1. Introduction
This digital multimeter is designed and manufactured in compliance with IEC-61010 safety requirements on electronic measuring instruments and handheld digital multimeters. It is compliant with IEC-61010 requirements pertaining to 600V CAT II and requirements on pollution degree of 2.
Please read carefully this user Manual and pay attention to the safety guidelines before operating this meter.
The meter is equipped with a backlight to facilitate users reading the measurement display in a dark place.
2. Safety Information
2.1. Safety Instructions
Before operating this meter, the operator must observe all standard safety procedures as below:
- Safety procedures against electric shock
- Safety procedures against unintended use
To ensure your personal safety, please use the test lead that is compatible with the meter. Before operating this meter, ensure that the test lead is flawless.
2.2. Safety Notices
- When the meter is used around some equipment that produces strong electromagnetic interference, the readings on the meter may be unstable and even cause serious errors.
- Don't operate the meter or pen-shaped meter when the appearance is damaged.
- The safety function of the meter will become null if the meter is not properly operated.
- The meter must be operated with great care when working in the vicinity of an exposed conductor or bus line.
- The meter is prohibited from being used in the vicinity of any explosive gas, vapor or dust.
- The measurement must be made with correct input terminals and functions and within the allowable measuring range.
- To prevent the meter from being damaged, the value to be input shall not exceed the extremes permitted by each measuring range.
- When the meter has already been connected to the line being measured, the operator is forbidden to touch the input terminal not in service.
- When the measured voltage is higher than DC60V or AC36V RMS, you should be careful to avoid electric shock.
- When making measurements with a test lead, put your fingers behind its protective ring.
- When switching to another measuring range, be sure that the test lead has already been removed from the measured circuit.
- For all DC functions, to prevent potential electric shock that resulting to incorrect reading, please first use AC functions to check the absence of any AV voltage. Then, select DC voltage measuring range equivalent to or greater than that for AC voltage.
- Before any test on electric resistance, diode, capacitor or continuity, the operator must cut off the power to the circuit to be measured, and discharge all high-voltage capacitors contained in the circuit to be measured.
- The electric resistance measurement or continuity test cannot be carried out in any live electrical circuit.
- Before the current measurement, the operator must first examine the protective tube of the meter. Before connecting the meter to the circuit to be measured, the operator must first power off the aforesaid circuit.
- Before repairing TV sets or measuring power switching circuit, the operator must be careful enough to prevent Voltage impulse with high amplitude from damaging the meter.
2.3. Safety Symbols
| Symbol | Explanation |
|---|---|
| ⚠ | Warning, an important safety symbol. |
| ~ | AC |
| ⎯ | DC |
| ⏚ | Ground |
| ⏚ | Double insulation protection |
| ⏚ | Fuse |
| CE | Compliant with European Union Directive |
| ⚡ | High voltage warning |
| CAT II | Class II 600V over-voltage protection |
2.4. Safe Maintenance Habits
- Pull out the test lead before opening the multimeter case or removing the battery cover.
- Use appropriate replacement parts to repair the multimeter.
- Before turning on the meter, all relevant power must be disconnected.
- The calibration and maintenance of the instrument must be operated by professionals.
- If any abnormality of the instrument is observed, please stop using it immediately and send it for repair, and any use is forbidden until it is qualified.
3. Product Overview
3.1. Multimeter Diagram (T21E Model)

Figure 1: NJTY T21E Digital Multimeter with test leads.
3.2. Panel Indication and Buttons

Figure 2: Multi-Function Back support frame design with icons for features.
- Flashlight
- NCV detection area
- LED display
- Buzzer
- Backlight/Flashlight button
- "FUNC/Hz" function selection button
- Rotary switch
- VΩmA°C input socket
- COM input socket
- 10A input socket
- Hold button
- NCV indicator
3.3. Symbols on the Display Unit
| Symbol | Explanation |
|---|---|
| 🔋 | Indicates that the battery is low (replace battery as soon as possible). |
| ⌛ | Auto power off indicator |
| - | Negative input polarity |
| AC | Input voltage AC |
| DC | Input voltage DC |
| ♫ | Multimeter in beep mode |
| →|← | Multimeter in diode test mode |
| AUTO | Automatic range measurement mode |
| H | Multimeter in data hold mode |
| °C | Unit of temperature (°C: Celsius; °F: Fahrenheit) |
| NCV | Non-contact AC voltage detection mode |
| V, mV | The unit of volt |
| A, mA, μA | The unit of current |
| MkHz | The unit of frequency |
| mF,μF, nF | The unit of capacitance |
3.4. Functional Keys
| Key | Description of functions |
|---|---|
| HOLD | Multimeter in data hold mode |
| 🔆 / 🔦 | Press: backlight on; Long press 2S: backlight and illumination light on. |
| FUNC/Hz | Selects AC/DC in A/mA/V/mV position; selects °C/°F in temperature position; selects Hz or % mode. |
3.5. Input Sockets
| Input socket | Description |
|---|---|
| COM | Common input terminal (connect black test lead). |
| VΩmA♫Hz°C | Input for voltage, resistance, diode, continuity, temperature, frequency, duty cycle (connect red test lead). |
| 10A | 10A positive input terminal (connect red test lead). |
4. Setup
4.1. Battery Installation
This meter uses 3x1.5V AAA batteries. Ensure they are correctly installed into the battery compartment. The 🔋 symbol will be displayed on the screen when the battery power is insufficient, indicating it's time to replace the batteries for accurate measurement.
4.2. Test Lead Connection
Always connect the black test lead to the COM input socket. Connect the red test lead to the appropriate input socket (VΩmA°C or 10A) based on the measurement function you intend to use.
5. Operating Guidelines
5.1. General Operation
Hold Mode
In hold mode, the reading can be maintained on the display unit. Change the measurement function position or press the HOLD key again to exit the hold mode.
- To enter: Press the FUNC/Hz key, and the reading will be held and the symbol "H" will appear on the LCD screen.
- To exit: Press the FUNC/Hz key again to reset the meter for normal measurement.
Backlight & Flashlight
The meter is equipped with backlight and illumination functions for convenient reading in dark situations.

Figure 3: Multimeter with backlight and flashlight activated.
- Short press 🔆 / 🔦 key: Turns on backlight. Short press again to turn off. Automatically turns off after 20 seconds of no operation.
- Long press 🔆 / 🔦 key (2 seconds): Turns on illumination function and backlight simultaneously. Short press again to turn off illumination. Automatically turns off after 30 seconds of no operation.
Auto Power Off
If no operations are made for 15 minutes, the meter will sound to remind the operator to automatically cut off power supply and enter a dormancy state.
5.2. Measurement Guidelines
Video 1: Demonstration of various measurement functions including voltage, resistance, continuity, NCV, and temperature.
AC Voltage and DC Voltage Measurement
To avoid electric shock or damage, do not attempt voltage measurement if the voltage is over 600V (DC/AC).
- DC voltage range: 600mV, 6V, 60V, 600V
- AC voltage range: 600mV, 6V, 60V, 600V
Steps:
- Turn the rotary switch to the "~mV" or "~V" position. Press "FUNC/Hz" to switch between DC and AC voltage function.
- Connect the black test lead to COM and the red test lead to VΩmA°C.
- Measure the voltage of the circuit (in parallel connection).
- Read the measured voltage on the LCD screen. For DC voltage, the display will show the voltage polarity.
Notes:
- Within the measuring range of DC 600mV or AC 6V, if there is no input, the meter may display some information. Short circuit "V - Ω" and "COM" terminals to reset to zero.
- Within AC voltage function, press "FUNC/Hz" to measure the frequency of the AC voltage source.
- The AC voltage measured is True RMS, accurate for sine, square, triangular, and step waves.
Electrical Resistance Measurement
Before measuring resistance, ensure all power sources are cut off and high-voltage capacitors are discharged.
- Measuring ranges: 600Ω, 6kΩ, 60kΩ, 600kΩ, 6MΩ, 60MΩ.
Steps:
- Turn the rotary switch to the "Ω" position. Press "FUNC/Hz" to select "Ω" range.
- Connect test leads to COM (black) and VΩmA°C (red).
- Read the measured resistance value on the LCD screen.
Notes:
- Measured circuit resistance may differ from rated value due to parallel resistance.
- For low resistance, short circuit test leads first, record reading, then subtract from measured value.
- At 60MΩ, readings may take a few seconds to stabilize.
- "OL" indicates over-range or open circuit.
Diode Test
Before diode testing, ensure all power sources are cut off and high-voltage capacitors are discharged.
Steps:
- Turn the rotary switch to the "→|←" position. Press "FUNC/Hz" to select the diode function.
- Connect test leads to COM (black) and VΩmA°C (red).
- Connect black test lead to negative pole and red test lead to positive pole of the diode.
- The meter displays the forward bias value. If polarity is reversed, "OL" is displayed.
A normal diode typically shows a forward voltage drop of 0.5V to 0.8V.
Beep Continuity Test
Before continuity testing, ensure all power sources are cut off and high-voltage capacitors are discharged.
Steps:
- Turn the rotary switch to the "♫" position. Press "FUNC/Hz" to select the continuity function.
- Connect test leads to COM (black) and VΩmA°C (red).
- Measure the resistance of the circuit.
If the measured resistance is no more than 50Ω, the beeper will sound continuously.
Capacitance Measurement
Before capacitance measurement, ensure all power sources are cut off and high-voltage capacitors are fully discharged (use DC voltage position to confirm discharge).
- Measuring ranges: 6nF, 60nF, 600nF, 6µF, 60µF, 600µF, 6mF, 100mF.
Steps:
- Turn the rotary switch to the "→|←" position.
- Connect test leads to COM (black) and VΩmA°C (red).
- Measure the capacitance of the capacitor.
Notes: Large capacitance measurements require a stabilization period. Pay attention to polarity for polarized capacitors.
Frequency Measurement
To avoid electric shock or damage, do not attempt frequency measurement if voltage is over 250V (DC/AC).
Steps:
- Turn the rotary switch to the "Hz%" position.
- Connect test leads to COM (black) and VΩmA°C (red).
- Measure the frequency of the circuit.
- Press "FUNC/Hz" to display frequency and duty ratio.
Current Measurement
Caution: Do not measure current if voltage between open-circuit voltage and ground is over 250V. If the fuse blows, damage or injury may occur. Always check the meter's protective tube before current measurement. Use correct input sockets, function positions, and measuring ranges. When a test lead is plugged into the current input socket, do not connect the other end to any circuit in parallel.
- DC current ranges: 60mA, 600mA, 10A
- AC current ranges: 60mA, 600mA, 10A
Steps:
- Turn the rotary switch to the appropriate current position. Press "FUNC/Hz" to switch DC/AC current function.
- Connect black test lead to COM. Connect red test lead to mA input for current < 600mA; connect to 10A input for current 600mA ~ 10A.
- Connect black test lead to negative pole of load, red test lead to positive pole of load.
- Connect power to circuit and read displayed value. If "OL" is shown, input is over range; switch to a higher range.
- Within AC current function, press "FUNC/Hz" to measure frequency.
NCV Test (Non-Contact Voltage)
Turn the rotary switch to NCV position. Place the top of the meter (NCV detection area) near the conductor. If AC voltage is detected, signal density indicators (high, medium, low) will light up, and the beeper will sound at different frequencies.
Note: Voltage may be present even without indication. Do not rely solely on NCV for voltage presence. Detection can be affected by socket design, insulation thickness, and external interference (e.g., flashlight, motor).
Temperature Measurement
Steps:
- Turn the rotary switch to the "°C/°F" position. The instrument displays the ambient temperature.
- Insert the red plug of the K-type thermocouple into VΩmA°C and the black plug into COM. Read the temperature value after stabilization.
Note: The maximum measuring temperature for the K-type thermocouple is 300°C.
6. Maintenance
This section provides basic information on maintenance, including replacement of protective tubes and batteries. Do not attempt maintenance unless you are experienced and have read calibration and performance test information.
6.1. Battery Replacement
- Turn off the multimeter.
- Unplug all test leads from the sockets.
- Unscrew the screw on the back cover.
- Open the back cover.
- Remove the old battery.
- Replace with new 3x1.5V AAA batteries.
- Close the back cover and tighten the screws.
6.2. Fuse Replacement
- Turn off the multimeter.
- Unplug all test leads from the sockets.
- Remove the rubber cover.
- Unscrew the rear screws with a screwdriver.
- Gently push the fuse back in, then replace it with a new one (usually 0.6A fuse).
6.3. General Care
Keep the multimeter clean and dry. Use a soft, damp cloth with mild detergent for cleaning. Do not use abrasives or solvents. Ensure input sockets are clean to prevent oxidation and inaccurate readings.
7. Troubleshooting
- Unstable Readings: May occur near strong electromagnetic interference. Move away from such equipment.
- "OL" Display: Indicates the measured value is over the selected measuring range or an open circuit. Select a higher range or check connections.
- Low Battery Indicator (🔋): Replace batteries promptly to ensure accurate measurements.
- No Display/Damaged Appearance: Do not operate the meter if it appears damaged. Send for repair.
- Incorrect Readings: Ensure test leads are properly connected to the correct input terminals and the rotary switch is set to the appropriate function and range.
8. Specifications
8.1. General Technical Parameters
- Operating Environment: 600V CAT II, Pollution level: 2, Altitude < 2000 m.
- Working Temperature & Humidity: 0~40°C (<80%RH)
- Storage Temperature & Humidity: -10~60°C (batteries removed when RH < 70%).
- Ambient Temperature for Testing/Calibration: 20°C ±2°C
- Allowable Max Voltage (terminal to ground): AC600V
- Protective Tube: mA position: F600mA/250V; A position: F10A/250V.
- Rotation Rate: Approximately 3 revolutions/S.
- Display Unit: 6000 counts LCD screen, automatic unit display.
- Outrange Indication: "OL" on LCD screen.
- Low Battery Indication: 🔋 appears when battery voltage is below normal.
- Input Polarity Indication: "-" appears automatically.
- Power: 3 x 1.5V AAA batteries (not included).
- Dimensions: 147mm (L) × 71mm (W) × 45mm (H).
- Weight: Approximately 220g (exclusive of batteries).

Figure 4: Product size diagram.
8.2. Precision Indicators
DC Voltage
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 600mV | 0.1mV | ± (0.5% reading + 5 digits) |
| 6V | 1mV | |
| 60V | 10mV | |
| 600V | 100mV |
Input resistance: 10MΩ; Maximum input voltage: DC600V
AC Voltage (True RMS)
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 600mV | 0.1mV | ± (1% reading + 4 digits) |
| 6V | 1mV | |
| 60V | 10mV | |
| 600V | 100mV |
Input resistance: 10MΩ; Maximum input voltage: AC600V; Frequency response: 40Hz - 1KHz TRMS
Frequency
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 9.999Hz | 0.001Hz | ± (1.5% reading + 5 digits) |
| 99.99Hz | 0.01Hz | |
| 999.9Hz | 0.1Hz | |
| 9.999KHz | 0.001KHz | |
| 99.99KHz | 0.01KHz | |
| 999.9KHz | 0.1KHz | |
| 9.999MHz | 0.001MHz |
Input voltage range: 200mV-10V AC RMS; Overload protection: 250V DC/AC
Electrical Resistance
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 600Ω | 0.1Ω | ± (0.8% reading + 5 digits) |
| 6KΩ | 1Ω | |
| 60KΩ | 10Ω | |
| 600KΩ | 100Ω | |
| 6MΩ | 1KΩ | |
| 60MΩ | 10KΩ |
Overload protection: 250V DC/AC; Open-circuit voltage: 2.4V
Diode and Buzzer
| Function | Measuring Range | Resolution | Test Condition |
|---|---|---|---|
| →|← | 0-3V | 0.001V | Forward DC current: approx. 1mA; Open-circuit voltage: approx. 3.2V. Displays approximated diode forward voltage drop. |
| ♫ | 600Ω | 0.1Ω | Buzzer sounds & yellow indicator on if internal resistance < 40Ω; red indicator on if 40-60Ω. |
Overload protection: 250V DC/AC
Capacitor
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 60nF | 0.01nF | ± (4.0% reading + 5 digits) |
| 600nF | 0.1nF | |
| 6µF | 1nF | |
| 60µF | 10nF | |
| 600µF | 100nF | |
| 6mF | 1µF | |
| 100mF | 10µF |
Overload protection: 250V
DC Current
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 60mA | 0.01mA | ± (1.2% reading + 5 digits) |
| 600mA | 0.1mA | ± (3% reading + 5 digits) |
| 10A | 10mA |
Overload protection: mA range with F600mA/250V; 10A range fuse with F10A/250V. For current > 5A, test time < 10 seconds, 1 min rest after measurement.
AC Current
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| 60mA | 0.01mA | ± (1.5% reading + 5 digits) |
| 600mA | 0.1mA | ± (3% reading + 5 digits) |
| 10A | 10mA |
Overload protection: mA range with F600mA/250V; 10A range fuse with F10A/250V. Max input current: mA range: 600mA AC TRMS; 10A range: 10A AC TRMS. For current > 5A, test time < 10 seconds, 1 min rest after measurement. Frequency response: 10Hz-1KHz TRMS.
Temperature
| Measuring Range | Resolution | Accuracy |
|---|---|---|
| °C (-20°C~1000°C) | 1°C | ± (1.0% + 3 digits) |
| °F (-4°F~1932°F) | 1°F | ± (1.0% + 3 digits) |
8.3. T21 Series Model Comparison

Figure 5: Comparison of features across T21 Series models.
9. User Tips
- Always double-check your test lead connections and rotary switch settings before taking any measurement to prevent damage to the meter or the circuit.
- For accurate resistance measurements, especially low values, consider shorting the test leads first to zero out any lead resistance.
- The magnetic back support frame is very convenient for hands-free operation, allowing you to attach the multimeter to metal surfaces while working.
- Utilize the backlight and flashlight features in dimly lit environments to ensure clear readings and safe operation.
- Remember to replace the batteries promptly when the low battery indicator appears to maintain measurement accuracy.
10. Warranty and Support
For detailed warranty information and further support, please refer to the official user manual PDF available at: User manual (PDF).