1. Introduction
The SWEVY Digital Refractometer SW-12S is a high-precision instrument designed for measuring salinity and concentration. It features a wide measurement range of 0-100‰ (parts per thousand) with automatic temperature compensation, making it suitable for various applications such as aquaculture, food processing (e.g., kimchi), laboratory analysis, and hydroponics. Its portable design and digital display ensure ease of use and accurate, instant results.

Figure 1: SWEVY Digital Refractometer SW-12S with highlighted features.
Video 1: Product demonstration of the SWEVY Digital Refractometer, showing measurement and features.
2. Specifications
| Parameter Name | Value |
|---|---|
| Product Model | SW-12S |
| Measurement Range | 0-100‰ |
| Resolution | 1‰ |
| Measurement Accuracy | ±2‰ |
| Temperature Compensation | 10-40°C |
| Ambient Temperature | 10-40°C |
| Battery | 2*AAA (Batteries not included) |
| Size | 114*58*31mm |
| Weight | 127g |

Figure 2: Detailed product parameters.
3. Parts Overview
Familiarize yourself with the components of your SW-12S Digital Refractometer:

Figure 3: SW-12S Refractometer body size and labeled parts.
- Measuring Slot: Where the sample is placed for measurement.
- Prism: The optical component for refractive index measurement.
- LCD: Liquid Crystal Display for showing measurement results, temperature, and refractive index.
- Switch Key (Power Button): Press for 1 second to turn the device on or off.
- Zero Key: Press for 1 second to perform zero calibration.
- Measuring Key: Initiates the measurement process.
- USB Charging Interface: For charging the device (ensure DC6V limit is not exceeded).

Figure 4: Close-up of the high-precision sample slot.
4. Setup
4.1 Battery Installation
The SW-12S refractometer requires 2 AAA batteries (not included). To install:
- Locate the battery compartment on the back of the device.
- Open the battery compartment cover.
- Insert 2 AAA batteries, ensuring correct polarity (+/-).
- Close the battery compartment cover securely.

Figure 5: Battery installation process.
4.2 Initial Power On
Press the power button (Switch Key) for 1 second to turn on the device. The LCD will light up.
5. Operation
5.1 Zero-Setting (Calibration)
Before each use, or if accuracy is in doubt, perform a zero-setting calibration using clean water or pure water (0.0‰ salinity).
- Ensure the measuring slot is clean and dry.
- Drop at least 0.3 ml of clean water or pure water into the measuring slot. Ensure the water covers the prism completely.
- Wait for approximately 20 seconds to allow for temperature compensation. The instrument and the measuring solution need to maintain the same temperature for accurate readings.
- Press the "Zero" key for 1 second. The display will show "Cleared".
- Check if the measured value is 00.0‰. If not, repeat the zero-setting process.

Figure 6: Visual guide for zero-setting and measurement steps.
5.2 Measurement
After successful zero-setting, you can proceed with sample measurement:
- Clean the measuring slot thoroughly and dry it.
- Drop at least 0.3 ml of your sample solution into the measuring slot. Ensure the sample covers the prism completely.
- Wait for approximately 20 seconds for temperature compensation.
- Press the "Meas" key to display the measurement value. The LCD will show concentration, temperature, and refractive index.
- For best accuracy, suck out the measuring solution and measure several times to ensure data reliability.

Figure 7: LCD displaying concentration, temperature, and refractive index.
5.3 Language Switching
To switch between Chinese and English interface:
- While the device is powered on, simultaneously hold down the power button and the "Zero" button. The language will toggle.
6. Maintenance
6.1 Cleaning
- Regular Cleaning: If there is dirt in the measuring slot or if the instrument has not been used for a long time, it should be cleaned.
- Cleaning Procedure: Drip clean water into the measuring slot and wash with paper towels more than three times. Always discard the towels after wiping.
- Stubborn Residue: If the measuring slot contains oiliness or is difficult to clean, use ethanol or soapy water.
- Important: Do not use metal or hard objects to clean the prism, as this may scratch the optical surface.
6.2 Storage
- When not in use, please store the refractometer in a dry place.
- Avoid collision and falls to prevent damage to the precise optical components.
- The USB charging voltage should not exceed DC6V.
7. Troubleshooting
| Error Message / Status Notice | Reason / Solution |
|---|---|
| Cover with hand | External light is too strong to measure. Ensure the prism is not exposed to direct strong light. |
| High temperature | Temperature excess in measuring slot. Allow the device and sample to cool down to the operating temperature range (10-40°C). |
| Low temperature | Temperature too low in measuring slot. Allow the device and sample to warm up to the operating temperature range (10-40°C). |
| Re zero | Please follow the zero-clearing step again. Ensure sufficient clean water is used and the process is completed correctly. |
| Less solution | Too little solution in measuring slot. Ensure at least 0.3 ml of sample is used. |
| Less water | Too little clean water in measuring slot (during zeroing). Ensure at least 0.3 ml of clean water is used. |
| Out of range | Excessive concentration. The sample's concentration exceeds the device's measurement range. |
| Low voltage | Charge with 5V charger or computer USB. |
| Call maintenance | Possible damage to components. Contact customer support for assistance. |
8. User Tips
- When using the refractometer, especially for home or non-laboratory use, it can be highly effective for "relative" before/after comparisons or for monitoring known solutions. While the 0-100‰ range is specified, understand that pure salt in the sensing bowl would not yield a proper reading. Focus on the refractive index (RI) readings for a more scientific understanding of concentration.
- Always ensure the sample slot is clean and dry before each measurement to prevent contamination and ensure accuracy.
- Allow sufficient time (approx. 20 seconds) for temperature compensation, especially when testing samples at different temperatures than the device, to achieve stable and accurate results.
9. Warranty and Support
9.1 Warranty Information
The manufacturer provides a one-year warranty service for non-man-made damage from the date of purchase. This warranty covers manufacturing defects and malfunctions under normal use.
Exclusions from Warranty:
- Artificial damage or damage caused by improper use and custody.
- Failure or damage caused by dismantling and assembling without authorization from the company.
- Instruments purchased through informal channels.
9.2 Further Documentation
For a detailed user manual in PDF format, please refer to the official document: SWEVY Digital Refractometer User Manual (PDF).