1. Introduction
Thank you for choosing the Victor 8045mini Digital Multimeter. This compact desktop multimeter is designed for precise electrical measurements, featuring a 3.5-inch TFT screen, 30000 counts display, and a range of functions including voltage, current, resistance, capacitance, frequency, and temperature. It supports dual display, data recording, mathematical operations, and USB communication for enhanced usability. This manual provides essential information for safe and effective operation of your device.

2. Safety Information
Please read and understand all safety warnings and operating instructions before using this multimeter. Failure to do so may result in injury or damage to the instrument.
- Always ensure the correct function and range are selected before making measurements.
- Do not exceed the maximum input limits for any range.
- Exercise extreme caution when working with live circuits.
- Inspect test leads for damage before each use.
- Do not operate the multimeter if it appears damaged or is not functioning properly.
- Keep the device dry and clean.
- Ensure the battery compartment is securely closed before operation.
3. Key Features
- Display: 3.5-inch TFT color screen with 480x320 resolution.
- Measurement Counts: Maximum display of 30000 counts.
- Operating Mode: Manual Range.
- Dual Display Function: Capable of displaying two measurement functions simultaneously.
- Measurement Functions: DC Voltage (up to 1000V), AC Voltage (up to 1000V), DC Current (up to 10A), AC Current (up to 10A), Resistance (up to 300MΩ), Capacitance (up to 300mF), Frequency (up to 300MHz), Temperature (-40 to 1000°C).
- Data Recording: Supports automatic/manual recording with up to 255 points, adjustable time step (1.0s to 10.0s).
- Mathematical Operations: Includes maximum value, dB/dBm, and relative value operations.
- Power Supply: Rechargeable DC5V lithium battery, can also be powered by an external power adapter via Type-C interface.
- Connectivity: USB Device Communication Interface for connection to computers.
- User Interface: Switchable display language (Chinese and English), adjustable backlight brightness, timer shutdown function.
- Safety: Built-in current fuse (600mA 250V / 10A 250V).
- Dimensions: 200mm x 62mm x 91mm.
- Weight: Approximately 0.45kg.

4. Product Components and Accessories
4.1. Panel Layout

- Display screen
- Menu selection key
- Operation key
- Range/direction key
- Power button
- Measurement function button
- Signal input terminal
- USB Device Communication Interface
- External Power Interface
- Indicator Light
4.2. Packaging Accessories

- Host (Multimeter Unit)
- Color Box
- Instruction Manual
- Certificate of Conformity
- USB Data Cable
- Type-C Charging Cable
- Test Probes
- K-type Temperature Probe
5. Initial Setup and Power
5.1. Powering the Device
The Victor 8045mini is equipped with a rechargeable DC5V lithium battery. It can be charged via the Type-C interface using the provided Type-C charging cable and an appropriate power adapter (not included). The device can also be powered directly by an external power adapter through the Type-C interface.

5.2. USB Communication
The multimeter features a USB Device Communication Interface (8 on the panel layout) allowing it to be connected to a computer for remote control and data transfer. Use the provided USB data cable for this connection.

6. Operating Instructions
6.1. Basic Operation
Press the Power button (5) to turn the device on or off. Use the Measurement function buttons (6) to select the desired measurement mode (e.g., V~ for AC Voltage, V- for DC Voltage, Ω for Resistance). The Range/direction key (4) allows for manual range selection if not in auto-range mode.
6.2. Dual Display Function
The 3.5-inch TFT screen supports a dual display function, enabling the simultaneous display of results from two different measurement functions. This is particularly useful in AC/ACA test mode, where a secondary display of the AC signal can be selected.

6.3. Data Recording (Recorder Function)
The multimeter supports both automatic and manual data recording. The maximum number of recorded points is 255. The automatic recording time step can be adjusted from 1.0 seconds to 10.0 seconds. Refer to the on-screen menu (Menu selection key 2) for detailed settings.
6.4. Mathematical Operations
Perform mathematical operations on measurement results, including calculating maximum values, dB/dBm, or relative value operations. Use the 'Math' button (3) to access these functions.
6.5. Temperature Measurement
To measure temperature, connect the K-type temperature probe to the appropriate input terminals. The display will show 'overload' or 'open circuit' if no probe is connected. You can select the unit of measurement between °C, °F, and K.
7. Technical Specifications
The following tables detail the measurement ranges, resolutions, and accuracies for various functions of the Victor 8045mini Digital Multimeter.
7.1. DC Voltage (DCV)
| Range | Resolution | Accuracy |
|---|---|---|
| 300.00mV | 0.01mV | ±(0.1%+5) |
| 3.0000V | 0.0001V | ±(0.05%+5) |
| 30.000V | 0.001V | ±(0.05%+5) |
| 300.00V | 0.01V | ±(0.5%+10) |
| 1000V | 0.1V | ±(0.5%+10) |
7.2. AC Voltage (ACV)
| Range | Frequency Range | Resolution | Accuracy |
|---|---|---|---|
| 300.00mV | 40Hz ~ 400Hz | 0.01mV | ±(1.0%+25) |
| >400Hz ~ 10kHz | ±(2.5%+60) | ||
| 3.0000V | 40Hz ~ 400Hz | 0.0001V | ±(0.8%+25) |
| >400Hz ~ 10kHz | ±(2.0%+25) | ||
| 30.000V | 40Hz ~ 400Hz | 0.001V | ±(0.8%+25) |
| >400Hz ~ 10kHz | ±(2.0%+25) | ||
| 300.00V | 40Hz ~ 400Hz | 0.01V | ±(1.2%+25) |
| >400Hz ~ 5kHz | ±(2.0%+25) | ||
| 750.0V | 40Hz ~ 200Hz | 0.1V | ±(1.0%+25) |
| >200Hz ~ 1kHz | ±(1.5%+25) |
7.3. DC Current (DCA)
| Range | Resolution | Accuracy |
|---|---|---|
| 300.00uA | 0.01uA | ±(0.15%+20) |
| 3.0000mA | 0.0001mA | ±(0.15%+20) |
| 30.000mA | 0.001mA | ±(0.15%+20) |
| 300.00mA | 0.01mA | ±(1.2%+50) |
| 3.0000A | 0.0001A | ±(1.2%+50) |
| 10.000A | 0.001A | ±(1.2%+50) |
7.4. AC Current (ACA)
| Range | Frequency Range | Resolution | Accuracy |
|---|---|---|---|
| 300.00uA | 40Hz ~ 400Hz | 0.01uA | ±(0.8%+30) |
| >400Hz ~ 10kHz | ±(3.0%+30) | ||
| 3.0000mA | 40Hz ~ 400Hz | 0.0001mA | ±(0.8%+30) |
| >400Hz ~ 10kHz | ±(3.0%+30) | ||
| 30.000mA | 40Hz ~ 400Hz | 0.001mA | ±(0.8%+30) |
| >400Hz ~ 10kHz | ±(3.0%+30) | ||
| 300.00mA | 40Hz ~ 400Hz | 0.01mA | ±(1.2%+30) |
| >400Hz ~ 5kHz | ±(3.0%+30) | ||
| 3.0000A | 40Hz ~ 400Hz | 0.0001A | ±(1.5%+30) |
| >200Hz ~ 1kHz | ±(2.5%+25) | ||
| 10.000A | 40Hz ~ 400Hz | 0.001A | ±(1.5%+30) |
| >200Hz ~ 1kHz | ±(2.5%+25) |
7.5. Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 300.00Ω | 0.01Ω | ±(0.5%+10) |
| 3.0000kΩ | 0.0001kΩ | ±(0.5%+10) |
| 30.000kΩ | 0.001kΩ | ±(0.5%+10) |
| 300.00kΩ | 0.01kΩ | ±(0.5%+10) |
| 3.0000MΩ | 0.0001MΩ | ±(0.8%+25) |
| 30.000MΩ | 0.001MΩ | ±(1.5%+50) |
| 300.0MΩ | 0.1MΩ | ±(5.0%+50) |
7.6. Capacitance
| Range | Resolution | Accuracy |
|---|---|---|
| 30.000nF | 0.001nF | ±(5.0%+50) |
| 300.00nF | 0.01nF | ±(5.0%+50) |
| 3.0uF | 0.1nF | ±(4.0%+50) |
| 30.000uF | 0.001uF | ±(4.0%+50) |
| 300.00uF | 0.01uF | ±(4.0%+50) |
| 3.0mF | 0.001mF | ±(10%+50) |
| 30.000mF | 0.001mF | ±(20%+50) |
| 300.00mF | 0.01mF | ±(20%+50) |
Note: Measurement speed for capacitors below 30mF, maximum 6s.
7.7. Frequency
| Range | Resolution | Accuracy |
|---|---|---|
| 10Hz ~ 300MHz | 0.01Hz - 0.01MHz | ±(0.1%+5) |
7.8. Temperature
| Range | Resolution | Accuracy |
|---|---|---|
| (-40 ~ 1000)°C | 0.1°C | ±(1.5%+3°C) |
7.9. General Characteristics
| Characteristic | Value | Characteristic | Value |
|---|---|---|---|
| Display | 3.5-inch color screen, 480 * 320 | TRMS | Yes |
| Precision | 30000 words | Interface | USB Device |
| Coefficient | ≤ Basic accuracy x0.1°C | Power Supply | Built in 2 lithium batteries |
| ON-OFF/Diode | Yes | Size | 200*91*62mm |
| Work Environment | 0°C~40°C | Weight | About 0.45kg |



8. Maintenance
- Cleaning: Wipe the device with a soft, damp cloth. Do not use abrasive cleaners or solvents.
- Storage: Store the multimeter in a cool, dry place away from direct sunlight and extreme temperatures.
- Battery Care: Recharge the lithium battery regularly to maintain its lifespan. If storing for extended periods, charge it to about 50% and recharge every few months.
- Fuse Replacement: If the current measurement function fails, the internal fuse may need replacement. Refer to a qualified technician for fuse replacement to ensure proper specifications are met.
9. Troubleshooting
- No Display: Check if the device is powered on. Ensure the battery is charged or connected to an external power source.
- Inaccurate Readings: Verify that the correct function and range are selected. Check test lead connections and ensure they are not damaged. Calibrate the device if necessary (professional service recommended).
- 'OL' or 'Overload' Display: This indicates that the measured value exceeds the selected range. Switch to a higher range or verify the circuit conditions.
- USB Connection Issues: Ensure the USB cable is securely connected to both the multimeter and the computer. Install any necessary drivers if prompted by your operating system.
10. User Tips
- Utilize the dual display feature to monitor two parameters simultaneously, which can be very efficient for diagnostics.
- Take advantage of the data recording function for logging measurements over time, especially for intermittent issues or long-term monitoring.
- The adjustable backlight brightness is useful in various lighting conditions, extending battery life in well-lit environments.
- For precise temperature measurements, ensure the K-type probe is properly inserted and making good contact with the surface being measured.
11. Warranty and Support
The Victor 8045mini Digital Multimeter is CE certified, indicating compliance with European safety standards.

For specific warranty details, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.
12. Factory Production and Testing
Watch this video to see the manufacturing and testing processes that ensure the quality and precision of your Victor 8045mini Digital Multimeter.
Factory video showing the manufacturing and testing process of the multimeter.