hailicare SK22

Contents

Finger Rehabilitation Exerciser Robot Gloves User Manual

Model: SK22

1. Introduction

The hailicare SK22 Finger Rehabilitation Exerciser Robot Gloves are designed to assist individuals in regaining hand mobility and flexibility, particularly for those recovering from stroke, cerebral infarction, or experiencing hand stiffness due to conditions like arthritis or prolonged office work. This device offers guided flexible training through various modes and adjustable pressure settings to support muscle strengthening, neurological recovery, and improved blood circulation.

2. Product Overview

2.1 Components

The package includes the following items:

  • 1 x Charging Main Unit
  • 1 x Training Glove (Left Hand or Right Hand, depending on selection)
  • 1 x Trainer (for Mirror Mode)
  • 1 x English Instruction Manual
  • 1 x Type-C Charging Cable (approx. 1.2m)
Packing List for SK22 Rehabilitation Robot Gloves
Figure 1: Included Components

2.2 Control Panel

The main unit features an intuitive PET panel with mechanical buttons for easy operation. The display shows the current mode, gear level, and battery status.

Control Panel Layout with button descriptions
Figure 2: Control Panel Overview
  • Auto Mode: Indicated on the display.
  • Mirror Mode: Indicated on the display.
  • Massage Mode: Indicated on the display.
  • Soft Mode: Indicated on the display.
  • Battery Level Display: Shows current battery status.
  • Level Display: Shows the current pressure intensity level (1-9).
  • On/Off Button: Powers the device on or off.
  • Mode Selection Button (M): Cycles through the 4 operating modes.
  • Level Adjust Button (+/-): Adjusts the pressure intensity level.

3. Specifications

FeatureDetail
Model NumberSK22 (103604)
Shell MaterialABS
Display TypeLED Screen
Main Unit BatteryLithium Polymer Battery #103040 3.7V, 2400mAh
Trainer BatteryCR2032
Air Pump LifespanApprox. 400 hours
Charging TimeApprox. 3 hours
Usage TimeApprox. 2 hours
ColorWhite (Main Unit), Orange/Black (Glove)
Rated Voltage5V
Rated Current2A
Positive Pressure≥105kPa
Negative Pressure≤-55kPa
Single Session Time15 minutes
Operating Modes4 (Automatic, Mirror, Massage, Soft)
Pressure Levels9 adjustable settings (default level 4)
CertificationsCE, RoHS
Voltage<50V AC
Batteries IncludedYes
MaterialNon-woven

3.1 Product Size Table

Product Size Table for Rehabilitation Gloves
Figure 3: Recommended Size Chart

Please refer to the size table above to select the appropriate glove size based on your hand width, hand length, and middle finger circumference.

4. Setup and Wearing Instructions

Follow these steps to properly wear and set up your rehabilitation gloves:

  1. Step 1: Insert Fingers. Insert each affected finger fully into the glove, pulling it snugly up the wrist. Ensure your fingers reach the top of each finger sheath.
  2. Step 2: Secure Wrist Strap. Secure the wrist strap firmly with the Velcro fastener. Pull the inner lining of the glove backward for a secure fit.
  3. Step 3: Fasten Auxiliary Strap. Fasten the auxiliary strap. Ensure no gap remains at the wrist when making a fist.
  4. Step 4: Connect Air Nozzle. Insert the glove air nozzle into the round port of the main unit.
  5. Step 5: Power On. Long press the On/Off button on the main unit to turn it on.
  6. Step 6: Select Mode and Adjust Level. Short press the Mode Selection button (M) to choose your desired mode. Short press the Level Adjust buttons (+/-) to choose your desired pressure intensity.
Steps for Wearing and Using Rehabilitation Gloves
Figure 4: Visual Guide for Wearing and Setup

4.1 Precautions for Wearing Gloves

  • Select the appropriate size and specification before use to ensure optimal effectiveness.
  • If fingers are stiff or difficult to wear gloves, gently massage fingers for 5 minutes before use.
  • It is recommended to wear gloves while sitting or lying down.

5. Operating Modes

The device offers 4 professional operating modes to cater to different rehabilitation needs:

  • Automatic Mode: Provides a comprehensive training sequence.
  • Mirror Mode: The affected finger is trained synchronously through the mirror trainer. The individual wears the trainer on their unaffected hand to assist in training the paralyzed hand. If both hands are paralyzed, simultaneous training of both hands is not possible; they must be trained separately. In such cases, an assistant must help by putting on the trainer to facilitate the training.
  • Massage Mode: Realizes interval training and helps reduce pain for patients when they first start training. Suitable for people with serious hand injuries.
  • Soft Mode: Less powerful, suitable for mild patients.
Mirror Training Mode Demonstration
Figure 5: Mirror Training Mode in action
Air Valve Design for targeted finger exercise
Figure 6: Air Valve Design for Single, Two, or Multi-Finger Exercise

5.1 Video Demonstration

Play video
Watch video online

Video 1: Product demonstration showing setup and various modes.

6. Maintenance

  • Cleaning: Wipe the main unit and gloves with a soft, damp cloth. Do not immerse the main unit in water. Ensure the gloves are dry before storage.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
  • Battery Care: Charge the device regularly, even if not in use, to maintain battery health. Avoid fully discharging the battery for extended periods.

7. Troubleshooting

  • Device not turning on: Ensure the battery is charged. Connect the Type-C charging cable and try again.
  • Glove not inflating/deflating: Check if the air nozzle is securely connected to the main unit. Ensure there are no kinks or blockages in the air tube.
  • Insufficient pressure: Adjust the pressure intensity level using the '+' button on the main unit. Ensure the glove is worn correctly and snugly.
  • Discomfort during use: Reduce the pressure intensity level. If discomfort persists, discontinue use and consult a healthcare professional.

8. User Tips

  • For optimal results, consistency is key. Use the device regularly as recommended by your healthcare professional.
  • Always select the correct glove size to ensure proper fit and effectiveness. Refer to the size chart in Section 3.1.
  • If you experience hand stiffness, a gentle massage before wearing the gloves can improve comfort and training effectiveness.
  • The device is portable and can be used cordlessly thanks to its large-capacity rechargeable battery. Take advantage of this for therapy on the go.
  • When using Mirror Mode, ensure the trainer is securely attached to the unaffected hand for accurate synchronous training.

9. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.

You can also refer to the official user manual (PDF) for additional details: View User Manual (PDF)

© 2024 hailicare. All rights reserved.

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.