NJTY T6015

Contents

NJTY T6015 Pointer Insulation Resistance Meter User Manual

Model: T6015

1. Overview

The NJTY T6015 Pointer Insulation Resistance Meter is a professional, handheld insulation resistance tester (megohmmeter) designed for electricians and technicians to verify the integrity of wiring and electrical insulation. It provides reliable performance for a wide range of electrical testing and maintenance applications.

The T60 series pointer-type insulation testers include models such as T6015, T6010, and T6005. These devices feature a modern design, compact size, and portability. They are fully functional, highly accurate, and easy to operate reliably. The T6015 supports 3-range insulation testing and AC/DC voltage testing, the T6005 supports 500V insulation testing, and the T6010 supports 1000V insulation testing. These products are suitable for measuring the insulation resistance of various electrical equipment and insulating materials, such as transformers, motors, cables, switches, and appliances. They are ideal for routine maintenance, testing, and verification of electrical equipment.

Key Features

  • Three Test Voltages: Offers versatile testing with selectable output voltages (100V, 500V, 1000V) to meet various international standards and application requirements.
  • Wide Measurement Range: Accurately measures insulation resistance from 0 to 2000 MΩ, suitable for both low and high-resistance diagnostics.
  • Built-in Safety: Features a low-resistance alarm for each voltage range, a maximum short-circuit current limited to 0.75mA for user protection, and an internal discharge resistor for safer operation.
  • Dual-Functionality: Not only an insulation tester but also a digital multimeter capable of measuring both AC and DC voltages up to 500V.
  • User-Friendly Design: Compact, portable, and powered by a standard rechargeable 18650 lithium-ion battery, which includes a clear low-battery warning (3.3V undervoltage indication).
  • Robust Construction: The unit itself is well-insulated, with the internal circuitry protected by a distance of over 50 meters from the case, ensuring durability and safety.

This tester is an ideal all-in-one diagnostic tool for ensuring electrical safety and preventing equipment failure in industrial, commercial, and residential settings.

Video: Product demonstration of NJTY T6015 Insulation Resistance Meter features and usage.

2. Safety Rules and Precautions

This manual contains instructions and safety terms that users must adhere to in order to ensure safe operation and maintenance of the instrument. Please read these operating instructions carefully before using the instrument. The symbol "⚠" on the instrument indicates that users must follow the relevant safety instructions in this manual. Pay special attention to the safety rules and warning notes in this manual to avoid electric shock and damage to the instrument.

  • Read this manual carefully before use.
  • Do not use the instrument if the battery cover is not securely closed, as this poses an electric shock hazard.
  • Before use, ensure the test leads' insulation is intact and undamaged.
  • To avoid damaging the instrument, perform insulation measurements only on disconnected circuits.
  • Do not press the "Test Button" when measuring voltage.
  • To avoid electric shock, do not touch the test leads or the circuit being tested during measurement.
  • Avoid touching the tested circuit immediately after testing, as residual charge may cause electric shock.

3. Setup

3.1 Panel Diagram

Diagram of the NJTY T6015 Pointer Insulation Resistance Meter front panel, showing all controls and input terminals.

Figure 1: Front Panel Layout of T6015

  1. Charging indicator light.
  2. Work indicator light.
  3. Gear position rotary switch (T6015 model).
  4. EARTH input terminal (connected to the ground terminal of the circuit under test).
  5. LINE input terminal (connected to the circuit under test).
  6. Test button.
  7. Pointer zero adjustment.
  8. Insulation meter reading dial.

3.2 Mechanical Zero Adjustment

Before starting insulation measurement, check whether the pointer is consistent with the "∞" (zero position on the left) mark on the insulation meter scale. If not, use a screwdriver to rotate the "pointer zero adjustment" on the meter to adjust the zero position.

Image showing a hand using a screwdriver to adjust the pointer to zero on the meter's dial.

Figure 2: Adjusting the pointer to 0 Ω before measurement.

3.3 Test Lead Connection

Insert the red and black test leads into the L and E ports of the meter respectively.

Image showing how to connect alligator clips to the test leads.

Figure 3: Connecting alligator clips to test leads.

4. Operation Instructions

4.1 Insulation Resistance Measurement

Starting and Shutting Down: Press and hold the "test button" for 1 second to turn on the meter. The green "working" indicator light on the upper right corner of the dial will light up. After normal startup, it will automatically shut down after 15 minutes. Press the "test button" three times within 5 seconds after powering on, the meter switches to continuous working state, and the two green indicators in the upper right corner light up, and the automatic shutdown function is canceled. You can manually shut down the meter 5 seconds after powering on, and press and hold for 1 second to shut down. When working in the undervoltage state, the working indicator light is red, and it will automatically shut down after 15 minutes.

Reading Measurement: After the insulation measurement is started, the test leads touch the insulator to be measured and read the measured value according to the pointer scale.

Image showing the meter performing an insulation resistance test on a circuit board.

Figure 4: Performing an Insulation Resistance Test.

4.2 AC and DC Voltage Measurement

The T6015 needs to be switched to the corresponding voltage range for measurement, while the T6010 and T6005 can be measured directly. Confirm that no insulation measurement has been started, i.e. the working indicator light is not on. Use the test leads to touch the two ends of the voltage to be measured. Read the measured value according to the pointer scale. When T6015 measures DC voltage, insert the red test lead into "+" and the black test lead into "-".

Image showing the meter measuring AC/DC voltage with test leads connected to a power source.

Figure 5: AC/DC Voltage Measurement.

4.3 Transistor Withstand Voltage Measurement

Connect the transistor to be tested to the meter according to polarity. Press the test button to directly read the BV(V) voltage scale. The read value is the withstand voltage of the transistor (the DC load voltage value L(V) at both ends of the material being tested). The maximum withstand voltage measurement value of T6015 and T6010 insulation meters is 1000V. The maximum withstand voltage measurement value of T6005 insulation meter is 500V.

Image showing the meter performing a transistor withstand voltage test.

Figure 6: Transistor Withstand Voltage Test.

4.4 Low Resistance Alarm

The meter features a low resistance buzzer alarm. When the measured resistance is less than the alarm setting value, the buzzer will sound to alert the user.

Image showing the meter indicating a low resistance alarm with a 'Di~ Di~ Di' sound graphic.

Figure 7: Low Resistance Alarm indication.

4.5 Battery Performance Detection

The meter can detect battery performance. When the working indicator light is red, it means the battery power supply is undervoltage, prompting that charging or replacement of batteries is required.

Image showing the meter detecting battery performance of a 9V battery.

Figure 8: Battery Performance Detection.

5. Charging and Replacing Batteries

  • To prevent electric shock, please remove the test leads from the instrument before charging.
  • When the instrument is working normally, the working indicator light is green, indicating that the battery is sufficient. When the working indicator light is red, it means that the battery power supply is undervoltage, prompting that charging or replacement of batteries is required.
  • Use 5V USB power supply (including power bank and car charger, etc.), and the charging interface of the instrument is Type-C.
  • The charging indicator light turns red during charging and automatically turns green when fully charged.
  • Do not start the test during charging.
  • If the battery fails, please replace it with a battery of the same specification.
Image showing the meter being charged via a USB Type-C cable connected to a wall adapter.

Figure 9: USB Type-C Rechargeable Function.

6. Specifications

ParameterT6015T6005T6010
Rated Test Voltage1000V, 500V, 100V500V1000V
Low-Resistance Alarm<2MΩ (100V), <1MΩ (500V), <200KΩ (1000V)<2MΩ<2MΩ
Transistor Withstand Voltage Test1000V, 500V, 100V500V1000V
Output Short-Circuit CurrentMax 0.75 mAMax 0.65 mAMax 1.3 mA
Measurement Range0-2000MΩ0-1000MΩ0-2000MΩ
Measurement Accuracy1M-2000MΩ ±10%1M-1000MΩ ±10%1M-2000MΩ ±10%
Voltage Measurement RangeDCV500V, 50V; ACV500V (50HZ)ACV500V (50HZ)
Insulation Test Output Voltage90%-120% of Rated Voltage
Discharge Resistor10MΩ
Voltage Measurement Error±5% (full scale)
Battery Detection3.3V undervoltage indication
InsulationBetween the movement circuit and the case: >50 meters

General Specifications

  • Power Supply: 1 × 3.7V 18650 rechargeable lithium battery.
  • Dimensions: 161 × 96 × 45 mm.
  • Weight: Approximately 534 g (including accessories).
  • Operating Temperature: 0°C-40°C, Relative Humidity <75% (non-condensing).
  • Storage Temperature: -10°C-50°C, Relative Humidity <85%.
  • Calibration Temperature: 23°C ±5°C, Relative Humidity <70%.
Image showing the dimensions of the NJTY T6015 meter: 16cm (6.3inch) length, 9.5cm (3.7inch) width, 4cm (1.6inch) height.

Figure 10: Product Dimensions.

7. Maintenance

This instrument is a precision electronic device and should be maintained with care.

  • If the instrument is not used for an extended period, recharge it every two months.
  • Store the instrument in a dry place and avoid humid environments.
  • Avoid strong vibrations or impacts, as they may damage the meter head or circuitry.
  • Do not modify the internal circuitry. If a malfunction occurs, contact us for assistance.

8. Troubleshooting

  • Meter does not turn on: Check the battery level and ensure it is properly charged.
  • Inaccurate readings: Verify the mechanical zero adjustment. Ensure test leads are securely connected and the circuit under test is disconnected from power.
  • Low resistance alarm: If the alarm sounds, the measured resistance is below the set threshold for the selected voltage range. Investigate the circuit for potential insulation breakdown.
  • No display/indicator lights: Check battery charge. If charging indicator is red, charge the device. If green, and still no lights, contact support.

9. User Tips

  • Regular Calibration: For optimal accuracy, consider performing regular calibration checks. The device has shown very good correctness in calibration tests.
  • Serial Number Location: If you need to locate the serial number, check the device's back panel or original packaging.
  • Verify Package Contents: Upon receiving your device, carefully check all items against the package contents list to ensure nothing is missing, such as alligator clip adapters.
  • Accurate Voltage Measurement: The device is noted for its exact voltage measurement capabilities.

10. Package Contents

The following items are included in your package:

  • T60 Series Insulation Resistance Tester × 1
  • Portable Tool Bag × 1
  • Insulation Resistance Tester Test Leads × 1 pair
  • User Manual × 1
  • 18650 Rechargeable Lithium Battery × 1
  • USB Charging Cable × 1
Image showing all items included in the product package: meter, test leads, USB cable, user manual, and carrying bag.

Figure 11: Package Contents.

11. Warranty and Support

This instrument is a precision electronic device. If a malfunction occurs or you require assistance, please contact the seller or manufacturer for support. Do not attempt to modify the internal circuitry yourself.