ASAIR AMC2000

Contents

ASAIR AMC2000 Mass Flow Controller User Manual

Model: AMC2000

1. Product Overview

The ASAIR AMC2000 Mass Flow Controller (MFC) is an advanced device designed for precise measurement and control of gas flow. It incorporates a self-developed MEMS mass flow chip, ensuring high accuracy, stability, and fast response times. This controller is factory-calibrated to guarantee its specified range, accuracy, and repeatability.

ASAIR AMC2000 Mass Flow Controller front view
Figure 1: Front view of the ASAIR AMC2000 Mass Flow Controller.

2. Key Features

  • Precise measurement and control of gas flow.
  • Utilizes self-developed MEMS mass flow chip for high accuracy and stability.
  • Excellent linearity and fast response time (≤2s).
  • High and low temperature resistance.
  • Measures flow and temperature simultaneously.
  • Supports standard Modbus RTU communication.
  • Good repeatability (0.5% F.S.).

3. Applications

The AMC2000 is suitable for monitoring and controlling the mass flow of various dry, clean, and non-corrosive gases, including air, nitrogen, argon, carbon dioxide, and oxygen (excluding flammable and explosive gases). Its versatility makes it ideal for a wide range of applications:

  • University scientific research
  • Fire protection systems
  • Environmental monitoring
  • Tobacco industry
  • Smart agriculture
  • Food processing
  • Medical applications
  • Other industrial uses
ASAIR AMC2000 Mass Flow Controller infographic showing features, applications, and specifications
Figure 2: Detailed infographic of AMC2000 features, applications, and specifications.

4. Specifications

The following table details the technical specifications for the AMC2000 Mass Flow Controller:

ParameterDescription
ModelAMC2000 (various ranges: R00-R04)
Measurement Range0~100mL/min (standard, other ranges up to 5L/min available)
Accuracy±3% F.S.
Repeatability0.5% F.S.
Response Time≤2s
Power Supply24V DC
OutputRS485 / Analog voltage
Working Pressure0.15~0.35 MPa
Calibrated GasAir (standard pressure)
Communication InterfaceDB9 male connector
Power Consumption≤3W
Connector2-G 1/4
Weight269g

Dimensions: The device has a width of approximately 70mm and a height of approximately 89mm. The gas inlet/outlet connectors are 2-G 1/4 threads, and the electrical connection is via a DB9 male connector.

ASAIR AMC2000 Mass Flow Controller showing DB9 connector and gas inlet
Figure 3: Close-up of the DB9 communication port and gas inlet.

5. Setup

  1. Mounting: The AMC2000 is designed for screw mounting. Secure the device firmly in the desired location, ensuring it is stable and free from excessive vibration.
  2. Gas Connections: Connect your gas lines to the 2-G 1/4 threaded inlet and outlet ports. Ensure all connections are leak-free and can withstand the specified working pressure of 0.15~0.35 MPa. The device is calibrated for air (standard pressure) by default.
  3. Power Supply: Connect a stable 24V DC power supply to the appropriate pins on the DB9 connector. Refer to the wiring diagram (if provided separately) for exact pin assignments.
  4. Communication/Output: Connect your control system or data acquisition device to the DB9 male connector for RS485 communication or analog voltage output. Ensure correct wiring for your chosen communication mode.

Important: Always ensure the gas being measured is dry, clean, and non-corrosive. Do not use with flammable or explosive gases unless explicitly stated by the manufacturer for a specific model variant.

6. Operating Instructions

  1. Power On: Once all connections are secure, apply 24V DC power. The device's display should illuminate, showing the current flow rate or other relevant parameters.
  2. Monitoring: The AMC2000 can display real-time flow and temperature measurements. Use the buttons on the device (if present, as indicated in Figure 4) to navigate display modes or adjust basic settings.
  3. Control: For precise flow control, integrate the AMC2000 with your control system via the RS485 (Modbus RTU) or analog voltage output. Refer to the communication protocol documentation for detailed commands and data formats.
  4. Gas Flow Adjustment: Adjust the upstream pressure or a connected valve to achieve the desired flow rate. The AMC2000 will provide feedback on the actual flow.
ASAIR AMC2000 Mass Flow Controller top view showing display and control buttons
Figure 4: Top view of the AMC2000, highlighting the display and control interface.

7. Maintenance

  • Regular Inspection: Periodically check all gas and electrical connections for tightness and signs of wear or damage.
  • Cleaning: Keep the exterior of the device clean and free from dust and debris. Use a soft, dry cloth for cleaning. Do not use abrasive cleaners or solvents.
  • Gas Quality: Ensure the gas supplied to the MFC is consistently clean and dry to prevent sensor contamination and maintain accuracy.
  • Recalibration: For critical applications, periodic recalibration by a qualified service center is recommended to ensure continued accuracy.

8. Troubleshooting

  • No Power/Display: Check the 24V DC power supply and ensure all power connections are correct and secure.
  • Inaccurate Readings: Verify that the gas type matches the calibration, ensure the gas is clean and dry, and check for any leaks in the gas lines. Confirm working pressure is within the specified range.
  • No Communication: Check the RS485 wiring and ensure the communication parameters (baud rate, parity, stop bits) are correctly configured in your control system.
  • No Flow Control: Verify that the control signal (RS485 or analog voltage) is being sent correctly and that the valve connected to the MFC is functioning properly.

If issues persist after performing these checks, please contact technical support.

9. User Tips

  • For optimal performance, allow the device to stabilize for a few minutes after power-up before taking critical measurements.
  • When integrating with a control system, it is advisable to test communication and control functions with a known, stable flow rate first.
  • Always ensure proper ventilation when working with gases, even non-flammable ones.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact your supplier or the manufacturer directly. Ensure you have your product model (AMC2000) and any relevant purchase details available when seeking support.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.