NJTY T26C

Contents

NJTY T26C Microcurrent Leakage Clamp Meter

User Instruction Manual

1. Product Overview

The NJTY T26C is a stable, safe, and reliable 3 1/2-digit digital microcurrent clamp meter designed for electricians. It is capable of measuring AC current, AC/DC voltage, and is particularly suited for measuring small AC currents with high resolution.

NJTY T26C Clamp Meter overview
Figure 1: NJTY T26C Clamp Meter, highlighting its capabilities for microcurrent and voltage measurements.

Key features include:

  • Microcurrent Leakage Current Measurement (mA, µA)
  • Suitable for AC Current measurement with 0.01mA resolution
  • Manual Range Switching: 20mA, 200mA, 2A, 60A
  • AC/DC Voltage 600V Test
  • Compact and easy to carry design

2. Setup and Preparation

2.1 Package Contents

Before using the device, please check the package contents:

  • NJTY T26C Clamp Meter
  • Test Leads (Red and Black)
  • Instrument Bag
  • User Manual (this document)
Package contents of NJTY T26C Clamp Meter
Figure 2: Included items: Clamp Meter, Manual, Test Leads, and Instrument Bag.

2.2 Battery Installation

The NJTY T26C requires 2 AAA batteries for operation. Follow these steps to install the batteries:

  1. Locate the battery compartment cover on the back of the meter.
  2. Use a screwdriver (if necessary) to open the battery compartment.
  3. Insert 2 AAA batteries, ensuring correct polarity (+ and -).
  4. Close the battery compartment cover securely.
Battery installation for NJTY T26C
Figure 3: Battery compartment showing installation of 2 AAA batteries.

2.3 Connecting Test Leads

For voltage measurements, connect the test leads to the appropriate input jacks:

  • Insert the red test lead into the 'V' (Voltage) input jack.
  • Insert the black test lead into the 'COM' (Common) input jack.

3. Operating Instructions

3.1 Power On/Off and Function Selection

To power on the meter, rotate the central rotary switch from the 'OFF' position to the desired measurement function (e.g., 20mA, 200mA, 2A, 60A, 600V). To power off, rotate the switch back to 'OFF'.

3.2 AC Current Measurement

To measure AC current using the clamp function:

  1. Rotate the function switch to the desired AC current range (20mA, 200mA, 2A, or 60A).
  2. Open the clamp jaw by pressing the lever.
  3. Place the clamp around a single conductor (wire) of the circuit you wish to measure. Ensure the jaw is fully closed.
  4. Read the current value displayed on the LCD screen.
AC Current Measurement with NJTY T26C
Figure 4: Performing an AC Current Measurement by clamping around a single conductor.

The maximum jaw opening diameter is 30mm.

Clamp jaw maximum diameter
Figure 5: The maximum diameter for clamping is 30mm.

3.3 AC Voltage Test

To measure AC voltage:

  1. Connect the test leads as described in Section 2.3.
  2. Rotate the function switch to the 'AC600V' position.
  3. Carefully touch the red and black test probes to the points where you want to measure the AC voltage.
  4. Read the voltage value displayed on the LCD screen.
AC Voltage Test with NJTY T26C
Figure 6: Performing an AC Voltage Test using the test leads on a power outlet.

3.4 DC Voltage Test

To measure DC voltage:

  1. Connect the test leads as described in Section 2.3.
  2. Rotate the function switch to the 'DC600V' position.
  3. Carefully touch the red test probe to the positive (+) point and the black test probe to the negative (-) point of the DC source.
  4. Read the voltage value displayed on the LCD screen.
DC Voltage Test with NJTY T26C
Figure 7: Performing a DC Voltage Test on a 9V battery.

3.5 HOLD Function

Press the 'HOLD' button to freeze the current reading on the display. Press it again to release the hold and resume live readings.

3.6 SELECT Function

The 'SELECT' button may be used to switch between different measurement modes within a single rotary switch position, if applicable (e.g., AC/DC, or different sub-functions). Refer to the display for active mode indication.

3.7 Automatic Shutdown

The meter features an automatic shutdown function after approximately 15 minutes of inactivity to conserve battery life.

4. Maintenance

4.1 Cleaning

To clean the meter, wipe the case with a damp cloth and a mild detergent. Do not use abrasives or solvents. Ensure the meter is powered off and disconnected from any circuits before cleaning.

4.2 Battery Replacement

When the battery indicator appears on the display, replace the batteries as described in Section 2.2. Always use 2 new AAA batteries.

4.3 Storage

When not in use for an extended period, remove the batteries to prevent leakage. Store the meter in its instrument bag in a cool, dry place, away from direct sunlight and extreme temperatures.

5. Troubleshooting

  • No Display/Weak Display: Check battery installation and replace batteries if necessary.
  • Incorrect Readings: Ensure the correct function and range are selected. Verify test lead connections are secure. For current measurements, ensure the clamp is fully closed around a single conductor.
  • Meter Does Not Power On: Ensure the rotary switch is moved from the 'OFF' position to a measurement function. Check batteries.

If problems persist, contact customer support.

6. Specifications

The following table details the functional parameters and general specifications of the NJTY T26C Clamp Meter:

Functional Parameters Table for NJTY T26C
Figure 8: Detailed functional parameters and physical specifications.
Functional Parameters - NJTY T26C
Model Number T26C
Range Resolution Accuracy
20mA 0.01mA ± (2.5%+5)
200mA 0.1mA
2A 1mA
60A 100mA ± (3%+8)
DC600V 1V ± (1.5%+5)
AC600V 1V
Body Dimensions (at the widest point): 183*63*29mm
Maximum Jaw Opening: 30mm
Net Weight of the Body: 136g (excluding battery)
Automatic shutdown: after about 15 minutes
Batteries: Uses 2 AAA batteries

7. User Tips

  • For accurate current measurements, always ensure the clamp is around only one conductor. Clamping around a cable containing both live and neutral wires will result in a zero reading as the currents cancel each other out.
  • When measuring microcurrents, ensure there are no other strong magnetic fields nearby that could interfere with the reading.
  • Always start with the highest range when measuring an unknown current or voltage to prevent overloading the meter.
  • The compact size makes it ideal for tight spaces, but always prioritize safety and ensure proper insulation.

8. Warranty and Support

For any technical assistance, troubleshooting beyond this manual, or inquiries regarding your product, please contact the seller or manufacturer's customer support. Keep your purchase receipt as proof of purchase.

Please refer to the product packaging or seller's information for specific warranty details and contact methods.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.