1. Introduction
This manual provides essential information for the safe and efficient operation of your Hantek Digital Oscilloscope. The Hantek DSO2000 series offers a range of models with varying bandwidths and features, including optional waveform generation. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage.
2. Safety Information
Always observe the following safety precautions to avoid injury and prevent damage to the instrument or other connected devices.
- Hazardous Voltage: Do not remove the cover of the oscilloscope. Hazardous voltage is present inside. Servicing should only be performed by qualified personnel.
- Power Source: Ensure the power source voltage matches the requirements specified on the rear panel of the instrument.
- Grounding: The instrument must be properly grounded to prevent electric shock.
- Environment: Do not operate the oscilloscope in wet or damp conditions. Avoid operating in explosive atmospheres.
- Probes: Use only probes and accessories supplied or recommended by Hantek. Ensure probes are correctly rated for the voltage being measured.
3. Package Contents
Verify that your package contains all the following items:
- Hantek Digital Oscilloscope Unit
- Certificate of Conformity
- Alligator Clip Cable
- Passive Probe (PP500B)
- Power Cable
- Square USB Cable
4. Product Overview
The Hantek Digital Oscilloscope series features a 7-inch capacitive multi-touch screen with 800x480 pixel resolution. It offers 2 analog input channels and various advanced functionalities for electrical measurements and analysis.
Key Features:
- Bandwidth: Available in 200MHz, 350MHz, and 500MHz options.
- Sampling Rate: Up to 2GSa/s (single-channel) / 1GSa/s (all channels) real-time sampling.
- Memory Depth: 80Kpts (single-channel) / 40Kpts (all channels).
- Impedance Switching: Supports 1MΩ/50Ω impedance switching for signal integrity.
- Waveform Generator (DSO2D Series Only): Built-in arbitrary waveform generator capable of outputting 5 standard waveforms (Sine, Square, Ramp, Exp, Noise, DC) and custom arbitrary waveforms, with burst output support. Max frequency 25MHz.
- Protocol Trigger & Decoding: Five standard serial protocol triggers and decoding, supporting UART, LIN, CAN, SPI, IIC.
- Measurement Functions: 32 automatic measurement and threshold measurement functions, with statistical analysis. Includes digital voltmeter and hardware frequency meter functions.
- Remote Control: Standard SCPI remote control instructions for building testing systems.
- Data Management: Save and export functions for settings, CSV data, images, reference waveforms, and waveforms.
- Sampling Methods: Normal, average, peak, and high-precision.
- User Interface: Supports 14 operating languages.
- Cursor Measurement: Manual and tracking modes, with Math function support.
- Vertical Measurement: 500uV-10V vertical measurement gear, 300VCAT II withstand voltage input.
- Display: FFT scale display, XY mode for dual window display.
5. Specifications
Model Comparison
| Model | Analog Channel | Analog Bandwidth | Real time sampling rate | Memory Depth | Waveform Generator | Decode |
|---|---|---|---|---|---|---|
| DSO2C20 | 2 | 200MHz | 2GSa/s | 80K | - | 5 types |
| DSO2C35 | 2 | 350MHz | 2GSa/s | 80K | - | 5 types |
| DSO2C50 | 2 | 500MHz | 2GSa/s | 80K | - | 5 types |
| DSO2D20 | 2 | 200MHz | 2GSa/s | 80K | 1 | 5 types |
| DSO2D35 | 2 | 350MHz | 2GSa/s | 80K | 1 | 5 types |
| DSO2D50 | 2 | 500MHz | 2GSa/s | 80K | 1 | 5 types |
Detailed Specifications
| Category | Specification | |||||
|---|---|---|---|---|---|---|
| LCD Size and Type | 7-inch capacitive multi-touch screen | |||||
| Display Resolution | 800 × 480 | |||||
| Vertical System Analog Channel | ||||||
| Input Coupling | DC, AC, GND | |||||
| Input Impedance | 1MΩ±1%, 50Ω±1% | |||||
| Input Capacitance | 17pF±3pF | |||||
| Maximum Input Voltage | 1MΩ | CAT I 300 VRMS, 400 Vpk; Transient Overvoltage 1600 Vpk | ||||
| 50Ω | 5VRMS | |||||
| Vertical Resolution | 8bit | |||||
| Vertical Sensitivity Range | 1MΩ | 500uV/div ~ 10V/div | ||||
| 50Ω | 500uV/div ~ 1V/div | |||||
| Offset Range | 1MΩ | ±1V (500uV/div ~ 50 mV/div) | ||||
| ±10V (100mV/div ~ 500 mV/div) | ||||||
| ±100 V (1V/div ~ 10 V/div) | ||||||
| 50Ω | ±1V(500uV/div ~ 50 mV/div) | |||||
| ±10V(100mV/div ~ 500 mV/div) | ||||||
| ±100V(1 V/div ) | ||||||
| Dynamic Range | ±5div ( 8bit ) | |||||
| Bandwidth Limit (Typical) | 20MHz,100MHz,200MHz,350MHz; selectable for each channel | |||||
| DC Gain Accuracy | ±3%FullScale | |||||
| DC Offset Accuracy | <200mV/div (±0.1div±2mV±1.5%of offset value) | |||||
| >200mV/div (±0.1div±2mV±1.0%of offset value) | ||||||
| Channel-to-Channel Isolation | 40dB, from DC to maximum rated bandwidth of each model | |||||
| ESD Tolerance | ±8kV (on input BNCs) | |||||
| Horizontal System Analog Channel | ||||||
| Range of Time Base | 200MHz: 2ns/div~1ks/div | 350MHz: 1ns/div~1ks/div | 500MHz: 500ps/div~1ks/div | |||
| Time Base Accuracy | ±50ppm | |||||
| Time Base Delay Range | before triggering | ≥1/2 screen width | ||||
| after triggering | 1 s or 100 div, whichever is greater | |||||
| Time Interval (ΔT) | Measurement±(1 sample interval)±(50ppm××readout)±50ps | |||||
| Horizontal Mode | YT | Default | ||||
| XY | X1 = Channel1, Y1 = Channel2 | |||||
| SCAN | Time base≥100ms/div, available to enter or exit the SCAN mode by rotating the Horizontal SCALE knob | |||||
| ROLL | Time base≥100ms/div, available to enter or exit the ROLL mode by rotating the Horizontal SCALE | |||||
| Acquisition System | ||||||
| Max. Sample Rate of Analog Channel | 2GSa/s (single-channel), 1GSa/s (all channels) | |||||
| Max. Memory Depth | 80Kpts (single-channel), 40Kpts (all channels) | |||||
| Acquisition Mode | Normal | Default | ||||
| Peak Detection | Capture 1ns glitches | |||||
| Average Mode | 2, 4, 8, 16…128 are available for you to choose, averaging point by point | |||||
| High Resolution | 12bits (max.) | |||||
| Arbitrary Waveform Generator (DSO2D Series Only) | ||||||
| Sample Rate | 200MSa/s | |||||
| Vertical Resolution | 12bit | |||||
| Max. Frequency | 25MHz | |||||
| Standard Waveform | Sine, Square, Ramp, Exp, Noise, DC | |||||
| Arb | Arb1, Arb2, Arb3, Arb4 | |||||
| Sine | Frequency Range | 0.1Hz~25MHz | ||||
| Square | Frequency Range | 0.1Hz~10MHz | ||||
| Ramp | Frequency Range | 0.1Hz~1MHz | ||||
| Exp | Frequency Range | 0.1Hz~5MHz | ||||
| Arb | Frequency Range | 0.1 Hz-25MHz | ||||
| Waveform Length | 4K | |||||
| Frequency | Accuracy | 100ppm(<10kH), 50ppm(>10kHz) | ||||
| Resolution | 100 mHz or 4 bits (whichever is greater) | |||||
| Amplitude | Output Range | 10mV~7Vpp(HighZ) | ||||
| 5mV~3.5Vpp(50Ω) | ||||||
| DC Offset Accuracy | Range | ±3.5V, HighZ | ||||
| ±1.75V, 50Ω | ||||||
| Resolution | 100 uV or 3 bits (whichever is greater) | |||||
| Accuracy | 2%(1 kHz) | |||||
| Output Impedance | 50Ω ± 1% | |||||
6. Setup
- Power Connection: Connect the provided power cable to the AC input on the rear panel of the oscilloscope and then to a suitable power outlet.
- Probe Connection: Connect the passive probe (PP500B) to the desired input channel (CH1 or CH2) on the front panel. Ensure a secure connection.
- Grounding: Attach the probe's ground clip to the circuit's ground point.
- USB Connection (Optional): For PC connectivity and data transfer, connect the square USB cable from the oscilloscope's USB port to your computer.
- Power On: Press the power button to turn on the oscilloscope.
7. Operating Instructions
Basic Operation:
- Channel Selection: Use the CH1 and CH2 buttons to enable or disable input channels.
- Vertical Controls: Adjust the vertical scale (Volts/Div) and position using the dedicated knobs for each channel.
- Horizontal Controls: Adjust the horizontal time base (Sec/Div) and position using the corresponding knobs.
- Trigger Settings: Use the TRIGGER LEVEL knob and related menu options to stabilize waveforms.
- AUTO Measurement: Press the 'AUTO' button for automatic waveform setup and measurement.
Advanced Functions:
- Waveform Generator (DSO2D Series): Access the waveform generator menu to select standard waveforms (Sine, Square, Ramp, Exp, Noise, DC) or load custom arbitrary waveforms. Adjust frequency, amplitude, and offset as needed.
- Measurements: Utilize the 32 automatic measurement functions for parameters like Vpp, Vmax, Vmin, Freq, Period, etc. Cursor measurement modes (manual and tracking) are also available.
- Decoding: Use the decode function for serial protocols such as UART, LIN, CAN, SPI, IIC.
- Save/Export: Use the save/export function to store settings, CSV data, screenshots, and waveform data to a USB drive.
- Display Modes: Switch between YT (time-domain), XY (phase comparison), SCAN, and ROLL modes as required for your analysis.
Video demonstrating the Hantek Digital Oscilloscope in operation.
8. Maintenance
- Cleaning: Clean the instrument's exterior with a soft, damp cloth. Do not use abrasive cleaners or solvents.
- Storage: Store the oscilloscope in a dry, dust-free environment when not in use.
- Calibration: Periodic calibration by qualified service personnel is recommended to maintain accuracy.
- Software Updates: Check the manufacturer's website for any available firmware updates to ensure optimal performance.
9. Troubleshooting
- No Power: Ensure the power cable is securely connected to both the oscilloscope and the power outlet. Check the power switch.
- No Waveform Display: Verify that probes are correctly connected to the input channels and the circuit under test. Check channel enable settings. Adjust vertical and horizontal scales.
- Unstable Waveform: Adjust the trigger level and trigger mode settings. Ensure the trigger source is correctly selected.
- Incorrect Measurements: Check probe attenuation settings (1X, 10X) on both the probe and the oscilloscope. Ensure proper grounding.
- Device Not Responding: Try restarting the oscilloscope. If the issue persists, contact technical support.
10. User Tips
- For precise measurements, always perform a probe compensation adjustment before use.
- When measuring high-frequency signals, use the 50Ω input impedance setting if your signal source is also 50Ω to minimize reflections.
- Utilize the 'Single' trigger mode for capturing non-repeating or transient events.
- Explore the various decoding options for serial communication protocols to streamline debugging of embedded systems.