1. Product Overview
The LEIRONG ZP300ST is an intelligent peristaltic pump designed for precise fluid transfer and dosing in various applications, including medical instruments, experimental scientific research, aquariums, and food machinery. It features an OLED display for clear parameter visibility, knob-adjustable settings, and a toggle lever for forward and reverse rotation. The pump is built with a durable stepper motor for precise control and a spring compensation structure to extend pump tube life.


2. Setup Instructions
2.1. Unpacking and Components
Carefully unpack the ZP300ST peristaltic pump and verify all components are present:
- 1 x ZP300ST Peristaltic Pump
- 1 x DC Power Adapter
- 1 x External Tube (2m)
- 1 x Packaging Box

2.2. Power Connection
- Ensure the pump is placed on a stable, level surface.
- Connect the DC power adapter to the power input port on the rear of the pump.
- Plug the power adapter into a suitable 100-240V AC power outlet. The pump operates on a safe low voltage of 24V DC.

2.3. Tube Installation
The ZP300ST pump features a transparent housing for visible pump operation and easy tube changes.
- Pinch the bottom sides of the pump cover to open the cover.
- Open the lever and bridge block.
- Remove the existing pump tube (if any).
- Replace with the new pump tube, ensuring it is properly seated within the pump head.
- Close the lever and bridge block, then close the pump cover until it clicks securely.

2.4. Optional Foot Switch Connection
An optional foot switch can be connected for hands-free operation. Connect the foot switch cable to the designated port on the pump.

3. Operating Instructions
3.1. OLED Display and Controls
The ZP300ST features an OLED display and a knob for intuitive operation.
- OLED Display: Shows various operating modes and parameters.
- Knob: Rotate to navigate menu options and adjust parameters. Press to confirm selections or exit menus.
- Toggle Lever: Adjusts forward and reverse rotation for bidirectional conveying.

3.2. Operating Modes
The pump supports four main operating modes:
- Continuous Mode: The pump operates continuously, displaying cumulative runtime. Flow rate/speed is adjustable.
- Time Dispensing (Allocation Mode): Set allocation duration, interval time, and flow rate.
- Volume Dispensing (Allocation Mode): Set liquid addition volume, interval time, and flow rate.
- Calibration Mode: Calibrate the set volume against the actual volume to reduce addition errors.



3.3. Special Functions
- Bidirectional Conveying: Switch freely between clockwise (CW) and counter-clockwise (CCW) rotation using the toggle lever to adjust flow direction.
- Suck-Back Function: Set suck-back speed and duration as needed to prevent dripping or backflow.
- Power-Off Memory Function: The pump remembers the last operation data after power-off and restarts with the previous settings.


3.4. Parameter Settings (Configuration Menu)
The Configuration menu allows setting various parameters:
- Power-off restart memory mode: Configure how the pump behaves after a power cycle.
- Factory reset: Option to restore default settings.
- Suck-back function: Enable/disable and adjust parameters for suck-back.
- External control mode: (Currently unavailable to the public; contact manufacturer for customization) Includes options for RS485, IO Switch, IO Direction, and Analog Signal.
IO Start/Stop Functionality
The foot switch (if connected) can be configured for different statuses:
- Pull-in (Normally Open, NO): Circuit connects only when the pedal is pressed.
- Disconnect (Normally Closed, NC): Circuit stays on by default and cuts off only when the pedal is pressed.
- Trigger: A brief signal is sent when the pedal is pressed, then it springs back.
4. Maintenance
4.1. Pump Tube Selection and Replacement
The choice of pump tube material affects its lifespan and compatibility with different liquids. Refer to the specifications for details on tube types (Silicone, Norprene, PharMed BPT) and their features.
To replace the pump tube, follow the steps outlined in Section 2.3. Regular inspection and replacement of the pump tube are crucial for maintaining performance and preventing leaks.
4.2. General Cleaning
Wipe the exterior of the pump with a soft, damp cloth. Do not use abrasive cleaners or immerse the pump in water. Ensure the pump head is clean and free of debris, especially after handling corrosive liquids.
4.3. Suggested Working Conditions
For optimal performance and longevity, operate the pump within the following conditions:
- Ambient temperature: 0–50°C
- Relative humidity: <80%
5. Troubleshooting
If you encounter issues with your ZP300ST pump, consider the following:
- Inaccurate Flow Rate: Use the Calibration Mode to calibrate the pump. Flow rate can vary depending on liquid viscosity, pipeline pressure, and tube length.
- No Flow/Intermittent Flow: Check the pump tube for kinks, blockages, or wear. Ensure the power adapter is securely connected and the pump is receiving power. Verify the operating mode and parameters are correctly set.
- Unusual Noise: Inspect the pump head for any foreign objects or misaligned components. Ensure the pump tube is correctly installed.
- Display Issues: If the OLED display is not functioning correctly, try restarting the pump. If the issue persists, contact customer support.
6. Specifications
| Item | Value |
|---|---|
| Model Number | ZP300ST |
| Brand Name | LEIRONG |
| Power Source | Electric (100-240V AC Input, 24V DC) |
| Motor Type | Stepper motor |
| Theory | Rotary Pump |
| Structure | Peristaltic Pump, Reduction gear |
| Pressure | Low Pressure |
| Control Mode | Knob speed/timing/quantitative |
| Rollers Quantity | 3 |
| Noise | ≤60dB |
| Power | 40W |
| Lifetime | 1000H (pump tube life varies) |
| Material | PA66+PC/Stainless steel |
| Certification | CE, RoHS, ISO9001 |
| Dimensions (L×W×H) | 126mm × 84mm × 89mm |
| Weight | 1510g |
6.1. Flow Rate by Tube Species
| Tube Species | ID×OD | Flow rate (ml/min) |
|---|---|---|
| S32/N32 | 3.2X6.4 | 0.1-250 |
| S40/B40 | 4X7.2 | 0.1-410 |
| S48/N48 | 4.8X8 | 0.2-500 |
Note: The above data is obtained from tests in LEIRONG's laboratory and is for reference only. Actual flow rates may vary depending on different liquids, pressures, and pipeline lengths.

7. User Tips
- Optimize Tube Life: The slower the rotational speed of the peristaltic pump, the longer the service life of the pump tube. Adjust speed according to your application needs to maximize tube longevity.
- Customization: For specific control modes or external control requirements not publicly available, contact LEIRONG for customization options.
- Regular Calibration: Utilize the Calibration Mode frequently, especially for applications requiring high precision, to minimize addition errors and ensure accurate dispensing.
8. Warranty and Support
For any technical assistance, troubleshooting beyond this manual, or inquiries regarding customization, please contact LEIRONG Official Store customer support. You can also refer to the detailed user manual in PDF format for more in-depth information: LEIRONG ZP300ST User Manual (PDF).
LEIRONG is committed to providing stable and reliable fluid solutions. Our products are exported to over 80 countries and hold certifications such as CE, RoHS, and ISO9001.