1. Introduction
The Freshliance G100 system provides real-time temperature and humidity monitoring for various environments such as cold rooms, laboratories, small warehouses, and other small areas. This comprehensive system integrates BlueTag series data loggers, the G100-S gateway, a router, a cloud platform, and a mobile application.
BlueTag series data loggers record on-site temperature and humidity, transmitting data continuously to the G100-S WiFi gateway via Bluetooth. The G100-S then relays this data to the cloud in real-time using WiFi communication. The system supports connections for up to 20 configurable Bluetooth loggers. Users can remotely access real-time temperature and humidity data, view live changes, and review data history through the mobile app or PC platform. The system also provides SMS or email alerts for any deviations. For challenging or hard-to-reach locations, BlueTag series loggers can be equipped with one or two external probes.

2. Setup
- Powering the Gateway: Connect the G100 WiFi Gateway to a power source using the provided DC 5V/1A Mini-USB adapter.
- WiFi Connection: Connect the G100 Gateway to your local WiFi network. Follow the instructions in the mobile application for initial network configuration.
- Sensor Pairing: Pair your BlueTag Bluetooth sensors with the G100 Gateway via Bluetooth communication. Refer to the BlueTag sensor's individual manual for specific pairing instructions. The G100 can support up to 20 sensors.
- External Probes (Optional): If using external probes with BlueTag sensors, ensure they are securely connected to the sensor unit.

3. Operating Instructions
Once the G100 Gateway and BlueTag sensors are set up and connected, you can begin monitoring your environment.
- Real-time Monitoring: Access the mobile application (Android and iOS) or the web platform to view real-time temperature and humidity data from all connected sensors.
- Data Review and Management: The platform allows for comprehensive data analysis, filtration, and graphing of live changes and historical data.
- Alarm Notifications: Configure alarm thresholds for temperature and humidity. The system will send SMS or email alerts if these thresholds are exceeded.
- Offline Data Caching: In case of temporary network loss, the system features persistent caching. Data recorded during offline periods will be automatically re-uploaded to the cloud once network connectivity is restored, ensuring no data loss.
Application Scenarios



4. Maintenance
- Cleaning: Gently wipe the G100 Gateway and BlueTag sensors with a dry, soft cloth. Do not use liquid cleaners or abrasive materials.
- Battery Replacement (BlueTag T10 Logistics Sensor): The BlueTag T10 uses a 3.0V CR2032 lithium battery with a battery life of approximately 1 year. Replace the battery when indicated by the device or app.
- Rechargeable Battery (BlueTag Storage Sensors): BlueTag storage sensors are equipped with rechargeable lithium batteries. Ensure they are charged as needed to maintain continuous operation.
- Firmware Updates: Periodically check the mobile application or web platform for available firmware updates for the G100 Gateway and BlueTag sensors to ensure optimal performance and access to new features.
5. Troubleshooting
- No Data on App/Web:
- Check if the G100 Gateway is powered on and connected to your WiFi network. The gateway has persistent caching, so data should re-upload once the network is restored.
- Ensure BlueTag sensors are powered on and within Bluetooth range of the G100 Gateway.
- Verify that the mobile app or web platform has an active internet connection. - Inaccurate Readings:
- Ensure sensors are placed in appropriate locations, away from direct heat sources, cold drafts, or excessive moisture.
- Calibrate sensors if necessary, following instructions in the app or sensor manual. - Sensor Disconnection:
- Check the battery level of the BlueTag sensor.
- Ensure the sensor is within the 100m (open area) Bluetooth range of the G100 Gateway. Obstacles can reduce this range. - Gateway Not Connecting to WiFi:
- Verify your WiFi network is 2.4GHz, as the G100 operates on this band.
- Check your WiFi password and signal strength.
6. Specifications
G100 WiFi Bluetooth Gateway
| Feature | Specification |
|---|---|
| Model | G100 WiFi Bluetooth Gateway |
| Antenna Type | External antenna*2 |
| Antenna Gain | 6dBi |
| Power | DC 5V/1A, Mini-USB |
| Standby Current | ≤40mA |
| Maximum Current | ≤90mA |
| Maximum Connections | 20 sensors |
| Communication Network | WiFi/Bluetooth |
| WiFi Frequency Band | 2.4GHz |
| Bluetooth Range | 100m (open area) |
| Operation System | APP of Android and iOS and Web |
| Working Temperature | -20℃~+55℃ |
| Working Humidity | 5%~95%RH (non-condensing) |
| Storage and Transportation | -40℃~+70℃ |
| Dimensions | Device: 93mm*62mm*21mm; Antenna: 19.5mm*13mm |
| Weight | 97g |
Bluetooth Sensors for Storage (BlueTag Series)
| Feature | BlueTag TH10R | BlueTag TH10R-I | BlueTag TH30R | BlueTag TH30R-I | BlueTag TH30R-ETU |
|---|---|---|---|---|---|
| Internal Probe Type | Temperature and humidity | ||||
| External Probe | Optional | √ | |||
| External Probe Channels | Max. 2 probes | ||||
| External Probe Type | See External Probe List | ||||
| Internal Temperature Range | -30℃~+70℃ | ||||
| Internal Temperature Accuracy | ±0.5℃ F.S. | ||||
| Temperature Units | ℃, ℉ | ||||
| Alarms | Max. 6 Alarm Points | ||||
| Report | PDF/CSV | ||||
| Connection | Bluetooth 5.2 | ||||
| Start Delay | 0 minute~72 hours | ||||
| Logging Interval | 1 minute ~99 minutes | ||||
| Memory Capacity | 20000 readings each probe | ||||
| Device Dimensions | 83mm*68mm*20.5mm | ||||
| Battery | Rechargeable lithium battery | ||||
| Installation | Hanging, Folding stand | ||||
| Weight | 85g (Main device) | 90g (Main device) | |||
External Probe List
| Feature | Type 1 | Type 2 | Type 3 | Type 4 | Type 5 | Type 6 | Type 7 | Type 8 |
|---|---|---|---|---|---|---|---|---|
| External Probe Type (Temperature Range) | -30℃~+70℃ | -30℃~+70℃ | -40℃~+85℃ | -80℃~+200℃ | -90℃~+400℃ | -90℃~+80℃ | -90℃~+200℃ | -196℃~+200℃ |
| Temperature Accuracy | ±0.5℃ (-90℃~+70℃); ±1.0℃ (Others) | |||||||
| Cable Length | 2m, 5m, 10m, 15m or 20m (5 adhesive cable clips every 2m, for cable management) | |||||||
Bluetooth Sensors for Logistics (BlueTag T10)
| Feature | Specification |
|---|---|
| Model | BlueTag T10 |
| Memory Capacity | Max. 65,000 readings |
| Shelf Life | 1 year |
| Start Delay | 0 minute-72 hours |
| Battery Life | 1 year |
| Logging Interval | 1 minute-99 minutes |
| Wireless Connection | Bluetooth 5.2 |
| Battery | 3.0V CR2032 lithium battery |
| APP Language | Chinese, English, French, German, Portuguese, Russian, Spanish |
| APP OS | Android, iOS |
| Temperature Unit | ℃, ℉ |
| Report Format | PDF & CSV |
| Temperature Range | -30℃~+70℃ |
| Temperature Accuracy | ±0.5℃ F.S. |
| Weight | 7.8 g |
| Dimension | 48mm*31mm*5mm |
| Installation | 3M adhesive tape |
| Certification | CE, RoHS, EN12830 |
7. User Tips
- Cloud Storage: Benefit from 2 years of free data cloud storage, ensuring your historical data is always accessible.
- Flexible Probes: Utilize external probes with your BlueTag sensors for monitoring in hard-to-reach areas or for specific temperature ranges not covered by internal sensors.
- Regular Monitoring: Regularly check the app or web platform to stay informed about your environmental conditions and respond promptly to any alerts.
8. Warranty and Support
The Freshliance G100 WIFI Gateway and associated BlueTag sensors come with a 2-year warranty from the date of purchase.
For technical support, inquiries, or warranty claims, please contact Freshliance customer service:
- Email: sales@freshliance.com
- Phone: 0086-371-89908620
- Website: www.freshliance.com