Introduction
The SWEVY SW38A is a precise LCD digital refractometer designed for quick and accurate measurement of Brix (sugar content) in various liquids. Featuring automatic temperature compensation (ATC) and IP67 waterproof protection, it is suitable for a wide range of applications, from fruit cultivation to beverage production.
Parts Identification
- Measuring Slot: Where the sample liquid is placed.
- Prism: Optical component for measurement.
- LCD: Digital display for temperature, Brix%, and refractive index.
- ZERO Button: Used for zero-setting calibration.
- START Button (Power Button): Turns the device on/off and initiates measurement.
Setup
Battery Installation
This instrument is powered by two AAA batteries. To install or replace batteries, remove the battery cover on the back of the device and insert the batteries according to the polarity markings. Ensure the cover is securely closed after installation.
Initial Power On
Press the START button to turn on the device. The LCD will display the current temperature and '0.0%' or a previous reading.
Operation
Zero-Setting (Calibration)
Before each use, or if accuracy is critical, perform a zero-setting calibration with clean water or pure water.
- Ensure the measuring slot is clean and dry.
- Place clear water (distilled or pure water) into the measuring slot. Ensure the water covers the prism completely, about 1mm deep, with a capacity of approximately 0.3ml.
- Press the ZERO button. The display should show '0.0%'. If not, press ZERO again to reset to zero.
- Wait for the device to stabilize (approximately 20 seconds for temperature compensation).
Measurement Steps
- After zero-setting, clean and dry the measuring slot.
- Drop the test sample (fruit juice, beverage, etc.) into the measuring slot. Ensure the sample covers the prism completely, about 1mm deep, with a capacity of approximately 0.3ml.
- Press the START button. The measurement results will be displayed within 3 seconds.
- For best accuracy, you may extract the solution, clean the slot, and repeat the measurement several times.
Language Switch
To switch the display language between Chinese and English, press and hold the ZERO button for 2 seconds to enter the settings interface. Use the START button to navigate and ZERO to select the desired language (LANG [EN] or [CH]).
Maintenance and Cleaning
Cleaning the Measuring Slot
After each measurement, or if the instrument has not been used for a long time, the measuring slot should be cleaned thoroughly. The device has an IP67 waterproof rating, allowing it to be fully flushed.
- Drip clean water onto the prism and gently wipe with a soft paper towel. Repeat this process at least three times to ensure all residue is removed.
- If the slot contains oily substances or is difficult to clean with water, use ethanol or soapy water, then rinse thoroughly with clean water.
Cautions for Measurement
- Cold or Hot Measuring Solution: Drop the solution and wait 20 seconds before measuring. The instrument requires time for temperature compensation to ensure accuracy.
- Solutions with Suspended Substances: Remove suspended solids using a filter spoon or gauze. Ensure the solution is mixed evenly and avoid scratching the optical prism with metal or hard objects, as uneven solutions or suspended solids can lead to inaccurate measurements.
Troubleshooting
| Error Message | Reason | Solution |
|---|---|---|
| Excessive light | External light is too strong to measure. | Move to a less brightly lit area or shield the measuring slot from direct light. |
| High Temp / Low Temp | The temperature in the measuring slot is out of range (10-40°C). | Allow the instrument and sample to reach room temperature (10-40°C) before measuring. |
| Re zero | Zero-clearing step was not performed correctly. | Follow the zero-clearing steps again carefully. |
| Less solution | Too little solution in measuring slot. | Add more sample to ensure the prism is fully covered (approx. 0.3ml). |
| Less water | Too little clean water in measuring slot during zero-setting. | Add more clean water for zero-setting. |
| Out of range | Excessive concentration. | The sample's Brix value exceeds the instrument's measurement range (0-38%). Dilute the sample if necessary. |
| Low voltage | Low battery voltage. | Replace the two AAA batteries. |
| Fault | Possible damage to components. | Contact customer support for assistance. |
Specifications
| Feature | Specification |
|---|---|
| Model Number | SW38A |
| Measurement Range | 0-38% Brix |
| Brix Min. Division / Resolution | 0.1% |
| Measurement Accuracy | ±0.2% |
| Temperature Compensation | 10-40°C |
| Weight | 113g |
| Size (L*W*H) | 117mm * 55mm * 29mm |
| Power Supply | 2 x AAA batteries (not included) |
| Protection Level | IP67 (Waterproof) |
| Certification | CE, RoHS |
| Reset Calibration Function | Yes |
User Tips
- Users have found this refractometer to be highly accurate and reliable, with measurements matching analog devices.
- The device is noted for its ease of use, making it suitable for both beginners and experienced users.
- The quick response time (results in 3 seconds) enhances convenience during repeated measurements.
Warranty and Support
For detailed information regarding warranty, service, or additional support, please refer to the official user manual PDF. A digital version of the user manual is available for download: