yieryi PHS-26C

Contents

Yieryi PHS-26C 6-in-1 Water Quality Monitor

Instruction Manual

Brand: Yieryi | Model: PHS-26C

Introduction

The Yieryi PHS-26C is a versatile 6-in-1 benchtop water quality monitor designed for accurate measurement of pH, EC (Electrical Conductivity), TDS (Total Dissolved Solids), CF (Conductivity Factor), ORP (Oxidation-Reduction Potential), and Temperature. This instrument is ideal for a wide range of applications including aquaculture, drinking water analysis, swimming pools, aquariums, food processing, laboratories, and the paper industry.

NOTE: While some PHS-26C models on the market may not include an ORP probe, this specific set is equipped with an ORP probe, allowing full functionality. The ORP cable length is 1.2 meters.

Yieryi PHS-26C 6-in-1 Water Quality Monitor with probes and display

Figure 1: Yieryi PHS-26C 6-in-1 Water Quality Monitor

Key Features

  • Comprehensive Water Quality Testing: Measures pH, EC, TDS, ORP, CF, and Temperature.
  • Large LCD Screen with Dual Display: Shows two values simultaneously for maximum efficiency.
  • High Precision with Automatic Temperature Compensation (ATC): Ensures accurate results across a temperature range of 0℃ to 50℃.
  • Multiple Applications: Suitable for aquaculture, drinking water, laboratories, agriculture, swimming pools, and aquariums.
  • Durable and Easy to Use: Features high-quality probes and a user-friendly operation panel.
  • High-Performance Bracket: Stable, firm, and easy to assemble/disassemble.
Diagram showing the 6-in-1 measurement capabilities of the water quality detector

Figure 2: 6-in-1 Water Quality Detector Capabilities

Images depicting various applications of the water quality monitor including drinking water, aquariums, agriculture, aquaculture, sewage treatment, and laboratory use

Figure 3: Diverse Applications of the PHS-26C Monitor

Close-up of the dual digital display showing PH and EC readings simultaneously

Figure 4: Dual Digital Display

Close-up images of the PH probe, temperature probe, and EC/TDS probe, highlighting their high-precision design

Figure 5: High-Precision Probes

Technical Specifications

ParameterRangeResolutionAccuracy
PH0.00 ~ 14.00 PH0.01 PH±0.01 PH
EC0.00 ~ 19.99 ms/cm0.01 ms/cm±2% F.S
CF0.0 ~ 199.9 CF0.1 CF±2% F.S
TDS10 ~ 19990 ppm10 ppm±2% F.S
ORP-1999 ~ 1999 mV1 mV±1 mV
Temperature0 ~ 100℃0.1℃±0.5℃
  • Display: PH: 4-digital LCD, Temp: 3-digital LCD
  • Automatic Temperature Compensation: 0℃ ~ 50℃
  • Operating Temperature: 0℃ ~ 50℃
  • Calibration: Manual two-points
  • Power Supply: AC220V
  • Plug type: European standard (other plug types may be available depending on region)

Package Contents

The package typically includes the following items:

  • 1 x PHS-26C 6-in-1 Water Quality Monitor
  • 1 x PH Probe
  • 1 x ORP Probe
  • 1 x EC Probe
  • 1 x Temperature Probe
  • 1 x Screwdriver (for calibration adjustments)
  • 1 x Power Plug (European standard, specific to region)
Image showing all components included in the PHS-26C package: the main unit, various probes, power adapter, screwdriver, and manual

Figure 6: Package Contents

Front Panel Description

Familiarize yourself with the controls and connections on the device:

Diagram of the PHS-26C front and back panels with numbered labels for each component

Figure 7: Front and Back Panel Layout

  1. Screen display
  2. ORP function key (mV)
  3. PH function key (pH)
  4. EC/CF/TDS function key (EC)
  5. Temperature function key (°C)
  6. EC/CF/TDS calibration knob (CAL)
  7. PH4 calibration knob
  8. PH7 calibration knob
  9. DC or RB switch (Power)
  10. AC input terminal
  11. EC electrode input terminal
  12. PH/ORP electrode input terminal
  13. Temp electrode input terminal

Setup

  1. Connect the AC power supply to the instrument, or press the "DC/RB" switch to turn on the self-locking power.
  2. Connect the appropriate electrodes (PH, ORP, EC, Temperature) to their respective jacks on the back of the instrument.
  3. Remove the protective caps from the electrodes before use.

Operation

  1. Immerse the electrodes into the solution to be tested.
  2. Stir the electrodes gently and wait for the reading to stabilize on the display.
  3. Select the desired measurement mode (pH, ORP, Temperature, EC, CF, or TDS) using the corresponding function keys (PH, mV, °C, or MODE button).
  4. Read the displayed value once it has stabilized.

Calibration

For accurate measurements, regular calibration is essential. The instrument features manual two-point calibration.

Calibration Video Guide

This video demonstrates the calibration process for EC, pH, and ORP functions.

EC Calibration

  1. Connect the power or press the "DC/RB" button.
  2. Press the "MODE" key to select the "EC" mode.
  3. Insert the EC electrode into the 12.88 ms/cm standard solution and gently shake it.
  4. Use the small screwdriver to adjust the CAL correction button until the displayed value reaches 12.88.
  5. Wash the electrodes with distilled water and dry them with filter paper.
  6. Insert the instrument into the 1.43 ms/cm standard solution. The displayed value should be within the allowable error range when compared with the value of the standard solution.
  7. Wash the electrodes with distilled water and dry them with filter paper.

PH Calibration

  1. Turn on the power switch and press the "PH" key. The instrument will be in PH measurement mode.
  2. Insert the PH electrode into the standard buffer solution with a pH of 6.86 and gently shake the electrode.
  3. After the displayed value stabilizes, use a small screwdriver to adjust the potentiometer on the left until the displayed value reaches 6.86.
  4. Wash the electrodes with distilled water and dry them with filter paper.
  5. Insert the electrode into the standard buffer solution with a pH of 4.01 and gently shake the electrode.
  6. After the displayed value stabilizes, use a small screwdriver to adjust the potentiometer on the left until the displayed value reaches 4.01.
  7. Wash the electrodes with distilled water and dry them with filter paper.
  8. Insert the electrode into the standard buffer solution with a pH of 9.18 and gently shake the electrode. Wait until the display value stabilizes. The display value should be within the allowable error range compared to the pH value of the buffer solution.
  9. Wash the electrodes with distilled water and dry them with filter paper.

ORP Calibration

  1. Place the ORP electrode in the ORP standard solution.
  2. Wait until the display value stabilizes. The display value should be within the allowable error range compared to the pH value of the buffer solution.
  3. Wash the electrodes with distilled water and dry them with filter paper.

Important Note for Calibration:

  • For the greatest accuracy, frequent calibration of the instrument is recommended.
  • The instrument must be recalibrated whenever: the pH electrode is replaced, after testing aggressive chemicals, where extreme accuracy is required, or at least once a month.
  • For a particularly accurate calibration, it is advised to use two beakers for each buffer solution: one for rinsing the electrode and one for the actual calibration. This minimizes the risk of contaminating the buffer solution.

Maintenance

  • After each use, turn off the meter.
  • Rinse the electrodes thoroughly with tap water or distilled water to minimize contamination.
  • Gently dry the electrodes with filter paper.
  • Always replace the protective cap on the electrodes after use to prevent damage and keep them moist.
  • When operating, stir the electrode gently and ensure no air bubbles (blebs) are on the electrode surface.
  • If abnormal phenomena occur, such as high measured deviation, blinking measured value, or no zero resetting, immerse the electrode in alcohol for cleaning.

Troubleshooting

If you encounter issues with your PHS-26C monitor, please review the calibration steps carefully. Ensure all probes are correctly connected and clean. If problems persist after following the calibration and maintenance guidelines, refer to the warranty information for support.

User Tips

  • Upon receiving your product, carefully inspect the packaging and contents to ensure everything is intact and undamaged.
  • For long-term use, consider inquiring with the seller about the availability of replacement probes, as electrodes can wear out or become defective over time.

Warranty and Support

Warranty

These instruments are warranted from all defects in material and manufacturing for a period of one year from the date of purchase. If during this period, the repair or the replacement of parts is required where the damage is not due to negligence or erroneous operation by the user, please return the parts to either the dealer or our offices, and the repair will be effected free of charge.

Additional Resources

For a detailed user manual and further documentation, please refer to the official PDF document: User manual (PDF)

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.