Electrolysis Epilator Removal Needle Instruction Manual

Model: TG01, TG03, TG09, K2, K4, K5

Brand: Yingcai

1. Introduction

1.1 Product Overview

The Electrolysis Hair Removal Machine is a professional beauty device designed for permanent hair removal. This system is widely used in beauty salons, aesthetic clinics, and professional hair removal centers. It provides stable output, flexible treatment modes, and high safety control through pedal operation.

1.2 Main Applications

Electrolysis is a proven method that destroys the hair follicle structure through controlled energy output. It is suitable for almost all hair colors, including light or fine hair.

Suitable Areas:

  • Upper lip
  • Chin
  • Face hair
  • Ear
  • Underarm
  • Bikini line
  • Legs
  • Arms
  • Chest and back
Diagram showing various body parts suitable for electrolysis hair removal, including armpit, face, bikini, body, leg, and arm.
Figure 1: Areas suitable for electrolysis treatment.

2. Safety Information

2.1 General Safety Requirements

  • Before using the machine, always wear an anti-static wrist strap to protect the device and ensure safety.
  • This device is for professional use only.
  • Do not use if packaging is opened or damaged.
  • Do not use if packaging is opened or damaged.
Warning signs for professional use only and not to use if packaging is damaged.
Figure 2: Important safety warnings.

2.2 Contraindications

Do not perform electrolysis if the client has heart disease, hypertension, diabetes, skin inflammation, scar constitution, pregnancy, or lactation. The operation is not recommended if any contraindication exists.

3. Product Components

3.1 Probe System Overview

The machine is designed with four functional probes to support different treatment modes. Each probe connects to a dedicated port on the back panel. By selecting the correct probe and working mode, the operator can perform professional skin treatment and hair removal functions.

The 4 probes include:

  1. Electrolysis Needle Handle (Hair Removal Probe)
  2. Foot Switch Pedal (Control Probe)
  3. Grounding Wrist Strap (Safety Probe)
  4. Power Input Connector (Power Probe)

Each probe has a specific purpose to ensure stable performance and safe operation.

3.2 1. Electrolysis Needle Handle (Treatment Probe)

This is the main working handpiece of the machine. It delivers micro electrical energy to the treatment area through the needle tip. The current targets the hair follicle root and damages the growth structure, which helps achieve permanent hair reduction.

How to Connect:

  • Locate the metal circular connector port on the back panel.
  • Align the handpiece connector with the port.
  • Push in firmly and twist to lock (if it is a locking connector).
Electrolysis needle handle in use for hair removal.
Figure 3: Electrolysis Needle Handle in operation.

3.3 2. Foot Switch Pedal (Foot Control Probe)

The foot pedal is used to activate the energy output without using hands. When the operator presses the foot pedal, the machine releases current output to the handpiece. Energy works only when the pedal is pressed. Releasing the pedal immediately stops output. This ensures better safety and accuracy, especially for delicate facial treatment.

How to Connect:

  • Plug the pedal cable into the large round connector port (usually bigger than the handpiece port).
  • Tighten the metal ring to secure the connection.

3.4 3. Grounding Wrist Strap (Safety Probe)

The blue wrist strap is a grounding safety accessory. It creates a stable circuit and reduces electrical instability during treatment. It helps to stabilize output flow, reduce discomfort, improve safety performance, and prevent static interference. The client wears the wrist strap during treatment.

How to Connect:

  • Insert the blue cable plug into the blue port on the machine.
  • Ensure the connector is fully inserted.

3.5 4. Power Input Connector (Power Probe)

This is the main power supply connection for the machine. The power adapter converts the wall electricity into a stable voltage that the machine can safely use.

How to Connect:

  • Plug the power cable into the DC power socket (right side with black connector).
  • Then connect the adapter into the wall socket.

4. Setup Guide

4.1 Step-by-Step Setup (Professional Operation)

Follow this order for best performance:

  1. Place the machine: Put the machine on a flat, stable table with enough airflow around ventilation holes.
  2. Connect the handpiece: Insert the electrolysis needle handle connector into the handpiece port.
  3. Connect the foot pedal: Insert the foot pedal cable into its designated port.
  4. Connect the grounding wrist strap: Plug the blue strap cable into the blue port, then ensure the client wears it.
  5. Connect the power supply: Plug the adapter into the machine, then into wall electricity.
  6. Turn on the machine: Switch ON the device and select the treatment mode on the screen.
Image showing compatible electrolysis machines: TG01, TG03, TG09.
Figure 4: Compatible Electrolysis Machines.

5. Operating Instructions

5.1 Control Panel Functions

Control panel of the electrolysis machine with various buttons and displays.
Figure 5: Electrolysis Machine Control Panel.

Button Descriptions:

  • START: Starts the treatment and time counting.
  • PAUSE: Temporarily pauses treatment and time.
  • RESET: Resets treatment time.
  • POWER: Adjusts current intensity for FLASH and THERMOLIZ modes (gear: 1-9).

Mode Explanations:

  • Flash Mode (Intermittent Energy Output or point to point): Technology principle is Thermolysis. Energy is released in short pulses, suitable for fast treatment and large areas.
  • THER or Continuous Mode (Thermolysis): Technology principle is Thermolysis, providing continuous energy output, stronger stimulation, and higher heat concentration.
  • Blend Mode (FLASH + BLEND / THER + BLEND): Technology principle combines Thermolysis + Electrolysis. It combines thermal energy and electrochemical reaction, resulting in lower skin stimulation compared to single modes. This mode is often described as low-pain or painless.

Note:

  • Single modes produce stronger skin stimulation.
  • Continuous mode has the highest stimulation.
  • Blend mode significantly reduces pain and discomfort.

Display Parameters for Each Mode:

  • Flash Mode (Thermolysis):
    • PULSE: Pulse energy level (1 – 9)
    • TIME: Pulse output duration (1 – 9)
    • TIMER: Silent key (inactive)
    • POWER: Total energy / current intensity (1 – 9)
  • Continuous Mode (Thermolysis):
    • PULSE: Silent key
    • TIME: Silent key
    • TIMER: Continuous energy output time (1 – 9)
    • POWER: Total energy / current intensity (1 – 9)
  • Blend Mode (FLASH + BLEND / THER + BLEND):
    • PULSE: Pulse energy (1 – 9)
    • TIME: Pulse output duration (1 – 9)
    • TIMER: Silent key
    • POWER: Total energy / current intensity (1 – 9)

5.2 Pre-Operation Preparation

Equipment & Consumables Inspection:

  • Check the integrity of the electrolysis hair removal machine (main unit, wires) and insulated needles.
  • Conduct a power-on test to ensure stable current output and normal operation.
  • Prepare consumables: insulated needles, medical alcohol, iodophor, sterile gauze, disposable gloves, conductive gel, physiological saline, hemostatic cotton swabs, etc.

Client Communication & Evaluation:

  • Inquire about the client's health status and screen for contraindications.
  • Confirm the hair removal area (e.g., lip hair, armpits, bikini line) and check skin condition; avoid areas with damage, redness, swelling, or allergy.
  • Test the client's current tolerance with a low gear on a small local area.

Skin Disinfection:

  • Wipe the hair removal area 2-3 times with medical alcohol cotton balls to remove oil and dirt.
  • For sensitive skin: disinfect with iodophor first, then remove iodine residue with physiological saline to avoid irritation.
  • Ask the client to remove all metal jewelry from the hair removal area to prevent abnormal current conduction.

5.3 Electrolysis Operation

Hair Follicle Positioning & Needle Insertion:

  • Observe hair follicle position and determine hair growth direction with a magnifying glass or special lighting equipment (needle insertion along the growth direction reduces pain and skin damage).
  • Hold the insulated needle, align the needle tip with the hair follicle opening, and gently rotate to insert it into the hair follicle root (depth: 0.5-1mm, adjusted by hair thickness; avoid over-insertion to prevent dermal damage).
  • Ensure the needle tip reaches the hair bulb (the key target of electrolysis).
Diagram illustrating how electrolysis works by inserting a needle into the hair follicle and applying current.
Figure 6: How Electrolysis Works.

Current Parameter Setting & Operation:

  • Set parameters based on the hair removal area and client's tolerance.
  • Pause and adjust parameters if the client reports obvious pain.
  • After completing electrolysis for a single hair follicle: turn off the current, gently pull out the needle, and press the hair follicle opening with a hemostatic cotton swab for a few seconds to prevent minor bleeding.

Sequential Operation & Density Control:

  • Operate in the order of top to bottom, left to right to avoid omissions.
  • Keep an interval of 1-2mm between adjacent hair follicles to prevent excessive skin irritation.
  • For dense hair areas: operate in batches; single operation duration ≤20 minutes to avoid skin redness and swelling caused by long-term current stimulation.

Post-Treatment Care:

  • After treatment, apply soothing gel, aloe gel, or a cold compress to reduce redness and irritation.
Image showing application of soothing gel after hair removal.
Figure 7: Post-treatment cooling and skin protection.

6. Maintenance

Regular maintenance is crucial for the longevity and optimal performance of your electrolysis machine. After each use, ensure all probes and accessories are cleaned and disinfected according to professional hygiene standards. Store the device in a clean, dry place away from direct sunlight and extreme temperatures. Periodically check cables for any signs of wear or damage.

7. Troubleshooting

If you encounter any issues with your electrolysis machine, please refer to the following basic troubleshooting steps:

  • No Power: Ensure the power cable is securely connected to both the machine and the wall socket. Check if the power outlet is functional.
  • No Energy Output: Verify that the foot pedal is correctly connected and being pressed during operation. Ensure the needle handle is properly inserted and locked. Check the selected mode and power settings on the control panel.
  • Needle Insertion Difficulty/Pain: Re-evaluate skin stretching and needle angle. Ensure the needle is inserted along the natural hair growth direction. Do not force the needle. Adjust parameters to a lower setting if pain persists.
  • Excessive Redness/Irritation: Review treatment parameters (pulse, time, power) and ensure they are appropriate for the client's skin sensitivity and hair thickness. Ensure proper skin preparation and post-treatment care are followed.

For persistent issues, please contact customer support.

8. Specifications

8.1 Needle Specifications

  • Material: 24ct Gold-Plated Surface
  • Diameter: 0.100mm (Ultra-Fine)
  • Shank Size: 0.80mm
  • Features: Superior electrical conductivity, hypoallergenic & corrosion-resistant, pre-sterilized & single-use.
  • Models: TG01, TG03, TG09, K2, K4, K5
Image highlighting key features of the electrolysis needles: 24ct Gold-Plated Surface, Ultra-Fine 0.100mm Diameter, 0.80mm shank size.
Figure 8: Key Features & Benefits of the Needles.

8.2 Machine Specifications

  • Treatment Modes: Flash, THER (Continuous), Blend (Flash + Blend / THER + Blend)
  • Output Control: Pedal operation
  • Power Input: Standard wall electricity via adapter

8.3 Package Contents

  • Electrolysis Needles (e.g., 50 filaments per box, 5 boards of 10 needles)
  • User Manual (PDF available: View PDF)
Diagram showing machine packaging with 5 boards of 10 needles, totaling 50 needles per box.
Figure 9: Machine Packaging and Needle Quantity.

8.4 Difference between K2 and K5 needles

The K2 and K5 needles, like other models (TG01, TG03, TG09, K4), are designed for electrolysis hair removal. While specific differences in taper or tip design are not detailed, they generally vary in their suitability for different hair types (fine vs. coarse) and skin sensitivities. Always select a suitable disposable needle size according to the hair thickness and treatment area.

9. User Tips

  • Users have noted that the needles penetrate easily, making them easy to use for effective treatment.
  • Always ensure the packaging is sealed and undamaged before use to maintain sterility and safety.
  • For optimal results and minimal discomfort, ensure the needle is inserted correctly into the hair follicle, following the natural growth direction, and activate the current only after proper insertion.

10. Warranty and Support

For any product inquiries, technical support, or warranty claims, please contact the seller directly through the platform where the product was purchased. Ensure you have your purchase details readily available.

11. How Electrolysis Works (Video)

Video demonstrating the process of electrolysis hair removal using the needle.
Documents and downloads