Hantek DSO2D50

Contents

Hantek DSO2D50 Digital Oscilloscope User Manual

Model: DSO2D50 | Brand: Hantek

1. Introduction

This manual provides comprehensive instructions for the Hantek DSO2D50 series digital oscilloscopes. It covers the instrument's features, setup procedures, operational guidelines, troubleshooting steps, and maintenance advice. Please read this manual thoroughly to ensure safe and effective use of your oscilloscope.

The Hantek DSO2D50 is a 2-channel digital oscilloscope with a 500MHz bandwidth and a 2GSa/s sampling rate, featuring a built-in arbitrary waveform generator. It is designed for electrical measurements and analysis.

For the most up-to-date information and a downloadable PDF version of this user manual, please visit the official Hantek website: http://www.hantek.com. A PDF version of the user manual is also available here: User Manual (PDF).

2. Safety Information

Read the following safety precautions carefully to avoid injury and to prevent damage to this product or any connected products. Always use this product in accordance with the regulations.

2.1 Summary of General Security Issues

  • Only professionally authorized personnel can perform repairs.
  • Use the right power cable: Use only the power cable approved by the country in which the product is used.
  • Connect and disconnect correctly: Before connecting the probe to the circuit being measured, connect the probe to the oscilloscope first. Before disconnecting, disconnect the probe and the circuit under test first.
  • Ground the product: To avoid electric shocks, the product is grounded through a grounding conductor of the power cable. Ensure the grounding conductor is properly connected to the ground before connecting input or output terminals.
  • Connect the probe properly: The ground wire of the probe has the same ground potential. Do not connect the ground wire to high voltage.
  • View all terminal rating values: Check all rating values and signs on the product to avoid fire or excessive current.
  • Do not operate with the cover open.
  • Avoid circuit exposure: Do not touch exposed connectors and components after power is switched on.
  • Do not operate if the product is suspected to be faulty.

2.2 Security Terms and Signs

Security terms in this manual:

WARNING: Indicates that the operation may not cause immediate damage to you.

NOTE: Indicates that the operation may cause damage to the product or other property.

Safety terms on products:

WARNING: Indicates a potential hazard may be caused to you if you do not perform this operation.

Safety signs on the product:

Hazardous Voltage ⚠ Safety Warning ⚠

2.3 Measurement Category

This instrument can be used for measurement under Class I.

WARNING: This instrument is only allowed to be used in the specified measurement class.

Measurement class definition:

  • Class I: Refers to measurements taken on a circuit not directly connected to the main power supply.
  • Class II: Refers to measurements taken on a circuit directly connected to low-voltage instruments (e.g., household appliances).
  • Class III: Refers to measurements taken on construction equipment (e.g., switchboards, circuit breakers).
  • Class IV: Refers to measurements taken at the source of low-voltage equipment (e.g., electricity meters).

2.4 Ventilation Requirement

The oscilloscope uses a fan for forced cooling. Ensure air intake and exhaust areas are free from obstructions and have free air. Provide at least 10 cm clearance beside, above, and behind the instrument for adequate ventilation.

NOTE: Inadequate ventilation may cause an increase in temperature, leading to damage. Keep the instrument well ventilated and inspect the air outlet and fan regularly.

2.5 Working Environment

Temperature:

  • Operating: 0°C to 50°C
  • Non-operating: -30°C to 70°C

Humidity:

  • Operating: Below +30°C: ≤95%RH (without condensation); +30°C to +40°C: ≤75%RH (without condensation); +40°C to +50°C: ≤45%RH (without condensation)
  • Non-operating: Below 65°C: ≤95% RH (without condensation)

WARNING: To avoid short circuit or electric shock, do not operate the device in a damp environment.

Altitude:

  • Operating: below 3 km
  • Non-operating: below 15 km

Installation (Overvoltage) Category: This product is powered by mains conforming to installation (overvoltage) category II.

WARNING: Ensure that no overvoltage (e.g., from lightning) reaches the product. Otherwise, the operator may be in danger of receiving electric shock.

Pollution Degree: Pollution Degree 2 (Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected).

2.6 Care and Cleaning

Care: Do not expose the LCD to direct sunlight for a long time when storing or placing the oscilloscope.

Cleaning:

  1. Use a lint-free cloth to remove dust from the outside of the oscilloscope and probe. Avoid scratching the display filter material.
  2. Clean the oscilloscope with a soft cloth soaked in water. For a more thorough cleaning, use an aqueous solution of 75% isopropyl alcohol.

NOTE: To avoid damaging the surface, do not use any corrosive reagent or chemical cleaning reagent.

WARNING: Before powering on the device again, ensure that it is dry enough to avoid electrical short circuit or personal injury caused by moisture.

2.7 Environmental Considerations

The product complies with WEEE Directive 2002/96/EC. For equipment recovery, appropriate methods should be used to recycle this product to ensure that most materials can be correctly reused.

3. Product Features

The Hantek DSO2D50 digital oscilloscope offers a comprehensive set of features for various measurement and analysis tasks:

  • Supports 1MΩ/50Ω impedance switching for signal integrity.
  • Sampling rate of 2GSa/s with a maximum bandwidth of 500MHz.
  • Built-in arbitrary waveform generator, capable of outputting 5 standard waveforms, custom arbitrary waveform output, and burst output.
  • Protocol Trigger: Includes five standard serial protocol triggers and decoding (UART, LIN, CAN, SPI, IIC) for serial bus data analysis.
  • 32 automatic measurement and threshold measurement functions with statistical analysis.
  • Two sets of digital voltmeter functions and hardware frequency meter functions.
  • Standard SCPI remote control instructions for building testing systems.
  • Save and export function for settings, CSV data, images, reference waveforms, and waveforms.
  • Four sampling methods: normal, average, peak, and high-precision.
  • 14 operating languages, supporting over 90% of countries and regions worldwide.
  • Two cursor measurement modes: manual and tracking, with Math function support.
  • 500uV-10V vertical measurement gear, 300VCAT II withstand voltage input.
  • FFT scale display for convenient reading of results.
  • Manual, single time, one-click AUTO measurement for simple and fast waveform measurement.
  • XY mode supports dual window display for understanding waveform phase changes.

4. Setup

4.1 General Examination

Upon receiving the oscilloscope, perform the following checks:

  1. Check the shipping package: Inspect for any transportation damage. If the packaging is severely damaged, retain it until electrical and mechanical testing is complete.
  2. Check the accessories: Refer to the 'Accessories' section (Appendix A) for a detailed list. Contact your dealer if any accessories are missing or damaged.
  3. Check the machine: If the instrument is damaged, not working properly, or fails performance tests, contact your dealer.

4.2 Appearance and Dimension

The oscilloscope has a compact design for ease of use and portability.

Hantek DSO2D50 Front View
Figure 4.1: Front view of the Hantek DSO2D50 Digital Oscilloscope.
Hantek DSO2D50 Side View
Figure 4.2: Side view of the Hantek DSO2D50 Digital Oscilloscope.

4.3 Preparation Before Use

  1. To Adjust the Supporting Legs: The oscilloscope has a bracket design at the bottom. Adjust the bracket to tilt the oscilloscope slightly upwards for better observation and operation. Close the bracket when not in use.
  2. To Connect to AC Power: The AC power supply specifications are ~100-120V, 50/60/400Hz; ~100-240V, 50/60Hz; 50Watts MAX. Connect the oscilloscope to the power supply using the provided power cord.
Hantek DSO2D50 Rear View with Power Connector
Figure 4.3: Rear view showing the power cord connector.

WARNING: To avoid electric shock, ensure that the instrument is correctly grounded.

4.4 Function Inspection

  1. Turn-on Checkout: Plug in the oscilloscope and press the power button. All front panel button lights will illuminate briefly, and the oscilloscope will perform self-checks. The startup screen will appear upon completion.
  2. Connect the Probe:
    a. Click on [Default Setup] to restore factory settings.
    b. Connect the grounding crocodile clamp wire of the probe to the grounding terminal.
    c. Connect the probe to the input terminal of channel 1 and the compensation signal output terminal (typically marked 5V@1KHz ∩).
    (Visual representation of compensation signal output terminal and ground terminal is typically provided in the manual.)
  3. Observe Square Waveform Signal: Set the probe attenuation ratio to X10 and click [AUTO SET]. Observe the waveform on the screen. A normal square wave signal should be displayed.
    (Visual representation of a square waveform signal is typically provided in the manual.)

WARNING: To avoid electric shock when using the probe, ensure the insulated wire is in good condition. Do not touch metallic parts of the probe when connected to a high voltage source.

4.5 Probe Compensation

When connecting a probe to an input channel for the first time, compensation is required to match the probe with the input channel. Uncompensated or offset probes can lead to measurement errors.

  1. Perform the first four steps of the 'Function Check' (Section 4.4).
  2. Compare the displayed waveform with the following ideal states:
    (Visual representations of 'Perfectly compensated', 'Over compensated', and 'Under compensated' waveforms are typically provided in the manual.)
  3. If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is 'Perfectly compensated'.
    (Visual representation of adjusting probe capacitance is typically provided in the manual.)

5. Front Panel Overview

The front panel of the DSO2D50 features various controls and a display for operation.

Hantek DSO2D50 Front Panel
Figure 5.1: Front panel of the Hantek DSO2D50 Digital Oscilloscope.

5.1 Front Panel Description

No.Description
1Power Key
2Bracket
3USB HOST Interface
4Menu Selection Key
5Vertical Control System, Waveform Control System
6Analog Channel Input
7Horizontal Control System
8Signal Source Output (limited to models with signal source)/External Trigger Input Channel
9Probe Compensation Signal Output Terminal/Ground Terminal
10Signal Source (limited to models with signal source)
11Trigger Control System
12Quick Keys for Operating Mode
13Menu Function Buttons
14Function Shortcut Keys
15Multi Functional Knob
16Display

5.2 Front Panel Function Overview

Horizontal Control:

  • Horizontal Time Base Knob: Rotate to modify the horizontal time base. Clockwise reduces, counter-clockwise increases. Press to switch between coarse/fine adjustment.
  • Horizontal Displacement Knob: Rotate to modify horizontal displacement (trigger displacement). Press to quickly reset displacement.

Vertical Control:

  • CH1 MENU and CH2 MENU Analog Input Channel Switches: Two channels labeled with different colors, corresponding to waveform colors.
  • Channel Vertical Offset Knob: Rotate to modify vertical offset. Press to quickly reset offset to zero.
  • Channel Vertical Gear Knob: Modify vertical gear. Rotate to decrease/increase gear. Press to switch between coarse/fine adjustment.
  • MATH MENU: Opens mathematical operation menu (A+B, A-B, A×B, A/B, FFT).

Wave Gen:

  • WAVE GEN Button: Press to open the signal generator channel.
  • BURST Button: Burst function.

Trigger Control:

  • Trigger Knob: Modify trigger level or threshold level. Press to quickly set trigger level to 50% of peak-to-peak value.
  • TRIG MENU: Opens trigger operation menu.
  • FORCE TRIG: Forces the oscilloscope to generate a trigger signal.

Quick Keys for Operating Mode:

  • AUTO SET: Automatically adjusts vertical gear, horizontal time base, and trigger mode for optimal waveform display.
  • SINGLE SEQ: Sets triggering mode to single shot.
  • RUN/STOP: Toggles oscilloscope operation between 'Run' (green backlight) and 'Stop' (red backlight).

Menu Function Buttons:

  • SAVE/RECALL: Saves and retrieves settings.
  • MEASURE: Enters measurement settings menu.
  • ACQUIRE: Enters sampling settings menu (time base mode, acquisition method, storage depth).
  • UTILITY: Enters system function settings menu (sound, language, testing, self-calibration).
  • CURSOR: Enters cursor measurement menu (manual, tracking).
  • DISPLAY: Enters display settings menu (type, afterglow time, waveform brightness).

Multi-functional Knob: Used to modify setting parameters by rotating.

6. Rear Panel Overview

The rear panel provides connectivity options for the oscilloscope.

Hantek DSO2D50 Rear Panel
Figure 6.1: Rear panel of the Hantek DSO2D50 Digital Oscilloscope.

6.1 Rear Panel Description

No.Description
1USB DEVICE
2Power Socket
3Handle

6.2 Rear Panel Function Overview

  1. USB DEVICE: This interface connects the oscilloscope to a computer, allowing users to send SCPI commands or customize programming via upper computer software.
  2. Power Socket: The power input terminal. Connect the oscilloscope to the power supply using the provided power cord.
  3. Handle: Used for easy access to the oscilloscope. Pull up vertically for use, press down when not in use.

7. User Interface

The user interface provides a visual representation of measurements and settings.

Hantek DSO2D50 User Interface
Figure 7.1: Example of the Hantek DSO2D50 User Interface.

7.1 User Interface Elements

  1. Hantek logo: Represents the Hantek trademark.
  2. Running status: Displays the oscilloscope's operating status (READY, TD, STOP, AUTO).
  3. Horizontal time base: Represents the time length per horizontal grid. Adjustable via the horizontal knob.
  4. Sampling rate: Displays the current sampling rate of the analog channel, which changes with horizontal time base variation.
  5. Storage depth: Displays the current storage depth of the simulated channel, which changes with horizontal time base variation.
  6. Horizontal displacement: Adjusts parameters via the horizontal knob to display specific horizontal position values.
  7. Trigger level: Displays the current trigger level of the analog channel.
  8. Operation menu: Press the menu button to open the corresponding menu.
  9. USB: If lit, indicates a connected USB drive.
  10. Signal generator: Displays the open status of the signal generator.
  11. Message box: Displays prompt messages.
  12. CH2 status label: Displays CH2 status, vertical gear, offset, channel coupling, and bandwidth limitations.
  13. CH1 status label: Displays CH1 status, vertical gear, offset, channel coupling, and bandwidth limitations.
  14. Simulate channel labels/waveforms: Different channels are labeled with different colors, consistent with waveform colors.

7.2 Help System

The built-in help system provides instructions for various front panel function buttons and menu keys.

Operation steps:

  • Click the [HELP] button to enter the help module.
  • Press other keys to obtain help information.
  • Press the [HELP] button again to exit the help module.

For example, after clicking [Help], clicking [UTILITY] will display an explanation menu for [UTILITY].

8. Operating the Vertical System

This series of oscilloscopes provides 2 input channels (CH1-CH2), each with an independent vertical control system. This section details the settings and operations for the vertical system.

8.1 To Enable or Disable the Analog Channel

  1. Enable analog channel: After connecting a signal to CH1, press the [CH1] button. The button light will turn on. The channel status label (e.g., 1DC 1V) will display current settings.
  2. Disable analog channel: If CH1 is open and selected, press [CH1] once to close it. If open but not selected, press [CH1] once to select it, then press again to close.

8.2 To Adjust the Vertical Scale

Vertical gear (V/div) refers to the voltage value per vertical grid. Adjusting it changes waveform amplitude. Rotate the corresponding knob to adjust the vertical gear. The probe ratio is 1X by default, and input impedance is 1MΩ. Vertical gear range: 500uV/div~10V/div.

8.3 To Adjust the Vertical Offset

Vertical offset is the deviation of the waveform's zero point from the screen center. Adjusting the offset knob moves the waveform up and down.

8.4 Channel Coupling

Setting the coupling method filters unwanted signals. Open the channel vertical menu and select the coupling type:

  • DC: Both DC and AC components pass through.
  • AC: DC component is blocked.
  • Grounded: Both DC and AC components are blocked.

NOTE: If input impedance is 50 ohms, coupling mode can only be DC.

8.5 Bandwidth Limit

Bandwidth limitation reduces noise. Options include Off, 20M, 100M, 200M, 350M. When enabled, a 'BW' flag is displayed.

8.6 Input Impedance

Two input impedance modes are available: 1MΩ (high impedance, current ignored) and 50Ω (matches device output impedance). Select via the [Impedance] menu tab.

8.7 Probe Ratio

Set the probe attenuation ratio via the [Attenuation] menu. Default is 1X. Correct setting is crucial for accurate measurements. Range: 0.01X-1000X.

MenuAttenuation Ratio (display amplitude of the signal: actual amplitude of the signal)
1X1:1
10X10:1
50X50:1
100X100:1
500X500:1
1000X1000:1

8.8 Waveform Invert

Open [CH1], click the inverting menu to turn on/off waveform inversion. Default is off. Inversion flips the waveform vertically, affecting mathematical operations and measurements.

Hantek DSO2D50 Waveform Invert Off
Figure 8.2: Example of a waveform with 'Invert' function turned off (normal display).
Hantek DSO2D50 Waveform Invert On
Figure 8.3: Example of a waveform with 'Invert' function turned on (inverted display).

8.9 Fine Tuning

Open [CH1], click the [Fine] adjustment menu. This allows for fine adjustment of the vertical gear within a small range, changing vertical resolution for detailed waveform observation. Default is coarse adjustment.

9. Operating the Horizontal System

This section provides detailed information on setting and operating the oscilloscope's horizontal system.

9.1 To Adjust the Horizontal Position

Horizontal displacement refers to the position of waveform trigger points relative to the screen center. Adjust by rotating the horizontal displacement knob.

9.2 To Adjust the Horizontal Time Base

Horizontal time base (s/div) represents the time value per horizontal grid. Range: 2ns/div~100s/div (default 1us/div). Adjust by rotating the horizontal time base knob.

9.3 Pan or Zoom Single Acquisition or Stopped Acquisition

After stopping the oscilloscope, the displayed data can be panned and scaled. This applies to data collected in a single or stopped acquisition.

9.4 Delayed Sweep

Delayed sweep horizontally magnifies a waveform segment for detailed viewing. Click the corresponding button to divide the screen into two display areas: an upper area for the waveform before amplification (main time base) and a lower area for the horizontally extended delayed scanning waveform (delayed scanning time base).

Hantek DSO2D50 Delayed Sweep Mode
Figure 9.1: Example of Delayed Sweep Mode on the Hantek DSO2D50 User Interface.

9.5 Horizontal Mode

Click [HORIZ MENU] to select horizontal modes: YT, XY, and Roll.

9.5.1 YT Mode

In YT mode, the Y-axis represents voltage and the X-axis represents time. When the horizontal time base is 100ms/div or slower, the oscilloscope enters scanning mode, collecting data before and after the trigger point.

9.5.2 XY Mode

In XY mode, both X and Y axes represent voltage levels. This mode is useful for phase difference measurement using the Lissajous method. The Lissajous figure (an ellipse) allows calculation of phase difference based on its shape. For example, a circle indicates a 90° phase difference.

9.5.3 Roll Mode

Click [HORIZ MENU] -> Mode -> Roll to select scroll mode. The waveform scrolls from right to left, refreshing the display. Horizontal gear range: 100ms to 100s.

10. Operating the Sample System

Click [Acquire] to enter the settings menu for the sample system.

10.1 Acquire Type

Open [Acquire], click [Acquire type] to select:

  • Real-time sampling: Uses real-time digital technology to collect waveforms.
  • Equivalent sampling: Reconstructs waveforms using equivalent sampling techniques.

10.2 Acquire Mode

Open [Acquire], click [Acquire Mode] to select Normal, Average, Peak, or High Resolution. Default is Normal.

  • Normal Mode: Samples the signal at equal time intervals for best display effect.
  • Average Mode: Averages multiple sampled waveforms to reduce random noise and improve vertical resolution. Higher average frequency reduces noise but slows response to waveform changes. Set average times (4, 8, 16, 32, 64, 128).
  • Peak Mode: Collects maximum and minimum values within the sampling interval to capture signal envelopes or narrow pulses, avoiding aliasing. Displays all pulses at least as wide as the sampling period.
  • High Resolution: Uses supersampling to average adjacent points, reducing random noise and producing a smoother waveform.

10.3 Memory Depth

Memory depth refers to the number of waveform points an oscilloscope can store in a single trigger acquisition. It reflects the storage capacity of the acquisition memory. The relationship between memory depth (MDepth), sample rate (SRate), horizontal time base scale (TScale), and number of horizontal divisions (HDivs) is: MDepth = SRate * TScale * HDivs.

A higher memory depth ensures a higher sample rate under the same horizontal time base scale. Open [Acquire], click [Mem Depth] to select storage depth options: 4K, 40K, 80K (single channel mode) or 4K, 40K (full channel mode).

11. Operating the Trigger System

Triggering involves setting conditions for the oscilloscope to capture and display waveforms. Correct trigger settings are essential for stable and meaningful waveform displays. Enter the trigger menu by clicking [TRIG MENU].

11.1 Trigger LEVEL

The trigger level is the signal voltage corresponding to the set trigger point. It moves up and down with the trigger level. For slope triggering, levels A and B can be set.

11.2 Trigger Source

Click [TRIG MENU] -> Source to select data sources: CH1~CH2, External, Line.

  • Analog channel: CH1-CH2 input signals can be trigger sources.
  • External source: [EXT TRIG] terminal on the front panel (0-3.3V CMOS waveform).
  • Line: Triggered at 50% level of AC power signal.

NOTE: Do not input signals exceeding 3.3V to [EXT TRIG/GEN OUT] connector.

11.3 Trigger Mode

The acquisition memory consists of a pre-trigger buffer and a post-trigger buffer. The oscilloscope provides Normal, Auto, and Single trigger modes. Default is Auto.

  • Normal mode: Updates displayed waveform only with a valid trigger.
  • Auto mode: Runs collection freely without effective triggering. Completes triggered acquisition when conditions are met, or non-triggered acquisition otherwise.
  • Single mode: Ends collection and enters stop state only with a valid trigger.
  • Forced trigger: Pressing the force trigger button generates a trigger signal in Normal and Single modes.

11.4 Trigger Type

The oscilloscope offers multiple trigger functions. Click [TRIG MENU] -> Type to select the trigger type.

11.4.1 Edge Trigger

Triggers at the specified threshold level on a rising, falling, or either edge. Set Type, Source, Slope Type (Rising, Falling, Either), Mode (Automatic, Normal), 50% (sets trigger level to vertical midpoint), and Holdoff (waiting time between triggers).

11.4.2 Pulse Trigger

Triggers on a specified width of positive or negative pulses. Set Type, Source, Polarity (Positive, Negative), When (Equal, Not Equal, Greater Than, Less Than), Width, 50%, Mode, and Holdoff.

11.4.3 Video Trigger

Captures complex standard analog and high-definition video signals. Supports NTSC and PAL. Set Type, Source, Polarity, Standard, Sync (ScanLine, LineNum, OddField, EvenField, AllFields), Line, Mode, and Holdoff.

Video StandardFrame FrequencyScan TypeTV Scan Line
NTSC30Interlaced Scan525
PAL25Interlaced Scan625

11.4.4 Slope Trigger

Triggers on positive or negative slope within a specified time. Set Type, Source, Slope Type (Rising, Falling), Level, When, Time, Mode, and Holdoff.

11.4.5 Overtime Trigger

Triggers when the time interval from a rising/falling edge to the adjacent falling/rising edge through the trigger level is greater than the set timeout time. Set Type, Source, Polarity, Time, 50%, Mode, and Holdoff.

11.4.6 Window Trigger

Triggers when the input signal passes high or low trigger levels. Provides high and low trigger levels. Set Type, Source, Level Select (Level 1, Level 2, Level 12), Mode, and Holdoff.

11.4.7 Pattern Trigger

Triggers on specified code patterns (combination of channel logic 'and'). Each channel can be H (high), L (low), or X (ignored). Can specify rising/falling edge. Set Type, Logic (AND, OR), Pattern (x, 1, 0, Rising, Falling, Either), Level, Mode, and Holdoff.

11.4.8 Interval Trigger

Triggers when the interval between two consecutive rising/falling edges satisfies a set time condition. Set Type, Source, Slope, When, Time, Mode, and Holdoff.

11.4.9 Under Amp Trigger

Triggers on a pulse that crosses one trigger level but not another. Set Type, Source, Polarity, When, Width, Level, Mode, and Holdoff.

11.4.10 UART Trigger

Triggers on UART serial protocol signals (frame start, frame end, data, verification errors). Set Type, Source, Idle Level, When, Baud Rate, Parity, Data Bits, Data, and Mode.

11.4.11 LIN Trigger

Triggers on synchronous field of LIN signals, designated identifiers, data, or frames. Set Type, Source, Idle Level, Identifier, Data, Baud Rate, Data Mask, When, and Mode.

11.4.12 CAN Trigger

Triggers at the beginning of a CAN signal frame, specified frame type, or error frame. Set Type, Source, Baud Rate, Idle Level, Identifier, Data, Data Mask, When, and Mode.

11.4.13 SPI Trigger

Triggers when timeout conditions are met and searches for specified data on SPI bus. Specify CLK and MISO lines. Set Type, Source, Slope, Level, Data Width, Data, Overtime, and Mode.

11.4.14 I2C Trigger

Triggers on I2C serial bus signals. Set Type, Clock source (SCL), Data source (SDA), Address, Level, When, and Mode.

12. Math Operations

The oscilloscope supports algebraic and FFT operations between analog channel waveforms. Enter the Math menu by clicking [Math].

12.1 Arithmetic Operations

Supports A+B, A-B, A*B, A/B. Set Operator, Source 1, Source 2, and Offset.

12.2 FFT (Fast Fourier Transform)

Performs FFT operations. Set Operator, Source, Center frequency, Span, Ref Level, Scale, Vertical Units (dB or VRms), and Window function (Rectangular, Hanning, Hamming, Blackman-Harris, Triangle, Flattop).

Window FunctionCharacteristicsWaveforms Applicable to the Window Function
RectangularDedicated window for discontinuous waveforms.Transient and short pulse waveforms.
HanningBetter frequency resolution, poorer amplitude resolution.Period Waveform.
HammingSlightly better frequency resolution than Hanning.Transient and short pulse waveforms.
Blackman-HarrisBest amplitude resolution, worst frequency resolution.Single frequency signal, searching for higher harmonics.
TriangleBetter frequency resolution.Strong narrowband signal
FlattopBetter amplitude resolution, poorer frequency resolution.Period Waveform.

13. Measurements

The oscilloscope allows various measurement methods, including screen grid scale, cursor, or automatic measurement. Enter the measurement menu by clicking [Measure].

Hantek DSO2D50 All Measure Display
Figure 13.1: Example of 'All Measure' display on the Hantek DSO2D50 User Interface.

13.1 Source and Type

Click [Source] to select CH1 or CH2. Click [Type] to select measurement items:

  • PkPk: Peak-to-peak voltage (Vmax - Vmin).
  • Frequency: Reciprocal of period.
  • Average: Arithmetic mean of waveform.
  • Vmax: Max voltage to GND.
  • Vmin: Min voltage to GND.
  • Period: Time between middle threshold points of two consecutive, like-polarity edges.
  • Vtop: Flat top voltage to GND.
  • Vmid: Voltage corresponding to intermediate measurement threshold.
  • Vbase: Flat bottom voltage to GND.
  • Vamp: Voltage from top to bottom (Vtop - Vbase).
  • RMS: Root mean square value.
  • OverShoot: Ratio of difference between max rising edge value and top value to amplitude.
  • PreShoot: Ratio of difference between min falling edge value and bottom value to amplitude.
  • PerRms: RMS value within a period.
  • PrdAvg: Arithmetic mean of first cycle.
  • Rise Time: Time for signal amplitude to rise from lower to upper threshold.
  • Fall Time: Time for signal amplitude to fall from upper to lower threshold.
  • +Width: Time between rising edge threshold and next falling edge threshold.
  • -Width: Time between falling edge threshold and next rising edge threshold.
  • +Duty: Ratio of positive pulse width to period.
  • -Duty: Ratio of negative pulse width to period.
  • FRR: Time difference between first rising edge of source A and source B.
  • FFF: Time difference between first falling edge of source A and source B.
  • FOV: Ratio of difference between max falling edge value and top value to amplitude.
  • FPRE: Ratio of difference between min rising edge value and bottom value to amplitude.
  • Bandwidth: Time from first edge to last edge of data source.
  • FRF: Time difference between first rising edge of source A and first falling edge of source B.
  • FFR: Time difference between first falling edge of source A and first rising falling edge of source B.
  • LRR: Time between first rising edge of source A and last rising edge of source B.
  • LRF: Time between first rising edge of source A and last falling edge of source B.
  • LFR: Time between first falling edge of source A and last rising edge of source B.
  • LFF: Time between first falling edge of source A and last falling edge of source B.

13.2 DVM (Digital Voltmeter)

Enter the DVM settings menu. Enable CH1/CH2 and select Type (DC RMS, AC RMS, DC).

13.3 Statistics

Enable statistics to display current, average, minimum, maximum values, root mean square error, and count of opened measurement items.

13.4 Clear All

Click [Meas] -> Clear All Menu Tags to clear all measurement parameter statistical functions.

13.5 All Measure

Click [Meas] -> All Measurement to display all measurement items on the screen.

13.6 Settings

Open [Meas] -> Settings to configure data source B for delay measurements (FRR, FFF, FRF, FFR, LRR, LRF, LFR, LFF).

13.7 Gate

Open [Meas] -> Gate Control Menu to enable gate control measurement. Measurement results will only be displayed between cursor A and B.

13.8 Quick Measurement after AUTO

After connecting the oscilloscope and inputting a valid signal, press [Auto Scale] to automatically set the waveform and open the following functions:

  • Single cycle: Displays signals for a single cycle, measuring time and voltage parameters.
  • Multi cycle: Displays signals for multiple cycles, measuring time and voltage parameters.
  • Automatic range: Automatically sets scale (Horizontal vertical gear, Horizontal gear, Vertical gear).
  • Data source: Sets channels to be automatically set (Display only, All).

14. Cursor Measurement

Cursors measure X-axis (time, frequency) and Y-axis (voltage) values of selected waveforms. A stable waveform display is required. Click [Cursor] to enter the cursor settings menu. Modes: None, Manual, Tracking.

14.1 Manual Mode

Manually adjust cursors to measure values. Set Source (CH1, CH2, Math), AX (horizontal position of cursor A), BX (horizontal position of cursor B), AXBX (both horizontal positions), AY (vertical position of cursor A), BY (vertical position of cursor B), AYBY (both vertical positions).

Hantek DSO2D50 Manual Cursor Measurement Example
Figure 14.1: Example of Manual Cursor Measurement on the Hantek DSO2D50 User Interface.

14.2 Track Mode

Two cursors (A and B) can be adjusted to measure X and Y values of two different sources. The marker automatically positions on the waveform when moving cursors horizontally/vertically.

Hantek DSO2D50 Track Measurement (before Horizontal Expansion)
Figure 14.2: Example of Track Measurement (before Horizontal Expansion) on the Hantek DSO2D50 User Interface.
Hantek DSO2D50 Track Measurement (after Horizontal Expansion)
Figure 14.3: Example of Track Measurement (after Horizontal Expansion) on the Hantek DSO2D50 User Interface.

15. Display Settings

In the display settings menu, users can set waveform display type, afterglow time, waveform brightness, grid type, and brightness. Click [Display] to enter the menu.

15.1 Type

Select between Vector (connects sampling points) and Point (displays only sampling points) for waveform display.

15.2 Intensity

Set the waveform brightness (0% to 100%) using the multi-functional knob V0.

15.3 Grid

Select screen grid types: Dot, Line, or Close (no grid).

15.4 To Set the Grid Brightness

Set grid brightness (0% to 100%) via the [Gridicule] menu tab or numeric keypad.

15.5 To Set the Screen Brightness

Set screen brightness (0% to 100%) via the [Brightness] menu button or numeric keypad.

15.6 Persist

In the afterglow menu, select afterglow time (off, infinite, 1s, 5s, 10s, 30s).

  • Specific value: Observes slow changes or low probability events.
  • Infinite: Displays newly acquired waveforms with normal brightness, while previously acquired waveforms are displayed with lower brightness. Useful for measuring noise, jitter, and capturing occasional events.

16. System Utility Settings

Click [Utility] to enter system configuration and set related functional parameters or interfaces.

16.1 Language

Switch between 14 languages, including Chinese and English, via [Utility] -> Language.

16.2 Sound

Turn sound on or off via [Utility] -> Sound.

16.3 Update

Upgrade firmware for USB storage devices. The process takes approximately 5 minutes.

Steps:

  1. Insert USB storage device with upgrade package into USB HOST interface.
  2. Press [Utility] to enter auxiliary menu.
  3. Press Upgrade -> Upgrade Firmware.
  4. Rotate V0 to select upgrade file and press V0 to confirm. Press Start Upgrade.

After upgrade, the oscilloscope restarts and performs self-calibration.

16.4 Pass/Fail

Monitor signal changes. Enter [Utility] -> Pass/Fail to access the Pass/Fail function menu.

16.4.1 To Enable or Disable the Pass/Fail Test Function

Click [Pass/Fail] -> Enable to enable/disable the pass testing function.

16.4.2 To Select the Source

Select channels CH1 to CH2 via [Source] menu tab in the Test menu.

16.4.3 Regular

Create pass/fail rules via [Regular] menu tab.

  1. Vertical: Set vertical tolerance range (0.020div - 4.00div) using V0 knob.
  2. Level: Set horizontal tolerance range (0.025div - 8.00div) using V0 knob.
  3. Create: Create a rule template based on vertical and horizontal settings.
  4. Save: Select storage location (internal flash memory, 1-10 locations) and save rule settings.
  5. Transfer out: Retrieve waveform rule settings.

16.5 Message

Turn on/off display information for failed attempts via [Pass/Fail] -> Message.

16.6 Mode

Select Pass, Fail, Pass Ring, Fail Ring via [Pass/Fail] -> Mode.

  • Pass: Output negative pulse train during testing.
  • Fail: Output negative pulse train during failed test.
  • Pass Ring: Output negative pulse train with ringing when passing.
  • Fail Ring: Output negative pulse train with ringing during failed test.

16.7 Run

Run pass/fail test via [Pass/Fail] -> Run.

16.8 Stop

Stop pass/fail test via [Pass/Fail] -> Stop.

16.9 Out Stop

Select on/off via [Pass/Fail] -> Out Stop. Enter STOP state when there is output, or continue running otherwise.

16.10 System Info

Display software version, hardware version, serial number, and other information via [UTILITY] -> System Info.

16.11 Calibrate

Perform self-calibration to optimize signal path and maximize measurement accuracy. Run if ambient temperature changes by >5°C. Preheat for 20 minutes before calibrating. Disconnect probes/cables from front panel input connector, then select "Self Calibration" via [Utility].

16.12 FrontPanel Selftest

Test functionality of all buttons, knobs, and LED lights on the front panel via [Utility] -> FrontPanel Selftest.

Display source code license via [Utility] -> Legal Information.

17. Decode Functions

Protocol decoding is crucial for protocol analysis. This oscilloscope provides common protocol decoding for UART, I2C, SPI, LIN, and CAN. Enter the decoding menu by clicking [Decode].

17.1 UART Decode

Decodes UART serial bus data. Set Type, Source, Idle Level, Label Position, Baud Rate, Parity, Data Bits, and Data.

17.2 LIN Decode

Decodes LIN signals. Set Type, Source, Label Position, Baud Rate, Idle Level, Identifier, Data, Data Mask, and When.

17.3 CAN Decode

Decodes CAN or CAN-FD signals. Set Type, Source, Baud Rate, Idle Level, Identifier, Data, Data Mask, When, and Mode.

17.4 SPI Decode

Decodes SPI bus communication. Specify CLK and MISO lines. Set Type, Source, Slope, Label Position, Data Width, Data, Overtime, and Mode.

17.5 I2C Decode

Decodes I2C serial bus signals. Set Type, Clock source (SCL), Data source (SDA), Address, Level, When, and Mode.

18. Store and Load

Users can save oscilloscope settings, waveforms, screen images, and parameters to internal memory or an external USB storage device. Saved items can be reloaded. File management operations (creating, deleting, renaming) are also supported.

18.1 Internal SAVE and RECALL

Click [SAVE/RECALL] to open the menu.

  • Save settings: Select SetUp, then Save To (Internal), select location (No.1~No.9), and Save.
  • Save waveform: Select Waveform, then Save To (Internal), select location, and Save.
  • Save reference: Select Reference, then Save To (Internal), select location, and Save.
  • Recall SetUp/Waveform/Reference: Select Type, then Recall From (Internal), and Recall.

18.2 External SAVE and RECALL

Similar to internal save/recall, but uses an external USB storage device.

  • Save SetUp/Waveform/Reference/CSV/Image: Select Type, then Save To (External), and Save. For images, Picture Invert option is available.
  • Recall SetUp/Waveform/Reference: Select Type, then Recall From (External), and Recall.

18.3 File Management

This operation is valid for external storage (USB storage). Supports English input for file/folder names.

  • New File: Insert USB drive, press Save key under [SAVE/RECALL], select Type, Save To (External), then New soft key. Input name using V0 knob and Save.
  • Delete File: Select file/folder using V0 knob and press Delete soft key.
  • Rename File: Select file/folder using V0 knob and press Rename soft key.

19. Quick Keys

Quick keys provide automatic settings for oscilloscope controls to generate useful displays of input signals.

19.1 Auto Scale

Click [Auto Scale] to automatically adjust the oscilloscope to recognize waveform type and display the input signal. Settings adjusted include Acquire (Normal), Cursor (Close), Timebase Mode (YT), Display (Vector), Horizontal (Adjusted), SEC/DIV (Adjusted), Holdoff (20ns), Trigger Level (50%), Trigger Mode (Auto), Trigger Source (Adjusted), Trigger Slope (Adjusted), Trigger Type (Edge), Sync (Adjusted), Standard (Adjusted), Bandwidth (Close), Coupling (DC), VOLTS/DIV (Adjusted).

19.2 Default Setup

Click [Default Setup] to restore the oscilloscope to its default settings. This includes Horizontal Time (200us), Horizontal Position (0s), Delayed Sweep (Off), Timebase Mode (YT), Fine (Off), Horizontal Expansion (Center), Vertical Scale (1V), Vertical Offset (CH1: 2V, CH2: -2V), CH1/CH2 (CH1-CH2 On), Channel Coupling (DC), BW Limit (Off), Attenuation (1X), Input Impedance (1MΩ), Invert (Off), Channel Unit (V), Fine (Off), Ch-Ch Skew (0s), Offset Cal (0V), Acquisition Mode (Normal), Memory Depth (4K), Trigger Type (Edge Trigger), Source Selection (CH1), Edge Type (Rising), Trigger Holdoff (1us), Display Type (Vector), Persistence Time (Off), Intensity (100%), Grid (Line), Gird Brightness (60%), Brightness (80%), AWG (Off), Wave (Sine), Frequency (1KHz), Amplitude (1.5V), Offset (0V), Start Phase (0°), Setting Type (Off), Impedance (HighZ), Cursor Mode (Off), Math Operator (A+B), Operation (Off).

20. Arbitrary Waveform Generator

The oscilloscope has a built-in arbitrary waveform generator. Enter the signal generator settings interface by clicking [WAVE GEN].

20.1 Output Waveform Type

Select the basic wave output.

  • Waveform: Select Sine, Square, Ramp, Exp, Noise, DC, Arb1-Arb4.
  • Frequency: Set frequency via V0 knob.
  • Amplitude: Set amplitude via V0 knob.
  • Offset: Set offset via V0 knob.
  • Symmetry/duty cycle: Set waveform duty cycle (square wave) or symmetry (triangular wave).
  • Impedance: Select high impedance or 50Ω.

20.2 Modulation

Two modulation methods: amplitude modulation (AM) and frequency modulation (FM). Click [WAVE GEN] to enter modulation settings. Select Type (AM, FM, PM).

  • Amplitude Modulation (AM): Variation of carrier wave amplitude with modulation wave. Set Modulation waveform, Modulation frequency, Modulation depth.
  • Frequency Modulation (FM): Variation of carrier frequency with modulation wave. Set Modulation waveform, Modulation frequency, Deviation.

20.3 Burst

Outputs burst waveforms with a specified number of cycles. Supports internal or manual control, N-loop and infinite burst types. Enter burst setting menu by clicking [Burst].

  • Type of burst: Select multiple cycles or infinite.
  • Counting: Set number of cycles via V0 knob (for multiple cycles).
  • Trigger: Click Trigger menu tab and press trigger button to output pulse train.

21. Remote Control

The oscilloscope can be remotely controlled via user-defined programming, IO software, or USB.

21.1 Remote Control via USB

  1. Connect the device: Connect the oscilloscope (USB VIEW interface) to the PC (USB HOST interface) using a USB data cable.
  2. Install USB driver: The PC will automatically install the driver upon first connection.
  3. Search for device resources: Open IO software, it will automatically search for connected resources. Click Rescan if needed.
  4. View Device Resources: Resources (machine model, serial number, version, USB interface info) appear in the USB (USB0) directory.
  5. Control the instrument remotely: Click Interactive IO in the IO interface to open the remote command control panel to send commands and read data.

NOTE: Before connecting the communication cable, turn off the instrument to avoid damaging its communication interface.

22. Troubleshooting

This section lists common faults and troubleshooting methods. If issues persist, contact Hantek technical support.

22.1 After pressing the power switch, the oscilloscope did not turn on

  1. Check if the power cord is properly connected.
  2. Check if the power switch is pressed.
  3. Restart the instrument.
  4. If the issue persists, contact Hantek technical support.

22.2 Serious distortion of input signal

  1. Check if the oscilloscope probe is properly connected to the BNC channel.
  2. Check if the probe is properly connected to the phenomenon being measured.
  3. Check if the oscilloscope probe is calibrated properly. If not, follow calibration instructions.

22.3 No signal appears on the screen

  1. Check if the probe is properly connected to the BNC connector of the signal input channel.
  2. Check if the channel is open (CH1-CH2 menu buttons).
  3. Check if there is signal output from the tested signal channel.
  4. If it's a DC signal with large amplitude, increase the amplitude range.
  5. Press the automatic measurement button to automatically detect the signal.
  6. If no waveform displays, contact Hantek technical support.

22.4 USB flash drive cannot be recognized

  1. Check if the USB flash drive is functioning properly.
  2. Confirm capacity (recommended ≤16GB).
  3. Confirm format (FAT32).
  4. Open the oscilloscope again and insert the USB drive.
  5. If the issue persists, contact Hantek technical support.

22.5 Waveform cannot be triggered

  1. Check if the triggering data source is consistent with the signal input channel.
  2. Check if the trigger level is adjusted correctly. Press the trigger level knob to return to the signal center point.
  3. Check if the triggering method is correct. For different input signals, select the appropriate method.

23. Specifications

Key technical specifications for the Hantek DSO2D50 Digital Oscilloscope:

FeatureSpecification
ModelDSO2D50
Analog Channels2
Analog Bandwidth500MHz
Real-time Sampling Rate2GSa/s
Memory Depth80K
Waveform Generator1
Decode Types5 types (UART, LIN, CAN, SPI, IIC)
DIY SuppliesElectrical
Battery IncludedNo
High-concerned chemicalNone
OriginMainland China

23.1 Package Contents

The standard package includes:

  • 1x Hantek DSO2D50 Oscilloscope
  • 2x BNC to Crocodile Clip Test Line
  • 1x Oscilloscope Test Probe
  • 1x Power Cable (EU Plug)
  • 1x USB Cable
  • 1x English User Manual
Hantek DSO2D50 Package Contents
Figure 23.1: Example of Hantek DSO2D50 package contents.

24. User Tips

  • Function Generator Output: Be aware that the built-in function generator has a low output level of 3.3V. Plan your circuits accordingly.
  • Probe Type: Standard probes are typically 10x attenuation. If your probes lack a button for switching attenuation, ensure you configure the oscilloscope's probe ratio setting correctly (e.g., 10:1) to get accurate readings.
  • Language Manuals: While this manual is in English, Hantek may offer manuals in other languages (e.g., German). Check the official Hantek website for localized versions.

25. Warranty and Support

25.1 Warranty Summary

Qingdao Hantek Electronic Co., LTD. (Hantek) warrants that the host and accessories are free from material and process defects during the warranty period. Hantek will repair or replace defective products free of charge during this period. Refer to the Hantek official website for detailed warranty regulations. Hantek disclaims all other warranties, express or implied, including merchantability and fitness for a special purpose. Hantek is not liable for indirect, special, or consequential damages.

25.2 Contact Information

If you have any questions or require service and support, please use the following contact details:

  • Email: service@hantek.com, support@hantek.com
  • Website: http://www.hantek.com
  • Switchboard: 400-036-7077
  • Tel: (0086)532-55678770 & 55678772 & 55678773
  • Address: #35 Building, No. 780 Baoyuan Road, High-tech Zone, Qingdao, Shandong, China 266114
  • Zip code: 266114

Ask a question about this manual