Pintech P6018A High-Voltage Probe User Manual

Model: P6018A

1. Introduction

The Pintech P6018A is a high-voltage probe designed for accurate measurement of high DC and AC voltages up to 18kV DC and 12kV RMS AC. It features a 1000:1 attenuation ratio and a bandwidth of DC-100MHz, making it suitable for a wide range of high-voltage testing applications with oscilloscopes.

2. Safety Information

WARNING: High voltages can be lethal. Always exercise extreme caution when working with high-voltage equipment. Read and understand all safety instructions before using this probe.

  • Ensure the oscilloscope and the probe are properly grounded before making any connections.
  • Never exceed the maximum voltage ratings of the probe (18kV DC, 12kV RMS AC, 12kVp-p pulse).
  • Always connect the ground clip to the circuit's ground point before connecting the probe tip to the high-voltage source. Disconnect the probe tip first, then the ground clip.
  • Do not use the probe if it appears damaged or if any part of the insulation is compromised.
  • Operate the probe within the specified environmental conditions (temperature and humidity).
  • Keep hands and other body parts away from high-voltage points when the circuit is energized.

3. Package Contents

Verify that all items are present in your package:

  • Pintech P6018A High-Voltage Probe
  • BNC Cable (integrated)
  • Crocodile Clips (pair)
  • Test Sticks (pair)
  • Accessory Kit (includes hook, nut, flat clip)
Pintech P6018A High-Voltage Probe with all accessories
Figure 3.1: Pintech P6018A High-Voltage Probe and included accessories.
Close-up of accessory kit, crocodile clips, and test sticks
Figure 3.2: Detailed view of the accessory kit, crocodile clips, and test sticks.

4. Setup Instructions

  1. Connect to Oscilloscope: Connect the BNC connector of the P6018A probe cable to an input channel of your oscilloscope.
  2. Set Oscilloscope Attenuation: Due to the probe's 1000:1 attenuation ratio, you must set your oscilloscope's channel attenuation to 1000X (or 1000:1) to obtain accurate voltage readings. Refer to your oscilloscope's manual for instructions on adjusting channel attenuation.
  3. Attach Accessories:
    • For general measurements, attach the desired test stick or hook from the accessory kit to the probe tip.
    • For grounding, connect one of the crocodile clips to the ground lead of the probe.
  4. Grounding: Always ensure the probe's ground clip is securely connected to the ground reference of the circuit under test before making any high-voltage measurements.

5. Operating the Probe

  1. Power Off Circuit: Before connecting the probe to the high-voltage source, ensure the circuit under test is de-energized.
  2. Connect Ground: Securely attach the probe's ground clip to the circuit's ground point.
  3. Connect Probe Tip: Carefully connect the probe tip (with the appropriate accessory) to the high-voltage test point.
  4. Energize Circuit: Once all connections are secure and you are at a safe distance, energize the circuit under test.
  5. Observe Waveform: Observe the voltage waveform on your oscilloscope. The oscilloscope will display the actual voltage, compensated for the 1000:1 attenuation, if set correctly.
  6. De-energize and Disconnect: Before disconnecting the probe, de-energize the circuit. First, remove the probe tip from the high-voltage point, then disconnect the ground clip.

Frequency Response

The probe's bandwidth is DC-100MHz (-3dB). The maximum measurable voltage decreases as frequency increases. Refer to the frequency response graph for detailed performance characteristics.

Frequency response graph showing Voltage (DC+AC peak) vs. Frequency
Figure 5.1: P6018A High-Voltage Probe Frequency Response Curve.

6. Maintenance

  • Cleaning: Clean the probe body and cable with a soft, dry cloth. If necessary, use a cloth lightly dampened with water or a mild cleaning solution. Do not use abrasive cleaners or solvents. Ensure the probe is completely dry before use.
  • Storage: Store the probe in a clean, dry environment, away from direct sunlight and extreme temperatures. Protect the probe tip and cable from physical damage.
  • Inspection: Regularly inspect the probe cable, connectors, and housing for any signs of wear, cuts, or damage. Do not use a damaged probe.

7. Troubleshooting

No Signal on Oscilloscope:
  • Check all cable connections between the probe and the oscilloscope.
  • Ensure the oscilloscope channel is enabled and properly configured.
  • Verify the circuit under test is energized and providing a signal.
Incorrect Voltage Reading:
  • Confirm that the oscilloscope's channel attenuation is set to 1000X (1000:1) to match the probe's attenuation ratio.
  • Check the probe's ground connection. A poor ground can lead to inaccurate readings.
  • Ensure the input voltage does not exceed the probe's maximum ratings.
Distorted Waveform:
  • Check for proper grounding.
  • Ensure the probe is within its specified bandwidth for the frequency being measured.
  • Verify the oscilloscope's input impedance and coupling settings.

8. Specifications

ParameterValue
ModelP6018A
Division Ratio1000:1
Pulse Voltage12kVp-p
Input Resistance200MΩ
Input Capacitor1.5 pF
Max. DC Voltage18kV (for 30 sec)
Max. AC RMS Voltage12kV (for 30 sec)
Max. Loading Current90uA
BandwidthDC-100MHz (-3dB)
Rise Time3.5 ns
Signal/Noise>60dB at 1kHz; >50dB at 1MHz
DCV Accuracy≤3%
ACV Accuracy≤3% at 60Hz (0-10kV rms), 5% at 60Hz (10-20kV rms)
Working Temperature-10~55℃
Working Humidity≤85%RH
Weight460g
Dimensions (LWH)80 x 80 x 320 mm
Dimensional drawing of the P6018A probe and accessories with measurements in mm
Figure 8.1: Dimensional drawing of the P6018A High-Voltage Probe and accessories.

9. User Tips

  • Always double-check your oscilloscope's attenuation setting to ensure it matches the probe's 1000:1 ratio for accurate readings.
  • For stable measurements, ensure a solid and low-impedance ground connection.
  • When measuring AC voltages, be mindful of the frequency and refer to the probe's frequency response curve to understand potential voltage derating at higher frequencies.
  • Keep the probe tip and cable clean and free from dust or moisture, which can affect insulation and measurement accuracy.

10. Warranty and Support

This product is provided without an explicit warranty. For technical support or inquiries regarding the Pintech P6018A High-Voltage Probe, please contact the manufacturer or your point of purchase.

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.

Was this page helpful?