LEIRONG ZP3000ST

Contents

LEIRONG ZP3000ST Intelligent Peristaltic Pump

User Manual

1. Introduction

The LEIRONG ZP3000ST is an intelligent peristaltic pump designed for precise fluid transfer and metering. It features a large flow rate, an OLED display with multiple operating modes, and robust construction suitable for various applications including medical instruments, experimental scientific research, aquarium titration, food machinery, and distribution filling. This manual provides detailed instructions for the safe and efficient operation and maintenance of your ZP3000ST peristaltic pump.

Front view of the LEIRONG ZP3000ST Intelligent Peristaltic Pump
Figure 1: Front view of the ZP3000ST Peristaltic Pump
Side view of the LEIRONG ZP3000ST Intelligent Peristaltic Pump
Figure 2: Side view of the ZP3000ST Peristaltic Pump

2. Safety Information

Please read all safety instructions carefully before operating the device. Failure to follow these instructions may result in electric shock, fire, or serious injury.

  • Ensure the power supply voltage matches the specifications of the pump (100-240V AC input, 24V DC operating voltage).
  • Do not operate the pump in wet or hazardous environments.
  • Keep the pump away from heat sources and direct sunlight.
  • Always disconnect the power before performing any maintenance or changing tubes.
  • Do not attempt to disassemble or repair the pump yourself. Contact qualified personnel for service.
  • Ensure proper ventilation around the pump during operation.

3. Package Contents

Verify that all items listed below are included in your package:

  • 1 x ZP3000ST Peristaltic Pump
  • 1 x DC Power Adapter
  • 1 x External Tube (2 meters)
  • 1 x Packaging Box
  • (Optional) 1 x Foot Switch
Image showing the contents of the product package: Peristaltic Pump, DC Power adapter, External Tube, and Packaging Box.
Figure 3: Product Package Contents

4. Product Overview

The ZP3000ST pump is designed for ease of use and precision. Key features include:

  • OLED Display: Provides clear information and allows selection of multiple operating modes (Continuous, Allocation, Calibration, Configuration).
  • Knob Control: Simple and intuitive knob for parameter adjustment and menu navigation.
  • Two-speed Buckle: Adjustable tightness for the pump tube, ensuring optimal performance and tube life.
  • Bidirectional Conveying: Easily switch the direction of fluid flow (clockwise/counter-clockwise).
  • Spring Compensation Structure: Extends pump tube life and prevents liquid backflow.
  • Transparent Housing: Allows visual monitoring of the pump head operation.
  • Stepper Motor Drive: Ensures precise control and low noise operation.
OLED Display showing multiple operating modes on the peristaltic pump
Figure 4: OLED Display and Operating Modes
Close-up of the two-speed buckle on the peristaltic pump, highlighting adjustable tightness
Figure 5: Two-speed Buckle for Adjustable Tightness
Diagram showing bidirectional conveying capability of the pump with a CCW/CW switch
Figure 6: Bidirectional Conveying

5. Setup and Tube Installation

5.1. Initial Setup

  1. Place the pump on a stable, level surface.
  2. Connect the DC power adapter to the pump's power input and then to a suitable power outlet.
  3. Ensure the pump head is securely closed before operation.

5.2. Tube Installation/Replacement

The pump tube is a critical component and may need replacement over time. Follow these steps:

  1. Unscrew the screw: Locate and unscrew the retaining screw on the pump head cover.
  2. Open the lever and bridge block: Gently open the lever and the bridge block to release tension on the tube.
  3. Open the clips on both sides and remove the pump pipe: Unclip the side clips and carefully remove the old pump tube.
  4. Replace the new pump pipe: Insert the new pump tube, ensuring it is properly seated within the pump head. Close the clips, bridge block, and secure the screw.
Step-by-step guide on how to change the pump tube: 1. Unscrew the screw, 2. Open the lever and bridge block, 3. Open the clips and remove the pipe, 4. Replace the new pipe.
Figure 7: Pump Tube Replacement Steps

6. Operating Modes

The ZP3000ST offers several operating modes accessible via the OLED display and knob control.

6.1. Menu Navigation

The main menu provides access to:

  • Conti. (Continuous Mode): For continuous fluid transfer with adjustable flow rate/rotation speed.
  • Alloc. (Allocation Mode): For precise volume or time-based dispensing.
  • Cal. (Calibration Mode): To calibrate flow rate and minimize errors.
  • Config (Configuration): For parameter settings, including back suction, factory reset, and external control.

Knob Operation:

  • Rotate the knob: To switch menu options or adjust parameter values.
  • Press the knob: To confirm a selection, enter a parameter interface, or exit a menu.
Peristaltic pump with knob adjustment for parameters
Figure 8: Knob Adjustment for Parameter Setting

6.2. Continuous Mode

This mode allows the pump to run continuously. You can set the flow rate or rotation speed as needed. The display shows the cumulative runtime.

  • Select Conti. from the main menu and press the knob.
  • Adjust the flow rate or rotation speed using the knob.
  • Press the knob to start/stop the operation.
Schematic diagram of fluid operation in Continuous Mode
Figure 9: Continuous Mode Fluid Operation Schematic

6.3. Allocation Mode

This mode supports two dispensing methods: Time-based Dispensing (Titration) and Volume-based Dispensing (Filling). Both allow setting interval time and adjusting flow rate.

6.3.1. Time-based Dispensing

Set the liquid addition duration, interval time, and flow rate.

  • Select Alloc. from the main menu.
  • Navigate to Time-based Dispensing.
  • Set the desired liquid addition time (max 99h 59m 59s), interval time (max 99h 59m 59s), and flow rate.
  • Press the knob to start/stop.
Schematic diagram of fluid operation in Time-Based Allocation Mode
Figure 10: Time-Based Allocation Mode Fluid Operation Schematic

6.3.2. Volume-based Dispensing

Set the liquid addition volume, interval time, and flow rate.

  • Select Alloc. from the main menu.
  • Navigate to Volume-based Dispensing.
  • Set the desired liquid addition volume (max 9999.9ml), interval time, and flow rate.
  • Press the knob to start/stop.
Schematic diagram of fluid operation in Volume Allocation Mode
Figure 11: Volume Allocation Mode Fluid Operation Schematic

6.4. Calibration Mode

This mode allows you to calibrate the pump to reduce errors between the displayed flow value and the actual measured value.

  • Select Cal. from the main menu.
  • Adjust the operating speed using the knob.
  • Start the calibration process. The display will show a countdown.
  • After the countdown, measure the actual dispensed volume and input this value using the knob.
  • Confirm to finish calibration.
Image showing Calibration Mode and Power-Off Memory Function features
Figure 12: Calibration Mode and Power-Off Memory Function

6.5. Suck-Back Function

The suck-back function prevents dripping after dispensing by reversing the pump briefly. This can be configured in the Configuration menu.

  • Select Config from the main menu, then navigate to Suck.
  • Set the suck-back speed and duration as needed.
  • To disable, set the duration to 0.
Schematic diagram of fluid operation in Suck-Back Function Mode
Figure 13: Suck-Back Function Mode Fluid Operation Schematic

7. External Control

The ZP3000ST pump supports various external control methods, which can be configured in the Config menu. These include IO Start/Stop, CW/CCW direction control, RS485 communication, and Analog Signal control.

7.1. Foot Switch (Optional)

An optional foot switch can be connected for hands-free operation. The foot switch can be configured for Normally Open (NO), Normally Closed (NC), or Trigger modes.

  • Normally Open (NO): Circuit is open when not pressed, closes when pressed.
  • Normally Closed (NC): Circuit is closed when not pressed, opens when pressed.
  • Trigger: Sends a brief signal when pressed, then returns to its natural state upon release.
Image showing the optional foot switch and its male and female connectors
Figure 14: Optional Foot Switch

7.2. Control Methods Overview

  • IO Start/Stop: External input for starting and stopping the pump.
  • CW/CCW: External control for changing the pump's rotation direction.
  • RS485: For serial communication and integration into larger control systems.
  • Analog Signal: For controlling pump speed or flow rate using an analog input (e.g., 0-5V, 4-20mA).
Diagram showing the internal control board with connections for IO, CW/CCW, RS485, and Analog Signal
Figure 15: Control Methods and Internal Board

8. Maintenance

8.1. Cleaning

Regular cleaning of the pump's exterior and pump head is recommended. Use a soft, damp cloth. Do not use abrasive cleaners or immerse the unit in water.

8.2. Pump Tube Replacement

Refer to Section 5.2 for detailed instructions on how to replace the pump tube. The lifespan of the pump tube varies depending on the material, operating conditions, and fluid corrosiveness. Replace tubes when signs of wear, cracking, or reduced performance are observed.

9. Troubleshooting

If you encounter issues with your ZP3000ST pump, try the following common solutions:

  • No Power: Check power connections, ensure the adapter is plugged in correctly, and verify the power outlet is functional.
  • No Fluid Flow:
    • Ensure the pump tube is correctly installed and not kinked or blocked.
    • Check for air bubbles in the tubing.
    • Verify the pump head is securely closed.
    • Confirm the flow rate or speed settings are not set to zero.
  • Inaccurate Dispensing:
    • Perform a calibration using the Calibration Mode (Section 6.4).
    • Check the pump tube for wear; replace if necessary.
    • Ensure the fluid is not too viscous or has particles that could affect flow.
  • Unusual Noise:
    • Check if the pump tube is correctly installed and not rubbing against internal components.
    • Ensure the pump is on a stable surface.

If the problem persists, please contact customer support for assistance.

10. Specifications

Item Value
Model NumberZP3000ST
BrandLEIRONG
Flow Range0.6-2800 ml/min (Note: Some models may support up to 3100 ml/min)
ApplicationMedical instruments, Experimental scientific research, Aquarium titration, Food machinery, Distribution filling
PressureLow Pressure
Power SourceElectric
Motor TypeStepper motor
Control ModeKnob speed/timing/quantitative
Input Voltage100-240V AC
Operating Voltage24V
Weight2820g
Pump Tube (Default)Norprene (N96H)
Pump Tube (Optional)Silicone (S96H)
Tube ID×OD9.6mm × 14.4mm
Rollers Quantity3
Noise<60dB
Power72W
Dimensions (W×H×D)150mm × 115.5mm × 172.9mm
CertificationsCE, RoHS, ISO9001
Tube Lifetime (Norprene)≥1000 hours
Tube Lifetime (Silicone)<200 hours

Note: Suggested working conditions: Ambient temperature 0-50°C, relative humidity <80%. All data are obtained from tests in LEIRONG's laboratory and are for reference only. Flow rate may vary depending on liquid properties, pressure, and pipeline length.

Table of detailed specifications for the ZP3000ST pump
Figure 16: Detailed Product Specifications
Diagram showing the product dimensions: 150mm width, 115.5mm height, 172.9mm depth
Figure 17: Product Dimensions

11. User Tips

  • For highly viscous liquids or applications requiring large torque, it is recommended to reduce the operating speed (e.g., below 300 RPM) to prolong tube life and ensure optimal performance.
  • Regularly check the pump tube for wear, especially if using silicone tubes, as their lifespan is shorter than Norprene tubes.
  • Utilize the calibration mode frequently for applications requiring high precision to maintain flow accuracy.
  • The power-off memory function ensures that the pump remembers the last operation data after a power cycle, saving time during repetitive tasks.

12. Warranty and Support

LEIRONG products are manufactured under strict quality control standards. For warranty information, technical support, or service inquiries, please refer to the official LEIRONG website or contact your local distributor. A user manual in PDF format is available for download from the product page or by contacting support.

PDF User Manual: A detailed user manual in PDF format is available. Please refer to the product's official page or contact LEIRONG customer support for the download link: View User Manual (PDF)

13. Video Demonstration

Watch this video for a visual guide on the features and operation of the ZP3000ST Intelligent Peristaltic Pump.

Video 1: ZP3000ST Intelligent Peristaltic Pump Demonstration

© 2023 LEIRONG. All rights reserved.

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.