BESDATA RL-BS011

BESDATA Flexible USB Rhinolaryngoscope RL-BS011 User Manual

Model: RL-BS011 | Brand: BESDATA

1. Introduction

The BESDATA Flexible USB Rhinolaryngoscope is an advanced medical device designed for efficient and accurate rhinolaryngological examinations and diagnoses in the Ear, Nose, and Throat (ENT) field. Its flexible design, high-quality imaging, and USB connectivity provide a convenient and effective solution for healthcare professionals.

BESDATA Flexible USB Rhinolaryngoscope connected to a laptop
Figure 1: BESDATA Flexible USB Rhinolaryngoscope in use with a laptop.

2. Key Features

This rhinolaryngoscope offers superior visualization solutions to simplify your workflow:

  • Flexibility and Maneuverability: The flexible nature allows easy navigation through nasal passages and the throat, providing better access to hard-to-reach areas and reducing patient discomfort.
  • High-Quality Imaging: Equipped with advanced imaging technology for high-resolution images and videos (1280*720 resolution), ensuring a clear and detailed view of structures.
  • Portable and Easy to Use: Compact and lightweight design, connecting via USB to computers or compatible devices, making it suitable for various clinical settings.
  • Real-time Visualization: USB connectivity enables simultaneous viewing of examinations on a screen, enhancing communication and patient engagement.
  • Documentation and Sharing: Software allows for capturing and storing images and videos, valuable for patient records and professional consultations.
  • Ergonomic Design: Designed for comfortable and easy handling during examinations.
  • Photo/WB Button: Integrated button for quick image capture and white balance adjustment.
  • Angulation Lever: Provides 1:1 implementation of hand movements for precise control of the distal end (130° Up/Down).
Features of the Flexible Video Rhinolaryngoscope
Figure 2: Key features of the Flexible Video Rhinolaryngoscope.
Diagram showing USB Output, Angulation Lever, Photo/WB Button, and 130 degree Up/Down angulation
Figure 3: Detailed view of the rhinolaryngoscope handle with labeled controls.

3. Package Contents

Upon opening the package, please verify that all the following items are included:

  • Flexible USB Rhinolaryngoscope (Model RL-BS011)
  • USB Connection Cable
  • USB-C Adapter
  • USB-A Adapter
  • User Manual (this document)
Video 1: Unboxing the BESDATA Flexible USB Rhinolaryngoscope, showing the device and included accessories.
USB cable and adapters in product packaging
Figure 4: USB cable and adapters neatly packed in the product box.

4. Specifications

Table 1: Product Parameters
ParameterDescription
Model NumberRL-BS011 (Other models: RL-BS008, RL-BS009, RL-BS010)
Outer Diameter (OD)3.5mm (Variants: 2.8mm, 4.0mm)
Inner Diameter (ID)N/A (2.0mm for RL-BS009, 1.2mm for RL-BS010)
Working Length330mm
Angulation RangeU130°/D130° (Up 130°, Down 130°)
Resolution1280*720
Field of View130°
MaterialPlastic
Power SourceElectric
Warranty1 Year
CertificationCE, ISO13485
PackageStrong Carton Case
Product Parameters table for various models of BESDATA Rhinolaryngoscopes
Figure 5: Detailed product parameters for BESDATA Rhinolaryngoscope models.

5. Setup Instructions

  1. Unpack the Device: Carefully remove the Flexible USB Rhinolaryngoscope and all accessories from its packaging. Inspect for any visible damage.
  2. Connect USB Cable: Attach the provided USB connection cable to the rhinolaryngoscope handle.
  3. Connect to Display Device: Connect the USB cable to your compatible device (laptop, tablet, or smartphone) using the appropriate USB-A or USB-C adapter. The device supports Android, macOS, and Windows operating systems.
  4. Install Software/Drivers: Ensure that any necessary viewing software or drivers are installed on your display device. The rhinolaryngoscope is compatible with hardware that supports USB video input.
Diagram showing USB connection of the endoscope to a laptop, tablet, and smartphone
Figure 6: USB output compatibility with various devices.

6. Operating Instructions

  1. Power On: Ensure your connected display device is powered on and the viewing software is launched.
  2. Insert Device: Gently insert the flexible tip of the rhinolaryngoscope into the patient's nasal passage or throat as required for the examination.
  3. Control Angulation: Use the Angulation Lever on the handle to precisely control the direction of the distal end. This allows for 130° upward and 130° downward movement to navigate and view different areas.
  4. Capture Images/Adjust White Balance: Press the Photo/WB Button on the handle to capture still images or adjust the white balance for optimal viewing.
  5. Observe Real-time Feed: Monitor the real-time video feed on your connected screen for detailed examination and diagnosis.
  6. Record Video: If your software supports it, initiate video recording to document the examination.

7. Maintenance

  • Cleaning and Disinfection: After each use, thoroughly clean and disinfect the flexible tip and insertion tube according to standard medical sterilization protocols. Ensure no residue remains.
  • Handle Cleaning: Wipe the handle and cable with a soft, damp cloth. Avoid submerging the handle in liquids.
  • Storage: Store the rhinolaryngoscope in its original protective packaging in a clean, dry, and temperature-controlled environment when not in use. Protect it from direct sunlight and extreme temperatures.
  • Cable Care: Avoid kinking or sharply bending the USB cable to prevent internal damage.

8. Troubleshooting

  • No Image on Screen:
    • Ensure the USB cable is securely connected to both the rhinolaryngoscope and the display device.
    • Verify that the viewing software is correctly installed and running.
    • Check if your display device's operating system (Android/macOS/Windows) is compatible and updated.
    • Try connecting to a different USB port or a different compatible device.
  • Poor Image Quality:
    • Ensure the lens at the tip of the rhinolaryngoscope is clean and free from debris or condensation.
    • Adjust the white balance using the Photo/WB button or software settings.
    • Check ambient lighting conditions; excessive or insufficient light can affect image clarity.
  • Device Not Recognized:
    • Restart your display device and try reconnecting.
    • Reinstall the viewing software or any necessary USB video drivers.
    • Ensure the USB adapters are functioning correctly.

9. User Tips

  • Maximize Patient Comfort: Leverage the device's flexibility and ergonomic design to minimize discomfort during examinations.
  • Utilize HD Imaging: Take advantage of the high-definition image quality (1280*720) for precise diagnostic insights.
  • Engage Patients: Use the real-time visualization feature to show patients what is being seen, fostering better understanding and cooperation.
  • Document Thoroughly: Regularly capture images and videos for comprehensive patient records and to track progress over time.

10. Warranty & Support

The BESDATA Flexible USB Rhinolaryngoscope (Model RL-BS011) comes with a 1-Year Warranty from the date of purchase. This warranty covers manufacturing defects and ensures return and replacement services for eligible issues.

For technical support, warranty claims, or any inquiries, please contact:

  • E-mail: 2787063633@qq.com
  • Whatsapp: +86-15096131687
  • Wechat: +86 15096131687

11. Certifications

This product is certified to meet international quality and safety standards:

  • CE Certification: Conforms to European health, safety, and environmental protection standards.
  • ISO13485 Certification: Complies with the international standard for quality management systems for medical devices.
Certificates for CE and ISO13485
Figure 7: Product certifications including CE and ISO13485.

Related Documents - RL-BS011

Preview BESDATA Video Otoscope Instruction Manual: Operation and Maintenance Guide
This manual provides comprehensive instructions for the BESDATA Video Otoscope, a medical device for examining the ear, nasal cavity, and mouth. It covers setup, operation, specifications, cleaning, troubleshooting, and warranty information for healthcare professionals.
Preview Intelligent Visual Industrial Endoscope User Guide
User guide for the Intelligent Visual Industrial Endoscope. Learn about its features, installation, specifications, safety precautions, and troubleshooting for industrial mobile phone connectivity.
Preview OllyMurs JT-SH500 ENT Visual Ear Endoscope - Product Overview
Detailed overview of the OllyMurs JT-SH500 ENT visual ear endoscope, featuring a 12-inch HD display, 2.7mm lens, and 10W flashlight.
Preview Snake Scope Flexible Endoscope User Manual
Comprehensive user manual for the Snake Scope flexible endoscope. Provides detailed instructions on installation, operation, features, and troubleshooting for optimal use as a digital endoscope or webcam.
Preview SUNUO Smart Digital Ear Camera with Screen - User Manual & Specifications
Comprehensive user manual for the SUNUO Smart Digital Ear Camera with Screen. Learn about its features, functions, setup, troubleshooting, and technical specifications for ear, nasal, oral, and skin examinations.
Preview Firefly DE550/551 Wireless Otoscope User Manual
Comprehensive user manual for the Firefly DE550/551 Wireless Otoscope, detailing setup, operation, software usage, maintenance, specifications, and warranty information.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.