1. Introduction
This manual provides essential information for the safe and effective installation, operation, and maintenance of your UXCELL Water Flow Sensor. This device is designed to detect the flow of liquid in a pipeline and activate a switch based on the presence or absence of flow. It is commonly used in water heaters, boilers, water dispensers, coffee machines, and various water treatment and cooling systems.
2. Safety & Warnings
- Read all instructions before use to ensure safety and prevent damage.
- Warning: Keep the device and its cord out of reach of children under 3 years old.
- Do not use the sensor in environments with overpressure or over-temperature conditions.
- Keep the sensor away from strong vibration and corrosive substances.
- Do not disassemble the sensor at will to avoid damage or failure. Unauthorized disassembly will void any potential support.
3. Product Overview
The UXCELL Water Flow Sensor utilizes a magnetic reed switch mechanism to detect liquid flow. It is available in several models, each designed for specific connection types and flow ranges.
Figure 1: Overview of Water Flow Sensor Models
Model Variations:
- SEN-CL43W6: Features G3/4 external thread diameter, suitable for flow ranges of 2.5-30 L/min.
- SEN-CL21F: Features G1/2 external thread diameter, suitable for flow ranges of 1.2-12 L/min.
- SEN-CL08C: Features barbed hose connections, suitable for flow ranges of 1.2-12 L/min.
Dimensions and Connections:
Figure 2: SEN-CL43W6 Dimensions (G3/4)
Figure 3: SEN-CL21F Dimensions (G1/2)
Figure 4: SEN-CL08C Dimensions (Barbed)
Figure 5: G3/4" Thread Detail
Figure 6: G1/2" Thread Detail
4. Setup
- Pipeline Direction: Install the sensor correctly according to the flow direction indicated by the arrow on the sensor body. Incorrect orientation will result in no signal output.
- Connection Size: Ensure the sensor's connection size (G1/2, G3/4, or barbed) matches your pipeline.
- Sealing: Ensure all joints are sealed tightly to prevent leakage. Use appropriate sealing tape or compounds if necessary.
- Filtration: A filter screen must be installed at the inlet of the sensor to prevent particles and debris from entering and damaging the sensor or affecting its accuracy.
- Wiring: Connect the power supply and signal wire properly before starting operation. The sensor typically has two wires for simple integration into your control circuit.
Figure 7: Sensor Wiring Detail
5. Operating Instructions
Once the sensor is correctly installed and wired, it operates automatically. When liquid flows through the sensor at or above its minimum flow rate, the internal magnetic reed switch will activate, providing a signal. When the flow stops or drops below the minimum threshold, the switch will deactivate.
- Ensure the system is powered on.
- Initiate liquid flow through the pipeline where the sensor is installed.
- The sensor will provide an output signal corresponding to the detected flow.
6. Maintenance
- Leak Inspection: Periodically check all connections and joints for any signs of leakage. Tighten or re-seal as necessary.
- Filter Cleaning: If an inlet filter is installed, inspect it regularly for debris accumulation. Clean or replace the filter as needed to maintain optimal flow and sensor performance.
- Connection Integrity: Ensure electrical connections remain secure and free from corrosion.
7. Troubleshooting
- No Signal Output:
- Verify the sensor is installed in the correct flow direction (check the arrow).
- Ensure there is sufficient water flow to activate the sensor (above the minimum flow rate).
- Check for blockages in the pipeline or a clogged inlet filter.
- Inspect wiring for proper connection and integrity.
- Confirm the power supply to the control circuit is active.
- Leakage at Connections:
- Ensure all threaded connections are tightened properly.
- Re-apply thread sealant (e.g., PTFE tape) if necessary.
- Check for cracks or damage to the sensor body or fittings.
- Sensor Not Activating Consistently:
- Clean the inlet filter to ensure unobstructed flow.
- Check for air bubbles in the liquid stream, which can interfere with flow detection.
- Ensure the liquid temperature and pressure are within the specified operating limits.
8. Specifications
| Feature | Specification |
|---|---|
| Product Name | Water Flow Sensor |
| Material | Brass (Shell) |
| Liquid Temperature | Not More Than 60°C |
| Max Switch Voltage | DC 250V |
| Max Contact Rating | 70W |
| Max Current | 10mA |
| Maximum Water Pressure | 1.75MPa |
| Ingress Protection Rating | IP65 |
| Color | Brass Tone |
Model-Specific Specifications:
| Type | External Thread Diameter | Flow Range | Size (L*W*H) | Net Weight |
|---|---|---|---|---|
| SEN-CL43W6 | G3/4 | 2.5-30 L/min | 62mm x 31mm x 31mm | 157g |
| SEN-CL21F | G1/2 | 1.2-12 L/min | 62mm x 28mm x 29mm | 153g |
| SEN-CL08C | Barbed | 1.2-12 L/min | 81mm x 21mm x 26mm | 115g |
9. User Tips
- Pre-Installation Check: Before installing, always confirm the sensor's flow range and connection type match your system's requirements.
- Flow Direction is Crucial: Pay close attention to the flow direction arrow on the sensor during installation. Incorrect orientation is a common reason for the sensor not providing an output signal.
- Protect from Debris: Always install a filter screen at the sensor inlet. This simple step prevents particles and debris from entering the sensor, which can cause blockages, damage, or inaccurate readings.
10. Warranty & Support
For any issues, technical support, or warranty inquiries, please contact your seller directly. Provide them with your purchase details and the specific model number of your sensor.
Manufacturer Information:
- Manufacturer: Guangzhou Dajiali Network Technology Co., Ltd.
- Address: B395, Room 106, Rulin Street, Tianhe, Guangzhou, China
- Email: udb7910@126.com
EU Authorized Representative:
- Representative: SUCCESS COURIER SL
- Address: CALLE RIO TORMESNUM.1, PLANTAI, DERECHAOFICINA 3Fuenlabrada, Madrid, 28947 Spain
- Email: successservice2@hotmail.com