Introduction
The PINTECH N2008Apro is a high-performance differential oscilloscope probe designed for safe and accurate measurement of high voltages. It features a bandwidth of 200MHz, a maximum input voltage of 1kV, and selectable attenuation ratios of 1000X and 100X, with an accuracy of ±1%. This probe is ideal for various applications including high voltage measurement, switching power supply design, inverter, UPS power supply, motor drives, power devices, high power systems, electric vehicles, and other related fields.

Setup Instructions
- Unpack the Kit: Carefully remove all components from the packaging. Verify that all items listed in the package contents are present.
- Connect to Oscilloscope: Connect the BNC output cable from the N2008Apro probe to an input channel on your oscilloscope. Ensure a secure connection.
- Power Connection: The probe requires external power. Connect the provided 6V adapter to the probe's power input.
- Attach Input Leads: Connect the red and black input leads (banana line input) to the corresponding terminals on the probe. The red lead is for the positive input, and the black lead is for the negative input.
- Grounding: Ensure your oscilloscope and the circuit under test are properly grounded to prevent measurement errors and ensure safety.

Operating Instructions
The N2008Apro probe offers two attenuation ratios and a bandwidth limit function for versatile measurements.
Powering On/Off
- Press the ON/OFF button to power on the probe. The POWER indicator will illuminate.
- Press the ON/OFF button again to power off the probe.
Selecting Attenuation Ratio
Use the Ratio switch to select the desired attenuation:
- X1000: For measuring higher voltages (up to 1kV Vp-p, 500V DC, 350V AC RMS).
- X100: For measuring lower voltages (up to 100V Vp-p, 50V DC, 35V AC RMS).
Bandwidth Limit
The probe includes a 5MHz Bandwidth Limit function:
- Engage the 5MHz Bandwidth Limit switch to filter out high-frequency noise, which can be useful for clearer signal observation in certain applications.
- Disengage the switch for full bandwidth (DC-200MHz for X1000, DC-70MHz for X100) measurements.
Over Range Indicator
The OVER RANGE indicator illuminates when the input voltage exceeds the selected attenuation ratio's maximum limit. If this occurs, immediately disconnect the probe from the circuit and select a higher attenuation ratio (X1000) or reduce the input voltage.
Zero Adjustment
The probe features a Zero adjustment function. After powering on and before making measurements, ensure the probe is not connected to any voltage source and adjust the Zero control until the oscilloscope displays a stable zero voltage.
Maintenance
- Cleaning: Clean the probe and accessories with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- Storage: Store the probe in a dry, dust-free environment when not in use. Keep it away from extreme temperatures and direct sunlight.
- Fuse: The probe has a built-in fuse for overload protection. If the probe stops functioning after an overload, the fuse may need inspection or replacement by qualified personnel.
- Cable Integrity: Regularly inspect the BNC cable and input leads for any signs of wear, cuts, or damage to the insulation. Replace damaged components immediately to ensure safety and measurement accuracy.
Troubleshooting
- No Power:
- Check the power adapter connection.
- Ensure the power outlet is functional.
- Verify the ON/OFF button is pressed.
- No Signal on Oscilloscope:
- Check the BNC cable connection to both the probe and the oscilloscope.
- Ensure the input leads are securely connected to the probe and the circuit under test.
- Verify the circuit under test is active and producing a signal.
- Check oscilloscope settings (input coupling, vertical scale, time base).
- "OVER RANGE" Indicator On:
- The input voltage is too high for the selected attenuation ratio. Switch to X1000 ratio.
- Reduce the voltage of the circuit being measured.
- Inaccurate Readings:
- Perform a Zero adjustment.
- Ensure proper grounding of both the probe and the circuit.
- Check for damaged cables or leads.
- Verify the oscilloscope's input impedance is set to 1MΩ.
Specifications

Key specifications include:
- Model: N2008Apro
- Bandwidth (-3dB): 100X: DC-70MHz, 1000X: DC-200MHz
- Attenuation Ratio: 100X, 1000X
- Accuracy: ±1%
- Rise Time: ≥1.75ns
- Input Differential Vp-p: 100X: 100V, 1000X: 1kV
- Input Differential V AC RMS: 350V
- Input Differential V DC: 100X: 50V, 1000X: 500V
- Differential Mode Voltage (Vpeak): 100X: 50V, 1000X: 500V
- Common Mode Voltage (rms): 350V
- Input Impedance: 9MΩ//2pF (Single-ended), 18MΩ//1pF (Both ends)
- Noise RMS: 100X: <400mV, 1000X: <700mV
- 5MHz Bandwidth Limited: Switchable: FULL/5MHz
- Output Voltage: <7V
- Oscilloscope Input Impedance: 1MΩ
- Power Adaptor: 6V adapter
- CMRR (Typical): 60Hz: >80dB, 100kHz: >50dB
- Length of BNC Cable: 90cm
- Length of Input Leads: 30cm
- Temperature: 0-50°C
- Humidity: 20-75%RH
- Weight: 412g
- Dimension (L*W*H): 187×83×41mm

User Tips
To ensure optimal performance and safety:
- Always start with the highest attenuation ratio (X1000) when measuring unknown high voltages to prevent damage to the probe and oscilloscope.
- Utilize the 5MHz bandwidth limit when observing signals with significant high-frequency noise to get a cleaner waveform.
- Regularly check the input leads for any signs of wear or damage, especially the insulation, to ensure safety.
- Ensure the oscilloscope's input impedance is correctly set to 1MΩ for accurate measurements with this probe.
Warranty and Support
The PINTECH N2008Apro probe is certified with CE, indicating compliance with European safety standards. For warranty information, technical support, or service, please contact your authorized PINTECH dealer or visit the official PINTECH website. Keep your purchase receipt as proof of purchase for warranty claims.