BXT BXT-FP640 Series

Contents

BXT-FP640 Series Flame Photometer Instruction Manual

Portable Spectrophotometer for Alkali and Alkaline Earth Metal Analysis

1. Product Overview

The BXT-FP640 series flame photometers are advanced portable spectrophotometers designed for the precise quantitative analysis of alkali and alkaline earth metals. These instruments are ideal for various applications in laboratories, environmental monitoring, agriculture, and clinical settings. Featuring a user-friendly seven-inch capacitive touch screen, the series offers models capable of detecting Potassium (K), Sodium (Na), Lithium (Li), Calcium (Ca), and Barium (Ba) with high accuracy and efficiency.

Front view of the BXT-FP640 Flame Photometer with touch screen display
Figure 1: Front view of the BXT-FP640 Flame Photometer, showing the touch screen display with Potassium (K) and Sodium (Na) readings.

2. Setup and Installation

  1. Unpacking: Carefully remove the flame photometer and all accessories from the packaging. Inspect for any signs of damage.
  2. Placement: Place the instrument on a stable, level surface in a well-ventilated area, away from direct sunlight, strong magnetic fields, and sources of vibration.
  3. Power Connection: Connect the power cord to the instrument's power input (AC220V ±22V, 50Hz ±1Hz, 250W) and then to a grounded power outlet. The power switch is located at the lower right corner of the back of the instrument.
    Close-up of the pressure gauge and power switch at the rear of the instrument
    Figure 2: Location of the pressure gauge and power switch.
  4. Air Pump Connection: Connect the air pump interface to a standard air compressor (if not integrated). Ensure the filter air pipe is correctly installed.
    Flame photometer connected to a standard air pump (air compressor)
    Figure 3: Flame photometer connected to an external air compressor.
  5. LPG Connection: Connect the LPG interface to your LPG (liquefied petroleum gas) supply. Ensure all connections are secure and leak-free.
  6. Waste Liquid Outlet: Connect appropriate tubing to the waste liquid outlet and direct it to a suitable waste container.
    Rear view of the flame photometer with labeled connections
    Figure 4: Rear view of the flame photometer showing connection points for air pump, LPG, and waste liquid.
  7. Pressure Adjustment: Use the adjust pressure knob and pressure gauge to set the appropriate gas pressure as required for operation.

3. Operating Instructions

3.1 Basic Operation

  1. Power On: Flip the power switch located at the rear of the instrument. The touch screen will illuminate, displaying the primary menu.
    Front-left view of the flame photometer showing the touch screen displaying the Primary Menu
    Figure 5: Primary Menu displayed on the touch screen.
  2. Ignition: Press the 'Ignition' button or turn the 'LPG Valve' knob to ignite the flame. Observe the combustion in the injection chamber.
    Close-up of the touch panel and injection chamber
    Figure 6: Detail of the touch panel and injection chamber for sample observation.
  3. Flame Adjustment: Use the 'Fine Adjust' knob and 'LPG Valve' knob to control the flame size. The instrument features automatic flameout protection.
    Front view of the flame photometer illustrating automatic flameout protection
    Figure 7: Automatic flameout protection feature with flame size indicators.
  4. Sample Analysis: Introduce your sample into the minimum suction crucible. The instrument will automatically draw and spray the sample at a volume of less than 6ml/min. Readings for elements like Potassium (K) and Sodium (Na) will be displayed on the screen.
    Front-right view of the flame photometer, highlighting the Fine Adjust and LPG Valve controls
    Figure 8: Controls for fine adjustment and LPG valve during operation.
  5. Reading Results: The display will show the concentration of the analyzed elements. The response time is typically less than 8 seconds.

3.2 Advanced Features (Model Dependent)

  • Printer Function: Some models offer selective configuration for a built-in thermal printer to print results directly.
  • USB Data Output: Certain models support USB data output for transferring results to a computer.
  • Linear Regression: Advanced models include automatic calculation of linear regression for calibration.

4. Maintenance

  • Regular Cleaning: Keep the instrument's exterior clean using a soft, damp cloth. Avoid abrasive cleaners or solvents.
  • Injection Chamber: Periodically clean the injection chamber to prevent residue buildup that could affect accuracy. Refer to specific model guidelines for detailed cleaning procedures.
  • Waste Container: Empty the waste liquid container regularly to prevent overflow.
  • Air Filter: Check and clean or replace the air filter in the air pipe as needed to ensure a clean air supply to the burner.
  • Calibration: Perform regular calibration using standard solutions to maintain measurement accuracy. Follow the on-screen prompts or specific calibration procedures for your model.

5. Troubleshooting

  • No Power:
    • Check if the power cord is securely connected.
    • Ensure the power switch is in the 'ON' position.
    • Verify the power outlet is functional.
  • No Flame/Unstable Flame:
    • Check LPG supply and connections.
    • Ensure adequate air pressure from the air pump.
    • Clean the burner head if it appears clogged.
    • Adjust the 'LPG Valve' and 'Fine Adjust' knobs.
  • Inaccurate Readings:
    • Perform instrument calibration with fresh standard solutions.
    • Ensure samples are prepared correctly and free from contamination.
    • Check for blockages in the sample aspiration system or injection chamber.
    • Verify the air filter is clean.
  • Touch Screen Unresponsive:
    • Restart the instrument.
    • Ensure the screen is clean and dry.

6. Specifications

Feature BXT-FB6450 BXT-FB6440 BXT-FB6431 BXT-FB6430 BXT-FP6410 BXT-FP640 BXT-6400A
Display Mode Seven inch capacitive touch LCD screen Seven inch capacitive touch screen Light intensity
Display Result Concentration 0.000-999.9 Concentration 0000-9999
Reading Range Potassium/sodium/lithium/calcium/barium Potassium/sodium/lithium/calcium Potassium/sodium/lithium/calcium Lithium/potassium/sodium 0.000-999.9 N/A
Number of Channels 5 4 3 3 N/A
Detection Limit (Potassium) ≤0.004mmol/L
Detection Limit (Sodium) ≤0.008mmol/L
Detection Limit (Lithium) ≤0.015mmol/L O O ≤0.015mmol/L ≤0.005mmol/L N/A
Detection Limit (Calcium) ≤0.050mmol/L O ≤0.050mmol/L O N/A
Detection Limit (Barium) ≤0.044mmol/L ≤0.044mmol/L O N/A
Linear Error (Potassium) ≤0.021mmol/L ≤0.03mmol/L
Linear Error (Sodium) ≤0.075mmol/L ≤0.03mmol/L
Linear Error (Lithium) ≤0.066mmol/L O O ≤0.021mmol/L N/A
Linear Error (Calcium) ≤0.066mmol/L O ≤0.075mmol/L O N/A
Linear Error (Barium) ≤0.066mmol/L O N/A
Response Time <8s <8s
Minimum Suction Crucible <6ml/min <6ml/min
Stability Continuous sampling with standard solution. Within 15s, the relative maximum change of the instrument indication value is less than 3%, measured once per minute, a total of 6 times. (For BXT-FP640/6410/6400A, max change is less than 15%)
Repeatability ≤3%
Print Function Selective configuration Selective configuration O (Optional)
Printer Type Built-in thermal printer Internal hot touch printer O (Optional)
USB Data Output Yes ● (Included) ● (Included) O (Optional)
Automatic Calculation of Linear Regression Yes ● (Included) O (Optional)
Main Engine Size 400mm x 250mm x 500mm
Net Weight 8KG
Power AC220V ±22V, 50Hz ±1Hz. 250W

Note: 'O' indicates optional or not applicable, '●' indicates included or applicable.

7. User Tips

  • Always use high-purity reagents and distilled water for preparing standard solutions to ensure accurate results.
  • Regularly check the gas and air supply lines for any kinks or leaks that could affect instrument performance.
  • When not in use, ensure the instrument is powered off and the gas supply is closed.
  • For models with USB output, ensure the correct drivers are installed on your computer for seamless data transfer.

8. Warranty and Support

This product is covered by a manufacturer's warranty against defects in materials and workmanship. For specific warranty terms and conditions, please refer to your purchase documentation.

For technical support, troubleshooting assistance, or any inquiries regarding your BXT-FP640 series flame photometer, please contact the seller directly through the AliExpress platform. Our support team aims to respond to all inquiries within 12 hours.

© 2024 BXT. All rights reserved.