1. Product Overview

The WinAPEX LCR-101A is a portable, tweezer-type digital LCR meter specifically designed for testing Surface Mount Device (SMD) components. It automatically identifies Inductance (L), Capacitance (C), Resistance (R), Equivalent Series Resistance (ESR), Dissipation Factor (D), and Quality Factor (Q).

This device features a built-in rechargeable lithium battery with Type-C charging, making it ideal for electronic maintenance, laboratory testing, and component screening. It boasts high precision and smart identification capabilities, simplifying the measurement of various component types.

WinAPEX LCR-101A Smart Tweezer LCR Meter
Figure 1: WinAPEX LCR-101A Smart Tweezer LCR Meter
LCR-101A on a circuit board, highlighting automatic component identification
Figure 2: LCR-101A automatically identifying components on a circuit board.

2. Package Contents

Upon unboxing, verify that all items are present and in good condition:

  • Digital bridge host product (LCR-101A Tweezer LCR Meter)
  • USB Type-C charging cable
  • Instruction manual
  • Exquisite packaging box
Video 1: Unboxing and initial power-on of the LCR-101A, showing package contents.
Product List image showing the LCR-101A, USB cable, and instruction manual in its packaging.
Figure 3: Contents of the LCR-101A package.

3. Key Features

  • Automatic Recognition: Automatically identifies component types (L, C, R, ESR, D, Q).
  • Multiple Test Frequencies: Offers 6 test frequencies: 20Hz, 100Hz, 120Hz, 1kHz, 10kHz, 100kHz.
  • Adjustable Test Signal Level: Test signal level adjustable from 0.1V to 1.0V.
  • Color IPS LCD Display: Clear and intuitive display for measurement data.
  • Rechargeable Battery: Built-in lithium battery with Type-C 5V charging.
  • PC Software Connectivity: USB interface for PC connection and WinAPEX software.
  • Data Retention: Press the 'M' key to lock current measured data.
  • Bilingual Interface: Supports both Chinese and English languages.
  • Auto Shutdown: Configurable auto-shutdown feature (5/15/30 minutes).
  • Metal Tweezer Head: Durable and precise metal probes for testing.
LCR-101A features: Intelligent digital bridge, auto-identify, Chinese/English, R/C/L, Diode, high-precision, portable.
Figure 4: Overview of LCR-101A's intelligent features.
New product, new upgrade image with icons for automatic recognition, three adjustable frequencies, 1.14-inch color screen, 250mAh lithium battery, measure capacitance, resistance, inductance, diode, Chinese/English switching, data retention, Type-C charge, Auto Shutdown (5/15/30 min).
Figure 5: Summary of new features and upgrades.

4. Technical Specifications

Item Parameter
Model Number LCR-101A
Test Frequencies 20Hz, 100Hz, 120Hz, 1kHz, 10kHz, 100kHz
Test Signal Level 0.1V ~ 1.0V adjustable
Measurement Items L/C/R, ESR, D(DF), Q, Phase Angle
Equivalent Circuit Auto Series/Parallel conversion
Basic Accuracy ±0.2%
Power Supply Built-in Li-ion battery, Type-C 5V charging
Display Color IPS LCD screen
Interface USB for PC WinAPEX software connection
Battery Included Yes
Certification CE, RoHS
Origin Mainland China

Measurement Ranges and Accuracy

Product Parameters table for capacitance, inductance, and resistance ranges and accuracies at 100Hz, 1KHz, 10KHz.
Figure 6: Detailed measurement ranges and accuracies.
Product Parameters table including diode, test frequency, test level, data retention, on-off, power button.
Figure 7: Additional product parameters and button functions.

5. Button Functions and Display

Button Function Definition:

  • AUTO: Automatic L/C/R identification and auto-ranging.
  • FREQ: Switches the test frequency point.
  • LEVEL: Adjusts the AC test signal voltage.
  • RANGE: Toggles auto/manual range lock.
  • EQU: Switches between Series / Parallel equivalent mode, and D/Q/phase value for secondary display.
  • REL: Relative zero offset for stray capacitance elimination.
  • CAL: Open/Short calibration for precision.
  • HOLD: Freezes current measured data.
  • SETUP: Accesses system settings (e.g., auto power off, backlight).
  • COM: Enables USB communication with WinAPEX PC software.

Display Overview:

The IPS color LCD screen provides clear readings. An example of the standard English display content is: AUTO DETECT LV:0.6V | FREQ:1KHz Rp:10.00Ω | D:0.649 | PAR REL:OFF | HOLD:OFF

Close-up of the LCR-101A display with labels for gear, level, frequency, power level, gear function symbol, unit symbol, Rs measurement value display, and parameter display (D/Z/Q/X).
Figure 8: LCR-101A display with labeled indicators.

Parameter Description:

  • D: Loss value
  • Q: Quality factor
  • Z: Impedance
  • OL: Series resistor
  • PF: Unit symbol
  • X: Reactance
Parameter Description table: D (Loss value), Q (Quality factor), Z (Impedance), OL (Series resistor), PF (Unit symbol), X (Reactance).
Figure 9: Explanation of display parameters.

6. Initial Setup

Powering On/Off:

  • To Power On: Short press the power button.
  • To Power Off: Long press the power button.

Charging the Device:

The LCR-101A has a built-in rechargeable lithium battery. When the battery is low, charge it using the provided Type-C USB cable and a 5V adapter.

Product details flawless image showing: 01. Short press to power on (Long press the power button to shut down M/Hold Settings Key), 02. Metal tweezers head (test resistance Inductance, Voltage level, etc), 03. Left and right adjustment function (S is the interface setting Left and Right Settings).
Figure 10: Power button and other physical features.

7. Standard Operation Guide

7.1 Pre-test Calibration (Required for precise measurement)

Calibration is crucial for accurate measurements. Perform both Open and Short calibrations before precise testing.

  1. Open Calibration: Keep the tweezer probes separated (not touching anything). Press the CAL button to finish OPEN calibration.
  2. Short Calibration: Short the two gold-plated probe tips tightly together. Press the CAL button to finish SHORT calibration. This removes residual stray inductance/capacitance from the test probes.

7.2 Component Measurement

  1. Power on the device. The default mode is AUTO, with a 1kHz test frequency and Parallel circuit.
  2. Gently clamp the SMD component between the two golden tweezer tips, ensuring a tight connection.
  3. The meter will automatically identify the component type and display its primary and secondary parameters on the screen.
  4. For testing tiny SMD components, press the REL button to eliminate fixture residual values, improving accuracy.
  5. To lock the current reading for data recording, press the HOLD button.
Four panels showing measurement of resistance, capacitance, inductance, and diode.
Figure 11: Examples of measuring different component types.
Image showing measurement of resistance, capacitance, inductance, and diode.
Figure 12: Visual guide to component measurement.

7.3 Test Level Selection

The device offers adjustable test signal voltage levels:

  • 0.3V: Safe and non-destructive, suitable for low impedance and easily saturated devices.
  • 0.6V: Offers strong universality, balancing accuracy and adaptability for greater measurement stability.
Bar chart comparing 0.3V and 0.6V test levels, describing 0.3V as 'Safe and non-destructive, suitable for low impedanceEasily saturated device' and 0.6V as 'Strong universality, balancing accuracy and adaptability Measuring for greater stability'.
Figure 13: Comparison of 0.3V and 0.6V test levels.

7.4 Measurement Frequencies

The LCR-101A supports three major measurement frequencies, each suited for different applications:

  • 100Hz: Ideal for large capacity electrolytic capacitors, power frequency phase turn-off components, and motor windings.
  • 1KHz: The industry standard testing frequency, suitable for universal thin-film capacitors, ceramic capacitors, and power inductors.
  • 10KHz: Best for small capacity ceramic capacitors, high-frequency inductors, and RF components.
Three major measurement frequencies diagram: 100Hz, 1KHz, 10KHz with their applications.
Figure 14: Applications for different measurement frequencies.

8. WinAPEX PC Software Instruction

8.1 Software Compatibility

The WinAPEX PC software supports Windows 7/10/11 (32-bit & 64-bit systems) for real-time data synchronization and batch testing.

8.2 Core Functions

  • Real-time synchronous display of all measuring data from the LCR-101A.
  • Multi-frequency continuous scanning measurement for batch components.
  • Data storage and export as CSV/Excel documents.
  • Upper/lower limit tolerance comparison with automatic PASS/FAIL judgment.
  • Remote control of tester frequency, voltage, and circuit mode via PC.

8.3 Connection Steps

  1. Install the WinAPEX driver on your PC.
  2. Connect the LCR-101A instrument to your computer via the USB cable.
  3. Launch the WinAPEX software, select the corresponding COM port, and click 'Connect'.
  4. After a successful connection, "COM ON" will be displayed on the tester screen.

9. Maintenance

  • Cleaning: Keep the gold-plated probe tips clean for accurate measurements. Use a soft, dry cloth.
  • Battery Care: When the battery is low, charge it promptly using a Type-C 5V adapter to ensure optimal performance and battery longevity.
  • Storage: Store the device in a safe place, away from extreme temperatures and humidity.

10. Safety & Caution

  • Do NOT test live powered electronic components to avoid device damage and potential electrical shock.
  • Avoid dropping or squeezing the instrument, as this can cause internal damage.
  • Ensure the probe tips are clean before each use to prevent inaccurate readings.
  • This electronic repair tool does not contain any high-concerned chemicals, ensuring safety during use.

11. User Tips

  • Always perform the Open and Short calibrations before taking precise measurements to ensure the highest accuracy.
  • If you are new to LCR meters, take some time to familiarize yourself with the instructions and calibration procedures.
  • The device is generally well-regarded for its functionality and ease of use.
  • Keep the instruction manual handy for reference, especially when exploring advanced features or troubleshooting.

12. Troubleshooting

If you encounter issues with your LCR-101A, consider the following:

  • Inaccurate Readings: Ensure the probe tips are clean and that you have performed both Open and Short calibrations recently. Check the component connection is tight.
  • Device Not Powering On: Verify the battery is charged. Connect the device to a 5V Type-C charger and allow it to charge for a period.
  • PC Software Connection Issues: Confirm that the WinAPEX driver is installed correctly on your PC and that you have selected the right COM port in the software.
  • Display Problems: Check system settings via the SETUP button for backlight adjustments.

13. Warranty and Support

For detailed warranty information and technical support, please refer to the official WinAPEX documentation or contact your retailer. A user manual in PDF format is available for download at WinAPEX LCR-101A User Manual (PDF).

© 2024 WinAPEX. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.