 5cm) |
| Weight | 0.1 kg |
| Origin | Mainland China |

5. Setup and Usage
The Xceldent Dental Bur Holder Block Sterilizer Case is designed for holding and sterilizing dental instruments. It does not require complex setup.
5.1. Initial Preparation
- Before first use, and for each subsequent use, ensure the case is clean and free from debris.
- Place the dental burs, reamers, files, or other small tools securely into the designated slots within the case.

5.2. Sterilization Process
The case is autoclavable. Follow these steps for sterilization:
- After placing instruments, close the case securely.
- Place the product in a sterilization pack (or foil) and then onto a sterilization tray, or use a burs stand designed for autoclave sterilization.
- Autoclave under the following conditions:
- Temperature: 121 degrees Celsius for 20 minutes or more.
- Temperature: 126 degrees Celsius for 15 minutes or more.
- Do not use high-pressure steam sterilizers that heat more than 200 degrees Celsius, including the drying process.
- Ensure sterilization is complete before removing the case from the autoclave.
6. Maintenance and Storage
6.1. Cleaning
- After each use, wash the case with a medical cleaning agent and brush to remove foreign substances, adherent body fluids, and tissues.
- Rinse thoroughly to remove all cleaning agent residue.
- When cleaning with an ultrasonic cleaner, place the case on a stand to avoid deterioration of the working parts.
6.2. Inspection
- Before each use, inspect the case for any deformations, scratches, cracks, or tarnishes.
- If any damage is found, dispose of the case and do not use it.
6.3. Storage
- Avoid storing at high temperature, high humidity, and direct sunlight.
- Keep liquids away from the case.
- Store the case at room temperature in a clean, dry environment.
- Do not store the product under a germicidal lamp to avoid deterioration.
7. Troubleshooting
This section addresses common issues you might encounter with the Xceldent Dental Bur Holder Block Sterilizer Case.
7.1. Case not closing properly
- Ensure no instruments are protruding or misplaced, preventing the lid from seating correctly.
- Check for any debris or foreign objects obstructing the closing mechanism.
7.2. Discoloration or corrosion after sterilization
- Verify that the correct sterilization parameters (temperature, pressure, time) were used, as specified in Section 5.2.
- Ensure that the case was thoroughly cleaned and rinsed of all cleaning agents before sterilization.
- Avoid using high-pressure steam sterilizers that exceed 200 degrees Celsius.
If issues persist or you encounter problems not listed here, please contact customer support.
8. User Tips
No specific user tips were available from reviews or Q&A for this product at this time. Always follow the instructions in this manual for optimal performance and safety.
9. Warranty and Support
9.1. Legal Manufacturer
IMOON DENTAL EQUIPMENT CO., LTD.ROOM 341.NO.650, CUIZHU STREET, HUAWAN ROAD LIWAN DISTRICT GUANGZHOU GUANGDONG PROVINCE.CHINA.
TEL: +86-20-81514759 +86-18988965069
FAX:+86-20-81514759
EMAIL: JIMOONDENTAL@163.COM
9.2. EU Representative
EC REP Obelis s.a.Bd. Général Wahis 53,
1030 Brussels, Belgium
Always keep this document near at hand. In case of loss, please contact the dental department of the manufacturer for a copy.
The specifications, structures, and materials are subject to change without prior notice for improvement purposes. The contents of this instruction manual are subject to revision without prior notice.