RCYAGO YY-1070

YAGO YY-1070 7-in-1 Water Quality Tester Instruction Manual

Model: YY-1070 | Brand: RCYAGO

1. Introduction

The YAGO YY-1070 is a versatile 7-in-1 multi-functional water quality tester designed for comprehensive analysis of various water sources. It accurately measures Free Available Chlorine (FAC), pH, Salinity, Oxidation-Reduction Potential (ORP), Total Dissolved Solids (TDS), Electrical Conductivity (EC), and Temperature. This device is ideal for monitoring water quality in swimming pools, hot tubs, spas, aquariums, hydroponic systems, drinking water, and more, ensuring optimal conditions and safety.

YAGO YY-1070 7-in-1 Water Quality Tester showing all functions
Figure 1: YAGO YY-1070 7-in-1 Water Quality Tester highlighting its multiple measurement capabilities.

2. Product Overview

Key Features:

What's in the Box:

Contents of the YAGO YY-1070 package including tester, calibration solutions, and manual
Figure 2: Package contents of the YAGO YY-1070 tester.

3. Setup

3.1 Battery Installation

  1. Unscrew the battery compartment cover located at the top of the device.
  2. Insert 4*1.5V LR44 button batteries, ensuring correct polarity.
  3. Securely screw the battery compartment cover back on.
Diagram showing the YAGO YY-1070 with labeled parts including battery compartment
Figure 3: Labeled components of the YY-1070, including the battery compartment.

3.2 Initial Check

  1. Press the 'ON/OFF' button to power on the device.
  2. The LCD display will light up, showing various measurement parameters.
  3. Press the 'MODE/CAL' button repeatedly to cycle through the different measurement modes (pH, TDS, Salt, EC, ORP, FAC, Temp) to familiarize yourself with the display.
  4. Remove the protective cap from the probe before use.
Video 1: Product demonstration showing power-on and mode switching.

4. Operating Instructions

4.1 Basic Measurement

  1. Ensure the device is powered on and the protective cap is removed from the probe.
  2. Immerse the probe into the water sample you wish to test. Ensure the probe is fully submerged.
  3. Press the 'MODE/CAL' button to select the desired measurement parameter (e.g., pH, FAC, Salt).
  4. Wait for the reading on the LCD display to stabilize. The reading will show the measured value and temperature.
  5. For accurate FAC readings, ensure the pool water pH is between 6.5-8 and ORP is between 487-840mV.
YAGO YY-1070 measuring water in a glass, with applications for pools, hot springs, and spas
Figure 4: The YY-1070 in use for water quality measurement.

4.2 Data Hold Function

4.3 Temperature Unit Conversion

Diagram showing short press for data hold and long press for temperature unit conversion on the YY-1070
Figure 5: Functions of the HOLD/TEMP button.

5. Calibration

The YAGO YY-1070 is factory-calibrated for all functions. It is recommended to recalibrate the meter every 2-4 weeks to maintain optimal accuracy, or if you suspect inaccurate readings.

5.1 Calibration Solutions

Use standard buffer solutions for calibration:

Calibration buffer solutions for pH and ORP
Figure 6: Calibration buffer solutions for pH and ORP.

5.2 Calibration Procedure

For detailed calibration steps for each parameter (pH, ORP, EC, TDS, Salinity), please refer to the specific instructions provided in the physical instruction manual included with your product. The general process involves:

  1. Prepare the appropriate calibration solution.
  2. Immerse the probe into the calibration solution.
  3. Enter calibration mode by long-pressing the 'MODE/CAL' button (or as specified in the manual).
  4. Adjust the reading to match the known value of the calibration solution.
  5. Confirm calibration and exit the mode.

6. Specifications

ParameterRangeResolutionAccuracy
pH0.1-14.0 pH0.1 pH±0.1 pH
FAC (Free Available Chlorine)0.1-4.0 ppm0.1 ppm±1 ppm
Salinity0-10000 ppm, 10.1-200.0 ppt, 0.01%-25.00%1 ppm, 10 ppm±2% F.S.
EC (Electrical Conductivity)0-10000 uS/cm, 10.01-19.99 mS/cm, 20.1-400.0 mS/cm1 uS/cm, 10 uS/cm, 0.01 mS/CM±2% F.S.
TDS (Total Dissolved Solids)0-10000 ppm, 10.1-200.0 ppt1 ppm, 10 ppm±2% F.S.
ORP (Oxidation-Reduction Potential)±999 mV1 mV±15 mV
Temperature0.0℃-60.0℃ (32.0℉-140℉)0.1℃/0.1℉±0.5℃

7. Maintenance

Close-up of the YY-1070 probe, showing it is highly sensitive, removable, and replaceable
Figure 7: Removable and replaceable probes for easy maintenance.

8. Troubleshooting

8.1 FAC Value Displays 0 or '----'

Problem: The Free Available Chlorine (FAC) value is displayed as 0 or '----'.

Solution: For the FAC value to display normally, the swimming pool water's pH should be between 6.5-8 pH, and the ORP should be between 487-840 mV. If either the pH or ORP value is outside this specified range, the FAC reading will not be displayed correctly.

ORP Meter explanation with remarks on pH and ORP ranges for normal FAC display
Figure 8: Conditions for normal FAC display based on pH and ORP values.

8.2 Discrepancy with Reagent Test Papers

Problem: The device detects 0 residual chlorine, but a reagent test paper shows a different value.

Solution: The detection principles of the YAGO YY-1070 digital tester and traditional test papers are different. Test papers often have a larger margin of error and cannot be accurately compared with the precise readings provided by electronic equipment.

9. User Tips

TDS value chart indicating water purity levels and examples of water sources
Figure 9: TDS value chart for water quality assessment.
Images showing various applications of the water quality tester, including drinking water, fish tanks, and hydroponics
Figure 10: Wide range of applications for the YY-1070 tester.

10. Warranty & Support

For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.


RCYAGO YY-1070 7-in-1 Digital Water Quality Tester: pH, Salt, TDS, EC, Temp, ORP, FAC Meter

RCYAGO YY-1070 7-in-1 Digital Water Quality Tester: pH, Salt, TDS, EC, Temp, ORP, FAC Meter

0:27 • 1280×720 • visual_overview

Related Documents - YY-1070

Preview DO-9100 Pen Type Intelligent Dissolved Oxygen Analyzer Instruction Manual
This document provides instructions for the DO-9100 pen type intelligent dissolved oxygen analyzer, covering its features, technical parameters, operation, maintenance, and troubleshooting.
Preview Noyafa Water Quality Meter Operation Manual
Operation manual for Noyafa pen-type water quality meters, covering troubleshooting, attention points, warranty, return authorization, and detailed parameter specifications for various models.
Preview Pen Type Water Quality Meter Operation Manual: pH, TDS, EC, SALT, ORP Measurement
Comprehensive operation manual for the pen-type water quality meter. Covers troubleshooting, calibration, parameters, and usage for measuring pH, TDS, EC, SALT, ORP, H2, and S.G. Includes detailed guides for models like EZ9901, C-600, and more.
Preview YY-103 Series Intelligent pH Meter Instruction Manual
Comprehensive instruction manual for the YY-103 Series Intelligent pH Meter, covering parameters, functions of different models, operation, app usage, and calibration procedures. Includes details on electrode types, specifications, and quality assurance.
Preview Manual Book - Ceiling Fan With Lights YY-2458 Series
This manual provides detailed instructions for the installation, operation, and maintenance of the YY-2458 series Ceiling Fan With Lights. It includes safety rules, parts list, assembly steps, remote control usage, app control instructions, and troubleshooting tips.
Preview Watts Safety Mixing Valves: Preventing Scalds and Legionella
Protect your home and facilities from hot water dangers. This guide details Watts Safety Mixing Valves, designed to prevent scalding injuries, thermal shock, and combat Legionella bacteria through precise temperature control and compliance with ASSE standards.