1. Product Overview
The ASAIR AGP3 Series is a MEMS Digital Display Resistance Pirani Vacuum Gauge designed for precise absolute pressure measurement in a wide range of applications. This series integrates advanced MEMS pressure sensor modules and MEMS Pirani sensor chips, offering high precision, excellent repeatability, and long-term stability. It is an ideal replacement for gauges like the Inficon PSG500.
Key features include a TFT color display for intuitive operation, 0-10V analog output, and standard Modbus RTU communication for seamless integration into various systems.
Figure 1: AGP3 Series Vacuum Gauge (RJ45 version) with digital display showing pressure reading.
Figure 2: AGP3 Series Vacuum Gauge (DB9 version) with digital display showing pressure reading.
2. Setup and Installation
2.1. Physical Installation
The AGP3 series vacuum gauges feature a standard KF16 flange pressure interface. The flange is made of 304 stainless steel with an electropolished surface, ensuring reliable high vacuum sealing. Its design allows for easy assembly and disassembly without deformation.
For compatibility with other systems, various KF16 adapters are available, including KF16-KF25, KF16-KF40, KF16-1/4VCR, KF16-1/2VCR, KF16-1/8NPT, KF16-1/4NPT, KF16-1/2NPT, KF16-1/4VCO, and KF16-1/2VCO. Please contact your supplier for specific adapter requirements.
Figure 3: KF16-KF25 Adapter Components. This image shows the KF16 clamp and sealing ring, and the KF16-KF25 adapter itself, along with a KF25 clamp and sealing ring.
2.2. Electrical Wiring
Ensure the power supply is disconnected before performing any wiring. Connect the gauge according to the specific model's communication interface and pin definitions.
2.2.1. AGP3000 / AGP3200 (DB9 Interface)
These models use a DB9 communication interface. The pin definitions are as follows:
Figure 4: DB9 Pinout Diagram and Wiring Harness for AGP3000/AGP3200.
| Pin # | Wire Color | Definition |
|---|---|---|
| 1 | Black | Signal Positive |
| 2 | Brown | 485 Communication A |
| 3 | Red | 485 Communication B |
| 4 | Orange | Power Positive |
| 5, 7, 8 | Yellow, Blue, Gray | GND |
| 6 | Green | Power Negative |
| 9 | White | Signal Negative |
2.2.2. AGP3100 (RJ45 Interface)
The AGP3100 model uses an RJ45 communication interface. The pin definitions are as follows:
Figure 5: RJ45 Pinout Diagram for AGP3100.
| Pin # | Description |
|---|---|
| 1 | 12~28VDC Power Positive |
| 2 | 12~28VDC Power Negative |
| 3 | Signal Positive |
| 4 | NC (No Connection) |
| 5 | Signal Negative |
| 6 | NC (No Connection) |
| 7 | 485 Communication B |
| 8 | 485 Communication A |
2.3. Initial Power On
After connecting the gauge, apply the specified power (9-28VDC for AGP3000, 12-28VDC for AGP3100/AGP3200). The gauge requires a preheating time of approximately 60 seconds to stabilize and provide accurate readings.
3. Operating Instructions
The AGP3 series vacuum gauges are designed for user-friendly operation with an integrated TFT color display and intuitive control buttons (Up, Down, and 'S' for Set/Select).
3.1. Display and Unit Selection
- The TFT color display shows the current vacuum pressure reading.
- Users can switch between pressure units: Pascals (Pa), Torr, and millibars (mbar) using the control buttons. Refer to the on-screen menu for unit selection.
3.2. Communication and Output
- Modbus RTU: The gauge supports standard Modbus RTU communication via its RS485 interface, allowing for integration with PLCs, SCADA systems, and other industrial control systems. Consult the Modbus communication protocol document for register addresses and commands.
- Analog Output: A 0-10VDC analog output signal is provided for continuous pressure monitoring. This output can be connected to data acquisition systems or analog controllers. The input impedance for this output should be greater than 500KΩ.
4. Maintenance
To ensure optimal performance and longevity of your AGP3 series vacuum gauge, follow these general maintenance guidelines:
- Keep the gauge and its connections clean and free from dust, debris, and moisture.
- Regularly inspect the KF16 flange and sealing rings for any signs of wear or damage. Replace sealing rings as needed to maintain vacuum integrity.
- Avoid exposing the gauge to extreme temperatures or harsh chemical environments beyond its specified operating conditions.
- Do not attempt to disassemble the gauge, as this may void the warranty and damage internal components. Refer all servicing to qualified personnel.
5. Troubleshooting
If you encounter issues with your AGP3 series vacuum gauge, consider the following general troubleshooting steps:
- No Display/Power: Check power connections and ensure the correct voltage is supplied. Verify that the power supply is functioning correctly.
- Incorrect Readings: Ensure the gauge has completed its 60-second preheating time. Check for leaks in the vacuum system. Verify that the correct pressure unit is selected on the display.
- Communication Issues: For Modbus RTU, check wiring connections (RS485 A/B), baud rate, parity, and slave ID settings. Ensure the communication cable is not damaged.
- Analog Output Problems: Verify the analog output wiring. Check the input impedance of the receiving device.
If problems persist after attempting these steps, please contact technical support.
6. Specifications
| Parameter | AGP3000 | AGP3100 | AGP3200 |
|---|---|---|---|
| Power Supply | 9-28VDC | 12-28VDC | 12-28VDC |
| Communication Interface | DB9 | RJ45 | DB9 |
| Measurement Range (N2 & Air) | 5x10-2 ~ 1x105 Pa | 5x10-2 ~ 1.2x105 Pa | 5x10-2 ~ 1.2x105 Pa |
| Accuracy (N2) |
| ||
| Repeatability | 1x10-1 ~ 1.2x105 Pa: ±2% of reading | ||
| Response Time | ≤300ms | ||
| Output Signal | RS485, 0-10VDC (Input Impedance >500KΩ) | ||
| Display Units | Pa, Torr, mbar (Switchable) | ||
| Pressure Interface | KF16 (Optional adapters available) | ||
| Power Consumption | <1.2W | ||
| Operating Temperature | 0-40°C | ||
| Preheating Time | 60s | ||
| Load Pressure | 130KPa | ||
6.1. Dimensions
The structural dimensions for the AGP3 series sensors are as follows (Unit: mm; Tolerance: ±0.5mm):
Figure 6: AGP3000, AGP3100, and AGP3200 Sensor Structural Dimensions. All models have a diameter of Ø30mm. The AGP3000 has a height of 59.9mm, while the AGP3100 and AGP3200 have a height of 64mm.
7. User Tips
- System Compatibility: When integrating with existing vacuum systems, ensure that the KF16 flange is compatible or use the appropriate adapter.
- Cable Management: Use high-quality shielded cables for communication (DB9 or RJ45) to minimize electrical interference, especially in industrial environments.
- Environmental Considerations: While robust, avoid installing the gauge in areas with excessive vibration, strong electromagnetic fields, or rapid temperature fluctuations to maintain measurement accuracy.
- Calibration: Although factory calibrated, periodic re-calibration by a qualified service provider is recommended for applications requiring the highest precision.
8. Warranty and Support
ASAIR products are manufactured to high quality standards. For specific warranty terms and conditions, please refer to the documentation provided at the time of purchase or contact your vendor. Technical support is available to assist with installation, operation, and troubleshooting queries.
Please retain your purchase receipt for warranty claims.