1. Introduction
The Hantek MPO6104D is a powerful 6-in-1 multifunctional desktop oscilloscope designed for a wide range of electronic testing and analysis. It integrates a 4-channel oscilloscope, a 16-channel logic analyzer, a 2-channel signal generator, a digital voltmeter, a serial protocol analyzer, and an FFT spectrum analyzer. This device features a 7-inch capacitive touch screen for intuitive operation and offers advanced capabilities for waveform capture and analysis.

Key Features:
- Six-in-One Functionality: Combines a 4-channel oscilloscope, 16-channel logic analyzer, 2-channel signal generator, digital voltmeter, serial protocol analyzer, and FFT spectrum analysis.
- High Performance: Waveform capture rate up to 400,000 frames per second (fast acquisition mode) with 256 levels of brightness and color temperature display.
- Intuitive Display: 7-inch WVGA (800x480) capacitive touch screen for easy navigation and control.
- Flexible Bandwidth: Available in 80MHz, 100MHz, and 200MHz models (MPO6104D is 100MHz).
- Deep Storage: Storage depth up to 64Mpts.
- Advanced Triggering: 16 triggering functions, including 5 protocol triggers, with options for Edge, Pulse, Video, Slope, Timeout, Window, Code, UART, LIN, CAN, SPI, and IIC triggers.
- Standard Interfaces: USB Device, USB Host, LAN for versatile connectivity and remote control.
2. Package Contents
Verify that all items listed below are included in your package. If any items are missing or damaged, please contact your supplier.
- 1 x Hantek MPO6104D Host Unit
- 1 x USB PC Communication Cable
- 4 x Oscilloscope Probes
- 1 x 110-230V Wide Voltage Power Cable
- 1 x Crocodile Clip Test Line
- 1 x BNC to BNC Signal Cable
- 4 x Logic Testing Modules
- 4 x Power Cables (for logic modules)
- 16 x Logic Test Heads
- 1 x Certificate
3. Initial Setup
3.1 Power Connection
- Ensure the oscilloscope is placed on a stable, flat surface with adequate ventilation.
- Connect the provided 110-230V wide voltage power cable to the power input on the rear panel of the oscilloscope.
- Plug the other end of the power cable into a suitable AC power outlet.
- Press the power button on the front panel to turn on the device.
3.2 Connecting Probes and Cables
- Oscilloscope Probes: Connect the oscilloscope probes to the BNC input connectors (CH1, CH2, CH3, CH4) on the front panel. Ensure a secure connection.
- Signal Generator Output: For signal generation, connect the BNC to BNC signal cable to the "Gen1 Out" or "Gen2 Out" ports on the front panel.
- Logic Analyzer: Connect the logic testing modules and logic test heads to the dedicated logic analyzer input ports.
- USB PC Communication: Connect the USB Device interface on the back panel of the oscilloscope to a USB port on your computer using the USB PC communication cable for data transfer and control.
- LAN Connection: Connect the LAN port on the back panel to your computer's network port using a standard network cable for network control.

4. Operating Instructions
4.1 7-inch LCD Touch Screen
The oscilloscope features a 7-inch WVGA (800x480) capacitive touch screen. This TFT widescreen display offers 256 levels of grayscale, realistic colors, and low power consumption. It supports a waveform capture rate of 60,000 wfms/s (dot display) in normal mode and up to 400,000 wfms/s in fast acquisition mode.


4.2 Pass/Fail Testing Function
The pass/fail testing function is used to evaluate signal quality in real-time. You can set specific rules for waveforms, and the oscilloscope will count the number of times tested, data frames passed/failed, and display the test results.
To enable Pass/Fail testing:
- Press the Utility button on the control panel.
- Navigate to 'Pass/Fail'.
- Select 'Test Enable'.
- Set the option to 'On'.


4.3 Triggering Functions
The MPO6104D offers 16 triggering functions, including 5 protocol triggers, to capture specific events in your signals. These include:
- Edge Trigger
- Pulse Trigger
- Video Trigger
- Slope Trigger
- Timeout Trigger
- Window Trigger
- Code Trigger
Optional triggers for serial protocols include UART, LIN, CAN, SPI, and IIC.

4.4 FFT Spectrum Analysis
The FFT (Fast Fourier Transform) spectrum analysis function uses simulated input channel waveforms to convert time-domain (YT) signals into their frequency components. This displays the relationship between dB V and frequency (Hertz), and the vertical reading changes from V to dB, allowing for detailed frequency domain analysis.


4.5 Segmented Collection Mode
Segmented storage allows for efficient data collection when multiple triggers are performed. Data from each trigger is stored in a dedicated segment. The oscilloscope supports 128Mpts and a segmented storage range of 1-8000 segments, with a single segment length of 32K, while maintaining a sampling rate of 1GSa/s.

4.6 Arbitrary Signal Generator
The integrated 2-channel signal generator can produce a variety of standard and arbitrary waveforms with a sampling rate of 200MSa/s and 12-bit vertical resolution. The frequency range is up to 25 MHz.
Standard Waveforms: Sine wave, Square wave, Pulse, Triangle wave, Noise, DC, Sinc, Exp, Haver Sine, Lorentz, Dual Tone Multi-frequency, Gauss, ECG signal.
Arbitrary Waveforms: Arb1, Arb2, Arb3, Arb4.
Refer to the "Specifications" section for detailed frequency characteristics of each waveform.
5. Interface and Controls
The MPO6104D features a combination of physical buttons, rotary encoders, and a touch screen for comprehensive control.
5.1 Front Panel Controls
The front panel includes dedicated buttons for common functions such as Menu, Default, Help, Cursors, Measure, Display, Wave Gen, Save/Recall, Run/Stop, Single, Auto Scale, and Utility. Rotary encoders are provided for adjusting Vertical, Horizontal, and Trigger settings for each channel.

5.2 Rear Panel Interfaces
The rear panel provides connectivity options for external devices and remote control:
- USB Device: For connecting to a PC for data transfer and software control.
- USB Host: For connecting USB storage devices (e.g., flash drives) to save waveforms or settings.
- LAN: For network connectivity, enabling remote control and data sharing over a local area network.
- Power Input: For connecting the AC power cable.

6. Specifications
6.1 Acquisition
| Parameter | Description |
|---|---|
| Acquisition method | Normal, average, peak, high resolution |
| Real-time sampling rate | 1 G Sa/s (single channel) 500 M Sa/s (dual channel) 250 M Sa/s (three / four channels) |
| Peak Detect | Analog Channel 4ns |
| Averaging mode | N samples for analog channels (N values: 2, 4, 8, 16, 32, 64, 128, 256, 512, and 1024) |
| High resolution | 12bit |
| Minimum detection pulse width | 8ns |
| Memory Depth | Single Channel 64M / 128M [EDU] Dual channel 32M / 64M [EDU] Three or four channels 16M / 32M [EDU] |

6.2 Input
| Parameter | Description |
|---|---|
| Number of channels | 4 analog channels, 16 (4 * 4) digital channels The number of analog and digital channels opened at the same time should be less than 4. |
| Input coupling | DC, AC or GND Digital channels are not supported |
| Input resistance | Analog channel: 25pF ± 3pF, 1MΩ ± 2% |
| DC coupling | Digital channels: (300KΩ ± 2%), (8 pF ± 3 pF) |
| Support probe attenuation coefficient | Analog channels 0.01X, 0.02X, 0.05X, 0.1X, 0.2X, 0.5X, 1X, 2X, 5X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X, 2000X, 5000X, 10000X |
| Voltage rating | 300V CAT II |
| Input voltage | Analog channels 300VRMS (10X) Digital channel -25V ~ 25V |
6.3 Horizontal
| Parameter | Description |
|---|---|
| Waveform Interpolation | Sine, Linear, Step |
| Horizontal scale | 2ns/div ~ 100s/div 1,2,5 steps |
| Time base mode | Y-T, X-Y, Roll |
| X-Y number | 1, 2 channel combination 1 XY 3, 4 channels combined 1 XY |
| Zero offset | ± 0.5 div × minimum time base scale |
| Sampling rate and delay | ±25ppm |
| Clock drift | ≤±5 ppm/year |
| Incremental time measurement accuracy (Full bandwidth) | ± (1 sampling interval + 100ppm × reading + 0.6ns) > 16 times average ± (1 sampling interval + 100ppm × reading + 0.4ns) Sampling interval = seconds / div ÷ 200 |
6.4 Vertical
| Parameter | Description |
|---|---|
| Bandwidth (-3db) | DPO6084: 80MHz DPO6104: 100MHz DPO6204: 200MHz |
| Vertical resolution | Analog channel 8bit Digital channel 1bit |
| Vertical scale | 500µV/div ~ 10V/div at input BNC 500µV/div to 128mV/div, ±1V 129mV/div to 1.28V/div, ±10V 1.29V/div to 10V/div, ±50V |
| Position range | ± 0.5 div × minimum time base scale |
| Selectable analog bandwidth limit, typically | 20MHz |
| Low frequency response (-3db) | ≤10Hz at BNC |
| Rise time at BNC, typical | DSO6084: ≤4.4ns DSO6104: ≤3.5ns DSO6204: ≤1.8ns |
| Vertical gain accuracy | In the "normal" or "average" acquisition mode, the accuracy from 10V/div to 10mV/div is ± 3%. In "normal" or "average" acquisition mode, the accuracy from 5mV/div to 500uV/div is ± 4%. |
| DC offset accuracy | ± 0.1div ± 2 mV ± 1% offset value |
| Channel Isolation | DC to Maximum Bandwidth: > 40 dB |
Note: The bandwidth is reduced to 6MHz when using the X1 probe.
6.5 Vertical (Digital Channel: for MPO6000D / MPO600EDU / DPO6000B / DPO6000C purchase digital probe LP104)
| Parameter | Description |
|---|---|
| Threshold | Adjustable threshold for 4 channels as a group |
| Threshold selection | TTL (1.4 V) 5.0 V CMOS (+2.5 V) 3.3 V CMOS (+1.65 V) 2.5 V CMOS (+1.25 V) 1.8 V CMOS (+0.9 V) ECL (-1.3 V) PECL (+3.7 V) LVDS (+1.2 V) 0V Custom |

6.6 Arbitrary Signal Generator (with signal source series)
| Parameter | Description |
|---|---|
| Number of channels | 2/3 channels (Arb1, Arb2, Arb3, Arb4) |
| Sampling rate | 200MSa/s |
| Vertical resolution | 12 bits |
| Frequency | 25 MHz |
| Standard waveform | sine wave, square wave, pulse, triangle wave, noise, DC, Sinc, Exp, Haver Sine, Lorentz, dual tone multi-frequency, Gauss, ECG signal |
| Arbitrary waveform | Arb1, Arb2, Arb3, Arb4 |
| Frequency characteristics | |
| Sine wave | 1uHz~25MHz |
| Square wave | 1uHz~15MHz |
| Pulse | 1uHz~5MHz |
| Triangle wave | 1uHz~1MHz |
| Noise | 1uHz~5MHz |
| Sinc | 1uHz~5MHz |
| Exp | 1uHz~5MHz |
| Haver Sine | 1uHz~5MHz |
| Lorentz | 1uHz~5MHz |
| dual tone multi-frequency | 1uHz~5MHz |
| Gauss | 1uHz~5MHz |
| ECG | 1uHz~5MHz |
| Arbitrary wave | |
| Wave Length | 8KSa |
| Frequency Range | 1uHz~75MHz (less than 10 kHz) 50 ppm (more than 10 kHz) |
| Frequency Resolution | 0.1 Hz or 4 bits, whichever is greater |
| Amplitude Output range | 10mV~7Vpp (high impedance) 5mV~3.5Vpp(50Ω) |
| DC offset Range | ± 7 V, high impedance ± 3.5 V, 50 Ω |
| DC offset Resolution | 100 µV or 3 bits, whichever is greater |
| DC offset Accuracy | 2% (1 kHz) |
| Output impedance | 50 Ω, high impedance |

6.7 General Technical Specifications
| Parameter | Description |
|---|---|
| Probe compensator output | Output voltage, typical: 2Vpp input ≥1MΩ load Frequency, typical: 1kHz |
| Power supply voltage | 100-120VACRMS (± 10%), 45Hz to 440Hz, CAT II 120-240VACRMS (± 10%), 45Hz to 66Hz, CAT II |
| Power consumption | less than 30W |
| Fuse | T, 3.15A, 250V, 5x20mm |
| Surroundings | |
| Operating temperature | 0~50 °C (32~122 °F) |
| Storage temperature | -40~+71 °C (-40~159.8 °F) |
| Humidity | ≤ + 104 °F (≤ + 40 °C): ≤90% relative humidity 106 °F ~ 122 °F (+ 41 °C ~ 50 °C): ≤60% relative humidity |
| Cooling method | Convection |
| Altitude | with and without operation: 3000m (10,000 feet) |
| Mechanical shock | |
| Random vibration | 0.31 g RMS from 50Hz to 500Hz, 10 minutes per axis |
| when not operating | 2.46g RMS from 5Hz to 500Hz, 10 minutes per axis |
| during operation | 50g, 11ms, half sine wave |
| Mechanical part | |
| Oscilloscope size | 318 x 140 x 150mm (length x width x height) |
| Weight | 2900g |

6.8 Display
| Parameter | Description |
|---|---|
| Display Type | 7" TFT LCD Diagonal |
| Display resolution | 800 (horizontal) * 480 (vertical) pixels |
| Display color | 16 million colors (24 bit true color) |
| Persist time | Min, 1 s, 5 s, 10 s, 30s, infinite |
| Display type | Point, vector |
| Display mode | Color grade |
| Display brightness | Adjustable |
| Grid Type | Optional |
| Grid brightness | Adjustable |
| Waveform brightness | Adjustable |
6.9 Interface
| Parameter | Description |
|---|---|
| Standard interface | USB Host, USB Device, LAN, WIFI (EDU model) |
| Optional interface | Aux trigger output / pass failure (MPO6000EDU model) UART HDMI |
7. Maintenance
7.1 General Care
- Keep the oscilloscope clean and free from dust. Use a soft, dry cloth for cleaning the exterior.
- Avoid exposing the device to extreme temperatures, humidity, or direct sunlight.
- Ensure proper ventilation around the unit to prevent overheating.
- Do not use abrasive cleaners or solvents that could damage the casing or screen.
7.2 Safety Precautions
- Do not attempt to open the device casing. There are no user-serviceable parts inside, and doing so may void your warranty and expose you to hazardous voltages.
- Always disconnect the power cable before cleaning or moving the device.
- Use only the provided power cable and accessories.
8. Troubleshooting
This section provides solutions to common issues you might encounter. If the problem persists, please contact technical support.
8.1 Device Does Not Power On
- Check if the power cable is securely connected to both the oscilloscope and the power outlet.
- Verify that the power outlet is functional by plugging in another device.
- Ensure the power switch on the oscilloscope is in the 'On' position.
- Check the fuse (T, 3.15A, 250V, 5x20mm) on the rear panel. Replace if blown, but only if you are qualified to do so.
8.2 No Waveform Display
- Ensure probes are correctly connected to the input channels and the circuit under test.
- Check probe compensation and attenuation settings.
- Adjust the vertical scale (Volts/Div) and horizontal scale (Time/Div) to an appropriate range for your signal.
- Press the 'Auto Scale' button to automatically adjust settings for a stable waveform display.
- Verify trigger settings. If the trigger level is outside the signal range, no stable waveform will be displayed.
8.3 PC Communication Issues
- Ensure the USB PC Communication Cable is correctly connected to the 'USB Device' port on the oscilloscope and a USB port on your computer.
- Install the necessary drivers for the oscilloscope on your computer.
- Verify that the control software on your PC is correctly configured to detect the device.
- For LAN connection, ensure the network cable is properly connected and network settings (IP address, subnet mask) are correct.
9. User Tips
- Probe Compensation: Regularly check and adjust probe compensation for accurate measurements, especially when using new probes or after significant temperature changes.
- Firmware Updates: Check the manufacturer's website periodically for firmware updates to ensure optimal performance and access to new features.
- Screen Calibration: If touch screen responsiveness seems off, check for a screen calibration option in the Utility menu.
10. Warranty and Support
10.1 Product Warranty
This product comes with a one-year warranty from the date of purchase. The warranty covers manufacturing defects and malfunctions under normal use. Please retain your proof of purchase for warranty claims.
Our products are manufactured under a strict quality control system, with a pass rate consistently maintained over 99.8%.
10.2 Technical Support
For technical assistance, troubleshooting, or any inquiries regarding your Hantek MPO6104D oscilloscope, please contact our support team:
- Whatsapp/Tel/Wechat: +8619902906050
- Web: http://www.thanksbuyer.com
- Address: Room 303 Building 2, Youpin Cultural and Creative Park, Zhangkeng, Minzhi Street, Longhua District, 518131, Shenzhen, China (Mainland)