MESTEK IR07B

Contents

MESTEK Digital Infrared Thermometer User Manual

Models: IR07, IR07A, IR07B

1. Safety Instructions

Please read this instruction manual carefully before using the instrument and keep it properly for future use.

Safety Symbols:

  • ▲ Important: Information that is dangerous.
  • CE: Compliant with European CE safety specifications. This instrument meets the following standards: EN60825-2014.

Warning!

Do not aim the laser at human eyes or reflective surfaces.

Notes:

  • When there is a temperature difference impact (caused by large or sudden changes in ambient temperature), wait about 20 minutes before use to allow the instrument to reach a stable temperature.
  • Avoid electromagnetic fields caused by electric welding and induction heating.
  • Avoid placing the instrument near or on hot objects.
  • Keep the instrument clean and prevent dust from entering the lens barrel.
  • Clean the lens: Use clean compressed air to clean the dust on the surface of the lens, then wipe the surface with a moist cotton swab.
  • Do not use solvent to clean the instrument.

2. Product Overview

The MESTEK Digital Infrared Thermometer is a professional high-temperature meter designed for various applications including barbecue, kitchen cooking, swimming pools, industrial production, mold measurement, steel/smelting, electricity/machinery, glass/ceramic, and heating/ventilation. It offers non-contact temperature measurement and supports K-type thermocouple input for liquid temperature measurement.

MESTEK Digital Infrared Thermometer with key features highlighted

Figure 2.1: MESTEK Digital Infrared Thermometer highlighting features like high/low temperature alarm, D:S ratio, Max/Min/Avg/Diff value, and temperature probe.

Key Features:

  • Support Laser Temperature Measure and Temperature Probe Liquid Temperature Measure.
  • Can Measure Temperature Max, Min, Average, Difference.
  • Color Display screen with backlight.
  • Adjustable Emissivity (0.1-1.0) for accurate measurements across different materials.
  • High and Low Temperature Alarm function.
  • Max Hold function.
  • Accuracy: ±2℃.
  • Object Distance Ratio (D:S): 30:1.
  • Resolution: 0.1℃/℉.
  • Response Time: 0.5 Seconds.
  • Response Wavelength: 8-14μm.
  • Celsius and Fahrenheit Temperature Unit Conversion.
  • Data Hold Display.
  • Positioning Laser On/Off.
  • Automatic Shutdown after 35 seconds without operation.
  • Power: 2 * 1.5V AA Battery (not included).

Product Appearance:

Diagram showing the dimensions and labeled parts of the MESTEK Infrared Thermometer

Figure 2.2: Product Appearance Introduction with key components labeled: Laser, Infrared sensor, Measure trigger, Battery Cover, K-type thermocouple socket, LCD display screen, Number minus button, Laser switch/MODE button, Number plus button.

LCD Display Panel:

Diagram of the LCD screen panel with numbered indicators and their descriptions

Figure 2.3: Introduction to Screen Panel with numbered indicators:

  1. Emissivity
  2. Measurement Indication
  3. Laser Indication
  4. Battery Low Voltage Indication
  5. Data Locking
  6. Irradiation Temp Value
  7. Temperature Unit
  8. Maximum Value Lock
  9. Minimum Value Lock
  10. Average Value
  11. Temperature Difference Value
  12. Alarm
  13. Hi High Alarm
  14. Low Low Alarm
  15. K-type Thermocouple Temp

Video 2.1: Unboxing and overview of the MESTEK Industrial Infrared Thermometer, showing the device and included accessories.

3. Setup

Battery Installation:

The thermometer requires two 1.5V AA batteries. When the battery power is insufficient, the screen displays the battery undervoltage symbol (🔋). At this time, the batteries must be replaced.

  1. Open the battery cover located at the bottom of the handle.
  2. Insert two new 1.5V AA batteries, ensuring correct polarity.
  3. Close the battery cover securely.
Diagram showing how to open the battery cover and insert AA batteries into the thermometer

Figure 3.1: Battery replacement procedure.

Initial Temperature Stabilization:

If the instrument experiences a significant or sudden change in ambient temperature, allow approximately 20 minutes for it to stabilize and reach the new ambient temperature before taking measurements. This ensures accurate readings.

4. Operation

General Measurement:

To take a measurement, press the measurement trigger. The screen will light up and display the current temperature. Release the trigger to hold the reading.

Mode Settings:

Press and hold the MODE key to enter the setting mode. Short press the MODE key to cycle through the following settings: Alarm Hi, Alarm Low, Emissivity (E), and Temperature Unit (C/T).

In setting mode, long press the ▲ or ▼ key to quickly increase or decrease the set value.

4.1. High Alarm Setting:

  1. Long press the MODE key to enter the high alarm setting (Alarm Hi) by default. The value will flash.
  2. Press the ▲ or ▼ key to set the desired high alarm value.
  3. Press the trigger to save the setting. When the measured temperature exceeds this value, the buzzer will sound an alarm.

4.2. Low Alarm Setting:

  1. Long press the MODE key to enter settings, then short press MODE to switch to the low alarm setting (Alarm Low). The value will flash.
  2. Press the ▲ or ▼ key to set the desired low alarm value.
  3. Press the trigger to save the setting. When the measured temperature falls below this value, the buzzer will sound an alarm.
Display showing high and low temperature alarm settings on the thermometer screen

Figure 4.1: High and Low Temperature Alarm Value Setting.

4.3. Emissivity Setting:

Emissivity is the ability of an object's surface to radiate infrared rays. The greater the emissivity, the stronger the radiation ability. The emissivity of most organic or metal oxidized surfaces is between 0.85 and 0.98. This thermometer's default emissivity is 0.95. When measuring, set the thermometer's emissivity to match the measured object's emissivity for accurate results.

  1. Long press the MODE key to enter settings, then short press MODE to switch to the emissivity setting (€). The emissivity value will flash.
  2. Refer to the Emissivity Reference Table below and press the ▲ or ▼ key to set the required emissivity value.
  3. Press the trigger to save the setting.
Diagram showing emissivity adjustment and a table of common material emissivities

Figure 4.2: Emissivity Adjustment and Reference Table.

Emissivity Reference Table:
Measured SurfaceEmissivityMeasured SurfaceEmissivity
Oxidized Metals0.5-0.9Carbon (unoxidized)0.8-0.9
Iron (rusty)0.5-0.7Graphite0.7-0.8
Iron (oxidized)0.6-0.95Carborundum0.9
Iron (casting, unoxidized)0.2Pottery and porcelain0.95
Iron (passivation/forging)0.9Clay0.95
Brass (oxidized)0.5Concrete0.95
Copper (oxidized)0.4-0.8Paper0.95
Zinc (oxidized)0.1Plastic0.95
Aluminum (A3003 alloy, oxidized)0.3Asbestos0.95
Aluminum (A3003 alloy, rough)0.1-0.3Asphalt0.95
Nickel (oxidized)0.2-0.6Basalt0.7
Steel (cold rolling)0.7-0.9Water0.93
Steel (polished)0.4-0.6Wood0.9-0.95
Glass plate0.85Ice0.98

4.4. Temperature Unit Setting:

  1. Long press the MODE key to enter settings, then short press MODE to switch to the temperature unit setting (C/T). The unit symbol (°C/°F) on the display screen will flash.
  2. Press the ▲ or ▼ key to set the required temperature unit (°C or °F).
  3. Press the trigger to save the setting.
Thermometer display showing temperature unit switching between Celsius and Fahrenheit

Figure 4.3: Temperature Unit Switching.

4.5. Laser Settings:

When the instrument screen is lit, short press the ▲ key to activate or turn off the laser aiming function.

Thermometer using its laser for non-contact temperature measurement in an industrial setting

Figure 4.4: Non-contact Temperature Measurement with Laser Positioning.

4.6. Max/Min/Average/Difference Measurement:

When the instrument screen is lit, repeatedly short press the ▼ key to cycle through and display the Max, Min, Avg, or Diff symbols. The instrument will then lock and display the corresponding measurement value (Maximum, Minimum, Average, or Temperature Difference).

Thermometer displaying MAX temperature while measuring bread in an oven

Figure 4.5: MAX Lock function, automatically locking the maximum value of the measurement.

4.7. K-type Thermocouple Temperature Measurement:

  1. Insert the K-type thermocouple plug into the thermocouple socket of the instrument, ensuring correct positive and negative pole signs.
  2. Pull the trigger. The instrument screen will display the 🌡️ symbol.
  3. During measurement, ensure the thermocouple probe contacts the measured object. The screen will display both the K-type thermocouple temperature and the infrared irradiation temperature.
Thermometer with K-type thermocouple inserted, measuring temperature in an engine bay

Figure 4.6: K-type Thermocouple Temperature Measurement.

4.8. Target Distance Ratio (D:S):

The instrument has a specific angle of view and field of view. It is crucial to ensure that the measured object completely fills the field of view of the thermometer, meaning the thermometer only 'sees' the measured object and not other surrounding objects. The larger the measured object, the farther the measurement distance can be. Conversely, the smaller the object, the closer the measurement distance must be. The ratio of the measuring distance (D) to the measured target size (S) is 30:1.

Diagram illustrating the target distance ratio (D:S) for accurate infrared temperature measurement

Figure 4.7: Target Distance Ratio (D:S) 30:1.

4.9. Water/Oil Temperature Measurement Method:

When measuring boiling liquids, do not let smoke or steam from boiling water come into contact with the temperature sensing zone. Maintain an angle of 45° to 60° to the surface of the liquid for accurate readings.

Illustration showing correct and incorrect methods for measuring water/oil temperature with an infrared thermometer

Figure 4.8: Correct and Incorrect Water/Oil Temperature Measurement Method.

5. Maintenance

Cleaning the Instrument:

  • Keep the instrument clean and prevent dust from entering the lens barrel.
  • To clean the lens, use clean compressed air to remove dust from the surface, then gently wipe with a moist cotton swab.
  • Do not use solvents or abrasive cleaners on any part of the instrument.

Battery Replacement:

Replace batteries when the low voltage indicator (🔋) appears on the display. Refer to the 'Battery Installation' section under Setup for detailed instructions.

6. Troubleshooting

If the device is not functioning as expected, please check the following:

  • No display or dim display: Check battery levels and replace if necessary. Ensure batteries are inserted with correct polarity.
  • Inaccurate readings:
    • Ensure the instrument has stabilized to ambient temperature (wait 20 minutes after significant temperature changes).
    • Verify the emissivity setting matches the material being measured.
    • Ensure the target object fills the field of view (D:S ratio).
    • Clean the lens if it is dirty.
    • Avoid strong electromagnetic fields.
  • Alarm not sounding: Check if the high/low alarm values are set correctly and if the alarm function is enabled.
  • Laser not working: Ensure the laser aiming function is activated (short press ▲ key).

For persistent issues, please contact customer support.

7. Specifications

Model Comparison:

Comparison table for MESTEK IR07, IR07A, and IR07B models showing temperature ranges

Figure 7.1: Model Comparison for IR07, IR07A, and IR07B.

Technical Parameters:

Detailed technical parameter table for MESTEK Infrared Thermometers

Figure 7.2: Technical Parameter Table.

ParameterIR07IR07AIR07B
Temperature Range-40°C ~ 1050°C (-40°F ~ 1922°F)-40°C ~ 1350°C (-40°F ~ 2462°F)-40°C ~ 1650°C (-40°F ~ 3002°F)
DisplayVA Reflective Color Screen
D:S30 : 1
Spectral Response8 - 14μm
Resolution0.1°C / 0.2°F
Emissivity0.10 ~ 1.00 (Adjustable)
Laser<1mW / 620-690nm Class 2
Response Time< 0.5 S
Auto Power OffAbout 35 seconds without operation
Irradiation Accuracy-40°C ~ 0°C: ± (3°C/6°F)
1°C ~ 800°C: ± (1.5%+2°C)
801°C ~ 1200°C: ± (3°C)
1201°C ~ 1650°C: ± (5°C)
K-Type Temp Range-40°C ~ 600°C (-40°F ~ 1112°F)
K-Type Accuracy± (1.5%+2°C/4°F)
Operating Environment0 ~ 50°C / 10% - 90% RH
Storage Temperature-20°C ~ 60°C
Dimensions173 x 102 x 54 mm
Power Supply2 x 1.5V AA Batteries
WeightApproximately 210g (without batteries)

8. Accessories

The standard package includes:

  • 1x Infrared Thermometer
  • 1x User Manual
  • 1x K-type Thermocouple
  • 1x Instrument Storage Bag
  • 2x 1.5V AA Batteries (Note: JSON indicates batteries are not included, but OCR lists them as accessories. Following OCR for accessories list.)

9. User Tips

No specific user tips were available from provided reviews or Q&A content.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the official MESTEK website or contact your retailer. You may also find additional guides and documents, including a PDF user manual, on the product's support page (e.g., User manual (PDF)).

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.