ZHURUI EPM5800-E

Contents

ZHURUI EPM5800-E AC+DC Electrical Parameter Tester Instruction Manual

Model: EPM5800-E (Tri-color)

1. Introduction

The ZHURUI EPM5800-E is a versatile AC+DC Electrical Parameter Tester designed to measure various electrical parameters for both AC and DC circuits. It is ideal for testing power supplies, drivers, and overall efficiency, particularly useful for LED driver testing. This manual provides detailed instructions for its installation, operation, and maintenance.

Front view of the ZHURUI EPM5800-E Electrical Parameter Tester displaying AC and DC voltage, current, power, AC power factor/frequency, and efficiency.
Figure 1: Front view of the EPM5800-E Electrical Parameter Tester

2. Main Technical Specifications

Meter Parameters
ParameterValueDescription
Meter Power Supply VoltageAC 85V~265VOperating voltage for the meter itself.
Power Consumption<2WPower consumed by the meter.
Sampling Rate2 times/secondMeasurement update frequency.
Operating Frequency45~65HzFrequency range for AC measurements.
Display Mode8 pieces of 5.6-inch nixie tubesDigital display type.
Overall Dimensions128.0mm x 135.0mm x 32.0mmPhysical dimensions including the panel.
Temperature and Humidity0~50°C, ≤85% RHRecommended operating environment.
Installation MethodScrew InstallationMounting method.

AC Measurement Specifications

ParameterMeasurement RangeResolutionAccuracy
AC Power0.1W~2500W0.01W [2]0.5% FS
AC VoltageAC 5V~300V0.1V0.5% FS
AC CurrentAC 0.005A~10A0.001A0.5% FS
Power Factor0~1.0000.0010.5% FS
Frequency45.0Hz~65.0Hz0.1Hz0.5% FS

DC Measurement Specifications

ParameterMeasurement RangeResolutionAccuracy
DC Power0.1W~3000W0.01W [2]0.5% FS
DC VoltageDC 1V~300V0.01V0.5% FS
DC CurrentDC 0.005A~10A0.001A0.5% FS

AC & DC Measurement Specifications (Efficiency)

ParameterMeasurement RangeResolutionAccuracy
Efficiency0~100%0.10.5% FS

Notes:
[1] This dimension includes the panel; the depth without the panel is 29mm.
[2] When the load power is less than 100W, the measurement resolution is 0.01W; when the power is 100W or more, the measurement resolution is 0.1W; and when the power is 1000W or more, the measurement resolution is 1W.

3. Panel Layout

The EPM5800-E features a clear display panel with dedicated sections for AC and DC electrical parameters, as well as efficiency. The panel is designed for embedded installation.

Side and back views of the EPM5800-E tester, showing the acrylic panel, screws, and PCB.
Figure 2: Side and Back Views of the EPM5800-E
Detailed view of the EPM5800-E's PCB (Printed Circuit Board) from the back.
Figure 3: EPM5800-E PCB (Back View)

The front panel displays the following:

  • Left Side (AC Parameters): AC Voltage (V), AC Current (A), AC Power (W), AC Power Factor/Frequency (Hz).
  • Right Side (DC Parameters): DC Voltage (V), DC Current (A), DC Power (W), Efficiency (%).

A button labeled "PF/FREQ" is located at the bottom center of the AC display section to switch between Power Factor and Frequency display.

4. Wiring Instructions (Viewed from the Back)

Proper wiring is crucial for the safe and accurate operation of the EPM5800-E. Refer to the diagram below and follow these steps carefully:

Wiring diagram for the ZHURUI EPM5800-E, showing connections for AC power supply, AC measurement, DC measurement, and DC load.
Figure 4: EPM5800-E Wiring Diagram (Back View)
  1. Meter Power Supply: Terminals 1 and 2 on the far left side of the meter's back are for the meter's own power supply. Connect these to an AC voltage within the range of AC 85V~265V.
  2. AC Measurement Input: Terminals 1 and 2 in the middle section of the meter are for AC measurement voltage input (AC 5V~300V). Terminals 3 and 4 are AC voltage output terminals, which should be connected to the input end of the driving power supply.
  3. DC Measurement Input: Terminals 1 and 2 on the far right side of the meter are for DC measurement voltage input (DC 1V~300V). Terminal 1 is positive and terminal 2 is negative. These should be connected to the output end of the driving power supply. Terminals 3 and 4 are DC voltage output terminals, which should be connected to the DC load (terminal 3 is positive and terminal 4 is negative).

Caution: Wiring must be done in accordance with the method shown in the above figure; otherwise, damage to the meter may occur. The EPM5800-E adopts a bare-board design. After the power is turned on, the board carries high voltage. It is forbidden to touch any part of the circuit board with any part of the body.

5. Operating Instructions

Key Setting Instructions

  • Power Factor/Frequency Switch: In the working state, press the left "Power Factor/Frequency" button on the front panel to switch the display between power factor and frequency.

Product Demonstration Video

Watch this video for a visual guide on connecting and operating the EPM5800-E tester, demonstrating how it measures AC and DC parameters and efficiency.

Video 1: EPM5800-E Electrical Parameter Tester Demonstration

6. Precautions

  1. Do not operate the meter in a high-temperature and high-humidity environment; use it within the normal temperature and humidity range.
  2. Wiring must be carried out as shown in the diagram; otherwise, damage to the meter may result.
  3. The meter's power supply voltage must not exceed AC 265V, the AC measurement voltage must not exceed AC 300V, and the DC measurement voltage must not exceed DC 300V.
  4. When the working current is greater than 10A or the load is short-circuited, the fuse may be burned out. A 10A fuse must be used when replacing the fuse.
  5. Do not expose the meter to environments with excessive dust, corrosive gases, or other harmful substances during storage and use.
  6. Do not touch any part of the meter except the panel during use, as the meter carries high voltage during operation, to avoid electric shock.

7. Troubleshooting

Common Problems

  • Meter does not light up after power-on: Check whether the power cord is properly connected.
  • All parameter displays show zero after power-on: The fuse may be burned out, and the fuse needs to be replaced.

8. Opening Dimensions

For embedded installation, ensure the opening dimensions are as follows (Unit: mm):

Diagram showing the opening dimensions for embedded installation of the EPM5800-E, with measurements 121mm x 114mm and screw holes 3.2x4mm.
Figure 5: Opening Dimensions for EPM5800-E (Unit: mm)
  • Panel Cutout: 121mm x 114mm
  • Screw Hole Diameter: 3.2mm (for 4 screws)

9. User Tips

No specific user tips were available from the provided data.

10. Warranty and Support

For warranty information, technical support, or any product-related inquiries, please contact the seller or manufacturer directly. Refer to your purchase documentation for contact details.

A user manual in PDF format is also available for download. Please refer to the product page or contact the seller for the link: View User Manual (PDF)