Veinasa XS-TBQ02A Total Radiation Sensor

Instruction Manual

1. Introduction

The Veinasa XS-TBQ02A Total Radiation Sensor is a high-precision short-wave total radiation sensor designed for solar radiation observation. It adheres to the latest ISO9060 and WMO technical standards, classifying it as a first-class total radiation sensor.

This sensor measures the solar radiation received by a horizontal plane from a 2π steradian solid angle (hemisphere to hemisphere), expressed in W/m². It operates passively, utilizing a thermopile sensor to generate an output voltage directly proportional to the incident radiation flux. The design incorporates two spherical glass covers to minimize measurement errors, particularly thermal offset, ensuring high accuracy.

The sensor is straightforward to use, requiring only a high-precision voltmeter to read measurement data. Incident irradiance is calculated by dividing the output voltage by the sensor's unique sensitivity, provided at the factory. It is compatible with most standard data acquisition systems and is suitable for meteorological observations, building physics studies, climate research, and solar energy acquisition experiments. A common application involves integrating it into weather stations for outdoor solar radiation measurement.

Veinasa XS-TBQ02A and XS-TBQ02 Total Radiation Sensors
Figure 1: Veinasa XS-TBQ02A (Class A) and XS-TBQ02 (Class B) Total Radiation Sensors.

2. Working Principle

The total radiation sensor operates on the principle of converting solar radiation into heat energy. A black coating on the sensor's surface absorbs incoming solar radiation. This absorbed energy generates a temperature difference across the thermopile within the sensor. This temperature difference, in turn, produces a voltage output signal that is directly proportional to the intensity of the incident solar radiation.

Diagram showing double-layer high-precision glass cover
Figure 2: Double-layer high-precision glass cover for enhanced accuracy.
Diagram illustrating light absorption and reflectivity reduction
Figure 3: Comparison of light absorption rates, highlighting the effectiveness of imported light-absorbing paint in reducing reflectivity for more accurate radiation measurement.

3. Specifications

Parameter Specification
Rated according to ISO 9060/WMO standardNational standard level 1
Response time (95% response)<10s
Zero offset (a) Corresponding 200W/m² net thermal radiation (ventilation)<15 W/m²
Zero offset (b) A response corresponding to a change in ambient temperature of 5K/h<4W/m²
Stability (change/year, percentage of full degree)±1.5%
Nonlinearity±1%(100~1000 W/m²)
Directional response±20 W/m²
Temperature response<4%(-10~40℃)
Tilt response<2%
Sensitivity range7-14μV/(W/m²)
Operating temperature-40~+80℃
Internal resistance<50Ω
Standard cable length3m
Measuring range0~4000 W/m²
Cable replacementThe user can remove and install the cable
Spectral range280~3000 nm (50% transmission point)
Read requirementA differential voltage channel or a single terminal voltage channel
Horizontal calibrationIncludes leveling bubble and adjusting foot
Uncertainty of daily exposure±2%
Weight (excluding cable)0.8 kg
Calibration traceabilityTrace WRR, procedure in accordance with ISO9847
Recommended calibration periodBiennially
Output signalOriginal output 0 ~ 20mV, Optional output RS485/4 ~ 20mA (non-external transmitter)
VoltageOthers (Power supply for XS-TBQ02 is 12-24VDC)
Size101*154 mm
OriginMainland China
Detailed technical parameters table for the sensor
Figure 4: Detailed Technical Parameters.
International standard product reference selection table
Figure 5: International Standard Product Reference Selection Table, comparing different classes of sensors.

4. Setup

  1. Unpacking and Inspection: Carefully unpack the sensor and inspect for any visible damage. Ensure all components, including the sensor and any specified cables, are present.
  2. Mounting Location: Choose a mounting location that provides an unobstructed view of the sky, free from shadows cast by buildings, trees, or other structures throughout the day. The sensor should be mounted horizontally.
  3. Horizontal Calibration: The sensor includes a leveling bubble and adjusting feet for precise horizontal calibration. Use these to ensure the sensor is perfectly level for accurate readings.
  4. Cable Connection: Connect the sensor's cable to your data acquisition system or high-precision voltmeter. The sensor offers original 0-20mV output, with optional RS485 or 4-20mA outputs available (these options may require a non-external transmitter or built-in module). Ensure the correct wiring for your chosen output type.
    Bottom view of the sensor showing signal output options
    Figure 6: Sensor bottom view with signal output options (0~20mV/RS485/4~20mA).
    XS-TBQ02 sensor showing signal output and power supply details
    Figure 7: XS-TBQ02 sensor with signal output (0-20mV) and power supply (12-24VDC) details. Note: Power supply details are indicative for XS-TBQ02 model and may vary for XS-TBQ02A depending on output configuration.
  5. Power Supply: If using an active output (e.g., RS485 or 4-20mA), ensure a stable power supply (typically 12-24VDC, refer to specific model documentation for exact requirements) is connected to the sensor.
  6. Data Acquisition System Configuration: Configure your data acquisition system to read the appropriate signal type (differential voltage, single terminal voltage, RS485, or 4-20mA) and apply the sensor's sensitivity for accurate irradiance calculations.

5. Operating Instructions

  1. Power On: Once the sensor is correctly installed and connected, power on your data acquisition system.
  2. Data Reading: The sensor will continuously measure solar radiation. Read the output voltage from your voltmeter or data acquisition system.
  3. Irradiance Calculation: To obtain the incident irradiance in W/m², divide the measured output voltage by the sensor's specific sensitivity value. This sensitivity is unique to each sensor and is provided with the product.
  4. Continuous Monitoring: The sensor is designed for continuous, all-weather data measurement. Ensure your data logging system is configured for continuous operation if required.
    Multiple sensors installed for all-weather data measurement
    Figure 8: Sensors deployed for accurate all-weather data measurement throughout the day.
Video 1: Product overview demonstrating key features like the built-in RS485 output module and double-layer glass cover.

6. Maintenance

  • Cleaning the Glass Cover: Regularly inspect and clean the double-layer high-precision glass cover to ensure it is free from dust, dirt, snow, or other obstructions that could affect measurement accuracy. Use a soft cloth and clean water or a mild cleaning solution.
  • Cable Inspection: Periodically check the sensor cable for any signs of wear, damage, or corrosion. The cable is designed to be user-replaceable if necessary.
  • Recalibration: A recommended calibration period for the sensor is biennially (every two years) to maintain optimal accuracy and traceability.

7. Troubleshooting

  • No Output/Incorrect Readings:
    • Check all cable connections to ensure they are secure and correctly wired according to the chosen output type (0-20mV, RS485, 4-20mA).
    • Verify that the power supply (if required for active outputs) is correctly connected and providing the specified voltage (e.g., 12-24VDC).
    • Ensure the data acquisition system is configured to read the correct signal type and that the input channels are functioning properly.
    • Inspect the glass cover for any obstructions or heavy soiling that might block solar radiation.
  • Inaccurate Readings:
    • Confirm the sensor is perfectly level using the built-in leveling bubble and adjusting feet.
    • Ensure there are no new obstructions (e.g., growing vegetation, new structures) casting shadows on the sensor.
    • Verify that the correct sensitivity value is being used in your data acquisition software for irradiance calculation.
    • Consider if the biennial calibration period has passed; recalibration may be necessary.

8. User Tips

  • For long-term outdoor deployments, consider installing a small fence or barrier around the sensor to prevent accidental damage from animals or human activity, while ensuring it does not cast shadows.
  • Keep a log of cleaning and calibration dates to ensure consistent data quality over time.
  • When comparing data with other radiometers, ensure all sensors are calibrated to the same standards and are mounted in identical environmental conditions for a fair comparison.

9. Support

For any technical assistance, inquiries, or after-sales service, please contact Veinasa Sensor Store directly. We are committed to providing professional support and ensuring your satisfaction.