1. Introduction
The Yieryi YY-200 is a versatile 4-in-1 digital meter designed for accurate measurement of Total Dissolved Solids (TDS), Electrical Conductivity (EC), and Temperature (in °C/°F) in various water sources. This compact and lightweight device is ideal for testing drinking water, aquariums, hydroponics systems, and more. It features a backlit LCD display for easy reading and a simple one-button mode switch.




2. Product Overview
The YY-200 meter is designed for ease of use and portability. Familiarize yourself with its components:

- Battery Compartment: Houses the LR44 batteries.
- LCD Display: Shows measurement readings and mode.
- ON/OFF Button: Powers the device on or off.
- Function Switch Button (SHIFT): Toggles between different measurement modes (TDS, EC, Temperature).
- Immersion Line: Indicates the maximum immersion depth for the electrode.
- Alloy Electrode: The sensing component for measurements.
- Electrode Protection Cover: Protects the electrode when not in use.
3. Setup
- Install Batteries: The device requires 2 x LR44 batteries (not included). Open the battery compartment and insert the batteries, ensuring correct polarity.
- Remove Protective Cap: Before use, carefully remove the electrode protective cap.
4. Operation
4.1 Taking Measurements
- Power On: Press the ON/OFF button to turn on the meter.
- Immerse Electrode: Dip the electrode into the solution you wish to test. Ensure the solution does not exceed the immersion line marked on the device.
- Select Mode: The instrument has four different modes: TDS (PPM), EC (µS/cm), Temperature (°C), and Temperature (°F). Press the SHIFT button to cycle through these modes until the desired measurement unit is displayed.
- Read Value: Wait for the reading to stabilize on the LCD display.


4.2 Calibration
The YY-200 meter comes factory-calibrated for daily water testing. However, if you require precise measurements or notice inconsistencies, you may perform a manual calibration. The calibration process for EC and TDS is similar.
EC Calibration:
- Prepare an EC calibration solution (e.g., 1413 µS/cm).
- Dip the meter into the EC calibration solution.
- Find the EC mode on the meter.
- Long press the "SHIFT" button until the EC value starts flashing.
- Press the ON/OFF button repeatedly (5 times) to increase the value.
- Press the SHIFT button repeatedly to decrease the value.
- Adjust the displayed value to match the known value of your EC calibration solution (e.g., 1413).
- Once the value is correct, long press the "SHIFT" button again until the flashing stops. Calibration is complete.
TDS Calibration:
- Prepare a TDS calibration solution.
- Dip the meter into the TDS calibration solution.
- Find the TDS mode on the meter.
- Long press the "SHIFT" button until the TDS value starts flashing.
- Press the ON/OFF button repeatedly (5 times) to increase the value.
- Press the SHIFT button repeatedly to decrease the value.
- Adjust the displayed value to match the known value of your TDS calibration solution.
- Once the value is correct, long press the "SHIFT" button again until the flashing stops. Calibration is complete.
5. Maintenance
- Cleaning the Electrode: After each use, wipe the electrode clean with a soft cloth or tissue.
- Power Off: Press the ON/OFF button to turn off the meter.
- Replace Protective Cap: Always replace the electrode protective cap to prevent damage and keep the electrode moist.
- Storage: Store the meter in a cool, dry place, away from direct sunlight and high temperatures.

6. Troubleshooting
- Inaccurate Readings: If readings appear inaccurate, ensure the electrode is clean and perform a calibration using a fresh calibration solution.
- No Display/Power Issues: Check if the LR44 batteries are correctly installed and have sufficient charge. Replace if necessary.
- Meter Not Responding: Ensure the device is not immersed beyond the immersion line, as this can cause a short circuit. Avoid throwing the instrument into the solution.
- Damage: The instrument is not waterproof. Avoid water ingress into the circuit board, battery leakage, or physical damage, as these are not covered under warranty.
7. Specifications
| Feature | Specification |
|---|---|
| Brand Name | Yieryi |
| Model Number | YY-200 |
| Functions | TDS, EC, Temperature (°C/°F) |
| TDS Range | 0–9990 Ppm |
| EC Range | 0–9990 µS/cm |
| Temperature Range | 0.1–80.0°C / 32.0–176.0°F |
| Accuracy | ±2% |
| Display Type | DIGITAL (with Backlight) |
| Power Supply | 2 × LR44 Batteries (Not Included) |
| Material | ABS + Alloy Electrode |
| Auto Power Off | After 5 Minutes |
| Operating Mode | Button Switch |
| Product Size | 150 × 20 × 20 MM |
| Net Weight | 23G |
| Key Features | 4 Modes, Backlit Display, Compact, Splash-Resistant |
| Certification | CE |
| Origin | Mainland China |
8. User Tips
- Pre-Calibrated for Convenience: The meter comes factory-calibrated, making it ready for immediate daily water testing without additional calibration.
- Simultaneous Temperature Reading: The device automatically detects and displays temperature alongside TDS or EC readings, offering comprehensive monitoring.
- Hydroponics Ready: With its EC mode (µS/cm) support, the YY-200 is perfectly suited for monitoring nutrient levels in hydroponics systems.
- Portability: Its lightweight (23g) and compact design make it easy to carry and use anywhere.
9. Warranty and Support
9.1 Warranty Scope
The warranty does not cover damages resulting from:
- Improper use (e.g., battery leakage, water ingress into the circuit board).
- Appearance damage.
- Torn labels.
As this is a professional instrument, it is recommended that non-professionals do not attempt to repair it to avoid further damage to the device.
9.2 Additional Resources
A detailed user manual in PDF format is available for download: User Manual (PDF)