1. Introduction
The ZOYI ZT-MD2 is a high-precision handheld LCR Tweezer Bridge Tester designed for accurate measurement and identification of various electronic components. It features smart auto-identification capabilities for voltage, resistance, capacitance, and inductance, making it an essential tool for electronics enthusiasts, engineers, and technicians. This manual provides detailed instructions for the safe and efficient use of your ZT-MD2 device.
2. Safety Information
Please read and understand all safety warnings and operating instructions before using this device. Failure to do so may result in injury or damage to the device or components being tested.
- Always ensure the device is in good working condition before use.
- Do not attempt to measure voltages exceeding the specified limits (DC Voltage up to 30V).
- Avoid contact with live circuits when handling the test probes.
- Do not operate the device in wet environments or near flammable gases.
- Keep the device away from strong electromagnetic fields.
- Only use the provided USB cable for charging the built-in lithium battery.
- Do not disassemble or modify the device. Refer all servicing to qualified personnel.
3. Package Contents
Verify that all items are present in your package:
- 1 x ZOYI ZT-MD2 LCR Tweezer Bridge Tester
- 1 x USB Charging Cable
- 1 x Calibration Board
- 1 x Carrying Case
- 1 x User Manual (this document)

4. Product Overview and Features
The ZT-MD2 is a compact and versatile LCR tester with advanced functionalities:
- Smart Auto-Identification: Automatically recognizes and measures resistance, capacitance, inductance, and voltage.
- Multi-Parameter Display: Shows primary and secondary parameters (D, Q, X, Z, P, Rs, Rp, Cs, Cp, Ls, Lp) for comprehensive component analysis.
- Wide Test Frequency Range: Supports 100Hz, 120Hz, 1KHz, 10KHz, and 100KHz test frequencies.
- DC Voltage Measurement: Capable of measuring DC voltage up to 30V, suitable for batteries and circuit boards.
- Signal Generator: Features 7 waveform outputs for various testing scenarios.
- Frequency Scanning: Automatically scans and measures component parameters at different frequencies.
- Data Hold Function: Allows for freezing the displayed measurement.
- OLED Display: High-definition screen with adjustable orientation for optimal viewing.
- Durable Construction: Encased in a durable aluminum alloy housing with gold-plated copper clips for reliable contact.
- Rechargeable Battery: Built-in lithium battery for portable use.
- Comparison Function: Preset a reference value to perform Pass/Fail testing.
- Data Logging: Supports recording data into Excel files for computer analysis.











5. Setup
5.1 Charging the Device
- Connect the provided USB charging cable to the ZT-MD2's charging port.
- Connect the other end of the USB cable to a standard USB power adapter (not included) or a computer's USB port.
- The battery indicator on the OLED display will show charging status. Charge until the battery icon indicates full charge.
5.2 Power On/Off
- To power on, press and hold the power button until the display illuminates.
- To power off, press and hold the power button until the display turns off.
5.3 Calibration
For optimal accuracy, it is recommended to perform a calibration periodically or if you suspect inaccurate readings.
- Ensure the device is powered on and the probes are clean.
- Carefully connect the probes to the provided calibration board according to the instructions on the board or device screen.
- Follow the on-screen prompts to complete the calibration process.
6. Operating Instructions
6.1 Basic Measurement (Auto Mode)
The ZT-MD2 features an intelligent auto-identification mode for quick measurements.
- Power on the device. It will typically start in Auto mode.
- Gently press the gold-plated copper clips onto the component you wish to measure.
- The device will automatically identify the component type (Resistor, Capacitor, Inductor, Diode) and display its primary and secondary parameters on the OLED screen.

6.2 Selecting Test Frequency
The ZT-MD2 supports multiple test frequencies. To change the frequency:
- While in a measurement mode, press the 'Freq' button (or equivalent on-screen control) to cycle through available frequencies (100Hz, 120Hz, 1KHz, 10KHz, 100KHz).
- Select the appropriate frequency based on the component's specifications or application.
6.3 DC Voltage Measurement
To measure DC voltage:
- Switch the device to Voltage measurement mode (if not in auto-identification).
- Carefully touch the probes to the positive and negative terminals of the DC voltage source (e.g., battery, circuit point).
- The voltage reading will be displayed on the screen. Ensure the voltage is below 30V.
6.4 Using the Comparison Function (Pass/Fail)
This feature allows for quick quality control by comparing a component's value against a preset reference.
- Enter the comparison mode (refer to on-screen menu or specific button).
- Set the desired reference value and tolerance percentage.
- Measure the component. The display will indicate "PASS" or "FAIL" based on the comparison.
6.5 Data Hold and Recording
- Data Hold: Press the 'HOLD' button (or equivalent) to freeze the current measurement on the display. Press again to release.
- Data Recording: Connect the ZT-MD2 to a computer using the USB cable. The device supports recording measurement data directly into Excel files for further analysis. Refer to the device's software interface (if applicable) for detailed recording instructions.
7. Maintenance
- Cleaning: Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
- Probe Care: Keep the gold-plated copper clips clean and free from debris to ensure good contact.
- Storage: Store the device in its carrying case in a cool, dry place when not in use. Avoid extreme temperatures and humidity.
- Battery: Recharge the battery regularly, even if the device is not used frequently, to maintain battery health.
8. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Device does not power on | Low battery or faulty power button. | Charge the device fully. If problem persists, contact support. |
| Inaccurate readings | Dirty probes, uncalibrated device, or component outside measurement range. | Clean probes. Perform calibration. Ensure component is within specified range. |
| Display not working correctly | Software glitch or hardware issue. | Try restarting the device. If problem persists, contact support. |
| Cannot connect to computer for data logging | Faulty USB cable, driver issue, or incorrect software settings. | Try a different USB port/cable. Ensure necessary drivers are installed. Check device software settings. |
9. Specifications
| Feature | Details |
|---|---|
| Brand Name | ZOYI |
| Model Number | ZT-MD2 |
| Operating Mode | Auto Range |
| Display Type | Digital Display (OLED) |
| Dimensions | 145*29*14mm |
| Battery Included | Yes (Built-in rechargeable lithium battery) |
| DIY Supplies | Electrical |
| DC Voltage Range | Up to 30V |
| AC Voltage Range | Up to 30V |
| Measuring Resistance Range | 10mΩ~10MΩ |
| Measuring Capacitance Range | 1pF~20mF |
| Measuring Inductance Range | 1uH~10uH |
| Test Frequencies | 100Hz, 120Hz, 1KHz, 10KHz, 100KHz |
| Test Voltage Levels | 0.1V, 0.3V, 0.6V |
| Parameters Measured | Rs/Rp/Cs/Cp/Ls/Lp/P/D/Q/X/Z |
| Diode Test | Yes |
| Origin | Mainland China |

10. User Tips
- For small surface-mount components, ensure the probes make firm and clean contact for accurate readings.
- When measuring components in-circuit, be aware that other components might affect the reading. For most accurate results, test components out-of-circuit.
- Utilize the adjustable screen orientation for comfortable viewing in various working positions.
- Regularly back up your recorded data to prevent loss.
11. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the contact information provided by your retailer or visit the official ZOYI website. Keep your proof of purchase for warranty claims.