1. Introduction
The ZHURUI EPM5800-E is an advanced AC+DC Electrical Parameter Tester designed for comprehensive measurement of various electrical parameters. It can accurately measure AC and DC voltage, current, power, power factor, frequency, and the efficiency of LED drivers. This manual provides detailed instructions for the safe and effective use of your device.


2. Main Technical Specifications
Meter Parameters
| Parameter | Value |
|---|---|
| Meter Power Supply Voltage | AC 85V~265V |
| Power Consumption | <2W |
| Sampling Rate | 2 times/second |
| Display Mode | 8 pieces of 5.6-inch nixie tubes |
| Operating Frequency | 45~65Hz |
| Overall Dimension (L×W×D) | 128.0mm × 135.0mm × 32.0mm (includes panel, depth without panel is 29mm) |
| Temperature and Humidity | 0~50°C, ≤85% RH |
| Installation Method | Screw Installation |
AC Measurement Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| AC Power | 0.1W~2500W | 0.01W (Note 2) | 0.5% FS |
| AC Voltage | AC 5V~300V | 0.1V | 0.5% FS |
| AC Current | AC 0.005A~10A | 0.001A | 0.5% FS |
| Power Factor | 0~1.000 | 0.001 | 0.5% FS |
| Frequency | 45.0Hz~65.0Hz | 0.1Hz | 0.5% FS |
DC Measurement Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| DC Power | 0.1W~3000W | 0.01W (Note 2) | 0.5% FS |
| DC Voltage | DC 1V~300V | 0.01V | 0.5% FS |
| DC Current | DC 0.005A~10A | 0.001A | 0.5% FS |
AC & DC Measurement Specifications (Efficiency)
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| Efficiency Measurement | 0~100% | 0.1 | 0.5% FS |
Notes:
[1] This dimension includes the panel; the depth without the panel is 29mm.
[2] When the load power is less than 100W, the measurement resolution is 0.01W; when the power is 100W or more, the measurement resolution is 0.1W; and when the power is 1000W or more, the measurement resolution is 1W.
3. Setup Instructions
Panel Layout

Wiring Instructions (Viewed from the Back)
- Terminals 1 and 2 on the left side of the meter's back are for the meter's power supply, which should be connected to an AC voltage within the range of AC 85V~265V.
- Terminals 1 and 2 in the middle of the meter are for AC measurement voltage input, with a voltage range of AC 5V~300V; terminals 3 and 4 are AC voltage output terminals, which should be connected to the input end of the driving power supply.
- Terminals 1 and 2 on the right side of the meter are for DC measurement voltage input, which should be connected to the output end of the driving power supply, with a voltage range of DC 1V-300V (terminal 1 is positive and terminal 2 is negative); terminals 3 and 4 are DC voltage output terminals, which should be connected to the DC load (terminal 3 is positive and terminal 4 is negative).

Caution: The EPM5800-E adopts a bare-board design. After the power is turned on, the board carries high voltage. It is forbidden to touch any part of the circuit board with any part of the body.
Opening Dimensions (Unit: mm)

The overall dimensions are 128.0mm x 135.0mm x 32.0mm, including the panel. The depth without the panel is 29mm. The mounting holes are 3.2mm x 4mm.
4. Operating Instructions
- Power On: Ensure all wiring is correctly connected according to the instructions. Power on the meter first, then power on the load. The display will show zero voltages for AC and DC before the load is powered.
- Reading Parameters: The meter features eight display windows. The four windows on the left show AC parameters (AC Voltage, AC Current, AC Power, AC Power Factor/Frequency). The three windows on the right show DC parameters (DC Voltage, DC Current, DC Power). The bottom right window displays the driver's efficiency.
- Power Factor/Frequency Switch: In the working state, press the left button labeled "PF/FREQ" to switch the display in the bottom-left window between Power Factor (PF) and Frequency (Hz).
5. Maintenance
- Do not operate the meter in high-temperature and high-humidity environments. Use it within the normal temperature and humidity range (0~50°C, ≤85% RH).
- Do not expose the meter to environments with excessive dust, corrosive gases, or other harmful substances during storage and use.
- Keep the device clean and free from debris.
6. Troubleshooting
- Meter does not light up after power-on: Check whether the power cord is properly connected.
- All parameter displays show zero after power-on: The fuse may be burned out. A 10A fuse must be used when replacing the fuse.
- Incorrect readings or damage: Ensure wiring is carried out exactly as shown in the diagram. Incorrect wiring can lead to damage.
7. User Tips
- Always double-check wiring before powering on the device to prevent damage and ensure accurate readings.
- When testing, observe the stability of readings. If readings fluctuate unexpectedly, re-check connections and power sources.
- For precise measurements, ensure the device is placed on a stable, non-vibrating surface.
8. Warranty & Support
Specific warranty information for the ZHURUI EPM5800-E is not provided in the available documentation. For warranty claims, technical support, or further assistance, please contact the seller or manufacturer directly.
A detailed user manual in PDF format is available for download: Download User Manual (PDF)