East Tester ET5407A+

Contents

ET5407A+ Programmable DC Electronic Load User Manual

Model: ET5407A+ | Brand: East Tester

1. Introduction

The East Tester ET5407A+ is a high-precision programmable DC electronic load designed for comprehensive testing of various power sources and energy storage devices. It offers 1mV/10mV and 1mA/10mA resolution and precision, making it suitable for applications such as charger testing, switching power supply evaluation, linear power supply analysis, and battery capacity discharge and aging tests. This manual provides detailed instructions for the safe and effective operation of your device.

Front view of ET5407A+ and ET5406A+ programmable DC electronic loads

Figure 1.1: ET5407A+ and ET5406A+ Programmable DC Electronic Loads

Key Features:

  • High resolution and precision: 1mV/10mV, 1mA/10mA.
  • Multiple basic measurement modes: Constant Current (CC), Constant Voltage (CV), Constant Resistance (CR), and Constant Power (CP).
  • Professional battery testing functions.
  • Dynamic test mode for power supply output performance.
  • Scan test mode for power output continuity within a range.
  • List mode to simulate various load state changes.
  • Short circuit test function.
  • 2.8-inch LCD segment code display with adjustable backlight.
  • User-friendly interface for simple and convenient operation.
  • Key lock function to prevent accidental operation during remote connection.
  • Built-in buzzer alarm.
  • Data storage function maintained after power-off.
  • Standard USB interface for remote operation (RS-232 and RS-485 optional).
  • Multiple safety protections: overcurrent, overvoltage, overpower, over-temperature, and input polarity reverse prompt.
  • Intelligent fan speed control for reduced noise.

2. Safety Information

WARNING: Please read all safety information carefully before operating the device to prevent injury or damage to the instrument.

  • Ensure a positive voltage is applied between the red (positive) and black (negative) terminals. Incorrect polarity will trigger a reverse connection notification and may damage the instrument or the tested power source.
  • Do not apply voltage to the input terminals until the instrument is powered on. The instrument performs a 2-millisecond short-circuit self-check upon power-on.
  • Always ensure the load and power source are properly connected before activating the load channel.
  • The device is equipped with overvoltage, overcurrent, overpower, and over-temperature protection. However, always operate within specified limits.
  • The maximum operating temperature is 0°C to 40°C, and storage temperature is -10°C to 70°C. Relative humidity should be less than 80%.
  • Only qualified personnel should perform maintenance or repairs.

3. Product Overview

3.1 Front Panel

Front panel of ET5407A+ with labeled controls and display

Figure 3.1: Front Panel Layout and Controls

The front panel features an LCD segment code screen for displaying numerical values and operational status. Key controls include:

  • LCD Segment Code Screen: Displays Vcc, A, W, and various status indicators (CC, CV, IR, VR, DYNA, LIST, SCAN, BATT, SHORT, OC, OV, OP, OT, LOCK, ON/OFF).
  • Mode Keys (CC, CV, I_R, V_R): Shortcut buttons for Constant Current, Constant Voltage, Current Range, and Voltage Range respectively.
  • Adjustment Knob: Used for adjusting parameter values.
  • ESC Key: Return key, exits editing or system setting states.
  • Arrow Keys (Up, Down, Left, Right): Navigation and increment/decrement.
  • ENT Key: Enter key, confirms settings.
  • MODE Key: Selects operation mode.
  • MENU Key: Accesses system settings.
  • SET Key: Parameter setting key.
  • CH Key: Channel switching (not activated for single channel models).
  • TRIG Key: Trigger key for dynamic and list modes.
  • ON Key: Activates/deactivates the load channel.
  • Positive Terminal Interface (Red): For positive connection.
  • Negative Terminal Interface (Black): For negative connection.

3.2 Rear Panel

Rear panel of ET5407A+ with labeled ports

Figure 3.2: Rear Panel Layout

  • Factory Serial Number: Unique identifier for the unit.
  • Power Outlet: AC power input.
  • Fuse Box: Contains the protective fuse.
  • USB Host Interface: Standard USB Type-B port for remote control.
  • Pin Definition / External Trigger Port: 6P 3.81mm pluggable terminal for R1 (Run1), P1 (Pass1), N1 (External Trigger), A/T (RS485 A / RS232 TXD), B/R (RS485 B / RS232 RXD), GND (Ground).
  • Cooling Port: Ventilation for heat dissipation.

4. Setup and Initial Operation

4.1 Unpacking and Inspection

Upon receiving your ET5407A+ electronic load, carefully unpack the contents and inspect for any signs of damage. The package should include:

  • 1 x ET5407A+ DC Electronic Load unit
  • 1 x Power Cord
  • 1 x User Manual (PDF version available here)

If any items are missing or damaged, contact your supplier immediately.

4.2 Power Connection

  1. Connect the provided power cord to the power outlet on the rear panel of the ET5407A+.
  2. Plug the other end of the power cord into a suitable AC power source (100Vac±10%~240Vac±10%, 50/60Hz).
  3. Ensure the device is placed on a stable surface with adequate ventilation around the cooling ports.

4.3 Connecting to a Device Under Test (DUT)

To ensure stable and safe operation, properly connect the positive (red) and negative (black) terminals of the ET5407A+ to the power source or device under test (DUT).

  • Connect the positive output of your DUT to the red positive terminal interface on the ET5407A+.
  • Connect the negative output of your DUT to the black negative terminal interface on the ET5407A+.
  • IMPORTANT: Ensure there is a positive voltage between the red and black terminals. Any negative voltage input will trigger a reverse connection notification and may cause damage.
  • Do not apply voltage to the input terminals until the ET5407A+ is powered on and has completed its self-check.

4.4 Power On

Press the red "Power" button on the front panel to turn on the device. The LCD screen will illuminate, and the device will perform a brief self-check.

5. Operating Instructions

5.1 Quick Start

The ET5407A+ features a user-friendly interface. The LCD screen displays voltage (Vcc), current (A), and power (W). Use the adjustment knob and arrow keys for navigation and parameter adjustment, and the ENT key to confirm selections.

Detailed front panel with button labels for quick start

Figure 5.1: Front Panel Controls for Quick Start

To activate the load channel, press the ON button. The "ON" indicator will appear on the display. To deactivate, press ON again.

5.2 Mode Selection

The ET5407A+ supports nine operation modes. Press the MODE button to enter the mode selection interface. Use the adjustment knob or arrow keys to select a mode, then press ENT to confirm or ESC to exit.

Available modes include:

  • CC (Constant Current): Discharges at a constant current.
  • CV (Constant Voltage): Maintains a constant input voltage.
  • CP (Constant Power): Maintains a constant power consumption.
  • CR (Constant Resistance): Maintains a constant resistance.
  • DYNA (Dynamic Mode): Switches between two set currents or voltages.
  • LIST (List Test): Executes a sequence of load conditions.
  • SCAN (Scan Mode): Examines continuous operation within a range.
  • BATT (Battery Mode): Performs professional battery discharge tests.
  • SHORT (Short Circuit Mode): Simulates a short circuit at the input.

Shortcut buttons CC and CV allow quick switching to these modes when the channel is inactive.

5.3 Range Switching

Both current and voltage have high and low ranges. In low range mode, current and voltage are limited (e.g., up to 3A and 20V). For higher values, switch to high range mode.

  • Press I_R to switch between high and low current ranges.
  • Press V_R to switch between high and low voltage ranges.

When switching ranges, associated limits (voltage/current) will reset to maximum values for the new range. Switching from high to low range will cap values exceeding the low range maximum.

5.4 System Settings

When the channel is inactive, press the MENU button to access system settings. Use arrow keys and ENT/ESC for navigation and confirmation.

  • Saving Parameters (SavE PArA Set): Save current parameter configurations (up to 1600 entries).
  • Loading Parameters (LoAd PArA Set): Load previously saved configurations.
  • Voltage/Current/Power Limits: Define protection thresholds. If exceeded, the device will trigger protection (OV, OC, OP) and lock buttons.
  • Trigger Source Selection (trIG): Choose between External (E tEr), Manual (MAnUA via TRIG button), or Software (SoFt via serial command) triggers.
  • Starting Voltage (v on) & Turn-off Voltage (v oFF): Define voltage thresholds for channel activation and deactivation.
  • Start Settings (StArt): Configure device behavior on startup (dEFAU for factory defaults, LASt for previous settings).
  • Buzzer (bEEP): Enable or disable the audible alarm.
  • Factory Reset (FACto): Resets all settings to factory defaults.
  • Backlight Brightness (LIGHt): Adjust LCD backlight from 0 (off) to 10 (max).
  • RS485 Address (485Ad): Set address for RS485 communication (0-31).
  • Baud Rate (bUAd): Configure communication speed (e.g., 9600).
  • Communication Mode (CoMMU): Select USB (standard), RS232 (optional), or RS485 (optional).
  • Qualification Test (qUALI): Enable/disable continuous monitoring against set limits.
  • System Information (InFo): View software version and serial number.

5.5 Detailed Mode Operations

5.5.1 Constant Current (CC) Mode

In CC mode, the electronic load draws a constant current regardless of input voltage changes. The main display shows measured voltage, current, and power. Press ENT to edit parameters like current setting and turn-off delay value.

5.5.2 Constant Voltage (CV) Mode

In CV mode, the load draws current to maintain the input voltage at a set value. Similar to CC mode, press SET to edit parameters like voltage setting and turn-off delay.

5.5.3 Constant Power (CP) Mode

In CP mode, the load consumes constant power, adjusting current as input voltage changes. Use SET to edit power setting and other parameters.

5.5.4 Constant Resistance (CR) Mode

In CR mode, the load maintains a constant resistance. The current adjusts with input voltage changes. Use SET to edit resistance setting and other parameters.

5.5.5 Dynamic Test (DYNA) Mode

This mode allows repeated switching between two set currents or voltages to test power source dynamic characteristics. Parameters include Dynamic Mode (Continuous, Pulse, Trigger), Load State (CC or CV), Value A, Value B, and Pulse Width Times. Parameters can only be edited when the channel is deactivated.

Video 5.1: Product demonstration showing various modes and functions.

5.5.6 LIST Test Operation

The LIST mode allows pre-configuring a sequence of load conditions for automated testing. Parameters include Test Steps (1-10), Step Mode (Continuous/Trigger), Looping (Enabled/Disabled), Modes (CC, CV, CP, CR, Open Circuit, Short Circuit), Setting Values, Delay, Comparison (OFF, Current, Voltage, Power, Resistance), and Upper/Lower Limits. Test results (PASS, FAIL, NONE) are displayed upon completion.

5.5.7 SCAN Test Operation

SCAN mode examines continuous operation within a specified range to capture critical parameters. Configurable parameters include Scan Type (CC, CV, CP), Start Value, End Value, Step Size, Step Delay, Threshold Types (Minimum Voltage, Voltage Transition, Drop), and Comparison Types (Current, Voltage, Power, None). Up to 255 data points can be recorded.

5.5.8 Battery Test (BATT) Mode

This mode is used to test battery discharge performance. It operates in CC or CR mode with cut-off conditions based on voltage (voLt), time (tIME), capacity (AH), or energy (WH). The device measures and displays discharge capacity and energy. Parameters include Discharge Modes (CC, CR), Cut-off Conditions, Cut-off Enable (1-3 settings), Discharge Setting Value, Cut-off Voltage, Cut-off Time, Cut-off Capacity, and Cut-off Energy.

Example: For CC discharge with voltage cut-off, you can set multiple discharge current and cut-off voltage configurations. For instance, discharging a 5-cell lithium battery (3000mAh) might involve setting a 3A discharge current with a 15V cut-off voltage. The instrument will automatically deactivate when the input voltage falls below the set cut-off voltage.

5.5.9 Short Circuit Test (SHORT) Mode

This mode simulates a short circuit at the input terminals. The actual short circuit current depends on the power supply's maximum output and should not exceed the device's current limit to avoid overcurrent protection. Basic operations are similar to CC mode.

6. Maintenance

6.1 General Care

  • Keep the instrument clean and dry. Avoid exposure to dust, moisture, or corrosive environments.
  • Do not block the cooling ports on the rear panel to ensure proper ventilation and prevent overheating.
  • Store the device within the specified temperature range (-10°C to 70°C) and relative humidity (< 80%).

6.2 Cleaning

Use a soft, dry cloth to clean the exterior of the instrument. For stubborn dirt, a slightly damp cloth with mild detergent can be used, ensuring no liquid enters the device. Do not use abrasive cleaners or solvents.

6.3 Fuse Replacement

If the device does not power on, check the fuse located in the fuse box on the rear panel. Replace it with a fuse of the same type and rating (Power fuse * 2 are included as standard accessories).

7. Troubleshooting

ProblemPossible CauseSolution
Device does not power on.No power supply; faulty power cord; blown fuse.Check power connection; replace power cord; replace fuse in the fuse box.
"OV", "OC", or "OP" indicator flashes, and buttons are locked.Overvoltage, overcurrent, or overpower protection triggered.Press ENT to release. Check input parameters and adjust limits in system settings.
"ON" indicator flashes, but no load operation.Open circuit state; input voltage below starting voltage.Ensure proper connection to DUT; check input voltage and adjust starting voltage setting.
"LOCK" icon displayed, buttons unresponsive.Keyboard locked (e.g., during remote operation or protection event).Press ENT briefly to unlock the keyboard.
"RV" (Reverse Polarity) prompt.Input polarity is reversed.Immediately disconnect power and correct the positive and negative connections.
Fan noise is high.High load or ambient temperature.The intelligent fan control adjusts speed based on temperature. Ensure adequate ventilation.

8. Specifications

The following table details the technical specifications for the ET5407A+ model:

Table of specifications for ET5406A+ and ET5407A+

Figure 8.1: Detailed Technical Specifications (ET5407A+ column is relevant)

ParameterET5407A+
Power200W
Rated Input Voltage0-180V
Rated Input Current0-30A
CV Mode Range0.1~19.999V, 0.1~180.00V
CV Mode Resolution1mV, 10mV
CV Mode Accuracy±(0.05%+0.02%FS)
CC Mode Range0~3.000A, 0~30.00A
CC Mode Resolution1mA, 10mA
CC Mode Accuracy±(0.05%+0.05%FS)
CR Mode Range0.05Ω~4.50kΩ
CR Mode Resolution10mΩ, 1Ω
CR Mode Accuracy±(0.1%+0.5%FS)
CP Mode Range0~200W
CP Mode Resolution10mW
CP Mode Accuracy±(0.1%+0.5%FS)
Tran Test ModeCC, CV
Tran Test T1&T20.05s~99.999s
Battery Test Discharge ModeCC, CR
Battery Test Max Discharge Capacity9999Ah
Overvoltage Protection (OV)>185V over voltage protection
Overcurrent Protection (OC)>31A over current protection
Overpower Protection (OP)210W
Over-temperature Protection (OT)85°C
Power Supply Voltage100Vac±10%~240Vac±10%, 50/60Hz
DisplayLCD screen with adjustable backlight
Operating Temperature0°C to 40°C
Storage Temperature-10°C to 70°C
Relative Humidity< 80%
InterfaceStandard USB, optional RS232 or 485
Dimensions (W×H×D)75mm × 155mm × 240mm

9. User Tips

  • For precise battery capacity testing, ensure the battery is fully charged before starting the discharge test in BATT mode.
  • When performing dynamic tests, carefully configure Value A, Value B, and Pulse Width Times to accurately simulate real-world load changes.
  • Utilize the parameter saving function to store frequently used test configurations, saving time and ensuring consistency for repetitive tasks.
  • If using remote control via USB, ensure the correct CH340 driver is installed on your computer.
  • Always double-check the voltage and current ranges before starting a test, especially when dealing with high power applications, to prevent accidental damage.

10. Support and Contact Information

For technical support, service, or inquiries, please contact East Tester:

  • Manufacturer: Hangzhou Zhongchuang Electron Co., Ltd / Zhejiang EastTester Measurement&Control Technology Co.,Ltd
  • Address: No 3, 4th floor, No 58 Yunhe Road, Leidian Town, Deqing Country, Huzhou, Zhejiang, China
  • Postcode: 313219
  • Telephone: 0086-0572-8832711 or 400-0000-332
  • Website: www.east-tester.com

Please have your model number (ET5407A+) and serial number ready when contacting support.