1. Introduction
The Yieryi YY-1090 is a versatile 5-in-1 digital water quality tester designed for high-precision measurement of pH, Total Dissolved Solids (TDS), Electrical Conductivity (EC), Salinity, and Temperature. This waterproof device (IP67 rated) features a backlit LCD display for easy reading and is suitable for various applications including aquariums, pools, hydroponics, and drinking water analysis.
2. What's in the Box
Upon opening the package, you should find the following items:
- 1 x Water Quality Tester (YY-1090)
- 1 x User Manual
- 1 x pH 4.00 Buffer Powder
- 1 x pH 7.00 Buffer Powder
- 1 x pH 10.01 Buffer Powder
3. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| Model | YY-1090 | ||
| pH | 0.0-14.0pH | 0.1pH | ±0.1pH |
| TDS | 0-9990ppm, 10.1-100.0 ppt | 1ppm; 10ppm; 0.1ppt | ±2% F.S |
| EC | 0-9990µS/cm, 10.01-200.0mS/cm | 1µS/cm; 10µS/cm; 0.01mS/cm | ±2% F.S |
| Salinity | 0-9990ppm, 10.1-100.0 ppt | 1ppm; 10ppm; 0.1ppt | ±2% F.S |
| Temperature | 0.0°C-50.0°C; 32.0°F-122°F | 0.1°C; 0.1°F | ±1°C ±1.8°F |
| Calibration | EC/TDS/SALT: 1413µS/cm; 12.88mS/cm; 111.8mS/cm calibration solution. pH: pH4.0/pH7.0/pH10.0 (Priority calibration of pH7.0 solutions). | ||
| Environment | 0.0°C-50.0°C; 32.0°F-122°F | ||
| Waterproof | IP67 | ||
| Size | 183*45*45mm | ||
| Display | LCD Backlight Display | ||
| Temp. ATC | 0.0°C-50.0°C; 32.0°F-122°F | ||
| Battery | 4*1.5V (LR44); Low battery indicator | ||
| Automatic shut-off | 5 minutes | ||
4. Setup and Installation
4.1. Battery Installation
The YY-1090 tester uses 4 x 1.5V LR44 button batteries. Ensure correct polarity when installing.
- Unscrew the battery protective cover at the top of the device.
- Remove the old batteries if present.
- Insert 4 new 1.5V LR44 batteries, paying attention to the positive and negative terminals.
- Screw the battery protective cover back on securely.
4.2. Probe Replacement
If the probe needs to be replaced due to damage or wear, follow these steps:
- Screw off the ring that secures the probe.
- Unplug the old probe from the main unit.
- Plug in the new probe.
- Screw the ring back on to secure the new probe.
Note: Replaceable electrodes have a 6-month warranty for non-human damage.
5. Operation
5.1. Basic Usage
- Remove the protective cap from the probe.
- Press the ON/OFF/HOLD button to turn on the instrument.
- Immerse the probe into the water sample you wish to test.
- Gently stir the solution and wait for the reading to stabilize on the LCD display.
- After use, rinse the probe with purified water and replace the protective cap.
5.2. Key Functions
- Lock Reading (HOLD): Press the ON/OFF/HOLD button briefly during measurement to lock the current reading. Press again to unlock.
- °C/°F Conversion: Press the TEMP/PH CAL button briefly to switch between Celsius (°C) and Fahrenheit (°F) temperature units.
- Mode Conversion: Press the MODE/EC CAL button briefly to cycle through the measurement modes: TDS (ppm), Salinity (ppm), EC (µS/cm), and pH.
6. Calibration
The tester is factory calibrated and ready to use. However, recalibration may be necessary over time or if readings appear inaccurate.
6.1. pH Calibration
Always perform pH 7.00 calibration first, then pH 4.00 and pH 10.01 if needed.
- **Prepare pH 7.00 solution:** Pour 1 packet of pH 7.00 buffer powder into a cup, add 250ml of purified or tap water, stir, and wait 5-10 minutes for the powder to dissolve completely.
- Turn on the tester by pressing the ON/OFF/HOLD button.
- Immerse the probe into the pH 7.00 calibration solution.
- Press and hold the TEMP/PH CAL button until "CAL" and "7.0" start blinking 3 times on the screen. This indicates successful calibration.
- Rinse the probe with purified water after calibration.
- Repeat the process for pH 4.00 and pH 10.01 solutions if necessary, using their respective buffer powders.
6.2. EC/TDS/Salinity Calibration
Use a 12.88mS/cm calibration solution for EC/TDS/Salinity calibration.
- Turn on the meter and switch to EC mode by pressing the MODE/EC CAL button until "µS/cm" is displayed.
- Immerse the probe into the 12.88mS/cm calibration solution and stir gently.
- Wait for the reading to stabilize.
- Press and hold the MODE/EC CAL button for 5 seconds and release. "CAL" will be displayed, and a value will flash 3 times.
- While the value is flashing, press the MODE/EC CAL button if the flashing value is higher than 12.88mS/cm, or the TEMP/PH CAL button if the flashing value is lower than 12.88mS/cm, to adjust the displayed value to 12.88mS/cm.
- Release the button and wait for the reading to stop flashing. The calibration is complete.
- Rinse the probe with purified water and replace the protective cap.
7. Maintenance
- **Cleaning:** After each use, always wash the electrode with purified water and keep it dry.
- **Probe Care:** The glass electrode is fragile. Handle it gently and avoid touching the bottom or edges of containers to prevent damage.
- **Storage:** Store the tester with the protective cap on to keep the probe moist and protected.
- **Battery Replacement:** Replace batteries when the low battery indicator flashes on the display. Use 4 x 1.5V LR44 batteries.
8. Troubleshooting
- **Jumping Readings:** When the test probe encounters air, it can generate resistance, causing the value to jump. This is normal. Gently stir the solution to remove air bubbles around the electrode.
- **Inaccurate Readings:** If readings become consistently inaccurate, recalibration (see Section 6) is recommended. Ensure calibration solutions are fresh and correctly prepared.
- **No Display/Power Issues:** Check battery installation and ensure batteries are not depleted. Replace if necessary.
9. User Tips
- For optimal accuracy, especially with pH measurements, ensure the probe is fully immersed in the solution and gently stirred to remove any air bubbles.
- Always use fresh calibration solutions for best results. pH buffer powders are single-use once mixed.
- Keep the protective cap on the probe when not in use to prevent it from drying out, which can affect its lifespan and accuracy.
- The device has an automatic shut-off feature after 5 minutes of inactivity to conserve battery life.
10. Warranty and Support
The Yieryi YY-1090 Water Quality Tester comes with a 1-year warranty for the main unit against non-artificial damage. The electrodes are considered consumables with a shelf life of 6 months, and it is recommended to replace them every 6 months for optimal performance.
For any questions or support regarding your purchase, please contact the seller or manufacturer directly.