Hantek DSO2D50/2D35/2D20/2C20/2C35/2C50

Contents

Hantek Digital Oscilloscope User Manual

Models: DSO2D50/2D35/2D20/2C20/2C35/2C50

Brand: Hantek

1. Introduction

This manual provides comprehensive instructions for the Hantek DSO2D20 series digital oscilloscopes, including models DSO2D50, DSO2D35, DSO2D20, DSO2C20, DSO2C35, and DSO2C50. It covers the front and back panels, user interfaces, and basic operation methods. Please read this document carefully to ensure proper and safe use of your device.

The latest version of this manual can be downloaded from the official Hantek website: http://www.hantek.com.

Document Number: 202506

Software Version: Software upgrades may introduce new functionalities. Please check the Hantek website for the latest version.

Document Format Conventions:

  • [Name]: Represents virtual keys and main interface icons (e.g., [Utility]).
  • Menu Text (bold + color): Represents a label or menu option (e.g., Sound).
  • Operation Steps: Uses "->" to indicate the next step (e.g., [Utility] -> Language).

The DSO2D20 series digital oscilloscope offers comprehensive functions and outstanding performance, with a sampling rate of 2GSa/s and a maximum bandwidth of 500MHz across the entire range. It supports 1MΩ/50Ω impedance switching for signal integrity and includes a built-in arbitrary waveform generator (for 'D' models) capable of outputting 5 standard waveforms and custom arbitrary waveform output. It features 9 triggering modes and 5 bus analysis and protocol decoding functions (UART, LIN, CAN, SPI, IIC), 32 automatic measurement functions, digital voltmeter, and hardware frequency meter functions. Standard SCPI remote control instructions are also supported for building testing systems.

2. Safety Requirements

Read the following safety precautions carefully to avoid injury and to prevent damage to this product or any connected products. Use this product in accordance with regulations.

2.1 Summary of General Security Issues

  • Only professionally authorized personnel can perform repairs.
  • Use the right power cable approved for the country of use.
  • Connect and disconnect correctly: Always connect the probe to the oscilloscope first, then to the circuit. Disconnect from the circuit first, then from the oscilloscope.
  • Ground the product: Ensure the product is properly grounded through the power cable's grounding conductor to prevent electric shocks.
  • Connect the probe properly: The probe's ground wire is at ground potential; do not connect it to high voltage.
  • View all terminal rating values: Check all rating values and signs on the product to avoid fire or excessive current.
  • Do not operate with the cover open.
  • Avoid circuit exposure: Do not touch exposed connectors and components after power is switched on.
  • Do not operate if the product is suspected to be faulty.

2.2 Security Terms and Signs

In this manual:

  • Warning: Indicates an operation that may not cause immediate damage to you.
  • Note: Indicates an operation that may cause damage to the product or other property.

On the product:

  • Warning: Indicates a potential hazard if the operation is not performed correctly.
  • Safety Signs: Hazardous Voltage, Safety Warning.

2.3 Measurement Category

This instrument is for measurement under Class I. It is only allowed to be used in the specified measurement class.

  • Class I: Measurements on circuits not directly connected to the main power supply.
  • Class II: Measurements on circuits directly connected to low-voltage instruments (e.g., household appliances).
  • Class III: Measurements on construction equipment (e.g., switchboards, circuit breakers).
  • Class IV: Measurements at the source of low-voltage equipment (e.g., electricity meters).

2.4 Ventilation Requirement

This oscilloscope uses a fan for forced cooling. Ensure air intake and exhaust areas are free from obstructions. Provide at least 10 cm clearance around the instrument for adequate ventilation.

Note: Inadequate ventilation can cause instrument damage due to increased temperature. Keep the instrument well-ventilated and regularly inspect the air outlet and fan.

2.5 Working Environment

  • Temperature: Operating: 0°C to 50°C; Non-operating: -30°C to 70°C.
  • Humidity: Operating: Below +30°C: ≤95%RH (non-condensing); +30°C to +40°C: ≤75%RH (non-condensing); +40°C to +50°C: ≤45%RH (non-condensing). Non-operating: Below 65°C: ≤95%RH (non-condensing).
  • Warning: To avoid short circuit or electric shock, do not operate the device in a damp environment.
  • Altitude: Operating: below 3 km; Non-operating: below 15 km.
  • Installation (Overvoltage) Category: This product is powered by mains conforming to installation (overvoltage) category II. Warning: Ensure no overvoltage (e.g., from lightning) reaches the product.
  • Pollution Degree: Pollution Degree 2 (normally only nonconductive pollution occurs; temporary conductivity due to condensation is expected).
  • Security Level: Class 1 - Grounded products.

2.6 Care and Cleaning

  • Care: Do not expose the LCD to direct sunlight for prolonged periods.
  • Cleaning: Use a lint-free cloth to remove dust. Clean with a soft cloth soaked in water or a 75% isopropyl alcohol solution. Note: Do not use corrosive reagents or chemical cleaning agents.
  • Warning: Before powering on again, ensure the device is dry to avoid electrical short circuits or personal injury.

2.7 Environmental Considerations

This product complies with WEEE Directive 2002/96/EC. For equipment recovery, use appropriate methods to recycle the product to ensure materials are reused and harmful substances are not released into the environment.

3. Product Features

The Hantek DSO2D20 series digital oscilloscopes are designed with advanced features for various applications:

  • Impedance Switching: Supports 1MΩ/50Ω impedance switching for signal integrity.
  • High Performance: 2GSa/s sampling rate and up to 500MHz bandwidth.
  • Arbitrary Waveform Generator (AWG): Built-in AWG (for 'D' models) capable of outputting 5 standard waveforms and custom arbitrary waveforms, with burst output.
  • Protocol Triggering: Five standard serial protocol triggers and decoding (UART, LIN, CAN, SPI, IIC) for easy serial bus data analysis.
  • Measurement Functions: 32 automatic measurement and threshold measurement functions with statistical analysis. Two sets of digital voltmeter and hardware frequency meter functions.
  • Remote Control: Standard SCPI remote control instructions for building testing systems.
  • Data Management: Save and export functions for settings, CSV data, images, and waveforms.
  • Sampling Methods: Four sampling methods: normal, average, peak, and high-precision.
  • Multi-language Support: 14 operating languages, supporting over 90% of countries and regions worldwide.
  • Cursor Measurement: Manual and tracking cursor measurement modes, with Math function support.
  • Vertical Measurement: 500uV-10V vertical measurement gear, 300VCAT II withstand voltage input.
  • Display: FFT scale display for convenient result reading.
  • Auto Measurement: Manual, single-time, one-click AUTO measurement for simple and fast waveform analysis.
  • XY Mode: Supports dual window display for understanding waveform phase changes.

4. Getting Started

4.1 General Examination

Upon receiving your oscilloscope, perform the following checks:

  1. Check the shipping package: Inspect for any transportation damage. If the packaging is severely damaged, retain it until the product passes electrical and mechanical testing.
  2. Check the accessories: Refer to the accessories list in Appendix A. Contact your dealer if any accessories are missing or damaged.
  3. Check the machine: If the instrument is damaged, not working properly, or fails performance tests, contact your dealer.

4.2 Appearance and Dimension

The oscilloscope has a compact design. Below is a front view illustrating its dimensions.

Front view of Hantek Digital Oscilloscope showing its display and controls.
Figure 4.1 Front view of the Hantek Digital Oscilloscope.
Side view of Hantek Digital Oscilloscope showing its depth.
Figure 4.2 Side view of the Hantek Digital Oscilloscope.

4.3 Preparation Before Use

  1. Adjust the Supporting Legs: The machine has a bracket at the bottom. Adjust it to tilt the oscilloscope slightly upwards for better stability and operation. Close the bracket when not in use.
  2. Connect to AC Power: The oscilloscope supports AC power supply specifications: ~100-120V, 50/60/400Hz; ~100-240V, 50/60Hz; 50Watts MAX. Use the provided power cord to connect the oscilloscope to the power supply.
Rear panel of the Hantek Oscilloscope with power cord connector and warning labels.
Figure 4.3 To Connect to AC Power. Note the warning about hazardous voltage inside.

Warning: To avoid electric shock, ensure that the instrument is correctly grounded.

  1. Turn-on Checkout: Plug in the machine and press the power button on the lower left of the front panel. During startup, all button lights will illuminate, and the oscilloscope will perform self-checks. The startup screen will appear after completion.
  2. Function Inspection:
    • Click [Default Setup] to restore factory settings.
    • Connect the grounding crocodile clamp wire of the probe to the grounding terminal.
    • Connect the probe to channel 1 input and the compensation signal output terminal.
Diagram showing how to connect the compensation signal output terminal and ground terminal.
Figure 4.4 To Use the Compensation Signal.

Set the probe attenuation ratio to X10 and click [AUTO SET]. Observe the waveform. If the square wave signal is not displayed normally, refer to the Probe Compensation section.

Warning: To avoid electric shock, ensure the probe's insulated wire is in good condition. Do not touch metallic parts of the probe when connected to a high voltage source.

Tip: The probe compensation signal is for adjustment only, not calibration.

4.4 Probe Compensation

When connecting a probe for the first time, adjust it to match the input channel. Uncompensated probes can lead to measurement errors.

  1. Perform the first four steps of the "Function Check" section.
  2. Compare the waveform with the ideal shapes: Perfectly compensated, Over compensated, Under compensated.
  3. If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is "Perfectly compensated."

5. Front Panel Overview

The front panel of the Hantek oscilloscope features various controls and inputs for operation.

Front panel layout of the Hantek Digital Oscilloscope with labeled controls.
Figure 5.1 Front Panel of the Hantek Digital Oscilloscope.
No.Description
1Power Key
2Bracket
3USB HOST Interface
4Menu Selection Key
5Vertical Control System, Waveform Control System
6Analog Channel Input
7Horizontal Control System
8Signal Source Output (for models with signal source)/External Trigger Input Channel
9Probe Compensation Signal Output Terminal/Ground Terminal
10Signal Source (for models with signal source)
11Trigger Control System
12Quick Keys for Operating Mode
13Menu Function Buttons
14Function Shortcut Keys
15Multi Functional Knob
16Display

5.1 Horizontal Controls

  • Horizontal Time Base Knob: Modifies the horizontal time base. Clockwise rotation reduces the time base, counter-clockwise increases it. Pressing the knob switches between coarse and fine adjustment.
  • Horizontal Displacement Knob: Modifies horizontal displacement (trigger displacement). Pressing the knob resets horizontal displacement.

5.2 Vertical Controls

  • CH1 MENU and CH2 MENU Analog Input Channel Switches: Activates/deactivates channels. Waveform color corresponds to channel color.
  • Channel Vertical Offset Knob: Modifies vertical offset. Clockwise increases, counter-clockwise decreases. Pressing resets offset to zero.
  • Channel Vertical Gear Knob: Modifies vertical gear. Clockwise decreases, counter-clockwise increases. Pressing switches between coarse and fine adjustment.
  • MATH MENU: Opens the mathematical operation function menu (A+B, A-B, A×B, A/B, FFT).

5.3 Wave Gen (for 'D' models)

  • WAVE GEN: Press to open the signal generator channel.
  • BURST: Burst button.

5.4 Trigger Controls

  • Trigger Knob: Modifies the trigger level. Clockwise increases, counter-clockwise decreases. Pressing sets trigger level to 50% of peak-to-peak value.
  • TRIG MENU: Opens the trigger operation menu.
  • FORCE TRIG: Forces the oscilloscope to generate a trigger signal.

5.5 Quick Keys for Operating Mode

  • AUTO SET: Automatically adjusts vertical gear, horizontal time base, and trigger mode for optimal waveform display.
  • SINGLE SEQ: Sets triggering mode to single shot.
  • RUN/STOP: Toggles between "Run" (green backlight) and "Stop" (red backlight) states.

5.6 Menu Function Buttons

  • SAVE/RECALL: Saves and retrieves settings.
  • MEASURE: Enters the measurement settings menu.
  • ACQUIRE: Enters the sampling settings menu (time base mode, acquisition method, storage depth).
  • UTILITY: Enters the system function settings menu (sound, language, advanced features, self-calibration).
  • CURSOR: Enters the cursor measurement menu (manual, tracking).
  • DISPLAY: Enters the display settings menu (display type, afterglow time, waveform brightness).

5.7 Function Shortcut Keys

  • DEFAULT SETUP: Restores factory default settings.
  • HELP: Provides instructions for function buttons and menu keys.
  • SAVE TO USB: Saves a screenshot to a USB drive.
  • DECODE: Opens the decoding settings menu.

5.8 Multi-functional Knob

Rotates to modify setting parameters.

6. Rear Panel Overview

The rear panel provides essential connectivity for the oscilloscope.

Rear panel of the Hantek Oscilloscope with power input, USB device port, and handle.
Figure 6.1 Rear Panel of the Hantek Digital Oscilloscope.
No.Description
1USB DEVICE
2Power Socket
3Handle
  1. USB DEVICE: Connects the oscilloscope to a computer for SCPI commands or custom programming.
  2. Power Socket: Input for AC power supply. Use the provided power cord.
  3. Handle: Pull up for easy carrying; press down when not in use.

Note: Before connecting the communication cable, turn off the instrument to avoid damaging its communication interface.

7. Vertical System Settings

This series of oscilloscopes provides two input channels (CH1-CH2), each with an independent vertical control system. This section details the settings, functions, and operations for the vertical system.

7.1 To Enable or Disable the Analog Channel

  1. Enable analog channel: Press the [CH1] button to open the channel. The button light will turn on.
  2. Disable analog channel: If the CH1 channel is open and selected, press [CH1] once to close it. If not selected, press [CH1] to select it, then press again to close.

7.2 To Adjust the Vertical Scale

Vertical gear (V/div) represents the voltage value per grid. Adjusting it changes the waveform amplitude. Rotate the corresponding knob to adjust the vertical gear. The probe ratio (default 1X) and input impedance (default 1MΩ) affect the vertical gear range (500uV/div~10V/div).

7.3 To Adjust the Vertical Offset

Vertical offset refers to the deviation of the zero point position of the channel signal. Rotate the offset knob to adjust the vertical offset, moving the waveform up or down.

7.4 Channel Coupling

Setting the coupling method filters unwanted signals. Open the channel vertical menu and select the coupling type:

  • DC: Both DC and AC components pass through.
  • AC: DC component is blocked.
  • Grounded: Both DC and AC components are blocked.

Tips: When input impedance is 50 ohms, coupling can only be DC.

7.5 Bandwidth Limit

Bandwidth limitation reduces noise. Turn off for high-frequency components to pass, or set to 20MHz, 100MHz, 200MHz, 350MHz to attenuate higher frequencies. Access via [CH1] -> BW menu.

7.6 Input Impedance

Two input impedance modes (1MΩ and 50Ω) reduce circuit load. Select via [CH1] -> Impedance menu.

7.7 Probe Ratio

Manually set the probe attenuation ratio (0.01X-1000X) via [CH1] -> Attenuation menu for accurate measurements. Adjust the channel attenuation ratio whenever the probe ratio changes.

MenuAttenuation Ratio (display amplitude of the signal: actual amplitude of the signal)
1X1:1
10X10:1
50X50:1
100X100:1
500X500:1
1000X1000:1

7.8 Waveform Invert

Open [CH1], click the inverting menu to turn on/off waveform inversion. When on, the waveform voltage value is inverted, affecting mathematical operations and measurements.

7.9 Fine Tuning

Open [CH1], click the Fine adjustment menu. Fine adjustment changes vertical resolution for detailed waveform observation. Coarse adjustment sets vertical gear in steps of 1 to 2 to 5.

8. Horizontal System Settings

This section provides detailed information on the oscilloscope's horizontal system, including settings and operations.

8.1 To Adjust the Horizontal Position

Horizontal displacement refers to the position of waveform trigger points relative to the screen center. Adjust by rotating the horizontal displacement knob.

8.2 To Adjust the Horizontal Time Base

Horizontal time base (s/div) represents the time value per grid. The adjustable range is 2ns/div~100s/div (default 1us/div). Adjust by rotating the horizontal time base knob.

8.3 Pan or Zoom Single Acquisition or Stopped Acquisition

After stopping, the display may contain collected data. Pan and zoom the data collected in a single or stopped acquisition.

8.4 Delayed Sweep

Delayed scanning magnifies a waveform segment for detailed viewing. Click the button in the button area to divide the screen into two display areas.

Delayed sweep mode showing waveform before and after amplification.
Figure 8.2 Delayed Sweep Mode.
  • Waveform before amplification: Upper half of the screen shows the main time base. Adjust horizontal displacement or base gear to move/expand the area.
  • Amplified waveform: Lower half shows the horizontally extended delayed scanning waveform. Delayed scanning time base improves resolution.

Tip: The delay scanning time base should be less than or equal to the main time base.

8.5 Horizontal Mode

Click [HORIZ MENU] to set the horizontal mode: YT, XY, and Roll.

8.5.1 YT Mode

Y-axis represents voltage, X-axis represents time. In YT mode, when the horizontal time base is 100 ms/div or slower, the oscilloscope enters scanning mode, collecting data before and after the trigger point.

Tips: For low-frequency signals in scanning mode, set "channel coupling" to "DC".

8.5.2 XY Mode

Both X and Y axes represent voltage levels. Useful for phase difference measurement using the Lissajous method.

Lissajous figure for phase difference measurement.
Figure 8.3 Lissajous Figure for Phase Difference Measurement.

Phase difference measurement: The Lissajous method measures phase difference between two signals of the same frequency. The phase difference angle (θ) can be calculated using sin θ = A/B or sin θ = C/D, where A, B, C, D are defined in the diagram. The X-Y function can test phase changes in a circuit network.

Tips: Reducing the sampling rate can improve the display effect of Lissajous graphics.

8.5.3 Roll Mode

Click [HORIZ MENU] -> Mode -> Roll. In this mode, the waveform scrolls from right to left. The horizontal gear range is 100ms to 100s.

9. Sample System Settings

Click [Acquire] to enter the settings menu for the sample system.

9.1 Acquire Type

Open [Acquire], click Acquire type, and select the collection type:

  • Real-time sampling: Uses real-time digital technology to collect waveforms.
  • Equivalent sampling: Reconstructs waveforms using equivalent sampling techniques.

9.2 Acquire Mode

Open [Acquire], click Acquire Mode, and select from Normal, Average, Peak, High Resolution. Default is Normal.

  • Normal Mode: Samples signals at equal time intervals for waveform reconstruction. Best for most waveforms.
  • Average Mode: Averages multiple sampled waveforms to reduce random noise and improve vertical resolution. Higher average frequency means less noise but slower response to waveform changes. Set average times (4, 8, 16, 32, 64, 128) via "Average Times" button.
  • Peak Mode: Collects maximum and minimum values within the sampling interval to capture signal envelopes or narrow pulses. Avoids aliasing but may have higher noise. Displays pulses at least as wide as the sampling period.
  • High Resolution: Uses supersampling to average adjacent points, reducing random noise and producing smoother waveforms. Different from "Average" mode.

9.3 Memory Depth

Memory depth refers to the number of waveform points stored in a single trigger acquisition. It reflects the storage capacity of the acquisition memory.

Diagram illustrating memory depth with pre-sample, trigger point, and delayed sample.
Figure 9.1 Memory Depth.

Memory depth (MDepth) is related to sample rate (SRate) and horizontal time base scale (TScale) by: MDepth = SRate * TScale * HDivs. Higher memory depth ensures a higher sample rate under the same horizontal time base scale.

Open [Acquire], click Mem Depth, and select storage depth options:

  • Single channel mode: 4K, 40K, 80K.
  • Full channel mode: 4K, 40K.

10. Trigger Settings

Triggering involves setting conditions for the oscilloscope to capture and display waveforms. Correct trigger settings convert unstable or blank screens into meaningful waveforms. Enter the trigger menu by clicking [TRIG MENU].

10.1 Trigger LEVEL

Trigger level is the signal voltage corresponding to the set trigger point. The trigger flag and level line move with changes. For slope, amplitude, and over-amplitude triggering, set levels A and B via Level Select in [TRIG MENU].

10.2 Trigger Source

Click [TRIG MENU] -> Source. Available sources: CH1~CH2, External, Line.

  • Analog channel: CH1-CH2 input signals can be trigger sources. Select a channel with connected signals for stable triggering.
  • External source: Triggered on [EXT TRIG] terminal (0-3.3V CMOS waveform).
  • Line: Triggered at 50% level of AC power signal.

Tips: Do not input signals exceeding 3.3 volts to [EXT TRIG/GEN OUT] connector.

10.3 Trigger Mode

The acquisition memory is divided into pre-triggered and post-triggered buffers. The oscilloscope first fills the pre-triggered buffer, then searches for a trigger. After a trigger, it fills the trigger buffer and displays data.

Schematic diagram of acquisition memory showing pre-trigger and post-trigger buffers.
Figure 10.1 Schematic Diagram of the Acquisition Memory.

Trigger modes: Normal, Auto, Single trigger (default Auto).

  • Normal mode: Updates waveform only with a valid trigger. Displays original waveform if no new trigger. Use for effectively triggered waveforms.
  • Auto mode: Freely runs collection without effective triggering. Allows untriggered scanning waveforms at 100ms/grid or slower. Completes triggered acquisition with valid trigger, or non-triggered acquisition if no valid trigger.
  • Single mode: Ends collection and enters stop state only with a valid trigger.
  • Forced trigger: In normal and single trigger modes, pressing the forced trigger button generates a trigger signal.

Set to "automatic" mode if signal characteristics are unknown to ensure waveforms are displayed, even if unstable.

10.4 Trigger Type

The machine has multiple triggering functions. Open [TRIG MENU], click Type, and select the trigger type.

10.4.1 Edge Trigger

Triggered at the trigger threshold specified by the input signal for the edge.

Diagram showing rising and falling edges with trigger level and trigger points.
Figure 10.2 Rising/Falling Edge Trigger.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Edge Trigger" from the Type menu.
  2. Source: Select CH1, CH2, external, or Line from the Source menu.
  3. Slope Type: Select Rising, Falling, or Either from the Slope menu.
  4. Mode: Select Automatic or Normal from the Mode menu.
  5. 50%: Sets trigger level to the vertical midpoint of the peak-to-peak value.
  6. Holdoff: Sets waiting time between triggers to ensure stable display of complex waveforms.

10.4.2 Pulse Trigger

Triggers on a specified width of positive or negative pulses. Set trigger source, polarity, limiting conditions, and pulse width in this menu.

Diagram showing positive and negative pulse widths with trigger level.
Figure 10.3 Positive Pulse Width/Negative Pulse Width.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Pulse" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Polarity: Select positive or negative polarity from the Polarity menu.
  4. When: Select Trigger Condition (equal, not equal, greater than, less than) from the When menu.
  5. Width: Set trigger release via multifunctional knob V0 or numeric keyboard.
  6. 50%: Sets trigger level to the vertical midpoint of the peak-to-peak value.
  7. Mode: Select Automatic or Normal from the Mode menu.
  8. Holdoff: Sets waiting time between triggers.

10.4.3 Video Trigger

Captures complex waveforms of standard analog and high-definition video signals. Detects vertical and horizontal intervals to generate a trigger. Supports NTSC and PAL.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Video Trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Polarity: Select positive or negative polarity from the Polarity menu.
  4. Standard: Select NTSC, PAL/SECAM from the Standard menu.
  5. Sync: Select field or line (ScanLine, LineNum, OddField, EvenField, AllFields) from the Sync menu.
  6. Line: Set line number to be triggered. Range 1-525 (NTSC) or 1-625 (PAL/SECAM).
  7. Mode: Select Automatic or Normal from the Mode menu.
  8. Holdoff: Sets waiting time between triggers.

10.4.4 Slope Trigger

Sets positive or negative slope triggering from one level to another within a specified time.

Diagram showing positive and negative slope times with upper and lower trigger levels.
Figure 10.4 Positive Slope Time/Negative Slope Time.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Slope Trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Slope Type: Select rising or falling edges from the Slope menu.
  4. Level: Select set level 1, level 2, or level 12 from the Level menu.
  5. When: Select Trigger Condition (less than, greater than, not equal, equal) from the When menu.
  6. Time: Set timeout period (8ns to 10s) via multifunctional knob V0.
  7. Mode: Select Automatic or Normal from the Mode menu.
  8. Holdoff: Sets waiting time between triggers.

10.4.5 Overtime Trigger

Triggered when the time interval (ΔT) from the rising or falling edge of the input signal to the end of the adjacent falling or rising edge is greater than the set timeout time.

Diagram showing overtime trigger condition with time out greater than delta T.
Figure 10.5 Overtime Trigger.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Overtime trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Slope type: Select Positive or Negative edges from the Slope Type menu.
  4. Time: Set timeout period (8ns to 10s) via multifunctional knob V0.
  5. 50%: Sets trigger level to the vertical midpoint of the peak-to-peak value.
  6. Mode: Select Automatic or Normal from the Mode menu.
  7. Holdoff: Sets waiting time between triggers.

10.4.6 Window Trigger

Provides high and low trigger levels. Triggers when the input signal passes the high or low trigger level.

Diagram showing window trigger with high and low trigger levels.
Figure 10.6 Window Trigger.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Window Trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Level Select: Adjust trigger level (Level 1, Level 2, Level 12) from the Level menu.
  4. Mode: Select Automatic or Normal from the Mode menu.
  5. Holdoff: Sets waiting time between triggers.

10.4.7 Pattern Trigger

Identifies triggering conditions by searching for specified code patterns (combination of channel logic 'and', H, L, X, rising/falling edge).

Diagram showing pattern trigger logic for CH1-CH4.
Figure 10.7 Pattern Trigger.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Pattern trigger" from the Type menu.
  2. Logic: Select AND, OR from the Logic menu.
  3. Pattern: Select x, 1, 0, Rising, Falling, or Either from the Pattern menu.
  4. Level: Select CH1 or CH2 from the Level menu.
  5. Mode: Select Automatic or Normal from the Mode menu.
  6. Holdoff: Sets waiting time between triggers.

10.4.8 Interval Trigger

Triggered when the interval between two consecutive rising (or falling) edges satisfies the set time condition (<, >, !=, =).

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Interval Trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Slope: Select rising or falling edge from the Slope menu.
  4. When: Select Trigger Condition (greater than, less than, equal, not equal) from the When menu.
  5. Time: Set trigger release via multifunctional knob V0 or numeric keyboard.
  6. Mode: Select Automatic or Normal from the Mode menu.
  7. Holdoff: Sets waiting time between triggers.

10.4.9 Under Amp Trigger

Triggers a pulse that crosses one trigger level but not another.

Diagram showing under amplitude trigger with trigger points and levels.
Figure 10.8 Runt Trigger.

Click [TRIG MENU] to open the trigger menu.

  1. Type: Select "Under Amp Trigger" from the Type menu.
  2. Source: Select CH1 or CH2 from the Source menu.
  3. Polarity: Select positive or negative polarity from the Polarity menu.
  4. When: Select Trigger Condition (greater than, less than, equal, not equal) from the When menu.
  5. Time: Set slope time (8ns to 10s) via multifunctional knob V0.
  6. Mode: Select Automatic or Normal from the Mode menu.
  7. Holdoff: Sets waiting time between triggers.

11. Math Operations

This series of oscilloscopes supports algebraic and FFT operations between various analog channel waveforms. Enter the Math menu by clicking [Math].

11.1 Arithmetic Operations

Supported operations: A+B, A-B, A*B, A/B.

  1. Operator: Select algebraic operations from the Operator menu.
  2. Source 1, Source 2: Select CH1 or CH2 as sources.
  3. Offset: Set vertical offset of the calculation result display window via multifunctional knob V0.

11.2 FFT (Fast Fourier Transform)

The FFT function analyzes the frequency components of a signal.

  1. Operator: Select FFT operation.
  2. Source: Select CH1 or CH2.
  3. Center: Set center frequency via multifunctional knob V0.
  4. Span: Set frequency span via multifunctional knob V0.
  5. Ref Level: Set vertical offset of the calculation result display window via multifunctional knob V0.
  6. Scale: Set vertical gear of the calculation result display window via multifunctional knob V0.
  7. Vertical Units: Select dB or VRms.
  8. Window: Select from 6 FFT window functions (Rectangular, Hanning, Hamming, Blackman-Harris, Triangle, Flattop) based on waveform characteristics.
  9. Show Only: Select whether to display the FFT function separately.
  10. AutoScale: Automatically adjusts vertical gear and offset for optimal display.

12. Measurements

The oscilloscope can measure displayed waveforms using screen grid scale, cursor, or automatic measurement. Access the menu by clicking [Measure].

12.1 DVM (Digital Voltmeter)

Enter DVM settings via the DVM menu tab.

  1. Enable: Turn on/off CH1 Enable or CH2 Enable.
  2. Type: Select display type (DC RMS, AC RMS, DC).
  3. Return: Go back to the previous menu.

12.2 Statistics

Click [Meas] -> Statistics to turn on/off statistics. Displays current value, average, minimum, maximum, root mean square error, and count of measurement items.

12.3 Clear All

Click [Meas] -> Clear All Menu Tags to clear all measurement parameter statistical functions.

12.4 All Measure

Click [Meas] -> All Measurement to display all measurement items.

12.5 Settings

Click [Meas] -> Settings to open the settings menu. Select 8 types of delay measurements (FRR, FFF, FRF, FFR, LRR, LRF, LFR, LFF). Set data source A and B.

12.6 Gate

Click [Meas] -> Gate Control Menu to open settings. Gate control measurement is only active when measurement type is enabled, measuring waveforms between cursors A and B.

12.7 Quick Measurement after AUTO

After connecting a signal and pressing [Auto Scale]:

  • Single cycle: Displays signals for a single cycle, measuring time and voltage parameters.
  • Multi cycle: Displays signals for multiple cycles, measuring time and voltage parameters.
  • Automatic range: Automatically selects scale (Horizontal vertical gear, Horizontal gear, Vertical gear).
  • Data source: Sets channels to be automatically set (Display only, All).

13. Cursor Measurement

Cursors measure X-axis (time) and Y-axis (voltage) values of selected waveforms. A stable waveform display is needed. Enter cursor settings via [Cursor].

  • X cursor: Vertical dashed line for horizontal adjustment, measures time (s) and frequency (Hz).
  • Y cursor: Horizontal dashed line for vertical adjustment, measures amplitude.

13.1 Manual Mode

Manually adjust cursors to measure signal source waveform values. Results change in real-time with cursor position.

  1. Source: Select CH1, CH2, MATH.
  2. AX: Select horizontal position of cursor A.
  3. BX: Select horizontal position of cursor B.
  4. AXBX: Adjusts both horizontal positions simultaneously, maintaining spacing.
  5. AY: Select vertical position of cursor A.
  6. BY: Select vertical position of cursor B.
  7. AYBY: Adjusts both vertical positions simultaneously, maintaining spacing.

13.2 Track Mode

Two cursors (A and B) measure X and Y values of two different sources. Markers automatically position on the waveform when moving cursors.

14. Display Settings

The display settings menu allows users to configure waveform display type, afterglow time, waveform brightness, grid type, and brightness. Enter via [Display].

14.1 Type

Select between Vector and Point. Vector connects sampling points with lines; Point displays only sampling points. Vector provides a more realistic waveform.

14.2 Intensity

Set brightness value (0% to 100%) via multifunctional knob V0.

14.3 Grid

Select screen grid types: Dot, Line, or Close.

14.4 To Set the Grid Brightness

Set grid brightness (0%~100%) via multifunctional knob V0 or numeric keypad.

14.5 To Set the Screen Brightness

Set screen brightness (0%~100%) via multifunctional knob V0 or numeric keypad.

14.6 Persist (Afterglow)

Select afterglow time (off, infinite, 1s, 5s, 10s, 30s).

  • Specific value: Observes burrs or low-probability events.
  • Infinite: Displays newly acquired waveforms without clearing previous ones, with lower brightness for older waveforms. Useful for measuring noise, jitter, and capturing occasional events.

15. System Utility Functions

Enter system configuration via [Utility]. Users can set system-related functional parameters or interfaces.

15.1 Language

Switch between 14 languages (including Chinese or English) via [Utility] -> Language.

15.2 Sound

Turn sound on or off via [Utility] -> Sound.

15.3 Update (Firmware)

Upgrade firmware for USB storage devices (takes ~5 minutes) via [Utility] -> Upgrade.

  1. Insert USB storage device with upgrade package into USB HOST interface.
  2. Press [Utility] to enter auxiliary menu.
  3. Press Upgrade -> Upgrade Firmware.
  4. Rotate V0 to select upgrade file and press V0 to confirm. Then press Start Upgrade.
  5. After upgrade, oscilloscope restarts, firmware updates, and self-calibrates.

15.4 Pass/Fail Test

Monitors signal changes. Enter via [Utility] -> Pass/Fail.

15.4.1 To Enable or Disable the Pass/Fail Test Function

Enter Pass Test Settings via Pass/Fail menu. Enable/disable pass testing via Enable menu.

15.4.2 To Select the Source

Select channels CH1 to CH2 via Source Selection menu.

15.4.3 Regular

Create pass/fail rules via Regular menu.

  1. Vertical: Set vertical tolerance range (0.020div - 4.00div) via Regular -> Vertical.
  2. Level: Set horizontal tolerance range (0.025div-8.00div) via Regular -> Horizontal.
  3. Create: Create a rule template based on vertical and horizontal settings.
  4. Save: Select storage location for rules.
  5. Transfer out: Retrieve waveform rule settings.

15.4.4 Message

Select to turn on/off display information for failed attempts via Pass/Fail -> Message.

15.4.5 Mode

Select Pass, Fail, Pass Ring, Fail Ring via Pass/Fail -> Mode.

  • Pass: Output negative pulse train during testing.
  • Fail: Output negative pulse train during failed test.
  • Pass Ring: Output negative pulse train with ringing when passing.
  • Fail Ring: Output negative pulse train with ringing during failed test.

15.4.6 Run

Run pass/fail test via Pass/Fail -> Run.

15.4.7 Stop

Stop pass/fail test via Pass/Fail -> Stop.

15.4.8 Out Stop

Select on/off via Pass/Fail -> Out Stop. Enters STOP state on output or continues running.

15.5 System Info

Display software version, hardware version, serial number, and other information via [UTILITY] -> System Info.

15.6 Calibrate

Press [UTILITY] -> Calibrate. Follow prompts for self-calibration. Run to optimize signal path for maximum accuracy, especially if ambient temperature changes by >5°C. Preheat for 20 minutes before calibrating. Disconnect probes/cables, select "Self Calibration" option.

15.7 FrontPanel Selftest

Test functionality of all buttons, knobs, and LED lights via [Utility] -> FrontPanel Selftest.

Display source code license via [Utility] -> Legal Information.

16. Quick Keys

Quick keys provide automatic settings for useful displays of input signals.

16.1 Auto Scale

Automatically sets oscilloscope controls to display input signal waveforms. Recognizes waveform type and adjusts controls.

ParameterSettings
AcquireNormal
CursorClose
Timebase ModeYT
DisplayVector
HorizontalAdjusted
SEC/DIVAdjusted
Holdoff20ns
Trigger Level50%
Trigger ModeAuto
Trigger SourceAdjusted
Trigger SlopeAdjusted
Trigger TypeEdge
SyncAdjusted
StandardAdjusted
BandwidthClose
CouplingDC
VOLTS/DIVAdjusted

16.2 Default Setup

Press [Default Setup] to display waveforms for four channels with default settings.

ParameterSetting
Horizontal
Horizontal Time Base200us
Horizontal Position0s
Delayed SweepOff
Timebase ModeYT
FineOff
Horizontal ExpansionCenter
Vertical
Vertical Scale1V
Vertical OffsetCH1: 2V, CH2: -2V
CH1/CH2CH1-CH2 On
Channel CouplingDC
BW LimitOff
Attenuation1X
Input Impedance1MΩ
InvertOff
Channel UnitV
FineOff
Ch-Ch Skew0s
Offset Cal0V
Acquire
Acquisition ModeNormal
Memory Depth4K
Trigger
Trigger TypeEdge Trigger
Source SelectionCH1
Edge TypeRising
Trigger Holdoff1us
Display
Display TypeVector
Persistence TimeOff
Intensity100%
GridLine
Grid Brightness60%
Brightness80%
AWG (for 'D' models)
AWGOff
WaveSine
Frequency1KHz
Amplitude1.5V
Offset0V
Start Phase0°
Setting TypeOff
ImpedanceHighZ
Cursor
ModeOff
Math
OperatorA+B
OperationOff

17. Arbitrary Waveform Generator (AWG)

This series of oscilloscopes (DSO2D models) includes a built-in arbitrary waveform generator, combining oscilloscope and signal generator functions for user convenience.

17.1 Output Waveform Type

Click [WAVE GEN] to enter signal generator settings and select basic wave output.

  1. Waveform: Select Sine, Square, Ramp, Exp, Noise, DC, Arb1-Arb4 from the Wave menu.
  2. Frequency: Set frequency via multifunctional knob V0. (DC and noise have no frequency parameters).
  3. Amplitude: Set amplitude via multifunctional knob V0.
  4. Offset: Set offset via multifunctional knob V0.
  5. Symmetry/Duty Cycle: For square waves, set duty cycle; for triangular waves, set symmetry.
  6. Impedance: Select output impedance (high impedance or 50Ω).

17.2 Modulation

The signal generator offers amplitude modulation (AM) and frequency modulation (FM). Modulated waveforms consist of a carrier wave and a modulation signal (sine, square, or triangular wave).

Click [WAVE GEN] to enter modulation settings. Select AM, FM, or PM (phase modulation) from the Type menu.

17.2.1 Amplitude Modulation (AM)

AM varies the carrier wave amplitude with the modulation wave.

Diagram illustrating Amplitude Modulation with modulating signal, carrier signal, and modulated waveform.
Figure 17.1 AM.
  • Modulation waveform: Select sine, square, or triangular wave from the Wave menu.
  • Modulation frequency: Set frequency via Mod Freq menu.
  • Modulation depth: Set depth (0-120) via Depth menu.

17.2.2 Frequency Modulation (FM)

FM varies the carrier frequency with the modulation wave.

Diagram illustrating Frequency Modulation with modulating signal, carrier signal, and modulated waveform.
Figure 17.2 FM.
  • Modulation waveform: Select sine, square, or triangular wave from the Wave menu.
  • Modulation frequency: Set frequency via Mod Freq menu.
  • Deviation: Set offset via Deviation menu.

17.3 Burst

Outputs burst waveforms with a specified number of cycles. Supports internal or manual control, with N-loop and infinite burst types. Access via [Burst].

  1. Type of burst: Choose between multiple cycles or infinite from the Type menu.
  2. Counting: Set number of cycles via count menu.
  3. Trigger: Press the Trigger button to output a pulse train with a specified number of cycles.

18. Remote Control

The oscilloscope can be remotely controlled via USB.

18.1 Remote Control via USB

  1. Connect the device: Connect the oscilloscope (USB VIEW interface) to the PC (USB HOST interface) using a USB data cable.
  2. Install USB driver: The PC automatically installs the driver upon first connection.
  3. Search for device resources: Open IO software and it will automatically search for connected resources. Click Rescan if needed.
  4. View Device Resources: Resources appear in the USB (USB0) directory, displaying model, serial number, version, and USB interface information.
  5. Control the instrument remotely: Click Interactive IO in the IO interface to open the remote command control panel to send commands and read data.

19. Troubleshooting

This section lists common faults and troubleshooting methods. If issues persist, contact Hantek technical support.

19.1 Oscilloscope Does Not Turn On

  1. Check power cord connection.
  2. Check if power switch is pressed.
  3. Restart instrument after checks.
  4. If still not turning on, contact technical support.

19.2 Serious Distortion of Input Signal

  1. Check probe connection to BNC channel.
  2. Check probe connection to phenomenon being measured.
  3. Check probe calibration.

19.3 No Signal Appears on the Screen

  1. Check probe connection to BNC connector.
  2. Check if channel (CH1-CH2) is open.
  3. Check for signal output from the tested channel.
  4. For DC signals with large amplitude, increase amplitude range.
  5. Press auto measurement button to detect signal.
  6. If still no waveform, contact technical support.

19.4 USB Flash Drive Cannot Be Recognized

  1. Check if USB flash drive is functioning.
  2. Confirm capacity (recommended ≤16GB).
  3. Confirm format is FAT32.
  4. Re-insert USB drive after opening oscilloscope.
  5. If still not recognized, contact technical support.

19.5 Waveform Cannot Be Triggered

  1. Check if triggering data source matches input channel.
  2. Check if trigger level is adjusted correctly (press trigger level knob to return to center).
  3. Check if triggering method is correct (default is edge triggering; select appropriate method for input signal).

20. Specifications

The Hantek DSO2D20 series oscilloscopes offer the following technical specifications:

ModelAnalog ChannelAnalog BandwidthReal time sampling rateMemory DepthWaveform GeneratorDecode
DSO2C202200MHz2GSa/s80K-5 types
DSO2C352350MHz2GSa/s80K-5 types
DSO2C502500MHz2GSa/s80K-5 types
DSO2D202200MHz2GSa/s80K15 types
DSO2D352350MHz2GSa/s80K15 types
DSO2D502500MHz2GSa/s80K15 types

Note: DSO2C20/DSO2C35/DSO2C50 models do NOT include Waveform Generation. DSO2D20/DSO2D35/DSO2D50 models DO include Waveform Generation.

ParameterSpecification
Rising Time (typical)≤1.75ns (200MHz), ≤1ns (350MHz), ≤700ps (500MHz)
No. of Input/Output Channels2 analog channel inputs; 1 signal generator channel output (DSO2D20 series)
EXT Channel Input1 EXT channel input
Sampling ModeReal-time sampling
Max. Sample Rate of Analog Channel2GSa/s (single-channel), 1GSa/s (all channels)
Max. Memory Depth80Kpts (single-channel), 40Kpts (all channels)
Peak DetectionCapture 1ns glitches
LCD Size and Type7-inch capacitive multi-touch screen
Display Resolution800 × 480 Pixels
Vertical System Analog Channel
Input CouplingDC, AC, GND
Input Impedance1MΩ±1%, 50Ω±1%
Input Capacitance17pF±3pF
Maximum Input Voltage (1MΩ)CAT I 300 VRMS, 400 Vpk; Transient Overvoltage 1600 Vpk
Maximum Input Voltage (50Ω)5VRMS
Vertical Resolution8bit
Vertical Sensitivity Range (1MΩ)500uV/div ~ 10V/div
Vertical Sensitivity Range (50Ω)500uV/div ~ 1V/div
Offset Range (1MΩ)±1V (500uV/div ~ 50 mV/div), ±10V (100mV/div ~ 500 mV/div), ±100 V (1V/div ~ 10 V/div)
Offset Range (50Ω)±1V (500uV/div ~ 50 mV/div), ±10V (100mV/div ~ 500 mV/div), ±100V (1 V/div)
Dynamic Range±5div (8bit)
Bandwidth Limit (Typical)20MHz, 100MHz, 200MHz, 350MHz; selectable for each channel
DC Gain Accuracy±3%Fullscale
DC Offset Accuracy<200mV/div (±0.1div±2mV±1.5%of offset value), >200mV/div (±0.1div±2mV±1.0%of offset value)
Channel-to-Channel Isolation40dB, from DC to maximum rated bandwidth of each model
ESD Tolerance±8kV (on input BNCs)
Horizontal System Analog Channel
Range of Time Base (200MHz)2ns/div~1ks/div
Range of Time Base (350MHz)1ns/div~1ks/div
Range of Time Base (500MHz)500ps/div~1ks/div
Time Base Accuracy±50ppm
Time Base Delay Range (before triggering)≥1/2 screen width
Time Base Delay Range (after triggering)1 s or 100 div, whichever is greater
Time Interval (ΔT) Measurement±(1 sample interval)±(50ppm××readout)±50ps
Horizontal Mode (YT)Default
Horizontal Mode (XY)X1 = Channel1, Y1 = Channel2
Horizontal Mode (SCAN)Time base≥100ms/div, available to enter or exit the SCAN mode by rotating the Horizontal SCALE knob
Horizontal Mode (ROLL)Time base≥100ms/div, available to enter or exit the ROLL mode by rotating the Horizontal SCALE
Acquisition System
Acquisition Mode (Normal)Default
Acquisition Mode (Peak Detection)Capture 1ns glitches
Acquisition Mode (Average Mode)2, 4, 8, 16…128 are available for you to choose, averaging point by point
Acquisition Mode (High Resolution)12bits (max.)
Arbitrary Waveform Generator (AWG)
Sample Rate200MSa/s
Vertical Resolution12bit
Max. Frequency25MHz
Standard WaveformSine, Square, Ramp, Exp, Noise, DC
ArbArb1, Abr2, Arb3, Arb4
Sine Frequency Range0.1Hz~25MHz
Square Frequency Range0.1Hz~10MHz
Ramp Frequency Range0.1Hz~1MHz
Exp Frequency Range0.1Hz~5MHz
Arb Frequency Range0.1 Hz-25MHz
Waveform Length4K
Frequency Accuracy100ppm (<10kH), 50ppm (>10kHz)
Frequency Resolution100 mHz or 4 bits (whichever is greater)
Amplitude Output Range (HighZ)10mV~7Vpp
Amplitude Output Range (50Ω)5mV~3.5Vpp
DC Offset Accuracy Range (HighZ)±3.5V
DC Offset Accuracy Range (50Ω)±1.75V
DC Offset Accuracy Resolution100 uV or 3 bits (whichever is greater)
DC Offset Accuracy Accuracy2%(1 kHz)
Output Impedance50Ω ± 1%

21. Accessories

The following accessories are included with your Hantek Digital Oscilloscope:

Order informationOrder number
Host model
2GSa/S, 200MHz 2-channel oscilloscopeDSO2C20
2GSa/S, 350MHz 2-channel oscilloscopeDSO2C35
2GSa/S, 500MHz 2-channel oscilloscopeDSO2C50
2GSa/S, 200MHz 2-channel oscilloscope + AWGDSO2D20
2GSa/S, 350MHz 2-channel oscilloscope + AWGDSO2D35
2GSa/S, 500MHz 2-channel oscilloscope + AWGDSO2D50
Standard accessories
USB cable--
Power Line--
Oscilloscope probe200MHz machine PP200B * 1, 350MHz machine PP300B * 1, 500MHz machine PP500B * 1
Crocodile Clip LineDSO2C20 series * 1, DSO2D20 series * 2
Accessories for DSO2D20/DSO2D35/DSO2D50 models including oscilloscope, certificate, alligator clip cable, passive probe, power cable, and square USB cable.
Figure 21.1 Accessories for DSO2D20/DSO2D35/DSO2D50 models.
Accessories for DSO2C20/DSO2C35/DSO2C50 models including oscilloscope, certificate, alligator clip cable, passive probe, power cable, and square USB cable.
Figure 21.2 Accessories for DSO2C20/DSO2C35/DSO2C50 models.

22. User Tips

  • If using the self-test generator, be aware it can be noisy; the internal 25 MHz generator offers better signal quality.
  • Consider acquiring a transport bag for easier portability and protection of your oscilloscope.

23. Warranty and Support

23.1 Warranty Summary

Qingdao Hantek Electronic Co., LTD. (Hantek) warrants that the host and accessories are free from material and process defects during the warranty period. Defective products will be repaired or replaced free of charge. Refer to the Hantek official website for detailed warranty regulations. For service, contact a Hantek repair center or local office.

Hantek disclaims all other warranties, express or implied, including merchantability and fitness for a special purpose. Hantek is not liable for indirect, special, or consequential damages.

23.2 Contact Information

If you have any questions, you can obtain service and support through the following ways:

  • Email: service@hantek.com, support@hantek.com
  • Website: http://www.hantek.com
  • Address: #35 Building, No. 780 Baoyuan Road, High-tech Zone, Qingdao, Shandong, China 266114
  • Switchboard: 400-036-7077
  • Tel: (0086)532-55678770 & 55678772 & 55678773
  • Zip code: 266114