1. Introduction and Safety Information
This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your ZIBOO 15B PRO or 17B PRO True RMS Digital Multimeter. Please read this manual thoroughly before use and keep it for future reference.
1.1 Safety Information
This product strictly adheres to the safety standards of European Electrotechnical Commission EN61010-1 and China GB T 13978. Adherence to these guidelines is crucial for safe operation.
- Always read the safety information before using the instrument and operate in strict accordance with the provisions.
- Measure according to the specified range and do not rotate the gear during measurement to avoid damage.
- To avoid possible electrical shock, do not touch live conductors exceeding 30V AC and 60V DC during measurement.
- When measuring AC and DC over 599V, a high voltage lightning symbol will display, and the buzzer will beep. Exercise extreme caution.
- Do not use the multimeter in flammable and explosive environments.
- If the meter displays a low battery symbol, replace the battery immediately to avoid measurement errors.
- Before measuring, check test leads and wires for damage or leakage. Do not use if damaged.
- Comply with national regulations and measurement safety standards. Wear insulating gloves or flame-retardant products when measuring hazardous voltage.
1.2 International Electrical Symbols
| Symbol | Explanation |
|---|---|
![]() | Be careful! Indicates a dangerous operation |
![]() | High current! Indicates a high voltage hazard warning |
![]() | European Union unified directive, European CE certification |
![]() | Grounding |
![]() | Double insulation or enhanced insulation protection |
![]() | Compliance with WEEE recycling instructions |
2. Product Overview
The ZIBOO 15B PRO and 17B PRO are new technology digital multimeters featuring a two-color protective cover, hand-held non-slip ergonomic design, and a large-screen LCD display. They offer multi-functionality, high precision, high performance, anti-drop durability, safety, and stability. These multimeters can accurately measure AC/DC voltage, AC/DC current, frequency, continuity, diode, resistance, capacitance, and include high voltage alarm, NCV (Non-Contact Voltage) and Live wire detection. They also feature misoperation circuit protection, making them suitable for professional electricians, electronic technology maintenance and repair, and DIY electronics enthusiasts.
The 17B PRO model extends DC Voltage measurement up to 1500V.
Video: Demonstration of ZIBOO 17B PRO Multimeter features including NCV, short-circuit test, AC current test, capacitance test, resistance test, AC voltage test, backlight function, and DC voltage test.
3. Key Features and Functions
Both ZIBOO 15B PRO and 17B PRO models offer a comprehensive set of features:
- Safety Rating: CAT III 600V (and CAT II 1000V for 15B PRO, CAT III 600V for 17B PRO) ensuring safety in residential, commercial, and industrial settings.
- Voltage Measurement: AC/DC Voltage up to 1000V (15B PRO) / 1500V (17B PRO).
- Current Measurement: AC/DC Current up to 10A. Optional AC 600A current clamp support.
- Resistance: Measures up to 60MΩ.
- Continuity: Buzzer with green LED flash for quick continuity checks.
- Capacitance: Measures a wide range of capacitance values.
- Diode Test: Quickly identifies faulty diodes.
- Frequency: Measures frequency up to 400kHz.
- Temperature Measurement (17B PRO only): Up to 1000°C with included Type K thermocouple.
- True RMS: Accurate readings for non-sinusoidal waveforms.
- VFD AC 600V (Variable Frequency Drive): For accurate measurements on VFD circuits.
- NCV (Non-Contact Voltage) / LIVE: Red LED flash and buzzing sound for non-contact voltage detection and live wire identification.
- Data Hold: Freezes the display for easier recording.
- Backlight: Bright white backlight for visibility in all lighting conditions.
- 6000 Counts Display: High resolution for precise readings.
- Auto Ranging: Automatically selects the correct measurement range.
- Intelligent Anti-Burning: Built-in double fuse with overload protection to prevent operational errors from damaging the meter.
3.1 Description of Appearance

Image: Front and back view of the ZIBOO 15B PRO Multimeter with dimensions.

Image: ZIBOO 15B PRO Multimeter with test leads.

Image: ZIBOO 17B PRO Multimeter with test leads.
The multimeter features a large LCD display, a central rotary dial for function selection, and several push buttons for additional features. Input jacks are located at the bottom for test leads.
3.2 LCD Display and Symbols
The LCD displays measurement readings, units, and various symbols indicating the current mode or status.
| Number | Symbol | Explanation |
|---|---|---|
| 1 | OFF | Turn off the power |
| 2 | ![]() | Auto power-off |
| 3 | ![]() | Low battery indication |
| 4 | mV | DC millivolt |
| 5 | TRUE RMS | True RMS measurement |
| 6 | V~ | AC voltage |
| 7 | Hz | Frequency |
| 8 | uA, mA, A | DC microampere, milliampere, ampere |
| 9 | uA~, mA~, A~ | AC microampere, milliampere, ampere |
| 10 | Ω, KΩ, MΩ | Resistance Ohm, Kiloohm, Megaohm |
| 11 | ![]() | Continuity beeper, Diode |
| 12 | nF, uF, mF | Capacitance nanofarad, microfarad, millifarad |
| 13 | REL | Clear the relative value |
| 14 | ![]() | Battery 1.5V 9V |
| 15 | NCV EF | Non-contact sensing signal |
| 16 | LIVE | AC Live wire detection |
| 17 | °C/°F | Temperature (17B PRO only) |
| 18 | ![]() | Current Clamp |
| 19 | VFD | AC variable frequency measurement |
4. Technical Specifications
4.1 General Technical Indicators
- Maximum Display: 6000 counts
- Sampling Rate: 3 times/second
- Overload Display: "OL"
- Auto Power Off: Approximately 15 minutes
- Precision: 23°C ± 5°C (18-28°C; Humidity <75%); 1 year
- Operating Temperature: 0°C to 40°C
- Storage Temperature: -10°C to 60°C
- Storage Humidity: ≤80%
- Altitude: 0-2000m
- Pollution Grade: 2
- Safety Codes: CE EN61010-1-2-030
- Electromagnetic Environment: EN61326-1-2
- Power: AA 1.5V x 3 batteries
- Dimensions: 183 x 91 x 45mm
- Weight: 402g (including battery)
4.2 Measurement Ranges and Accuracy
| Function | Range (15B PRO) | Range (17B PRO) | Resolution | Accuracy |
|---|---|---|---|---|
| DC Voltage (DCV) | 60.00mV, 600.0mV, 6.000V, 60.00V, 600.0V, 1000V | 60.00mV, 600.0mV, 6.000V, 60.00V, 600.0V, 1500V | 0.01mV to 1V | ±(0.5%+3) to ±(0.8%+5) |
| AC Voltage (ACV) | 6.000V, 60.00V, 600.0V, 1000V | 6.000V, 60.00V, 600.0V, 1000V | 0.001V to 1V | ±(0.8%+5) to ±(1.2%+5) (400Hz, 1KHz) |
| Frequency (Hz) | 10.00Hz, 100.0Hz, 1.000KHz, 60KHz | 10.00Hz, 100.0Hz, 1.000KHz, 60KHz | 0.01Hz to 0.01KHz | ±(0.5%+5) |
| Resistance (Ω) | 600.0Ω, 6.000KΩ, 60.00KΩ, 600.0KΩ, 6.000MΩ, 60.00MΩ | 600.0Ω, 6.000KΩ, 60.00KΩ, 600.0KΩ, 6.000MΩ, 60.00MΩ | 0.1Ω to 0.01MΩ | ±(1%+5) to ±(2.5%+5) |
| Capacitance | 6.000nF, 60.00nF, 600.0nF, 6.000uF, 600.0uF, 6000uF | 6.000nF, 60.00nF, 600.0nF, 6.000uF, 600.0uF, 60000uF | 0.001nF to 0.1mF | ±(1.5%+5) to ±(8%+5) |
| DC Current (DCA) | 600.0uA, 6.000mA, 60.00mA, 600.0mA, 6.000A, 10.00A | 600.0uA, 6.000mA, 60.00mA, 600.0mA, 6.000A, 10.00A | 0.1uA to 0.01A | ±(0.8%+5) to ±(2%+5) |
| AC Current (ACA) | 600.0uA, 6.000mA, 60.00mA, 600.0mA, 6.000A, 10.00A | 600.0uA, 6.000mA, 60.00mA, 600.0mA, 6.000A, 10.00A | 0.1uA to 0.01A | ±(1%+5) to ±(2%+5) (1KHz) |
| Temperature (°C/°F) (17B PRO only) | N/A | -50°C to 1000°C, -58°F to 1832°F | 1°C / 1°F | ±(1.5%+4) to ±(2%+5) |
| Continuity / Diode | Continuity, Diode | Continuity, Diode | N/A | 0.500V-0.800V (Diode) |
| NCV / LIVE | 50V-1000V | 50V-1000V | N/A | 0-5cm (Non-contact induction) |
| Current Clamp | 60.0A, 600.0A (AC/DC) | 60.0A, 600.0A (AC/DC) | 0.1A | ±(3%+5) |
5. Initial Setup
5.1 Battery Installation
The ZIBOO multimeter requires 3x AA 1.5V batteries (LR06-AA). Batteries are not included and must be installed before first use.
- Ensure the multimeter is turned OFF and disconnect all test leads from the circuit and the meter.
- Locate the battery cover on the back of the multimeter.
- Unscrew the retaining screw(s) on the battery cover.
- Remove the battery cover.
- Insert three AA 1.5V batteries, observing the correct polarity (+/-) as indicated inside the battery compartment.
- Replace the battery cover and securely fasten the screw(s).
Note: After replacing the battery, ensure that the battery cover is tightly screwed before the meter can be used normally.

Image: Internal view showing battery compartment and fuses.
6. Operating Instructions
To operate the multimeter, select the desired function using the rotary dial and connect the test leads appropriately. The meter features auto-ranging for most functions, simplifying operation.
6.1 Connecting Test Leads
- For most voltage, resistance, continuity, diode, and capacitance measurements, connect the red test lead to the "VΩmA" jack and the black test lead to the "COM" jack.
- For high current (up to 10A) measurements, connect the red test lead to the "10A" jack and the black test lead to the "COM" jack.
- For current clamp measurements, connect the current clamp output to the appropriate jacks as per the clamp's instructions (typically "VΩmA" or "Hz" for signal output).
6.2 Measurement Functions
Turn the rotary dial to the desired measurement function. Use the MODE button to cycle through sub-functions if available (e.g., AC/DC for voltage/current, diode/continuity).
- DC Voltage (DCV): Select the "V" position on the dial. The meter will automatically detect DC voltage. Connect leads in parallel with the DC source.
- AC Voltage (ACV): Select the "V~" position on the dial. The meter will automatically detect AC voltage. Connect leads in parallel with the AC source.
- Resistance (Ω): Select the "Ω" position. Connect leads across the component.
- Continuity: Select the "Ω" position and press MODE until the continuity symbol appears. A buzzer sounds for low resistance.
- Diode Test: Select the "Ω" position and press MODE until the diode symbol appears. Connect leads across the diode.
- Capacitance: Select the "nF" position. Connect leads across the capacitor.
- DC Current (DCA): Select the "uA/mA" or "A" position. Connect leads in series with the DC circuit.
- AC Current (ACA): Select the "uA/mA~" or "A~" position. Connect leads in series with the AC circuit.
- Frequency (Hz): Select the "Hz" position. Connect leads across the signal source.
- Temperature (°C/°F) (17B PRO only): Select the "°C/°F" position. Connect the thermocouple to the appropriate jacks.
- NCV (Non-Contact Voltage): Select the "NCV" position. Bring the top of the meter near a live conductor.
- LIVE Wire Detection: Select the "LIVE" position. Use the red probe to touch the suspected live wire.
- VFD (Variable Frequency Drive): Press and hold the MODE button for 3 seconds to activate VFD mode when in AC voltage measurement.
6.3 Button Functions
- HOLD: Short press to freeze the current reading on the display. Press again to release.
- RANGE: Press to switch between auto-ranging and manual ranging. In manual mode, press repeatedly to cycle through ranges.
- MAX/MIN: Press to record maximum and minimum values during a measurement.
- MODE: Press to switch between different sub-functions within a rotary dial position (e.g., AC/DC, Diode/Continuity).
- VFD: Long press MODE to activate VFD mode.
- Backlight (*): Press the backlight button to turn on the display backlight. The backlight automatically turns off after approximately 30 seconds.

Image: Measuring DC voltage.

Image: Measuring AC current with an optional clamp meter.

Image: Measuring resistance.

Image: Measuring capacitance.

Image: Demonstrating NCV function.
7. Maintenance
7.1 Replacing Fuses
To avoid electric shock, disconnect the test leads from the circuit and the meter before replacing fuses.
- Fuse Specification (15B PRO / 17B PRO):
- F1: 500mA/1000V, 6.3x31.5mm
- F2: 10A/1000V, 6.3x31.5mm
The fuses are located inside the meter, accessible after removing the battery cover and potentially additional screws. It is recommended that fuse replacement be performed by a qualified technician.
7.2 Replacing Batteries
To avoid electric shock, disconnect the test leads from the circuit and the meter before replacing batteries.
- Battery Specification: LR06-AA 1.5V x 3
Follow the steps in Section 5.1 "Battery Installation" to replace the batteries.
7.3 Repair and Service
Any repairs requiring the opening of the meter's cover should only be performed by an approved and professionally experienced technician. Improper maintenance can lead to damage to the instrument or risk of electric shock. If a fault cannot be resolved, please contact ZIBOO official after-sales service.
Contact Information:
E-Mail: sales-support@ziboo.online
8. Troubleshooting
If you encounter issues with your ZIBOO multimeter, refer to the following common problems and solutions:
- Meter does not turn on:
- Check if the batteries are installed correctly with the right polarity.
- Replace the batteries if they are depleted. - Inaccurate readings:
- Ensure the correct function is selected on the rotary dial.
- Check if the test leads are fully inserted and not damaged.
- Replace batteries if the low battery indicator is on.
- Ensure the measurement range is appropriate (though auto-ranging handles most cases). - No continuity beep:
- Ensure the meter is in continuity mode (check display symbol).
- Check test leads for breaks or poor connection. - No display or "OL" (Overload) displayed:
- "OL" indicates the measurement exceeds the selected range or the maximum capability of the meter. Switch to a higher range if in manual mode, or ensure the input is within the meter's limits.
- If no display, check battery and power. - Fuse blown during current measurement:
- This usually happens if the meter was connected in parallel for current measurement or if the current exceeded the fuse rating. Always connect in series for current measurement and ensure the expected current is within the meter's limits. Replace the appropriate fuse (see Section 7.1).
9. User Tips
- Accuracy: Users report excellent calibration and accuracy, often comparable to higher-end brands like Fluke. The ZIBOO multimeters are noted for being very precise and reliable.
- Build Quality: The meters are well-built with good internal protection, including ceramic fuses, MOVs, and quality resistors, providing enhanced safety. The rotary switch is described as chunky and sturdy.
- Display: Features a large, bright, and easy-to-read display with a white backlight. Note that the backlight typically stays on for about 30 seconds before automatically turning off to conserve battery.
- Test Leads: While the included test leads are decent, some users suggest upgrading to silicone test leads for better flexibility and durability.
- Battery Compatibility: The meter can operate with three rechargeable NiMh AA batteries.
- Capacitance Measurement: Noted for fast capacitance measurement.
- Portability: The meter is described as robust and suitable for on-the-go electrical system testing.