1. Introduction
The COXO C-Smart I Pilot is an advanced 5-in-1 dental endomotor system designed to meet the complex needs of various dental treatments. This versatile device integrates an endo motor, apex locator, endo activator, ortho polishing, and torque wrench functionalities into a single, intuitive unit. Featuring Bluetooth connectivity, an independent handpiece, a high-definition color LCD, and a powerful brushless motor, the C-Smart I Pilot offers precision, efficiency, and ease of use for dental professionals.

2. Safety Information
⚠ Disclaimer:
This item may be intended for use by professional health care providers only. The seller may ask you to provide evidence that you are a qualified health care provider and are entitled to buy and use certain items in your country.
Please read this manual thoroughly before operating the device. Improper use can lead to injury or damage to the equipment. Always follow standard dental safety protocols and local regulations.
- Ensure the device is used by qualified and trained dental professionals only.
- Inspect the device and accessories for any damage before each use.
- Do not attempt to repair or modify the device yourself. Contact authorized service personnel for assistance.
- Keep the device away from water and other liquids to prevent electric shock.
- Use only original COXO accessories and attachments.
3. Product Overview
The C-Smart I Pilot system consists of a main host unit and a wireless handpiece, designed for convenient and intelligent operation.

3.1 Key Features
- Bluetooth Connectivity: Automatic matching, stable data transmission, and two-way control for seamless operation.
- 5-in-1 Functionality: Integrates Endo motor, Apex Locator, Endo Activator, Ortho, and Polishing modes, along with a Torque Wrench function.
- Synchronized Root Canal Enlargement and Measurement: Combines functions for improved efficiency and accuracy.
- Personalized Customization: Stores data for up to four users with editable program names.
- Independent Handpiece: Nine root canal programs can be activated even when disconnected, ensuring continuous treatment.
- High-Definition Large Screen: Color LCD with an intuitive graphical interface and full touch screen.
- Multiple Torque Actions: Programmable motor actions (reverse/stop/reciprocating) when set torque is reached.
- High-Performance Brushless Motor: Max speed 3000 rpm, max torque 5 N·cm, extended service life.
- Enhanced Ease of Use: Root canal sensitivity settings, automatic start/stop, one-button speed adjustment on handle.
- Contactless Charging: Enhanced safety and convenience without a charging fork.
- Mini Head: Mini 6:1 bending file with a head height of only 10.2 mm and 8mm head diameter.

4. Specifications
| Parameter | Value |
|---|---|
| Brand Name | COXO |
| Model Number | C-SMART-I PILOT |
| Product Name | C-SMART-I PILOT Endo Motor |
| Adaptor Working Voltage | AC100~240V, 50/60Hz |
| Rotation Speed Range | 100~3000rpm |
| Torque Range | 0.6~5.0N.cm |
| Motion | Reciprocating and Rotary |
| Contra Angle Head | 360 degree rotation head, Mini 6:1 bender, headstock height 10.2mm |
| Head Diameter | 8 mm |
| Operating Mode | Short-term operation |
| Electric Shock Protection Level | Type B |
| Commodity Quality Certification | CE |
| Instrument Classification | Class II |
| Packing Material | Carton |
| Color | Black |

5. Setup
Before first use, ensure all components are present and undamaged. Refer to Figure 8 for package contents.

5.1 What's in the Box
- COXO C-Smart I Pilot Main Host Unit
- Wireless Handpiece
- Power Adaptor
- User Manual
- Various accessories (e.g., Apex Locator cable, contra-angle attachments, polishing cups, torque wrench attachments, endo activators)
5.2 Initial Charging
- Connect the power adaptor to the main host unit and a suitable power outlet.
- Place the handpiece onto the charging dock of the main unit. The contactless charging system will begin charging automatically.
- Ensure the device is fully charged before first use. The battery indicator on the handpiece and main unit display will show charging status.
5.3 Handpiece Connection
The handpiece connects to the main unit via Bluetooth. This connection is typically automatic upon powering on both devices. The system is designed for stable data transmission and two-way control.

6. Operating Instructions
The C-Smart I Pilot features a large color LCD with a full touch screen for easy navigation and control.

6.1 Power On/Off
- Press the power button on the main host unit to turn it on.
- The handpiece will automatically connect via Bluetooth.
- To power off, press and hold the power button on the main unit.
6.2 Mode Selection (5-in-1 Functionality)
The system offers five primary modes: Endo motor, Apex Locator, Endo Activator, Ortho, and Polishing. Select the desired mode via the touch screen interface on the main unit.

6.2.1 Endo Motor Mode
- Select 'Endo Motor' on the main unit's display.
- Choose from nine pre-set root canal programs or customize settings for speed (100-3000 rpm) and torque (0.6-5.0 N·cm).
- The handpiece features a screen and buttons for quick adjustment of speed and program switching.
- The brushless motor provides stable performance for root canal preparation.
- The handpiece can operate independently for nine root canal programs even if the wireless connection to the main unit is temporarily lost.
6.2.2 Apex Locator Mode
- Attach the apex locator cable and file clip to the handpiece and the patient.
- The system provides real-time root canal length measurement on the main unit's display.
- This mode can be synchronized with the Endo Motor for simultaneous enlargement and measurement.
6.2.3 Endo Activator Mode (C1-EA)
- Attach the C1-EA Endo Activator tip to the handpiece.
- Select 'Activator' mode on the main unit.
- Use for irrigation and activation of root canal disinfectants.
6.2.4 Ortho Mode (C-IB1)
- Attach the C-IB1 Ortho attachment to the handpiece.
- Select 'Ortho' mode on the main unit.
- Follow specific orthodontic procedures as required.
6.2.5 Polishing Mode
- Attach a polishing cup to the handpiece.
- Select 'Polishing' mode on the main unit.
- Adjust speed as needed for dental polishing procedures.
6.2.6 Torque Wrench Function (C-TW3)
- Attach the C-TW3 Torque Wrench attachment to the handpiece.
- Select 'Torque Wrench' mode on the main unit.
- Use for precise torque control during implant placement or other procedures requiring specific torque values.
6.3 Personalized Settings
- Access the 'User' menu on the main unit.
- Up to four user profiles can be stored.
- Edit user names and program names for easy identification and recall of preferred settings.
7. Maintenance
Proper maintenance ensures the longevity and optimal performance of your COXO C-Smart I Pilot system.
- Cleaning: After each use, clean the handpiece and attachments according to standard sterilization protocols for dental instruments. Use appropriate disinfectants.
- Sterilization: The contra-angle head and other autoclavable components should be sterilized in an autoclave as per manufacturer guidelines and local regulations.
- Battery Care: Keep the handpiece charged when not in use. Avoid fully discharging the battery frequently to prolong its lifespan.
- Storage: Store the device in a clean, dry environment, away from direct sunlight and extreme temperatures.
- Inspection: Regularly inspect cables, connections, and the handpiece for any signs of wear or damage.
8. Troubleshooting
This section addresses common issues you might encounter with your C-Smart I Pilot system.
| Problem | Possible Cause | Solution |
|---|---|---|
| Handpiece not charging | Not properly seated on charging dock; Power adaptor not connected; Faulty adaptor/unit. | Ensure handpiece is correctly placed; Check power connections; Try a different power outlet; Contact support if issue persists. |
| Bluetooth disconnection | Out of range; Interference; Temporary glitch. | Move handpiece closer to main unit (within 10m); Restart both devices; Ensure no strong electromagnetic interference. |
| Motor not rotating | Low battery; Incorrect mode/settings; Handpiece not properly attached. | Charge handpiece; Verify selected mode and parameters; Ensure contra-angle is securely attached. |
| Inaccurate Apex Locator readings | Poor contact; Dry canal; Faulty cable/clip. | Ensure good contact with file and patient; Ensure canal is moist; Check cable and clip for damage. |
| Touch screen unresponsive | Software glitch; Screen damage. | Restart the main unit; If persistent, contact support. |
9. User Tips
- Pre-program Settings: Take advantage of the personalized customization feature to save your frequently used settings for different procedures and users. This saves time during treatment.
- Regular Calibration: For optimal accuracy, especially with the Apex Locator, ensure regular calibration and check connections before each use.
- Ergonomics: The independent handpiece allows for greater freedom of movement. Position the main unit conveniently for easy touch screen access while maintaining an ergonomic working posture.
- Attachment Management: Keep all attachments organized and clean. Having them readily available and sterilized will streamline your workflow.
10. Warranty & Support
The COXO C-Smart I Pilot Advanced Endomotor System comes with a 1-year warranty. For any technical assistance, service, or warranty claims, please contact your local distributor or the manufacturer directly.
After-sale Service: Online Technical Support
Warranty Period: 1 Year
Shelf Life: 1 Year