AmScope T800 Series Biological Trinocular Compound Microscope

User Manual

Model: T800

Brand: AmScope

Important Safety Information

This instruction manual provides guidance for the operation, troubleshooting, and maintenance of the B800/T800 series inverted biological microscope. Please read this manual thoroughly before operating and keep it with the instrument for future reference. The manufacturer reserves the right to modify technical specifications due to technological advancements without prior notice, ensuring optimal operation.

1. Operation Notice
  1. As the microscope is a high-precision instrument, always operate it with care and avoid physical shock during operation.
  2. Do not expose the microscope to direct sunlight, high temperatures, damp conditions, dust, or acute vibrations. Ensure the worktable is flat and horizontal. The required operating environment is: Indoor temperature: 5°C ~ 40°C, Max relative humidity: 80%.
  3. When moving the microscope, use both hands to hold the handle at the back and the front end of the microscope body. Handle with care.
    Warning: Holding the stage, focusing knob, head, or light source when moving will damage the microscope.
  4. When operating, the surface of the light source will get hot. Ensure there is enough room for heat dissipation around the light source.
  5. Connect the microscope to ground to avoid lightning strikes.
  6. For safety, ensure the power switch is at "OFF" and power it off. Wait for the bulb to cool down completely before replacing the LED.
    Note: Bulb selected is a Single 3W LED.
  7. Wide voltage range is supported (100~240V AC). An additional transformer is not required. Ensure the power supply is within this range.
  8. Use the special wire supplied by our company.
  9. All power OFF devices have not been set in a position that is difficult to operate.
2. Maintenance
  1. Wipe the lens gently with a soft lens tissue. Carefully wipe off oil or fingerprints with tissue moistened with a little 3:7 mixture of alcohol and ether or dimethylbenzene.
    Warning: Alcohol and ether are flammable. Do not place these chemicals near fire or fire sources. Use them in a ventilated place when turning on/off electrical devices.
  2. Do not use organic solutions to wipe the surface of other components. Use neutral detergent if necessary.
  3. If the microscope is damped by liquid, immediately cut off the power and wipe it dry.
  4. Never disassemble the microscope. Doing so will affect its function or damage it.
  5. After use, cover the microscope with a dust cover.
3. Safety Signs
Sign Signification
Hot surface warning icon The surface gets hot and do not touch it with bare hand.
Read instructions icon Study the instructions before use. Unsuitable operation could lead to personal injury or instrument fault.
| Main switch ON
O Main switch OFF

Components Overview

The AmScope T800 Series Biological Trinocular Compound Microscope is a robust, high-performance instrument designed for professionals in medical and research fields. It features superior optics and a sturdy mechanical design.

  • A robust, high-performance microscope with superior optics and mechanical design.
  • The trinocular head provides a dedicated photo port for mounting a camera.
  • Super-widefield optics provide a larger field-of-view so you can see more specimen at any magnification.
  • The infinity-corrected optical system reduces distortion and improves color accuracy.
  • The quintuple lens turret is equipped with five plan-achromatic objective lenses producing high-contrast images with edge-to-edge sharpness.
  • The Koehler illumination system provides centered, evenly-distributed lighting with a field aperture for maximum contrast and definition.
AmScope T800 Series Microscope, front-left view

AmScope T800 Series Biological Trinocular Compound Microscope.

Microscope Components Diagram
Detailed exploded view of the compound microscope T800 components

Detailed exploded view of the AmScope T800 microscope components.

Key Parts Identification
Labeled diagram of AmScope T800 microscope components

Labeled diagram showing the main components of the AmScope T800 microscope.

Front View Components:

  • Eyepiece
  • Trinocular Head
  • Light Path Selecting Pole
  • Condenser up-down knob
  • Objective
  • Stage
  • Condenser Center Adjusting Screw
  • Condenser
  • Field Diaphragm Adjusting Ring
  • Y-axis Moving Knob
  • X-axis Moving Knob
  • Coarse Focusing Knob
  • Fine Focusing Knob
  • Tension Adjustment Ring
  • Power Switch

Rear View Components:

Rear view of the AmScope T800 microscope showing power input and light source

Rear view of the AmScope T800 microscope.

  • B800/T800 Body
  • Inner Hexagon Spanner
  • Power Socket
  • Light Source Group
  • Diopter Adjustment Ring
  • Nosepiece
  • Focus Arm
  • Fine Focusing Knob

Setup Instructions

Follow these steps to assemble your AmScope T800 microscope. Ensure the work area is clean and free of dust.

2-1 Assembling Scheme
Assembly diagram of the microscope components

General assembly scheme for the microscope components. Numbers indicate assembly order.

Important: Before assembling, ensure there is no dust, dirt, or other materials that could interfere. Assemble carefully and avoid scratching any part or touching glass surfaces.
2-2 Assembling Steps
2-2-1 Assemble the LED Light Source Group
Diagram showing LED light source assembly

LED Light Source Group assembly.

  1. Push the LED light source group into the light source base thoroughly.
  2. Rotate the LED light source group so the light source cable faces downward and is centered.
  3. Tighten the two lock screws using an M4 inner hexagon spanner.
2-2-2 Assemble the Stage
Diagram showing stage assembly

Stage assembly.

  1. Loosen the lock-screw on the stage fully.
  2. From the rear area of the rounded hole center on the base, carefully align the two "V" buttons on the bottom of the stage into the "V" rounded groove.
  3. Screw down the lock screw.
Note: The stage of different sizes and shapes is available according to customer specifications.
2-2-3 Assemble the Condenser
Diagram showing condenser assembly

Condenser assembly.

  1. Rotate the coarse focusing knob to raise the stage to its highest position.
  2. Rotate the condenser up-down knob to lower the condenser bracket to its lowest position.
  3. Loosen the condenser lock-screw fully.
  4. Swing out the front lens of the condenser with the scale facing forward. Align the lock screw of the condenser with the groove of the condenser stand. Push the condenser into the innermost part of the stand.
  5. Screw down the condenser lock-screw, and raise the condenser to the highest position with the condenser up-down knob.
2-2-4 Assemble the Head
Diagram showing head assembly

Microscope head assembly.

  1. Loosen the head lock-screw fully.
  2. From a slightly right position, insert the coattail interface on the bottom of the head into the hole of the microscope body with a slight left incline. Keep the two eyepiece tubes forward, then screw down the lock screw.
2-2-5 Assemble the Objective
Diagram showing objective assembly

Objective lens assembly.

  1. Rotate the coarse focusing knob to lower the stage.
  2. Install the objectives into the nosepiece from the lowest magnification to the highest in a clockwise direction.
Note: Search and focus the sample using a low magnification objective (5X or 10X) first. Then switch to higher magnifications as needed. When replacing objectives, rotate the nosepiece until it clicks ("ka-da") to ensure the objective is centered in the light path.
2-2-6 Assemble the Eyepiece
Diagram showing eyepiece assembly

Eyepiece assembly.

  1. Remove the cover of the eyepiece tube.
  2. Insert the eyepiece into the eyepiece tube until it touches the bottom.
2-2-7 Connect the Power Cord
Diagram showing power cord connection

Power cord connection.

  1. Ensure the power switch is at "O" (OFF).
  2. Insert one end of the power cord into the power socket of the microscope.
  3. Insert the other end of the power cord into the power supply socket.
Warning: Do not use strong force if the power cord is bent or twisted, as this will damage it. Use the special wire supplied by our company. If it is lost or damaged, choose one with the same specifications. Connect the power cord appropriately to ensure the instrument is connected to ground.

Operation Guide

3-1 Set Illumination
Diagram showing illumination adjustment knob

Illumination adjustment.

  1. Connect the power and turn the main power switch to "ON".
  2. Adjust the light adjustment knob until the illumination is comfortable for observation. Rotate clockwise to increase brightness, counterclockwise to decrease brightness.
Note: Using bulbs in a low-voltage state can extend bulb life.
3-2 Place the Specimen Slide
Diagram showing specimen slide placement

Placing a specimen slide.

  1. Push the wrench of the specimen holder backwards.
  2. Loosen the wrench, and clamp the slide by the clips with the cover glass facing up.
  3. Rotate the X and Y-axis knobs to move the specimen to the center (alignment with the center of the objective).
3-3 Adjust Focusing
Diagram showing focusing knobs

Focusing adjustment.

  1. Place the slice on the stage and hold it down with the clip. Shift the 4X objective into the light path.
  2. Loosen the up-limit knob, then observe through the right eyepiece with your right eye. Rotate the coarse focusing knob until the image appears in the view field, then lock the up-limit knob.
Note: Locking the up-limit knob prevents the objective from touching the slice during high-power objective coarse focusing. The up-limit knob does not affect the fine focusing knob.
  1. Rotate the fine focusing knob for clear details.
3-4 Adjust the Focusing Tension
Diagram showing focusing tension adjustment ring

Focusing tension adjustment.

If the coarse focusing handle feels too heavy or too loose, or if the specimen loses focus easily, adjust the tension adjustment ring. Rotate the ring in the direction of the arrow to tighten or loosen the focusing system.

3-5 Adjust the Diopter
Diagram showing diopter adjustment ring

Diopter adjustment.

After the image is clear in the right eyepiece, observe the left eyepiece with your left eye. Rotate the diopter adjustment ring until the image is clear. The diopter adjustment ring has a range of ±5 diopters. The value aligned with the scale is your eye's diopter. The dot "•" on the left side can also indicate this. Remember your eye's diopter for future use.

Note: If both eyepiece tubes have diopter adjustment rings, you can select one as a standard of reference and adjust the other. The standard eyepiece should be set to "0" before focusing.
3-6 Adjust the Interpupillary Distance
Diagram showing interpupillary distance adjustment

Interpupillary distance adjustment.

When observing with both eyes, hold the base of the two prism bases and rotate them around the axis until there is only one field of view. The dot "•" above the left eyepiece base points to the interpupillary distance indicator scale. The adjustable range is 50~76mm. Remember your interpupillary distance for future use.

3-7 Center the Condenser
Diagram showing condenser centering

Condenser centering process.

  1. Rotate the condenser up-down knob to raise it to the highest position.
  2. Rotate the spanner to move the front lens into the light path.
Note: Move the front lens of the condenser into the light path when the objective is beyond 20X.
  1. Move the 20X objective into the light path and focus the specimen.
  2. Rotate the field diaphragm adjustment ring to its smallest position. The image of the field diaphragm should be observable through the eyepiece.
  3. Rotate the condenser up-down knob to adjust the image to the clearest.
  4. Adjust the condenser center adjusting screw to center the image in the field of view.
  5. Gradually open the field diaphragm. If the image remains centered and inscribed to the field of view, the condenser is correctly centered.
  6. For use, you can slightly enlarge the field diaphragm to make the image circumscribed to the field of view.
3-8 Adjust the Field Diaphragm
Diagram showing field diaphragm adjustment

Field diaphragm adjustment.

By limiting the diameter of light entering the condenser, the field diaphragm prevents other light and strengthens image contrast. When the image is just at the edge of the field of view, the objective performs best and provides the clearest image.

3-9 Adjust the Aperture Diaphragm
Diagram showing aperture diaphragm adjustment

Aperture diaphragm adjustment.

The aperture diaphragm determines the numerical aperture of the illumination system. Matching the N.A. of the illumination system with the N.A. of the objective improves resolution, contrast, and depth of field. As contrast is usually low, it is advised to set the condenser aperture diaphragm to 70%-80% of the objective's N.A. Remove the eyepiece to observe from the tube if necessary. Adjust the aperture diaphragm adjusting ring until the proportion is correct (refer to Fig. 18 & 19 in the original manual for visual guidance, where ① is the image of the aperture diaphragm and ② is the edge of the objective).

Use of Scale: Set the scale of condenser N.A. to 80% of the objective's N.A. For example, with a 40X objective (N.A. 0.65), set the aperture diaphragm scale to 0.65 x 0.8 = 0.52.

3-10 Select the Light Path
Diagram showing light path selection pole

Light path selection.

For the trinocular head, the light path selecting pole controls the light-energy ratio between the binocular and trinocular paths. When the pole is pushed to the innermost position, all light enters the binocular head. When pulled to the outermost position, both binocular and trinocular observation are possible. Typically, push the selecting pole innermost for binocular observation, and pull it outermost for trinocular observation (TV & Photography).

3-11 Use the Eye-cap
Diagram showing eye-cap usage

Eye-cap usage.

  1. If you wear glasses, turn over the eye-cap to prevent glasses from touching and damaging the eyepiece.
  2. If you do not wear glasses, open the eye-cap to prevent stray light from disturbing observation.
3-12 Use the Color Filter
Diagram showing color filter placement

Color filter placement.

Color filters can adjust background light for better suitability and strengthen image contrast. Four colors are selectable: blue, green, yellow, and white.

Note: Place the rough side of the filter downwards.
3-13 Use the Oil Objective (100X)
Diagram showing oil objective usage

Oil objective (100X) usage.

  1. Use the 4X objective to focus the specimen.
  2. Place a drop of oil on the specimen.
  3. Rotate the nosepiece counterclockwise and rotate the oil objective (100X) into the light path. Then use the fine focusing knob to focus.
Warning: Ensure there are no air bubbles in the oil, as they will affect the image.

To check for air bubbles:

  • Move the eyepiece to examine for air bubbles. Open the aperture diaphragm and field diaphragm fully and observe the edge of the objective from the tube (it should appear round and light).
  • Rotate the nosepiece slightly and swing the oil objective a few times to remove any air bubbles.
  1. After use, wipe the front lens with a tissue moistened with a small amount of 3:7 mixture of alcohol and ether or dimethylbenzene. Wipe off the oil on the specimen at the same time.
Warning: Do not put another objective into the light path before the oil is wiped, to avoid wetting the dry objective. Too much dimethylbenzene would dissolve the lens's stickiness.
3-14 Assemble and Use the Simple Polarizing
Diagram showing simple polarizing assembly

Simple polarizing system assembly.

The simple polarizing system includes a polarizer and a 360° rotatable analyzer.

  1. Unplug the dust-cover from the socket of the front body, and insert the 360° rotatable analyzer into the socket.
  2. Put the polarizer into the condenser groove.
  3. Rotate the 360° rotatable analyzer to begin simple polarizing observation.
Note: When the view field of the eyepiece is darkest, the polarizer and 360° rotatable analyzer are in the polarized orthogonal state.
3-15 Assemble and Use the Disc Phase Contrast Condenser
Diagram showing disc phase contrast condenser assembly

Disc phase contrast condenser assembly.

Assembly of the Disc Phase Contrast Condenser refers to the "2-2-3 Assemble the Condenser" section. It features a magnification ring diaphragm and bright field "BF" hole positions on the phase contrast ring. For phase contrast observation, the ring diaphragm magnification should match the phase contrast objective, and the ring diaphragm should be centered. For bright field observation, rotate the phase contrast ring to the "BF" position. A "ka-da" sound indicates that a diaphragm or hole is rotated into the center of the light path.

3-15-1 Center the Halo
Diagram showing halo centering

Halo centering process.

  1. Rotate the 10X phase contrast objective into the light path, then rotate the ring diaphragm of the phase contrast ring to the 10X position.
  2. Turn the aperture diaphragm lever to the leftmost position. (Always keep the aperture diaphragm at maximum in phase contrast observation.)
  3. Place the specimen on the slice clip and focus it for clarity.
  4. Remove one observation eyepiece and replace it with a CT (centering telescope) into the tube without diopter adjustment.
  5. Loosen the lock screw of the centering telescope, move the telescope tube up and down to adjust the image of the halo and phase ring for clarity, then lock the screw.
  6. Adjust the phase contrast ring adjusting lever while observing from the centering telescope, and center the halo with the phase ring.
  7. Remove the CT (centering telescope) and insert the eyepiece to observe the phase contrast effect.
  8. Adjust phase contrast objectives of other magnifications and ring diaphragms following the above steps.
Warning: Optimal phase contrast observation cannot be achieved if the halo is not centered correctly. Center the brightest halo to the objective phase ring if a double image appears.
3-16 Assemble TV Device
Diagram showing TV device assembly

TV device assembly.

  1. Loosen the lock-screw on the trinocular head and remove the dust-cover.
  2. Remove the dust-cover caps of the TV adapter and the CCD. Insert the screw thread end of the TV adapter into the CCD, then install the TV device with CCD into the trinocular tube. Screw down the lock-screw tightly.
3-17 Assemble the Intermediate Spectrometer
Diagram showing intermediate spectrometer assembly

Intermediate spectrometer assembly.

  1. Fully loosen the lock screw.
  2. From a slightly right position, insert the coattail interface on the bottom of the intermediate spectrometer into the round hole of the microscope body with a slight left incline. Position the intermediate spectrometer camera device interface on the left side of the microscope body. Then tighten the lock screw.
  3. Fully loosen the lock screw.
  4. From a slightly right position, insert the coattail interface on the bottom of the observation head into the round hole of the microscope body with a slight left incline. Ensure the two tubes face forward. Then tighten the lock screw.
  5. Assemble the TV device into the intermediate spectrometer, referring to "3-16 Assemble TV Device".
  6. For binocular observation, after the image is clear, pull the light path selecting pole to the outermost position to observe the image. If the image is not clear, rotate the adjustment tube until the image is clear.
3-18 Use the Digital Head
Diagram showing digital head connection

Digital head connection.

Insert the connector of the USB data cable into the USB output interface on the back of the microscope head, and insert the other connector into the USB interface of the computer. View the microscope video using video capture and analysis software. Binocular observation is also possible.

Note: The digital USB interface has a voltage of 5V and current ≤ 500mA. Do not use strong force if the data cable is bent or twisted, as this will damage it.

Troubleshooting

This section provides solutions for common issues that may arise during the operation of your microscope.

1. Optical System Problems
Problem Cause Solution
The bulb is bright but the field of view is dark. Condenser is too low. Adjust the condenser.
Condenser is not centered. Center the condenser.
Light path selecting pole is in the trinocular observation position. Push the light path selecting pole to the binocular observation position.
The side of the field of view is dark or not even. The nosepiece is not in the right position. Turn the nosepiece into the right position.
Stain or dust is observed in the field of view. Stain or dust has accumulated on the lens (condenser, objective or eyepieces). Clean the lens.
Stains have accumulated on the specimen. Clean the specimen.
Stains have accumulated on the lens. Clean the lens.
Unclear image. No cover glass on the specimen slide. Add the cover glass.
The cover glass is not standard. Use a standard cover glass with thickness 0.17mm.
The cover glass faces down. Put the cover glass to face up.
The immersion oil has accumulated on the dry objective. Clean thoroughly.
The immersion oil is not used for oil objective. Use immersion oil.
Air bubble in the immersion. Get rid of the air bubble.
Use wrong immersion oil. Use a correct one (cedar oil).
The aperture diaphragm is not opened correctly. Adjust it.
Stain or dust has accumulated on the lens of eyepiece. Clean the lens.
One side of the image is dark or the image moves while focusing. Condenser is too low. Adjust the condenser.
The specimen slide is not fixed. Fix it with clips.
The nosepiece is not in the right position. Turn the nosepiece into the right position.
Condenser is not centered. Center the condenser.
The eyes feel tired easily. The right field of view doesn't superpose with the left one. Interpupillary distance is incorrect. Adjust the interpupillary distance.
Diopter adjustment is incorrect. Adjust the diopter.
The eyepiece for the right eye is different from the left one. Use the same eyepieces.
2. Mechanical System Problems
Problem Cause Solution
Cannot focus when using high magnification objective. The cover glass faces down. Put the cover glass to face up.
The cover glass is not standard. Use a standard cover glass with thickness 0.17mm.
The objective touches the cover glass while turning the nosepiece. The cover glass faces down. Put the cover glass to face up.
The cover glass is not standard. Use a standard cover glass with thickness 0.17mm.
Coarse focusing knob is too tight. Tension adjustment ring is too tight. Loosen it to an appropriate position.
Stage declines itself and cannot stay on the focal plane. Tension adjustment ring is too loose. Tighten it to an appropriate position.
Coarse focusing knob cannot rise. The up limit knob is locked. Loosen the up limit knob.
Coarse focusing knob can't decline. The condenser is too low. Raise the condenser.
Cannot move the slide smoothly. The slide is not fixed correctly. Adjust it correctly.
The movable specimen holder is not fixed properly. Adjust it correctly.
The image moves obviously when touching the stage. The stage is fastened incorrectly. Fasten the stage correctly.
3. Electrical System Problems
Problem Cause Solution
The bulb does not work. No power supply. Check the connection of the power cord.
The bulb is not installed correctly. Install it correctly.
The bulb is burnt out. Replace it.
The bulb burns out usually. A wrong bulb is used. Replace it with a correct one.
The field of view is not bright enough. A wrong bulb is used. Replace it with a correct one.
The use of light adjusting knob is wrong. Adjust it correctly.
The bulb flickers or the brightness is not stable. The bulb is burnt out soon. Replace it with a new one.
The wire doesn't connect well. Connect it correctly.

Technical Specifications

4-1 Key Technical Specifications of B800/T800 Series Biological Microscope
Component Specification
Optical System Infinity color corrected optical system
Head Gemel Binocular head, 30° inclined, 360° rotatable
Gemel Trinocular head, 30° inclined, 360° rotatable
Splitting ratio: Binocular Head 100%, Binocular Head/Trinocular Head 50%/50%
Gemel type of binocular head 30°~60° inclined adjustable, 360° rotatable
Eyepiece PL10X high eye-point plan eyepiece, line field of view: 22mm.
Nosepiece Inner location five holes nosepiece
Objective Infinity plan achromatic objective (4X, 10X, 20X, 40X, 100X),
Infinity plan phase contrast objective (10X, 20X, 40X, 100X)
Focus System Coaxial coarse & fine focusing system with up limit-stopper & tension adjustable. Travel range: 30mm. Height of stage bracket group adjustable, fine focusing precision: 0.002mm
Stage 185×165mm double-layer mechanical moving stage. Moving range: 80mmx55mm, precision: 0.1mm, superhard oxidized metal stage, left and right hand position optional.
175×145mm double-layer mechanical moving stage. Moving range: 75mmx50mm, precision: 0.1mm, left and right hand position optional.
Condenser NA1.2/0.22 Swing-out achromatic Condenser
NA1.25 Quintuple Disc Phase-contrast Condenser
NA0.9 Dry Dark Field Condenser
NA1.25 Oil Dark Field Condenser
Illumination System 6V2A external power adapter, transmission Koehler illuminator systems, output single 3W LED bulb, brightness continuously adjustable.
Polarizer Analyzer 360° rotatable, Polarizer and analyzer can be moved out of the light path.
Filter Yellow, green, blue, and neutral filter
4-2 Parameters of Objective
Type Magnification Numerical Aperture (N.A.) Operating Range (mm) Conjugate Distance (mm) Parfocal Distance (mm) Thickness of Cover Glass Magnification Mark (color ring)
Infinity plan achromatic objective 4X 0.10 11.9 ∞ 45 0.17 Red
10X 0.25 12.1 ∞ 45 0.17 Yellow
20X 0.40 1.56 ∞ 45 0.17 Light green
40X(S) 0.65 0.36S ∞ 45 0.17 Light Blue
100X(S) (OIL) 1.25 0.18 ∞ 45 0.17 White
Infinity plan phase contrast objective 10X 0.25 12.3 ∞ 45 0.17 Yellow
20X 0.40 1.56 ∞ 45 0.17 Light green
40X(S) 0.65 0.36 ∞ 45 0.17 Light Blue
100X(S) (OIL) 1.25 0.18 ∞ 45 0.17 White
Camera Specifications (Optional, depending on model)

The AmScope T800 series microscopes can be equipped with various USB 3.0 cameras. Below are the specifications for different camera options:

3.1MP USB 3.0 Camera
Specifications for 3.1MP USB 3.0 camera

Detailed specifications for the 3.1MP USB 3.0 camera option.

Feature Specification
Sensor Type Back-illuminated CMOS
Sensor Optical Format 1/2.8"
Active Pixels 3.1M (2048 x 1536)
Pixel Size 2.5µm x 2.5µm
Active Sensor Area 5.12mm x 3.84mm
Shutter electronic rolling shutter
Sensitivity 600mV @ 1/30s
Spectral Response 380-650nm with IR-cut filter
Capture Resolution and Maximum Framerate 50fps @ 2048x1536
Connectivity USB 3.0
Lens-mount C-mount
Power 5VDC over USB
6.3MP USB 3.0 Camera
Specifications for 6.3MP USB 3.0 camera

Detailed specifications for the 6.3MP USB 3.0 camera option.

Feature Specification
Sensor Type Back-illuminated CMOS
Sensor Optical Format 1/1.8"
Active Pixels 6.3M (3072 x 2048)
Pixel Size 2.4µm x 2.4µm
Active Sensor Area 7.37mm x 4.92mm
Shutter electronic rolling shutter
Sensitivity 425mV @ 1/30s
Spectral Response 380-650nm with IR-cut filter
Capture Resolution and Maximum Framerate 30 fps @ 3072x2048, 38fps @ 1536x1024
Connectivity USB 3.0
Lens-mount C-mount
Power 5VDC over USB
12MP USB 3.0 Camera
Specifications for 12MP USB 3.0 camera

Detailed specifications for the 12MP USB 3.0 camera option.

Feature Specification
Sensor Type Back-illuminated CMOS
Sensor Optical Format 1/1.7"
Active Pixels 12M (4000 x 3000)
Pixel Size 1.85µm x 1.85µm
Active Sensor Area 7.4mm x 5.55mm
Shutter electronic rolling shutter
Sensitivity 280mV @ 1/30s
Spectral Response 380-650nm with IR-cut filter
Capture Resolution and Maximum Framerate 25fps @ 4000x3000, 50fps @ 2048x1080
Connectivity USB 3.0
Lens-mount C-mount
Compensating Lens 0.75X
Power 5VDC over USB
20MP USB 3.0 Camera
Specifications for 20MP USB 3.0 camera

Detailed specifications for the 20MP USB 3.0 camera option.

Feature Specification
Sensor Type Back-illuminated CMOS
Sensor Optical Format 1"
Active Pixels 20M (5440 x 3648)
Pixel Size 2.4µm x 2.4µm
Active Sensor Area 13.06mm x 8.76mm
Shutter electronic rolling shutter
Sensitivity 462mV @ 1/30s
Spectral Response 380-650nm with IR-cut filter
Capture Resolution and Maximum Framerate 15fps @ 5440x3648, 50fps @ 2736x1824, 60fps @ 1824x1216
Connectivity USB 3.0
Lens-mount C-mount
Compensating Lens 1X
Power 5VDC over USB
Software Requirements and Features
Software OS and Hardware Requirements, and Features

Software operating system and hardware requirements, along with available features.

Category Details
OS Requirements Windows (32/64 bit) XP/Vista/7/8/10, Mac OS 10.8+, Linux kernel 3.13+
Hardware Requirements Intel Core2 2.8GHz or comparable processor, 4GB RAM
Features (Windows Version)
  • Live video preview
  • Captures still images and video
  • Time-lapse capture
  • Exposure and white-balance controls
  • Dark-field correction
  • Digital binning
  • Image processing
  • Measurement tools, including lines, arcs, polygons
  • Segmentation and counting tools
  • Image-stitching, extended depth-of-field, HDR
  • Fluorescence color-mapping
Features (Mac/Linux Versions)
  • Live video preview
  • Captures still images and video
  • Exposure control
  • Color adjustment
  • Measurement tools, including lines, arcs, polygons

Packing List

The standard package for the AmScope T800 Series microscope includes:

  • One trinocular head
  • One microscope body
  • One C-mount photo port adapter
  • Five objective lenses: 4X, 10X, 20X, 40X, 100X
  • One pair 10X eyepieces
  • One stage assembly
  • One swing condenser
  • One LED illuminator assembly
  • One AC power adapter (US and Canada standard)
  • One hex driver
  • Two blank filter inserts
  • One dust cover

User Tips

No specific user tips were available from provided reviews or Q&A content. However, general best practices for microscope use include:

  • Always start with the lowest magnification objective to locate your specimen, then gradually increase magnification.
  • Use immersion oil only with the 100X oil objective and ensure proper cleaning after use to prevent damage to other lenses.
  • Regularly clean optical components with appropriate lens cleaning solutions and tissues to maintain image clarity.
  • When not in use, cover the microscope with a dust cover to protect it from environmental contaminants.

Warranty and Support

For warranty information, technical support, or service inquiries, please contact AmScope customer service or refer to the official AmScope website. Keep your purchase receipt as proof of purchase for any warranty claims.