1. Introduction
The SWEVY SW93T Digital Brix Refractometer is a high-precision instrument designed for accurately measuring the sugar content (Brix) in various solutions, including fruit juices, beverages, milk tea, and other food products. Featuring a wide measurement range of 0-93% Brix, one-key zeroing, and automatic temperature compensation (ATC), this device provides reliable and easy-to-read results on its high-definition LCD screen. It is suitable for use in home kitchens, the catering industry, soybean product factories, and scientific research.
2. Safety Information
Please read all instructions carefully before using the device. Keep the device away from extreme temperatures and direct sunlight. Avoid dropping or subjecting the device to strong impacts. Do not attempt to disassemble or repair the device yourself, as this may void the warranty and cause damage.
3. Parts Identification
- Measurement Slot: Where the sample solution is placed.
- Optical Prism: The component that interacts with the sample for measurement.
- Display Screen: LCD screen showing temperature, Brix percentage, and refractive index.
- ZERO Button: Used for zero calibration and navigating settings.
- START Button (Power Button): Used to power on/off the device and initiate measurements.
4. Setup
4.1 Battery Installation
The instrument is powered by two AAA batteries. To install or replace batteries, remove the battery cover located on the back of the device and insert the batteries according to the polarity markings. Ensure the cover is securely reattached.
4.2 Initial Zeroing (Calibration)
Before first use or if measurements are inaccurate, perform a zero calibration using clean water (distilled water is recommended).
- Ensure the measurement slot is clean and dry.
- Place approximately 0.3 ml (about 1 mm depth) of clean water onto the optical prism in the measurement slot.
- Press the START button to take a measurement.
- Check if the displayed value is 0.0%.
- If the value is not 0.0%, press the ZERO button to reset the calibration.
4.3 Language and Refractive Index Settings
To access settings:
- Press and hold the ZERO button for 2 seconds to enter the setting interface.
- Use the ZERO button to select between 'LANG' (Language) and 'RI' (Refractive Index).
- Press the START button to change the selected item's value (e.g., switch language between English and Chinese, or enable/disable refractive index display).
- Press and hold the ZERO button for 2 seconds again to return to the main interface.
5. Operating Instructions
5.1 Measurement Procedure
- Ensure the device has been properly zero-calibrated (see Section 4.2).
- Clean the measurement slot thoroughly and dry it.
- Place approximately 0.3 ml (about 1 mm depth) of your sample solution onto the optical prism in the measurement slot.
- For cold or hot solutions, allow the solution to sit in the slot for about 20 seconds to allow for temperature compensation. The instrument and the solution need to reach the same temperature for accurate readings.
- Press the START button to display the Brix value.
- For best accuracy, it is recommended to take several measurements and average the results.
5.2 Handling Solutions with Suspended Substances
If your sample solution contains suspended solids, follow these steps to ensure accurate measurement:
- Step 1: Remove as many suspended solids as possible from the measuring solution using a filter spoon or gauze.
- Step 2: Ensure the measuring solution is mixed evenly. Avoid using metal or hard objects to mix, and be careful not to scratch the optical prism. Uneven solutions or suspended solids can lead to measurement uncertainty.
5.3 Product Demonstration Video
6. Maintenance
6.1 Cleaning
Regular cleaning of the measurement slot is crucial for accurate readings and device longevity. If there is dirt or oiliness, or if the instrument has not been used for a long time, it should be cleaned as follows:
- Drip clean water (or ethanol/soapy water for oiliness) into the measurement slot.
- Wash with paper towels.
- Repeat this process at least three times, ensuring to throw away the towels after each wipe to prevent cross-contamination.
- Ensure the slot is completely dry before storage or next use.
6.2 Storage
Store the refractometer in a cool, dry place away from direct sunlight and extreme temperatures. Keep it in its original packaging or a protective case to prevent dust accumulation and physical damage.
7. Troubleshooting
Refer to the table below for common issues and their solutions:
| Error Message | Reason | Solution |
|---|---|---|
| Excessive light | External light is too strong to measure. | Move to a less brightly lit area or shield the measurement slot from direct light. |
| High Temp / Low Temp | The temperature in the measuring slot is out of range (10-40°C). | Allow the device and sample to reach room temperature or within the operating range. |
| Re zero | Zero-clearing step was not followed correctly. | Repeat the zero-calibration process (Section 4.2). |
| Less solution | Too little solution in the measuring slot. | Add sufficient sample (approx. 0.3 ml). |
| Less water | Too little clean water in the measuring slot during zeroing. | Add sufficient clean water (approx. 0.3 ml) for zeroing. |
| Out of range | Excessive concentration of the sample. | The sample's Brix value exceeds the 0-93% range. Dilute the sample if necessary. |
| Low voltage | Low battery power. | Replace the two AAA batteries. |
| Fault | Possible damage to components. | Contact customer support for assistance. |
8. Specifications
| Feature | Specification |
|---|---|
| Model Number | SW93T |
| Measurement Range | 0-93% Brix |
| Measurement Accuracy | ±0.2% |
| Brix Min. Div. | 0.1% |
| Temperature Compensation | 10-40°C (Automatic Temperature Compensation - ATC) |
| Min. Div. | 0.1% |
| Size | 117 x 55 x 29 mm |
| Weight | 140g (without batteries) |
| Power Supply | 2 x AAA batteries (not included) |
| Certification | CE, RoHS |
| Language Switching | Chinese and English |
| Detection Functions | Temperature/Brix/Refractive Index |
| Low Battery Alert | Yes |
| Reset Calibration Function | Yes |
9. User Tips
- Always use distilled water for zero calibration to ensure the most accurate baseline.
- For consistent results, ensure the sample and the refractometer are at a stable temperature before measurement, especially for very hot or cold liquids.
- Clean the prism thoroughly after each use to prevent residue buildup, which can affect future readings.
- When measuring viscous liquids, allow extra time for the sample to settle and for temperature equalization.
- If the device shows a 'Low voltage' error, replace the batteries promptly to avoid interruptions during use.
10. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the documentation provided with your purchase or contact the seller directly. Additional resources, including a detailed user manual in PDF format, may be available. You can access the official user manual PDF here.